Keithley

/

Keithley 2440-C

SCPI COMMAND REFERENCE

Keithley 2440-C

Keithley 2440-C

The Keithley 2440-C is Keithley's source measure unit in the Series 2400 SourceMeter family. Its SCPI remote-control interface exposes 578 commands across 60 subsystems - 317 settable and 261 queries - and every command is documented here with its parameters, accepted values, return format, and a copy-paste PyVISA example for automating test and measurement.

578

Commands

317

Settable

261

Queries

60

Subsystems

SCPI

1996.0

Verified

2026-07-16

All commands

Programs the V-source protection value, which sets the maximum voltage level the SourceMeter can output. Available settings depend on the model, and the power-on default is NONE.

NameTypeValuesDescription
limitenum20V | 40V | 60V | 80V | 100V | 120V | 160V | NONESelects the V-source protection limit.
:SOUR:VOLT:PROT 20

Sets the I-source amplitude. The value is specified in amps.

NameTypeValuesDescription
nrealAI-source amplitude to set.

Sets the source output level for the voltage source. The manual describes it as the V-source amplitude, in volts.

NameTypeValuesDescription
nrealVVoltage source amplitude to set.
:SOURce:VOLTage:LEVel <n>
:SOUR:VOLT:LEV 10
:SOURce:FUNCtionset / query

Selects the source function. The allowed values are voltage or current.

NameTypeValuesDescription
nameenumVOLTage | CURRentSource function to select.
:SOURce[1]:DELayset / query

Sets the source delay for the selected source function. The command is listed among source memory saved configurations.

Sets or reports the source auto delay state for source channel 1. The manual lists it as a saved configuration command for sweep operation.

Selects the source function shape for the 2430 pulse function. The manual lists this command as a saved configuration item.

Sets the source pulse delay for the selected source. This command is listed as the remote command for 2430 pulse delay in the source memory saved configurations table.

Sets the sweep direction. UP sweeps from start to stop, and DOWn sweeps from stop to start.

NameTypeValuesDescription
nameenumUP | DOWnSweep direction.
:SOURce:SAVEset / query

Save setup in memory at the specified memory location.

NameTypeValuesDescription
ninteger-Memory location to save the setup in.
:SOURce:SAVE <n>

Queries the auto clear mode for the selected source. The mode can be ALWays or TCOunt.

enum - Returns the current auto clear mode.

Queries the state of auto clear for the selected source. For Model 2430 Pulse Mode, auto clear is always enabled and the command is ignored.

boolean - Returns whether auto clear is enabled or disabled.

Queries the I-Source mode. The mode selects fixed, sweep, or list current-source operation.

enum - Returns the current I-Source mode.

Queries whether autoranging is enabled for the I-Source current range. The command returns the current auto-ranging state.

enum - Returns the autoranging state for the current range setting.

Queries the selected I-Source range setting. The source range is part of the current-source configuration under :SOURce[:CURRent].

enum - Returns the current I-Source range setting.

Queries the state of auto settling time for the source delay. For the Model 2430 Pulse Mode, source delay is not used and this command is ignored.

boolean - Returns the auto settling time state.

Queries the source settling time. For Model 2430 Pulse Mode, source delay is not used and the command is ignored.

real (s) - Returns the source settling time in seconds.

Queries the output mode for the source function. Model 2430 only; the available modes are DC and PULSE.

enum - Returns the source output mode.

Queries the center point for the current sweep configuration.

real - Returns the current center point.

Queries the current sweep span. The span is a source current setting under the CURRent subsystem.

real - Returns the present current span setting.

Specifies the start level for the I-sweep. The corresponding query returns the start level for current sweep.

NameTypeValuesDescription
nreal0 - 0Start level for the I-sweep.
real - Returns the start level for current sweep.

Sets the current sweep step value. The query form returns the step value for the current sweep.

real - Returns the step value for the current sweep.

Queries the stop level for the current sweep.

real - Returns the stop level for the current sweep.

Sets the current level when triggered for the selected source path. The level value is specified as a numeric amplitude parameter.

NameTypeValuesDescription
nreal-999.9999e+18 - +999.999e+18Current level to apply when triggered.

Queries the triggered current level state in the current source path. The slice lists the associated setter as :STATe <b>, which enables or disables current scaling factor, and the query as :STATe? to query that state.

boolean - Returns the current scaling factor state for the triggered current level setting.

Queries the V-source mode. The mode can be FIXed, SWEep, or LIST.

enum - Returns the present V-Source mode.

Queries whether voltage autoranging is enabled or disabled. This command is under the source voltage path and is active when the V source is active.

enum - Returns the autoranging state.

Queries the V-source range setting. The command is part of the :SOURce[1]:VOLTage path.

returns - Returns the V-source range setting.

Queries the first sweep point soak time.

real (s) - Returns the soak time for the first sweep point.

Queries the center point for the voltage sweep. The command is part of the :SOURce[1]:VOLTage sweep setup branch.

real - Returns the center point for the voltage sweep.

Returns whether the voltage protection limit has been detected. It yields 1 for yes and 0 for no.

integer - Voltage limit detected: 1 (yes), 0 (no).

Sets or queries the voltage protection level for the selected source. The command defines the output voltage limit path under :VOLTage:PROTection.

NameTypeValuesDescription
leveloptrealMINimum - MAXimumVoltage limit level.
real (V) - Returns the present voltage limit level.

Queries the voltage span setting for the selected source voltage path. The span is used to define the sweep range.

real (V) - Returns the voltage span.

Queries the stop level for the voltage sweep. The value is the present V-sweep stop point.

real - Present stop level for the voltage sweep.

Queries the voltage scaling factor for the triggered voltage source level. The scaling factor applies to the triggered voltage level path under :SOURce[1]:VOLTage[:LEVel]:TRIGgered.

real - Returns the voltage scaling factor.

Sets the triggered voltage level for the source voltage function. This command is part of the voltage source level subsystem and uses the triggered level path.

NameTypeValuesDescription
nreal-999.9999e+18 - +999.9999e+18Voltage level to apply when triggered.

Queries the triggered voltage amplitude for the selected source. The command is part of the voltage source level hierarchy and corresponds to the triggered amplitude setting.

real (V) - Returns the specified voltage level in volts.

Queries the start point for the current source list sweep. The command belongs to the :LIST:CURRent subtree and returns the configured list sweep start point.

integer - Returns the list sweep start point.

Queries the I-Source list. The command is part of the LIST source mode under :SOURce[1]:LIST.

string - Returns the I-Source list.

Configure Source Memory Sweep.

NameTypeValuesDescription
ninteger1 - 100Specifies the number of sweep points.

Queries the number of sweep points in Source Memory sweep mode.

integer - Returns the number of sweep points in the source memory sweep.

Recalls settings from a source memory location.

NameTypeValuesDescription
ninteger1 - 100Memory location to recall settings from.

Configures the start location for Source Memory Sweep.

NameTypeValuesDescription
NRfreal1 - 100Start location for Source Memory Sweep.

Queries the abort-on-compliance state for source sweep mode. The setting controls whether sweep aborts occur never, early, or late on compliance.

enum - Returns the abort-on-compliance state.

Queries the sweep direction for the source sweep mode. The sweep direction is from start to stop (UP) or from stop to start (DOWN).

enum - Returns the sweep direction.

Queries the number of sweep points in the source sweep setup. The count is documented in the source sweep command summary.

integer - Number of sweep points in the sweep.

Queries the source ranging mode used for sweep source mode.

enum - Returns the present source ranging mode.

Queries the sweep spacing type for source sweep mode. The spacing can be linear or logarithmic.

enum - Returns the current sweep spacing type.

Queries the pulse delay for Pulse Mode on Source 1. The delay is part of the :PULSe configuration and is measured in seconds.

real (s) - Returns the pulse delay in seconds.

Queries the configured pulse width for Pulse Mode on Source 1. The width is set with :SOURce[1]:PULSe:WIDTh <n> and is documented as 0.00015 to 0.005 seconds.

real (sec) - Returns the pulse width.

Turns the source output off. The output turns off after all programmed source-measure operations are completed and the instrument returns to the idle state; if auto output-off is enabled, it turns off automatically.

NameTypeValuesDescription
immediateopt--Optional immediate form shown in the syntax line.

Controls auto output-off for the source. When enabled, an INITiate, READ?, or MEASure? starts source-measure operation and the output turns on at the beginning of each SDM cycle, then turns off after each measurement is completed.

NameTypeValuesDescription
benum1 | ON | 0 | OFFEnable or disable auto output-off.

Selects the auto mode for source auto output-off. The ALWAYS option turns the source off with each SDM cycle, and TCOunt turns it off after the trigger count expires.

NameTypeValuesDescription
nameenumALWAYS | TCOuntSelects the auto mode.

Selects the DC sourcing mode for the specified source. The source can be set to fixed, list, or sweep mode.

NameTypeValuesDescription
nameenumFIXed | LIST | SWEepDC sourcing mode to select.

Selects the source mode. VOLTage uses the V-Source, CURRent uses the I-Source, and MEMory enables memory sweep mode.

NameTypeValuesDescription
nameenumVOLTage | CURRent | MEMorySource mode to select.

Selects the DC sourcing mode for the specified source. The source can be set to fixed, list, or sweep mode.

NameTypeValuesDescription
nameenumFIXed | LIST | SWEepSelects the DC sourcing mode.

Selects the source current range for the I-Source. The range can be set numerically or with the keywords DEFault, MINimum, MAXimum, UP, or DOWN.

NameTypeValuesDescription
nreal-1.05 - 1.05Specifies the I-Source level used to select the range.
qualifieroptenumDEFault | MINimum | MAXimum | UP | DOWNSelects a default, minimum, maximum, next-higher, or next-lower range.

Manually selects the V-Source range by specifying an approximate source level; the instrument chooses the lowest range that can accommodate that level. The query form can be used to read the current, default, minimum, or maximum range.

NameTypeValuesDescription
nreal-210 - 210Specify V-Source level to select the range.
qualifieroptenumDEFault | MINimum | MAXimum | UP | DOWNSelect the default, lowest, highest, next higher, or next lower range.

Enables or disables auto range for the specified current source. When enabled, the instrument automatically selects the most sensitive range for the source level; when disabled, it uses the currently selected range.

NameTypeValuesDescription
stateenum0 | OFF | 1 | ONDisables or enables auto range.

Enables or disables auto range for the specified voltage source. When enabled, the instrument automatically selects the most sensitive range for the source level; when disabled, it uses the current range.

NameTypeValuesDescription
stateenum0 | OFF | 1 | ONEnables or disables auto range.

Sets the fixed source amplitude immediately. The allowable range depends on whether the source is current or voltage and on the instrument model.

NameTypeValuesDescription
nreal-1.05 - 1.05Set I-Source amplitude.
qualifieroptenumMINimum | MAXimum | DEFaultAllows setting the minimum, maximum, or default amplitude instead of a numeric value.

Sets the fixed voltage source amplitude immediately. The accepted range depends on the instrument model; DEFault, MINimum, and MAXimum are also supported.

NameTypeValuesDescription
nreal-210 - 210Set V-Source amplitude in volts.
qualifieroptenumDEFault | MINimum | MAXimumSets the amplitude to the default, minimum, or maximum allowed value.

Sets the fixed source amplitude used when triggered. The value depends on the source type and model, and may also be specified with DEFault, MINimum, or MAXimum.

NameTypeValuesDescription
nreal-1.05 - 1.05Set I-Source amplitude.
qualifieroptenumDEFault | MINimum | MAXimumOptional keyword selecting the default, minimum, or maximum triggered amplitude.

Sets the fixed source amplitude to be used when triggered. The command is not valid for list or sweep mode.

NameTypeValuesDescription
nreal-1.05 - 1.05Set I-Source amplitude for 2400/2400-LV/2401.
nreal-210 - 210Set V-Source amplitude for 2400/2400-LV/2401.
nreal-1.05 - 1.05Set I-Source amplitude for 2410.
nreal-1100 - 1100Set V-Source amplitude for 2410.
nreal-3.15 - 3.15Set I-Source amplitude for 2420.
nreal-63 - 63Set V-Source amplitude for 2420.

Selects the source ranging mode for sweeps. BEST selects a single fixed source range for the entire sweep, AUTO selects the most sensitive range for each sweep level, and FIXed holds the present range when the sweep starts.

NameTypeValuesDescription
nameenumBEST | AUTO | FIXedSource ranging mode for the sweep.

Selects the scale for the sweep.

NameTypeValuesDescription
nameenumLINear | LOGarithmicScale used for the sweep.

Specifies the start current level for a current sweep.

NameTypeValuesDescription
nreal-1.05 - 1.05Set I-Source level.
qualifieroptenumMINimum | MAXimum | DEFaultAllows setting the start current to the minimum, maximum, or default level.

Sets the stop current level for a current sweep. The allowed range depends on the instrument model; the value may also be specified with MINimum, MAXimum, or DEFault.

NameTypeValuesDescription
nreal-1.05 - 1.05Stop current level for the current sweep.
qualifieroptenumMINimum | MAXimum | DEFaultAlternative to a numeric value: sets the stop current to the minimum, maximum, or default level.

Specify the start voltage level for a voltage sweep.

NameTypeValuesDescription
nreal-210 - 210Set the source voltage level for the sweep start.
qualifieroptenumMINimum | MAXimum | DEFaultSelects the minimum, maximum, or default start voltage level.

Specifies the stop voltage level for a voltage sweep.

NameTypeValuesDescription
nreal-210 - 210Set V-Source level (volts).
qualifieroptenumMINimum | MAXimum | DEFaultAllows use of the minimum, maximum, or default voltage level keyword instead of a numeric value.

Specifies the center point of the current sweep. The value range depends on the instrument model and source mode.

NameTypeValuesDescription
nreal-2.1 - 2.1Center point of the current sweep.
qualifieroptenumDEFault | MINimum | MAXimumAllows setting the center point to the default, minimum, or maximum allowed value.

Specifies the span of the current sweep.

NameTypeValuesDescription
nreal-2.1 - 2.1Set I-Source level.
qualifieroptenumMINimum | MAXimum | DEFaultSets the span to the minimum, maximum, or default value.

Specifies the center point of the voltage sweep.

NameTypeValuesDescription
nreal-420 - 420Center point of the voltage sweep.
qualifieroptenumDEFault | MINimum | MAXimumSelects the default, minimum, or maximum voltage center value.

Specifies the span of the voltage sweep.

NameTypeValuesDescription
nreal-420 - 420Set V-Source level (volts).
qualifieroptenumDEFault | MINimum | MAXimumAllows setting the voltage span to the default, minimum, or maximum allowed value.

Specifies the step size for a current sweep. The manual also lists the corresponding voltage-sweep form of the command.

NameTypeValuesDescription
nreal-2.1 - 2.1Sets the I-Source level step size.
qualifieroptenumDEFault | MINimum | MAXimumAllows selection of the default, minimum, or maximum step size.

Specifies the step size for a voltage sweep.

NameTypeValuesDescription
nreal-420 - 420Step size for the V-source level; on 2400-LV and 2401 the range is -42 to 42 V.
qualifieroptenumDEFault | MINimum | MAXimumSelects the default, minimum, or maximum allowable voltage step size.

Sets the total number of source-measure points in a sweep. For linear sweeps the points are equally spaced between start and stop levels; for logarithmic sweeps they are equally spaced on a logarithmic scale.

NameTypeValuesDescription
ninteger1 - 2500Number of source-measure points in the sweep.
:SOURce[1]:SWEep:POINts <n>

Controls abort on compliance during a sweep. The sweep can be aborted never, at the start of the SDM cycle, or at the end of the SDM cycle.

NameTypeValuesDescription
nameenumNEVer | EARLy | LATEMode for aborting a sweep on compliance.

Sets the direction of a sweep. UP runs the sweep from start to stop, and DOWn runs it from stop to start.

NameTypeValuesDescription
nameenumUP | DOWnDirection of sweep execution.

Defines the I-Source list for list sourcing mode. The list can contain up to 100 source values, and the instrument will sequentially source each current value when operation is started.

NameTypeValuesDescription
<NRf list>real-1.05 - 1.05Comma-separated list of current source values.
:SOURce[1]:LIST:CURRent 0.01, 0.13, 0.005

Defines the V-source list for list sourcing mode. The list can contain up to 100 source values and is sourced sequentially during operation, with a measurement performed at each level.

NameTypeValuesDescription
NRf listreal-210 - 210Comma-separated list of V-source values.
:SOURce[1]:LIST:CURRent 0.01, 0.13, 0.005

Adds one or more values to an existing current source list. Up to 100 values can be appended at a time, and multiple appended lists can be used to build a list of up to 2500 points.

NameTypeValuesDescription
listreal-1.05 - 1.05One or more current source values to append to the list.

Determines the length of the specified current source list. The response message indicates the number of source values in the list.

integer - Number of source values in the current source list.

Adds one or more values to an existing voltage source list. Up to 100 values can be appended at a time, and appended lists are added to the end of the list.

NameTypeValuesDescription
listreal-210 - 210One or more voltage source values to append to the list.

Queries the length of the specified voltage source list. The response message indicates the number of source values in the list.

integer - Number of source values in the specified V-Source list.

Sets the starting point in the current list sweep. The start point is 1-based and must be less than or equal to the number of points in the sweep; the starting point is restored when the sweep direction is changed back to up.

NameTypeValuesDescription
ninteger1 - 100Set start point of list sweep.

Sets the starting point of the voltage list sweep. The start point is 1-based and must be less than or equal to the number of points in the sweep; the default is 1.

NameTypeValuesDescription
ninteger1 - 100Set start point of list sweep.

Saves the present instrument setup in the specified memory location. Up to 100 setups can be saved.

NameTypeValuesDescription
NRfinteger1 - 100Specify memory location
:SOURce[1]:MEMory:SAVE <NRf>

Specifies the number of sweep points to execute. For example, if setups are saved in memory locations 1 through 12, a 12-point sweep can be specified with this command.

NameTypeValuesDescription
NRfinteger1 - 100Number of sweep points.

Returns the SourceMeter to the setup stored at the specified memory location.

NameTypeValuesDescription
locationinteger1 - 100Specify memory location

Selects the starting location for a Source Memory Sweep. The starting location is specified as a memory location number.

NameTypeValuesDescription
nrfinteger1 - 100Memory location to start from.

Sets the triggered current scaling factor. The same command is also documented for triggered voltage scaling factor on the related voltage path.

NameTypeValuesDescription
nreal-999.9999e+18 - 999.9999e+18Scaling factor.

Sets the triggered scaling factor for the source level. The same command is documented for both current and voltage source levels.

NameTypeValuesDescription
nreal-999.9999e+18 - 999.9999e+18Scaling factor.
:SOURce[1]:SOAKset / query

Sets the multiple mode soak time. With SYST:RCMode set to MULTiple, it specifies how long the unit will sit after the first sweep point while autoranging to let a multiple SourceMeter configuration settle.

NameTypeValuesDescription
NRfreal0.000 - 9999.999Soak time.

:SENSe:FUNCtionset / query

Selects the measure function. The function can be voltage or current.

NameTypeValuesDescription
functionenumVOLTage | CURRentMeasure function to select.

Sets the current compliance for manual ohms measurements.

NameTypeValuesDescription
nreal-Current compliance value for manual ohms measurements.

Sets the voltage compliance for manual ohms measurements.

NameTypeValuesDescription
nreal-Voltage compliance value for manual ohms.

Selects the manual current measurement range. The range value is specified as n = range.

NameTypeValuesDescription
nreal-Current measurement range to select.

Enables or disables auto-ranging for the current measurement range. The state argument selects ON or OFF.

NameTypeValuesDescription
stateenumON | OFFEnables or disables auto amps range.

Selects the manual ohms measurement range.

NameTypeValuesDescription
nreal-Range setting value.

Selects the manual volts measure range.

NameTypeValuesDescription
nreal-Measurement range to select.

Enables or disables auto volts range selection.

NameTypeValuesDescription
stateenumON | OFFTurns auto volts ranging on or off.
:SENSe:AVERageset / query

Enable or disable the filter. The manual notes that for the Model 2430 Pulse Mode, the filter cannot be enabled.

NameTypeValuesDescription
stateenumON | OFFEnables or disables the filter.

Sets the filter count used by the measurement averaging filter.

NameTypeValuesDescription
countinteger1 - 10Filter count.

Selects the filter type for measurement averaging.

NameTypeValuesDescription
modeenumREP | MOVFilter type selection.

Sets the current integration rate for the specified sense function. It is the source-memory-saved configuration for current integration rate.

Sets the concurrent functions for the selected sense subsystem. It is listed as a source-memory-saved configuration mode.

Sets the resistance measurement integration rate by specifying the number of power line cycles. This command is listed as the remote command for resistance integration rate.

Sets the voltage integration rate (NPLC) for the selected sense function. This command is listed in the source memory saved configurations table.

Enables or disables current range holdoff.

NameTypeValuesDescription
stateenumON | OFFEnables or disables current range holdoff.

Returns only the most recent reading from SENSe[1] data.

string - Returns the most recent reading.

Queries the state of the specified measurement function. The instrument returns 1 if the function is on and 0 if it is off.

NameTypeValuesDescription
nameenumVOLTage[:DC] | CURRent[:DC] | RESistanceName of the measurement function whose state is queried.
integer - 1 if the specified function is on, 0 if it is off.

Selects measurement function(s) for Sense 1. The command accepts a function list or the ALL keyword to enable all functions when concurrent mode is enabled, or only ohms when concurrent mode is disabled.

NameTypeValuesDescription
function listenumVOLTage[:DC] | CURRent[:DC] | RESistanceFunctions to enable or disable.

Returns the list of measurement functions that are enabled. The command belongs to the SENSe[1]:FUNCtion subsystem.

enum - Returns the enabled measurement function list.

Queries the synchronization status of measure and compliance range synchronization for current protection. The command is part of the current protection subtree under :SENSe[1]:CURRent[:DC].

enum - Returns the range synchronization status.

Sets the auto-ranging lower limit for current measurement. The limit applies to the current range under :SENSe[1]:CURRent[:DC]:RANGe.:AUTO.

NameTypeValuesDescription
nreal-105e-6 - 105e-6Auto-ranging lower limit current.

Queries the auto-ranging lower limit for the current measurement range. Returns the lower limit used by the current auto range setting.

real (A) - The auto-ranging lower limit for current, in amperes.

Queries the present upper measurement range for current in the selected DC current sense path. The range is selected by specifying the expected current reading.

real - Returns the present current measurement range upper value.

Sets the lower limit for auto ranging on the voltage measurement range. The command belongs under the voltage range path and applies to DC voltage measurement configuration.

NameTypeValuesDescription
nreal0.21Lower limit for auto ranging.

Queries the lower limit used for auto-ranging on the voltage measurement range. The command is part of the voltage range configuration path.

real - Returns the auto-ranging lower limit for the voltage measurement range.

Queries the selected voltage measurement range. The range is selected by specifying the expected voltage reading.

real - Returns the selected voltage range.

Sets or queries the resistance range holdoff delay in seconds.

NameTypeValuesDescription
delayreal0 - 999.9999Holdoff delay in seconds.

Queries the resistance range holdoff delay. The delay is set under the resistance range holdoff command and is expressed in seconds.

real (s) - The current resistance range holdoff delay in seconds.

Enable or disable current range holdoff for source memory sweeps.

NameTypeValuesDescription
stateboolean-Turns current range holdoff on or off.

Sets the lower limit for resistance auto ranging. The command belongs to the resistance range configuration subtree.

NameTypeValuesDescription
countreal-Lower limit value for auto range.

Queries the lower limit of the resistance range. This is part of the resistance range configuration commands.

real - The present lower resistance range limit.

Sets the upper limit for auto range in resistance measurement. The detailed parameter values are referenced in the command reference notes.

NameTypeValuesDescription
nreal-Upper limit value for auto range.

Queries the present resistance measurement range upper setting. The command is part of the resistance range configuration path under :SENSe[:1]:RESistance:RANGe[:UPPer].

real - Returns the present resistance range upper value.

Queries range synchronization for voltage protection. The command is part of the voltage compliance/protection subsystem and reports whether measure and compliance range synchronization is enabled.

enum - Returns the range synchronization state.

Specifies the digital filter count for voltage DC average filtering. The count is adjustable from 1 to 100, with a default of 10.

NameTypeValuesDescription
ninteger1 - 100Filter count.

Queries the digital filter count for the voltage DC average filter path. The count is documented as a value from 1 to 100, with a default of 10 for the setter.

integer - Returns the filter count.

Specifies the digital filter type for voltage DC averaging. The filter type can be MOVing or REPeat.

NameTypeValuesDescription
nameenumMOVing | REPeatFilter type to use.

Enables or disables the digital filter state for the voltage DC averaging function. The filter path is under :SENSe[1]:VOLTage[:DC]:AVERage.

NameTypeValuesDescription
benumON | OFFEnable or disable the filter.

Queries the digital filter state for the voltage DC average function. For Model 2430 Pulse Mode, filtering is not used and this command is ignored.

enum - Returns the filter state.

Enables or disables concurrent measurements. When enabled, the instrument can measure more than one selected function simultaneously; when disabled, only one measurement function can be enabled.

NameTypeValuesDescription
benum0 | OFF | 1 | ONDisable or enable concurrent measurements.
enum - Returns the current state of concurrent measurements.

Specifies one or more measurement functions to be enabled when concurrent measurements are enabled. Each function in the list must be quoted and separated by commas.

NameTypeValuesDescription
function liststring-One or more measurement functions to enable.
[:SENSe[1]]:FUNCtion[:ON] <function list>
[:FUNCtion] "VOLTage", "CURRent"

Query functions that are enabled when concurrent measurements are enabled. Each function in the list must be enclosed in quotes and separated by commas.

string - Returns the enabled functions as a quoted comma-separated function list.
:FUNCtion “VOLTage”, “CURRent”
:FUNCtion:OFF ‘VOLTage’, ‘CURRent’

Disables all measurement functions. If concurrent measurements are disabled, only the ohms function will be enabled by the related enable-all command; this command turns all functions off.

Queries the number of measurement functions that are disabled. The response indicates how many functions are currently off.

integer - Number of functions that are disabled.

Queries the state of the specified measurement function. A returned response of "0" indicates the function is disabled, and "1" indicates it is enabled.

NameTypeValuesDescription
nameenumCURRent[:DC] | VOLTage[:DC] | RESistanceMeasurement function to query.
boolean - Returns 0 if the specified function is disabled and 1 if it is enabled.

Enables all measurement functions. If concurrent measurements are enabled, amps, volts, and ohms are performed simultaneously; if concurrent measurements are disabled, only the ohms function is enabled.

Queries the number of measurement functions that are enabled or disabled. When sent as :ON:COUNt?, the response indicates the number of enabled functions; when sent as :OFF:COUNt?, the response indicates the number of disabled functions.

integer - Number of measurement functions enabled for :ON:COUNt? or disabled for :OFF:COUNt?.

Queries the selected ohms measurement mode. The manual describes the related setter as selecting MANual or AUTO ohms mode.

enum - Returns the current ohms measurement mode.

Controls offset-compensated ohms measurement. A value of 1/ON enables offset compensation and 0/OFF disables it.

NameTypeValuesDescription
benum1 | ON | 0 | OFFEnable or disable offset compensation.

Queries the present current measurement range in DC mode. The returned value may also be requested as MINimum, MAXimum, or DEFault on this command family if supported by the instrument.

real (A) - Returns the present selected current range in amps.

Selects the upper measurement range for current in DC mode. The value can be a numeric expected reading or the keywords UP, DOWN, DEFault, MINimum, or MAXimum.

NameTypeValuesDescription
nreal-1.05 - 1.05Expected reading in amps used to select the range.
qualifieroptenumUP | DOWN | DEFault | MINimum | MAXimumAlternative range-selection keywords.

Selects the resistance measurement range. The command accepts a numeric range value or the keywords DEFault, MINimum, MAXimum, UP, and DOWN.

NameTypeValuesDescription
nreal0 - 2.1e8Expected reading in ohms.
qualifieroptenumMINimum | MAXimum | DEFault | UP | DOWNAlternate range selection keyword.

Selects the resistance measurement range. The range can be set to a numeric expected reading, or to DEFault, MINimum, MAXimum, UP, or DOWN.

NameTypeValuesDescription
nreal0 - 2.1e8Expected reading in ohms.
qualifieroptenumDEFault | MINimum | MAXimum | UP | DOWNSpecial range-selection keyword: DEFault uses the default range, MINimum selects the minimum range, MAXimum selects the maximum range, UP selects the next higher range, and DOWN selects the next lower range.

Returns the present measurement range for DC voltage. The response can be a numeric value or a range qualifier such as UP or DOWN, depending on the instrument model and selected range.

enum - Returns the present DC voltage measurement range, or a range-direction keyword when applicable.

Selects the measurement range for the specified voltage DC sense function. The range can be chosen by expected reading or by the UP/DOWN keywords; MINimum, MAXimum, and DEFault are also accepted.

NameTypeValuesDescription
nreal-210 - 210Expected reading used to select the most sensitive voltage range that accommodates it.
qualifieroptenumMINimum | MAXimum | DEFault | UP | DOWNSelects the minimum range, maximum range, default range, or the next higher/lower measurement range.

Controls auto ranging for amps on the current measurement function. When enabled, the instrument automatically goes to the most sensitive range; when disabled, it remains at the automatically selected range.

NameTypeValuesDescription
stateenum0 | OFF | 1 | ONEnables or disables auto range.

Sets the auto-ranging lower limit for the current measurement range. The lower limit must be less than or equal to the upper limit, and if the lower and upper limits are equal, auto ranging is effectively disabled.

NameTypeValuesDescription
nreal-105e-6 - 105e-6Auto range lower limit for amps.

Queries the auto ranging lower limit for current measurement. The lower limit is programmable and must be less than or equal to the upper limit.

real (A) - Returns the auto range lower limit for amps.

Controls auto ranging for current measurements. With auto ranging enabled, the instrument automatically selects the most sensitive range for the measurement.

enum - Returns the state of auto range.

Controls auto ranging for ohms. When auto ranging is enabled, the instrument automatically goes to the most sensitive range for the measurement; when disabled, it remains at the automatically selected range.

NameTypeValuesDescription
stateenum0 | OFF | 1 | ONEnable or disable auto range.

Sets the auto-ranging lower limit for resistance measurements. The lower limit must be less than or equal to the upper limit.

NameTypeValuesDescription
nreal-2.1e8 - 2.1e8Auto-ranging lower limit for ohms (2400 and 2410).

Queries the auto range lower limit for ohms. Auto range lower limits are intended primarily for SYST:RCM MULT support, and the lower limit must be less than or equal to the upper limit.

real (Ohms) - The auto range lower limit for ohms.

Controls auto ranging for ohms. When enabled, the instrument automatically goes to the most sensitive range to perform the measurement; when disabled, it remains at the automatically selected range.

enum - Returns the state of auto range.

Controls auto ranging for volts. When auto ranging is enabled, the instrument automatically goes to the most sensitive range to perform the measurement; when disabled, it remains at the automatically selected range.

NameTypeValuesDescription
stateenum0 | OFF | 1 | ONEnables or disables auto range.

Sets the auto ranging lower limit for volts. The lower limit must be less than or equal to the upper limit, and if the lower limit equals the upper limit auto ranging is effectively disabled.

NameTypeValuesDescription
nreal-21 - 21Volts lower limit.

Queries the auto-ranging lower limit for voltage ranges. The lower limit is programmable and must be less than or equal to the upper limit.

real (V) - Returns the auto-ranging lower limit for volts.

Controls auto ranging for volts. When enabled, the instrument automatically goes to the most sensitive range to perform the measurement; disabling it leaves the instrument at the automatically selected range.

NameTypeValuesDescription
stateenum0 | OFF | 1 | ONEnable or disable auto range.
enum - Returns the current auto-range state.

Queries the auto ranging upper limit for current in amps. Auto range upper limits are intended primarily for SYST:RCM MULT support.

real (Ohms) - The auto ranging upper limit for current. For current and voltage, the upper limit is controlled by the compliance range and is available only as a query.

Enables or disables current range holdoff. This feature is available only by remote and is used to speed up low-current measurements when sourcing voltage and measuring current.

NameTypeValuesDescription
benumON | OFFEnable or disable current range holdoff.

Sets the current range holdoff delay. This delay is used with current range holdoff, a feature available only for source memory sweeps and remote control.

NameTypeValuesDescription
NRfreal0 - 999.9999Delay in seconds.

Queries the current range holdoff delay. This feature is available only for current range holdoff in source memory sweeps.

real (s) - Returns the holdoff delay in seconds.

Sets the auto ranging upper limit for resistance. This limit is intended primarily for SYST:RCM MULT support; when autoranging is disabled, any range above the limit can be programmed manually (ohms only).

NameTypeValuesDescription
nreal-2.1e8 - 2.1e8Ohms upper limit for 2400 and 2410 models.

Queries the auto-ranging upper limit for resistance. The upper limits are intended primarily for SYST:RCM MULT support; for resistance, the limit is in ohms and depends on the instrument model.

real (Ohms) - Auto range upper limit for ohms.

Queries the auto-ranging upper limit for volts. For voltage and current, the upper limit is controlled by the compliance range and is available only as a query.

real (Ohms) - Auto ranging upper limit for volts.

Sets the current compliance limit for the selected sense function. The manual also lists corresponding voltage-compliance commands nearby, but this command is for current protection only.

NameTypeValuesDescription
nreal-1.05 - 1.05Current compliance limit; may also be set to DEFault, MINimum, or MAXimum.
qualifieroptenumDEFault | MINimum | MAXimumAlternative keyword for the compliance limit.

Sets the current compliance limit. The allowed range depends on the instrument model, and the limit can also be set to DEFault, MINimum, or MAXimum.

NameTypeValuesDescription
nreal-1.05 - 1.05Current compliance limit.
qualifieroptenumDEFault | MINimum | MAXimumSets the compliance limit to the default, minimum, or maximum allowed value.

Sets the voltage compliance limit. The allowable range depends on the instrument model, and the value may also be set to DEFault, MINimum, or MAXimum.

NameTypeValuesDescription
nreal-210 - 210Voltage compliance limit.
qualifieroptenumDEFault | MINimum | MAXimumSets the compliance limit to the default, minimum, or maximum value.

Sets the voltage compliance limit for the selected sense function. The allowable range and default/min/max values depend on the instrument model.

NameTypeValuesDescription
nreal-210 - 210Voltage compliance limit.
qualifieroptenumDEFault | MINimum | MAXimumSelects the default, minimum, or maximum voltage compliance limit.

Controls current measure and compliance range sync. Turning it on makes the measurement range follow the compliance range setting when autoranging is off, and changing one setting automatically changes the other.

NameTypeValuesDescription
benumOFF | ONEnables or disables range sync.

Returns whether the current source is in compliance. A returned value of 1 means the source is in compliance, and 0 means it is not in compliance.

boolean - Returns the current compliance state; 1 indicates the source is in compliance and 0 indicates it is not in compliance.

Controls voltage measure and compliance range sync. Turning it on makes the measurement range follow the compliance range setting when autoranging is off; changing one setting automatically changes the other.

NameTypeValuesDescription
benumOFF | ONDisable or enable range sync.

Queries whether the source is in voltage compliance. A returned value of "1" indicates compliance, and "0" indicates the source is not in compliance.

integer - Returns 1 if the source is in compliance and 0 if the source is not in compliance.

Sets the integration period (measurement speed) for current measurements in power-line cycles (PLC). The setting is global, so changing the speed for voltage, current, or resistance affects all measurement functions.

NameTypeValuesDescription
nreal0.01 - 10Power-line cycles per integration.
qualifieroptenumMINimum | MAXimum | DEFaultOptional keyword to query the default, minimum, or maximum PLC value.
real - Returns the programmed PLC value, or the default, minimum, or maximum PLC value when queried with DEFault, MINimum, or MAXimum.

Sets the integration period (speed) for current measurements in power-line cycles. It is a global command, so setting the speed for one measurement function affects voltage and resistance as well.

real (PLC) - Returns the programmed PLC value for current measurements. With a qualifier, returns the default, minimum, or maximum PLC value.

Queries the programmed NPLC value for resistance measurements. The query also supports the qualifiers DEFault, MINimum, and MAXimum to return the corresponding default or limit value.

NameTypeValuesDescription
qualifieroptenumDEFault | MINimum | MAXimumSelects the programmed value, the *RST default, the minimum, or the maximum PLC value to return.
real (PLC) - Returns the programmed PLC value, or the *RST default, minimum, or maximum value when queried with DEFault, MINimum, or MAXimum.

Sets the integration period (measurement speed) in power-line cycles for measurements. This is a global command; setting the voltage speed also sets current and resistance to the same PLC value.

NameTypeValuesDescription
nreal0.01 - 10Power-line cycles per integration.
qualifieroptenumMINimum | MAXimum | DEFaultSpecifies the minimum, maximum, or default PLC value.

Sets the integration period (measurement speed) in power-line cycles for voltage measurements. This is a global command, so setting voltage speed also affects current and resistance measurements.

NameTypeValuesDescription
nreal0.01 - 10Power-line cycles per integration.
qualifieroptenumMINimum | MAXimum | DEFaultOptional keyword to query the default, minimum, or maximum PLC value.
real - Returns the programmed PLC value. With DEFault, returns the *RST default PLC; with MINimum or MAXimum, returns the corresponding limit value.

Queries the averaging filter count. The manual also documents qualifier forms to query the *RST default count, the minimum allowable count, or the maximum allowable count.

NameTypeValuesDescription
qualifieroptenumDEFault | MINimum | MAXimumOptional query qualifier to return the *RST default, minimum allowable, or maximum allowable filter count.
integer - Returns the current averaging filter count, or the *RST default, minimum allowable, or maximum allowable filter count when queried with DEFault, MINimum, or MAXimum.

Selects the averaging filter type for the sense function. The filter type can be Repeat or Moving, and the filter count is set with :AVERage:COUNt; the command is ignored in Model 2430 Pulse Mode.

NameTypeValuesDescription
nameenumREPeat | MOVingSelects the averaging filter type.
enum - Returns the current filter type.

Enables or disables the digital averaging filter. When enabled, voltage, current, and resistance readings are filtered according to how the filter is configured.

NameTypeValuesDescription
stateenum0 | OFF | 1 | ONEnable or disable the digital filter.

:FUNCset / query

Selects the measurement function. In the ohms programming example, it sets the function to resistance.

NameTypeValuesDescription
functionenum"RES"Measurement function to select.

:RES:MODEset / query

Selects the ohms measurement mode. The example sets it to auto mode.

NameTypeValuesDescription
modeenumAUTOOhms measurement mode.
:RES:RANGset / query

Selects the ohms measurement range. The programming example shows setting the range to 20E3 for a 20 kΩ range.

NameTypeValuesDescription
rangerealΩOhms measurement range to select.
RES:RANG 20E3

:SYSTem:AZERoset / query

Controls auto zero for remote operation. Disabling auto zero can increase pulse speed, but zero drift may reduce pulse measurement accuracy.

NameTypeValuesDescription
stateenumDISABLE | ENABLESelects whether auto zero is disabled or enabled.
:SYSTem:GUARdset / query

Sets the guard type used by Source Memory saved configurations. It is listed in the Source memory saved configurations table as the remote command for Guard type.

:SYSTem:PREsetset / query

Places the SourceMeter into the idle state. The manual notes that :SYSTem:PREset can be executed under any circumstance while operating within the trigger model.

:SYSTem:ERRorset / query

Path to read messages in the error queue. The optional :NEXT?, :ALL?, and :COUNt? forms return and clear error information or count errors.

Path to return error code numbers only. Use [:NEXT]? to return and clear the oldest error code, or :ALL? to return and clear all error codes.

Queries the current guard type.

enum - Returns the selected guard type.

Queries the last pressed key.

integer - Returns the last pressed key number.

Query the current power-on setup.

enum - Returns the power-on setup setting.
:SYSTem:RESset / query

Specify the resistance threshold for contact check. Used with :SYSTem:CCHeck; the default is 50 ohm.

NameTypeValuesDescription
Nrfreal50Resistance threshold for contact check.

Controls auto zero and NPLC caching. The state setting enables auto zero, disables it, or forces an immediate update with ONCE.

NameTypeValuesDescription
stateoptenumOFF | ON | ONCEControls auto zero state.

Queries the auto zero state. Auto zero can be OFF (disabled), ON (enabled), or ONCE (force immediate update of auto zero).

enum - Returns the auto zero state.

Returns a list of NPLC values in the cache, ordered from oldest to newest.

real - A list of NPLC values in the cache from oldest to newest.

Controls NPLC caching.

NameTypeValuesDescription
stateoptenumON | OFF | 1 | 0Enables or disables NPLC caching.

Enables or disables auto line-frequency selection. The setting is not affected by *RST or :SYSTem:PRESet.

NameTypeValuesDescription
stateenumbEnable or disable auto frequency.
enum - Returns the state of auto frequency.

Query the selected line frequency.

real (Hz) - Returns the line frequency in Hz.

Queries the auto range on compliance mode setting. The mode controls whether auto range on compliance uses single or multiple mode.

enum - Returns the current auto range on compliance mode setting.
:SYSTem:TIMEset / query

Timestamp functions. Resets the timestamp to zero seconds and can enable or disable automatic timestamp reset when exiting idle.

NameTypeValuesDescription
autoenumOFF | ONEnable or disable timestamp reset when exiting idle.

Enable or disable timestamp reset when exiting idle.

NameTypeValuesDescription
stateoptenumON | OFFControls whether the timestamp is reset when the instrument exits idle.
:SYSTem:PRESetset / query

Returns the instrument to states optimized for front panel operation. :SYSTem:PRESet defaults are listed in the SCPI tables.

Controls the SourceMeter beeper at a specified frequency and duration. The beeper must be enabled to use this command.

NameTypeValuesDescription
freqreal65 - 2e6Specify frequency in Hz.
timereal0 - 7.9Specify time duration.
:SYSTem:BEEPer500, 1

Initializes battery-backed RAM. This clears trace data, resets buffer size and timestamp format, deletes user-defined math expressions, and reinitializes memory sweep and standard save setups.

Returns the list of NPLC values stored in the cache, from oldest to newest.

string - Returns the list of NPLC values stored in the cache, from oldest to newest.

Enables or disables NPLC caching. When enabled, the instrument caches A/D reference and zero values for recent NPLC settings.

NameTypeValuesDescription
stateenum1 | ON | 0 | OFFEnable or disable NPLC caching.

Enable or disable the beeper. When enabled, a short beep is provided to signal that a front panel key has been pressed.

NameTypeValuesDescription
stateenum1 | ON | 0 | OFFEnable or disable the beeper.

Manually selects the line frequency setting. The manual says to use 50 for 50 Hz or 400 Hz operation, and 60 for 60 Hz operation.

NameTypeValuesDescription
freqinteger50 - 60Line frequency setting to select.

Enables or disables automatic line frequency selection. When enabled, the instrument senses the line frequency on power-up and selects the appropriate setting; manually setting the line frequency disables auto frequency.

NameTypeValuesDescription
stateenum1 | ON | 0 | OFFEnables or disables auto line frequency selection.

Reads all error codes and messages from the Error Queue. All messages are sent to the computer when the SourceMeter is addressed to talk, and are then removed from the queue.

string - All error messages in the Error Queue are returned. All messages are removed from the queue.

Reads all errors from the Error Queue as codes only. It is identical to :SYSTem:ERRor:ALL? except that only the codes are returned and all errors are cleared from the queue.

integer - All error codes in the Error Queue are sent to the computer when the SourceMeter is addressed to talk. The actual messages are not returned.

Reads and returns the oldest error code from the Error Queue. The error is cleared from the queue after the query is sent.

integer - The oldest error code in the Error Queue; the message text is not returned.

Returns the number of messages in the Error Queue. After the command is sent and the SourceMeter is addressed to talk, a decimal number is returned.

integer - The number of messages currently in the Error Queue.

Reads the oldest error from the error queue, returning the code and message. After the command is sent and the instrument is addressed to talk, the oldest message is sent to the computer and removed from the queue.

string - Returns the oldest error as code and message.
:SYSTem:KEYset / query

Simulates a front-panel key press using a numeric key-press code. The query form returns the key-press code number of the last key pressed.

NameTypeValuesDescription
NRfinteger1 - 32Key-press code number to simulate.
integer - Key-press code number of the last key pressed.
:syst:key 15
:SYSTem:LOCalset / query

Removes the SourceMeter from remote state and enables the front panel keys. This action command can only be sent over the RS-232 interface.

:SYSTem:RWLockset / query

Enables or disables local lockout for the front panel keys during remote operation. When enabled, the keys are locked out; when disabled, they remain operational in remote. This command can only be sent over the RS-232 interface.

NameTypeValuesDescription
benum0 | OFF | 1 | ONDisable or enable local lockout.

Returns the current timestamp value.

string - The current timestamp value.

Reads the version of the SCPI standard being used by the SourceMeter.

string - Returns the SCPI standard version.
1996.0
:SYSTem:RCModeset / query

Controls the auto range change mode. In SINGle mode the SourceMeter auto ranges only after the first reading; in MULTiple mode it auto ranges up on compliance in the Delay phase of the Source-Delay-Measure cycle.

NameTypeValuesDescription
nameenumSINGle | MULTipleAuto range change mode.
enum - Returns the current auto range change mode.

Resets the absolute timestamp to 0 seconds. The timestamp also resets to 0 seconds every time the SourceMeter is turned on.

Enables or disables automatic timestamp reset. When enabled, the timestamp is automatically reset when exiting the idle layer of the trigger model.

NameTypeValuesDescription
benum1 | ON | 0 | OFFEnables or disables auto timestamp reset.
:SYSTem:POSetupset / query

Determines the power-on setup conditions of the SourceMeter.

Selects the contact-check threshold resistance. The threshold ranges from 0 to 60 and maps to 2W, 15W, or 50W threshold resistance depending on the numeric value.

NameTypeValuesDescription
resistancereal0 - 60Threshold resistance value used to select the contact-check threshold.

Query sense selection for contact check.

enum - Returns the current sense selection.
:SYSTem:CCHeckset / query

Enables or disables contact check. When contact check is used, remote sensing must be enabled with :SYST:RSEN ON.

NameTypeValuesDescription
benum0 | OFF | 1 | ONDisable or enable contact check.

Sets the contact check threshold resistance. The contact check function indicates a failure when the connection path resistance is greater than the programmed value.

NameTypeValuesDescription
NRfreal0 - 60Threshold resistance for contact check; the manual maps 0 to 2 as 2 Ω, >2 to 15 as 15 Ω, and >15 to 60 as 50 Ω.

Sets the contact check threshold resistance. The contact check function indicates a failure when the connection path resistance is greater than the programmed value.

real (ohm) - Returns the contact check threshold resistance value.

Enables or disables contact check. When contact check is used, remote sensing must be enabled (:SYST:RSEN ON).

NameTypeValuesDescription
benum0 | OFF | 1 | ONEnable or disable contact check.
enum - Returns the current contact check state.

Queries whether Message Exchange Protocol (MEP) is enabled. A return of 1 indicates SCPI protocol is selected; 0 indicates MEP is disabled and the 488.1 protocol is enabled.

integer - Returns 1 if MEP is enabled (SCPI protocol selected) and 0 if MEP is disabled (488.1 protocol enabled).
:SYSTem:MEP[:STATe]?

Controls NDAC hold-off in GPIB 488.1 protocol mode. The default is OFF.

NameTypeValuesDescription
stateenumON | OFFTurns NDAC hold-off on or off.

Enables or disables the measurement filter. In Pulse Mode, the command causes error +831 because filtering cannot be used.

NameTypeValuesDescription
stateenumON | OFFTurns the filter on or off.

Enables or disables concurrent measurements. In Pulse Mode, using this command causes error +831: Invalid in Pulse Mode.

NameTypeValuesDescription
stateenumON | OFFTurns concurrent measurements on or off.

Selects the ohms measurement function when used in Pulse Mode. The manual notes that this command will select the ohms function.

Enables or disables offset-compensated ohms measurements. In Pulse Mode this command is invalid and causes error +831.

NameTypeValuesDescription
stateenumON | OFFTurns offset-compensated ohms on or off.

Sets or queries the measurement function used for signal-oriented measurements. In Pulse Mode, :MEASure[:<function>]? is invalid and causes error +831.

NameTypeValuesDescription
functionoptenum-Measurement function name.

Performs an initiate operation and takes the instrument out of idle so measurements can be made. The command automatically turns the source output on and leaves it on after the instrument returns to idle.

string - Returns the measured reading.

Measures the selected function and returns the reading. The function is optional in the command syntax.

NameTypeValuesDescription
functionoptenum-Selected measurement function.
numeric - Returns the measured value.

:SOUR:VOLTset / query

Sets the source voltage level. The example shows the source set to 10 V with `:SOUR:VOLT 10`.

NameTypeValuesDescription
voltagerealVSource voltage level to output.
:SOUR:VOLT 10
:SOUR:VOLT:STARTset / query

Sets the sweep start voltage. The example shows it used with the volts sweep commands to define a linear staircase sweep.

NameTypeValuesDescription
voltagerealVStart voltage for the sweep.
:SOUR:VOLT:START 1 1V start voltage.
:SOUR:CURR:TRIGset / query

Sets the triggered source current level. The example shows it being used to output 100 mA when triggered.

NameTypeValuesDescription
currentrealATriggered source current level.
:SOUR:CURR:TRIG 0.1

Defines the math formula for a user-defined math function. Valid names are VOLTage, CURRent, RESistance, and TIME, and the formula may use the listed math operators.

NameTypeValuesDescription
formstring-Math formula to define for the user-defined function.
:CALCulate:MATH[EXPression] <form> Define math formula (form = formula).

Read the result of the math expression from CALC1. The command returns only the most recent math result when used as :LATest?.

returns - Returns the result of the math expression from CALC1.

Delete all user-defined math expressions in CALC1. This is the :MATH:DELete path for the CALCulate[1] subsystem.

NameTypeValuesDescription
ALLenumALLSelects deletion of all user-defined expressions.

Deletes a specified user-defined math expression from the CALC1 math subsystem.

NameTypeValuesDescription
nameoptstring-Name of the user-defined expression to delete.

Define a math expression using standard math operator symbols. This command belongs to the CALC1 MATH subsystem and creates or defines the user expression name.

NameTypeValuesDescription
formstring-Math expression to define using standard math operator symbols.

Queries the name created for the user-defined math expression.

string - Returns the created name for the user-defined math expression.

Enable or disable math expression state for CALC1.

NameTypeValuesDescription
benumOFF | ONEnable or disable the math expression.

Queries the state of math expression CALC1. Returns whether the expression is enabled or disabled.

enum - Returns the current state of the math expression.

Same as :EXPRession <form> command.

NameTypeValuesDescription
formoptstring-Math expression to define using standard math operator symbols.

Path to configure and control math expressions for CALC1. Use :EXPRession <form> to define a math expression using standard math operator symbols, or :EXPRession? to query the current expression.

Queries the math expression defined for CALC1. The expression uses standard math operator symbols.

string - The current math expression.

Queries the list of math expression names. The returned list includes built-in names and any user-defined expression names.

string - Returns a list of math expression names, including the built-in names POWER, OFFCOMPOHM, VOLTCOEF, VARALPHA, and %DEV, plus any user-defined expressions.

Selects the math expression name for CALC1. The name may be one of the built-in expressions or a user-defined name up to 10 ASCII characters.

NameTypeValuesDescription
namestring-Name of the math expression to select.

Queries the selected math expression name for CALCulate[1]. The expression can be one of the built-in names or a user-defined name up to 10 ASCII characters.

string - Returns the selected math expression name.

Deletes the specified user-defined math expression from the catalog. Once removed, the expression can no longer be selected; built-in math expressions cannot be deleted.

NameTypeValuesDescription
namestring-Name of the user-defined math expression to delete.
:DELete “math1”

Specifies the units suffix name for a user-defined math calculation. Use three ASCII characters; if fewer than three are used, pad with spaces on the right.

NameTypeValuesDescription
unitsstring-Units suffix name for the user-defined math calculation.
:calc:math:unit “Z ”
:calc:math:unit ‘Z’

Query the units suffix name for a user-defined math calculation. The units name is a three-character ASCII string; if fewer than three characters are used, pad with spaces on the right.

string - Returns the units suffix name for the selected user-defined calculation.
:calc:math:unit “Z ”
:calc:math:unit ‘Z’

Query the user-defined math expression for the selected calculation. After a math expression is defined, it is assigned to the name created with the :NAME command.

string - Returns the user-defined math expression currently selected for the calculation.

Defines a user-defined math formula for calculation 1. The formula may use instrument readings, numeric constants, and standard math operator symbols.

NameTypeValuesDescription
formstring-Mathematical formula using instrument readings, numbers, and standard math operator symbols.

Defines a math formula for a user-defined calculation. The expression can use instrument readings, numeric constants, standard math operators, and optional vector indices in brackets.

NameTypeValuesDescription
formstring-Mathematical formula using instrument readings, numbers, and standard math operator symbols.
:CALCulate[1]:MATH[:EXPression][:DEFine] <form>

Reads the result of the CALC1 calculation. For scalar math it returns results for all programmed source-measure points; for vector math it returns results only for the specified vector points.

real - The calculation result; valid results can be up to ±9.9e37. A NAN value of +9.91e37 indicates an invalid result due to an expression error, a disabled required measurement function, or CALC1 being disabled.

Enables or disables the CALC1 calculation. When enabled, the selected math expression is performed during triggered source-measure operations; when disabled, :CALC1:DATA? returns NAN.

NameTypeValuesDescription
benum0 | OFF | 1 | ONEnables or disables CALC1 calculation.

Queries the on/off state of CALC1 calculation. When enabled, the selected math expression is performed on triggered source-measure operations; when disabled, :CALC1:DATA? returns NAN.

enum - Returns the state of CALC1 calculation as on or off.

Request the raw readings stored in the buffer. The example shows it being used after a measurement sequence has been initiated.

definite_length_block - Returns the raw buffer readings.
::TRACE:DATA?
::TRACE:DATA?

Queries the buffer size.

integer - The current buffer size.

Queries the number of readings currently stored in the buffer.

integer - Number of readings stored in the buffer.

Queries the timestamp format used in the trace subsystem.

enum - Returns the timestamp format setting.
:TRACe:CLEarset / query

Clears the buffer of readings. If the buffer is not cleared, a subsequent store will overwrite the old readings.

Reads the contents of the buffer. When sent and the SourceMeter is addressed to talk, all readings stored in the data store are sent to the computer; the bus format is controlled by the :FORMat subsystem.

string - All readings stored in the data store are sent to the computer.

Reads the status of storage memory. When addressed to talk, the SourceMeter sends two comma-separated values: bytes available and bytes reserved for storing readings.

numeric - Two values separated by commas: the number of bytes of memory available and the number of bytes reserved to store readings.
:TRACe:FEEDset / query

Selects the source of readings to be placed in the trace buffer. Raw Sense readings, Calc1 results, or Calc2 readings can be stored.

NameTypeValuesDescription
nameenumSENSe[1] | CALCulate[1] | CALCulate2Source of readings to place in the buffer.

Starts or stops buffer storage by selecting how the trace buffer is controlled.

NameTypeValuesDescription
nameenumNEXT | NEVerSelects whether the buffer fills once and stops or disables buffer storage.
:TRACe:POINtsset / query

Specifies the size of the buffer. The command also supports querying the current size and the MINimum, MAXimum, and DEFault values.

NameTypeValuesDescription
ninteger1 - 2500Buffer size.
qualifieroptenumMINimum | MAXimum | DEFaultQueries the smallest allowable, largest allowable, or *RST default buffer size.

Selects the timestamp format for buffer readings. ABSolute references each timestamp to the first reading stored in the buffer, while DELTa returns the time between buffer readings.

NameTypeValuesDescription
nameenumABSolute | DELTaTimestamp format to use for buffer readings.
enum - Returns the current timestamp format.
:TRACe:TSTamp:FORMat <name>
:FORMat?

Sets the fail condition behavior for limit 1 compliance. The manual lists this command as the remote command for "Limit 1 fail conditions".

Sets the REL value used by CALC2 nulling. The table lists this command as the remote command for the REL value setting.

Enables or disables the CALC2 relative (NULL) function.

NameTypeValuesDescription
stateenumON | OFFTurns the CALC2 NULL function on or off.

Sets the memory location used for the next pass branch in Source Memory Sweep limit test branching. When NEXT is selected, the sweep proceeds to the next memory location regardless of pass or fail.

Sets the lower limit for Limit X. X is Current or Voltage, and Limit X can be Limit 2, 3, or 5-12.

NameTypeValuesDescription
dataoptreal-Lower limit value.

Enable or disable Limit X. The manual identifies Limit X as Limit 2, 3, or 5-12, with X standing for CURRent or VOLTage elsewhere in the table.

NameTypeValuesDescription
stateenumON | OFFEnables or disables Limit X.

Sets the source used for the Limit X upper bit pattern on Source Memory Sweep operations. X is CURRent or VOLTage.

NameTypeValuesDescription
source2enum-Source for the Limit X upper bit pattern.

Sets the upper limit data for Limit X. X is Current or Voltage, and Limit X refers to Limit 2, 3, 5-12.

NameTypeValuesDescription
dataoptreal-Upper limit value for Limit X.

Sets the source memory location used as the pass branch target for limit testing sweep branching. This option is available via remote for configuring sweep branching.

NameTypeValuesDescription
locationinteger1 - 100Source memory location to jump to on pass.

Specify the lower Limit X value, where X can be 2, 3, or 5-12.

NameTypeValuesDescription
nreal-The limit value to use for the lower Limit X setting.

Specifies the upper Limit X value for limit testing, where X may be 2, 3, or 5-12.

NameTypeValuesDescription
nreal-Limit value for the selected upper Limit X.

Specifies the fail source memory location. The location may be a memory number or NEXT for the next location.

NameTypeValuesDescription
locationenumNRf | NEXTFail source memory location; NRf means memory number and NEXT means next location.

Queries the result of Limit X test. For X = 2, 3, 5-12, it returns 0 for pass and 1 for fail.

enum - Limit test result: 0 = pass, 1 = fail.

Specify pass pattern for sorting mode.

NameTypeValuesDescription
NRfinteger-Bit pattern value.
NDNinteger-Bit pattern value.

Returns only the most recent CALC2 math result.

string - Returns only the most recent CALC2 math result.

Returns the result of Limit 5 through Limit 12 tests. The return is 0 for pass or 1 for fail.

integer - 0 (pass) or 1 (fail).

Specifies the output fail pattern used by Limit 5 to 12 tests in grading mode. The value is an <NRf> or <NDN> pattern, with the documented range depending on bit width.

NameTypeValuesDescription
patterninteger0 - 7Output fail pattern for grading mode in 3-bit form.
patterninteger0 - 15Output fail pattern for grading mode in 4-bit form.

Configure the lower limit for Limit 5 through Limit 12 tests. The lower limit data value may be specified with an optional :DATA node.

NameTypeValuesDescription
noptreal-9.999999e20 - 9.999999e20Specify the lower limit value.

Queries the lower limit data for Limit 5 through Limit 12 tests. Returns the present lower limit value.

NameTypeValuesDescription
qualifieroptenum-Optional data-node qualifier.
real - The lower limit value.

Queries the fail bit pattern for Limit 5 through Limit 12 pass testing. The manual says this returns the output fail pattern used for grading mode.

NameTypeValuesDescription
qualifieroptenum-Not documented for this query.
integer - Returns the fail bit pattern used for grading mode.

Enables or disables Limit 5 through Limit 12 tests.

NameTypeValuesDescription
bboolean-State of the Limit 5 to 12 tests.

Configure the output fail pattern used for grading mode for Limit 5 to 12 tests. The command takes a source pattern value and supports 3-bit or 4-bit fail patterns.

NameTypeValuesDescription
sourceinteger0 - 7Specify the output fail pattern for grading mode (3-bit).
sourceinteger0 - 15Specify the output fail pattern for grading mode (4-bit).

Queries the output fail pattern used for grading mode for Limit 5 through Limit 12 tests. The returned value is the fail bit pattern.

integer - Fail bit pattern used for grading mode.

Configure the upper limit for Limit 5 through Limit 12 tests. The value is stored in SCPI memory and defaults to 1.

NameTypeValuesDescription
nreal-9.999999e20 - 9.999999e20Upper limit value.

Queries the upper limit value for Limit 5 through Limit 12 tests. The upper limit is configured with :CALCulate2:LIMit5…12:UPPer[:DATA] <n>.

real - Returns the present upper limit value.

Enable or disable clearing of test results when :INITiate command is sent.

NameTypeValuesDescription
benumON | OFFEnable or disable auto clear of test results.

Queries the state of auto clear for CALCulate2. Auto clear controls whether test results are cleared when the INITiate command is sent.

enum - Returns the auto-clear state.

Clear the latest limit test result and reset the Digital I/O port back to :SOURce2:TTL settings.

NameTypeValuesDescription
[:IMMediate]opt--Clear the latest limit test result immediately.

Specifies the next FAIL Source Memory location for CALCulate2. The location can be set to a specific memory number or to NEXT.

NameTypeValuesDescription
locationinteger1 - 100FAIL Source Memory location or NEXT location.

Query the PASS memory location for CALCulate2 fail settings.

enum - Returns the PASS memory location setting.

Define fail digital output pattern. Used to specify the fail output pattern and the next FAIL source memory location.

NameTypeValuesDescription
<NRf>integer0 - 7Specify output fail pattern in 3-bit format.
<NDN>integer0 - 15Specify output fail pattern in 4-bit format.

Queries the fail digital output pattern for CALCulate2. The returned value is the fail bit pattern.

integer - Fail bit pattern.

Specifies the next PASS Source Memory location. The location may be set to NEXT or a numeric memory location from 1 to 100.

NameTypeValuesDescription
locationenumNEXTNext PASS Source Memory location.
locationinteger1 - 100PASS Source Memory location number.

Queries the PASS memory location for CALC2. The location can be NEXT or a numeric memory location from 1 to 100 when set via the corresponding setter.

enum - Returns the PASS memory location.

Specify the pass digital output source memory location for CALCulate2. The location can be the next location or a numbered location from 1 to 100.

NameTypeValuesDescription
source2integer1 - 100Pass Source Memory location.

Queries the pass bit pattern for CALCulate2. The returned format is set by :FORMat:SOURce2 <name>.

integer - Returns the pass bit pattern. The format of the returned value is set by :FORMat:SOURce2 <name>.

Queries the input path used for limit tests. The selected path determines whether limits compare against CALC1 results, measured voltage, current, or resistance.

enum - Returns the selected input path for limit tests.

Automatically acquires the null offset value. If no reading is available, the next available reading becomes the null offset value.

Queries the null offset value for the selected feed. When null offset is enabled, the CALC2 reading is the feed reading minus the offset value.

real - Returns the null offset value.
:CALCulate2:NULL:OFFSet?

Controls null offset for CALC2. When enabled, the CALC2 reading includes the null offset value; when disabled, it does not.

NameTypeValuesDescription
benum1 | ON | 0 | OFFEnables or disables null offset.
enum - Queries the state of null offset.

Queries when Limit 1 test failure occurs.

enum - Returns whether Limit 1 test failure occurs when the SourceMeter goes into compliance or comes out of compliance.

Queries the specified lower limit for limit test x. The query can also be qualified with DEFault, MINimum, or MAXimum to read the corresponding lower, lowest allowable, or largest allowable lower limit.

NameTypeValuesDescription
qualifieroptenumDEFault | MINimum | MAXimumOptional query qualifier selecting the default, minimum, or maximum lower limit.
real - Returns the specified lower limit value, or the *RST default, lowest allowable, or largest allowable lower limit when qualified accordingly.

Queries the specified upper limit for CALC2 limit tests. The query also accepts DEFault, MINimum, or MAXimum to query the *RST default, lowest allowable, or largest allowable upper limit.

NameTypeValuesDescription
qualifierenumDEFault | MINimum | MAXimumSelects the default, minimum, or maximum upper limit value to query.
real - Returns the specified upper limit value for the selected limit test.

Queries the programmed source value for the specified limit. The source value can be specified as binary, octal, decimal, or hexadecimal in the corresponding setter command.

numeric - Programmed source value for the specified limit.

Returns the programmed composite pass source value for the Digital I/O Port. The value may be specified as a 3-bit or 4-bit pattern in decimal, binary, octal, or hexadecimal form when set by the corresponding setter.

string - Returns the programmed source value.

Specifies the fail pattern for Limit 1 testing in sorting mode, using a 3-bit or 4-bit output pattern. The output value may be entered in decimal, binary, octal, or hexadecimal form.

NameTypeValuesDescription
patternnumeric0 - 73-bit or 4-bit fail pattern value; the manual allows decimal, binary, octal, or hexadecimal entry formats.

Specifies the fail pattern for sorting mode. For the Limit 1 test, it defines the 3-bit or 4-bit output pass pattern when Limits 2, 3, and 5-12 are disabled.

string - Returns the programmed source value.

Queries the pass or fail source memory location used during Source Memory Sweep limit testing. With PASS selected, the sweep branches to the specified memory location on success; with FAIL, it branches on failure. With NEXT selected, it proceeds to the next location in the list.

enum - Returns the current pass/fail source memory location setting.

Controls auto-clear for limit test results. When enabled, test results and Digital I/O output lines reset when :INITiate starts a new test sequence; when disabled, :IMMediate must be used to clear them.

NameTypeValuesDescription
stateenum1 | ON | 0 | OFFEnables or disables auto-clear for test results.

Queries the auto-clear state for limit test results. When enabled, test results clear and the Digital I/O port output lines reset when :INITiate starts a new test sequence; when disabled, :IMMediate must be used to clear them.

enum - Returns the state of auto-clear for test results.

Selects binning control for grading mode; immediate at failure or end of sweep.

NameTypeValuesDescription
modeenumIMMEDIATE | ENDBinning control mode for grading operation.

Selects the limit testing mode for contact check operation.

NameTypeValuesDescription
modeenumGRADING | SORTINGLimit testing mode to use.

Queries the current limit testing mode for contact check operation.

enum - Returns the selected limit testing mode.

Defines the digital output fail patterns for the Limit 4 test. The pattern may be entered as a decimal value or in binary, octal, or hexadecimal notation.

NameTypeValuesDescription
patternnumeric0 - 15Digital output fail pattern value for the Limit 4 test.

Queries the source value for the specified LIMIT 4 test. The command is part of the contact check function and relates to defining the digital output fail patterns for the LIMIT 4 test.

enum - Returns the source value for the specified limit.

:ROUTe:TERMinalsset / query

Selects the front or rear terminals for the source/measure connection path.

NameTypeValuesDescription
stateenumFRONt | REARChooses which terminals are active.

Queries the state of the selected source/measure terminal routing. The slice does not provide the returned format or values.

:TRIGger:DELayset / query

Sets the trigger delay. Not available when the Model 2430 is in Pulse Mode.

:TRIGger:COUNtset / query

Sets the trigger count. The count is the number of trigger cycles to perform.

NameTypeValuesDescription
ninteger-Trigger count to set.

Controls trigger bypass. The setting names the trigger direction as source or acceptor.

NameTypeValuesDescription
nameenumSOURce | ACCeptorTrigger direction or bypass control mode.
:TRIGger:ILINeset / query

Selects the trigger layer input line. The input line number is specified as NRf.

NameTypeValuesDescription
NRfinteger-Input line number.
:TRIGger:INPutset / query

Selects which trigger-layer events are generated at the trigger input layer. The event list can be SOURce, DELay, SENSe, or NONE.

NameTypeValuesDescription
event listenumSOURce | DELay | SENSe | NONETrigger input layer events to select.
:TRIGger:OLINeset / query

Selects the trigger layer output line.

NameTypeValuesDescription
NRfinteger-Select the trigger layer output line number.
:TRIGger:OUTPutset / query

Selects the trigger layer output events. The event list can be SOURce, DELay, SENSe, or NONE.

NameTypeValuesDescription
event listenumSOURce | DELay | SENSe | NONETrigger layer output events to enable.
:TRIGger:SOURceset / query

Specifies the trigger control source. The available sources are IMMediate or TLINk.

NameTypeValuesDescription
nameenumIMMediate | TLINkTrigger control source.

Selects the output trigger line in the Arm asynchronous trigger configuration. The selected line can be queried with the corresponding `:OLINe?` command.

NameTypeValuesDescription
ninteger1 - 4Selects the output trigger line (1, 2, 3, or 4).

Selects the control source for the Arm Layer sequence. The source can be immediate, timer, manual, bus, or one of the trigger link/test sources.

NameTypeValuesDescription
nameenumIMMediate | TIMer | MANual | BUS | TLINk | NSTest | PSTest | BSTestControl source for the Arm Layer sequence.

Queries the control source for the Arm Layer Sequence Layer 1. The control source is one of the documented trigger source keywords for that layer.

enum - Returns the control source for the Arm Layer Sequence Layer 1.

Configures whether the trigger bypass is enabled or disabled for the arm sequence. It is a setter command with the value accepting either source bypass or acceptor bypass.

NameTypeValuesDescription
nameenumSOURce | ACCeptorEnable (SOURce) or disable (ACCeptor) bypass.

Sets the timer interval in the Arm layer sequence. The interval is specified in seconds and can be used when the arm control source is TIMer.

NameTypeValuesDescription
nreal0.001 - 99999.99Timer interval in seconds.

Queries the trigger count for the Trigger Layer sequence. The value returned is the current count setting.

integer - Returns the trigger count; INFinite is represented as +9.9e37.

Queries the trigger delay in the Trigger Layer. For Model 2430 Pulse Mode, trigger delay is not used and the command is ignored.

real (s) - The source delay in seconds.

Queries the trigger layer control source for the specified sequence. The control source can be IMMediate or TLINk.

enum - Returns the trigger layer control source.

Selects the input trigger line for asynchronous output trigger configuration. The input trigger line can be one of four lines.

NameTypeValuesDescription
ninteger1 - 4Selects the input trigger line.

Configures input trigger event detectors for the asynchronous trigger subsystem. The enabled events can be SOURce, DELay, SENSe, or NONE.

NameTypeValuesDescription
event listenumSOURce | DELay | SENSe | NONEEnable the selected input event detectors.

Configure output triggers. Select when the output trigger occurs after SOURce, DELay, SENSe, or not at all (NONE).

NameTypeValuesDescription
event listenumSOURce | DELay | SENSe | NONEOutput trigger event selection.

Queries the state of bypass direction under trigger sequence configuration. The bypass is enabled with SOURce or disabled with ACCeptor.

enum - Returns the bypass state as the configured direction.
:TRIGger:CLEarset / query

Clears any pending latched input triggers immediately. The manual recommends sending it after :ABORt when using external triggering and before an initiate command at the start of a program.

Sets the trigger count for the trigger model.

NameTypeValuesDescription
ninteger1 - 2500Specifies the count.

Sets the delay used in the trigger layer. After the programmed trigger event occurs, the instrument waits until the delay expires before performing the Device Action.

NameTypeValuesDescription
nreal0 - 999.9999Delay in seconds.

Selects the trigger event control source. With IMMediate, operation continues immediately; with TLINk, operation continues when a trigger pulse is received via Trigger Link.

NameTypeValuesDescription
nameenumIMMediate | TLINk | TIMer | MANual | BUS | NSTest | PSTest | BSTestTrigger event control source.

Controls trigger bypass direction. Use SOURce to enable control source bypass or ACCeptor to disable it.

NameTypeValuesDescription
nameenumSOURce | ACCeptorSpecifies whether control source bypass is enabled or disabled.

Enables event detectors in the Trigger Layer when TLINk is the selected Trigger Layer control source. Event detectors are enabled by listing their names in the event list; disabled detectors do not hold up operation.

NameTypeValuesDescription
event listenumSOURce | DELay | SENSe | NONESpecifies which Trigger Layer event detectors to enable, or NONE to disable all event detectors.
:TRIGger:INPut SOURce, SENSe

Selects the input line for the Trigger Link on the trigger layer. For normal operation, Trigger Link input and output should not share the same line.

NameTypeValuesDescription
lineinteger1 - 4Input line number.
:TRIGger[:SEQuence[1]][:TCONfigure]:ILINe 1

Selects the Trigger Link output line for the trigger layer. For normal operation, Trigger Link input and output should not share the same line.

NameTypeValuesDescription
NRfinteger1 - 4Selects output line #1 through #4.

Specifies when trigger pulses occur on the selected output trigger line of the Trigger Link. For the trigger layer, you can select one to three events separated by commas; NONE disables the output trigger.

NameTypeValuesDescription
event listenumSOURce | DELay | SENSe | NONEEvent or events that cause output triggers for the trigger layer.
enum - Returns the selected output trigger event(s).

Programs the contact check timeout interval. When contact event detection is enabled, the SourceMeter waits this long before issuing an open lead failure and moving on to the next measurement; if set to 0, it waits indefinitely.

NameTypeValuesDescription
timeoutreal0.00000 - 999.99990Contact check timeout.
:TRIGger:SEQuence2:TOUT <NRf>
:TRIGger:SEQuence2:TOUT 0

Programs the contact check timeout interval. When contact event detection is enabled, the instrument waits this timeout before issuing an open lead failure and proceeding to the next measurement; if set to 0, it waits indefinitely.

NameTypeValuesDescription
timeoutreal0.00000 - 999.99990Contact check timeout.

:INITiateset / query

Takes the instrument out of the idle state and initiates trigger-model operation. The manual notes that it is one of the commands that perform an initiate operation.

Takes the SourceMeter out of idle state and initiates source-measure operation. :READ? and :MEASure? also perform an initiation.

:ARM:COUNtset / query

Sets the arm count. The parameter n specifies the count.

NameTypeValuesDescription
ninteger-Count value.
:ARM:DIRectionset / query

Controls arm bypass direction.

NameTypeValuesDescription
nameenumSOURce | ACCeptorSelects the arm bypass direction.
:ARM:ILINeset / query

Selects the arm layer input line.

NameTypeValuesDescription
NRfinteger-Input line number.
:ARM:OLINeset / query

Selects the arm layer output line. The parameter is the output line number.

NameTypeValuesDescription
NRfinteger-Output line number.
:ARM:OUTPutset / query

Selects the arm layer output events. The event list can be TENTer, TEXit, or NONE.

NameTypeValuesDescription
event listenumTENTer | TEXit | NONEArm layer output events to select.
:ARM:SOURceset / query

Specifies the arm control source. The available sources are IMMediate, TLINk, TIMer, MANual, BUS, NSTest, PSTest, or BSTest.

NameTypeValuesDescription
nameenumIMMediate | TLINk | TIMer | MANual | BUS | NSTest | PSTest | BSTestArm control source.
:ARM:TIMerset / query

Sets the timer interval for the :ARM subsystem. The query form returns the timer interval, and the query also accepts DEFault, MINimum, and MAXimum to return the *RST default value or the lower and upper limits.

NameTypeValuesDescription
nreal1 msec - 99999.99 secTimer interval to program.
qualifieroptenumMINimum | MAXimum | DEFaultReturns the lowest allowable value, largest allowable value, or *RST default value when used in the query form.
real (s) - Returns the timer interval. With DEFault, MINimum, or MAXimum, returns the *RST default value, the lowest allowable value, or the largest allowable value, respectively.
:ARM:TIMer 0.1
:ARM:TIMer DEFault
:ARM:TIMer MINimum
:ARM:TIMer MAXimum

Sets the arm count for the program trigger model. The count can be set from 1 to 2500, or set by the special keywords DEFault, MINimum, MAXimum, and INFinite as documented.

NameTypeValuesDescription
ninteger1 - 2500Specify the arm count.
qualifieroptenumDEFault | MINimum | MAXimum | INFiniteSets the count to 1 for DEFault or MINimum; MAXimum is noted as see NOTE; INFinite applies to ARM:COUNt only.

Specifies the arm event control source. The source determines what event causes the arm sequence to continue, such as immediate execution, Trigger Link, timer, manual trigger, bus trigger, or test pulses.

NameTypeValuesDescription
nameenumIMMediate | TLINk | TIMer | MANual | BUS | NSTest | PSTest | BSTestSelects the arm event control source.

Sets the interval for the arm layer timer. The timer is in effect only if it is the selected control source.

NameTypeValuesDescription
nreal0.001 - 9999.999Timer interval in seconds.
nreal10000.00 - 99999.99Timer interval in seconds.

Controls arm bypass direction. Use SOURce to enable control source bypass or ACCeptor to disable it.

NameTypeValuesDescription
nameenumSOURce | ACCeptorEnables or disables control source bypass.
enum - Returns the current bypass state.

Selects the Trigger Link input line for the arm layer. For normal operation, Trigger Link input and output should not share the same line.

NameTypeValuesDescription
nrfinteger1 - 4Selects the Trigger Link input line number.
integer - Returns the present input trigger line.

Selects the output line used for the Trigger Link in the arm layer. For normal operation, Trigger Link input and output should not share the same line.

NameTypeValuesDescription
nrfinteger1 - 4Output line number.

Selects the arm-layer output trigger events. The manual also lists the corresponding trigger-layer form of the command and the available event keywords.

NameTypeValuesDescription
event listenumTENTer | TEXit | NONEArm layer output trigger event selection.

Queries the actual output pattern of Digital I/O source 2. The command is not valid for the Model 2401.

string - The actual output pattern.
:SOURce2:TTLset / query

Controls the digital output line logic levels using a decimal value corresponding to the output line states.

NameTypeValuesDescription
valueinteger0 - 15Decimal value that sets the digital output line states.
:SOUR2:TTL 5

Specifies the delay for the digital output clear pass/fail pattern. The delay is given in seconds.

NameTypeValuesDescription
delayreal0.0000 - 60Specifies the pulse-width delay for the pass/fail pattern.

Queries the delay for the digital output clear function on Source 2.

real (sec) - Returns the current clear delay value.

Queries the BUSY and EOT polarity for digital output line 4. The setting is part of the :SOURce2:TTL4 digital output mode controls.

enum - Returns the BUSY and EOT polarity.

Queries the Digital I/O line 4 mode.

enum - Returns the Digital I/O line 4 mode.

Reads the actual digital output pattern for Source 2 TTL. The output pattern format is set by `FORMat:SOURce2 <name>`.

string - The actual output pattern as a digital output pattern value.

Specify the digital output pattern for Source 2 TTL output. The default output pattern may also be queried, and the actual output pattern can be read with :ACTual?.

NameTypeValuesDescription
<NRf>, <NDN>numeric-Digital output pattern values.

Specifies the digital output pattern for Source 2. The default output pattern can also be queried.

NameTypeValuesDescription
leveloptintegerNRf - NDNDefault output pattern value.
integer - Returns the default output pattern.

Sets the logic levels of the digital output lines for the Digital I/O port. When set high, the specified output line is approximately +5 V; when set low, it is 0 V.

NameTypeValuesDescription
leveloptreal0 - 7Specifies the digital output value to set.

Sets the polarity of the BUSY and EOT signal in 3-bit mode. The command controls whether the signal is active high or active low.

NameTypeValuesDescription
binteger0 - 1Sets the EOT/BUSY polarity to high or low.

Controls Digital I/O port line 4 mode. Line 4 can be used as an End-of-Test or Busy signal in 3-bit output mode.

NameTypeValuesDescription
nameenumEOTest | BUSYSelects whether line 4 is used as EOT or BUSY.

Sets the delay for digital output auto-clear. After the delay times out, the output returns to the pattern programmed by the :TTL:LEVel command.

NameTypeValuesDescription
nreal0 - 60Specify the auto-clear delay in seconds.
qualifieroptenumDEFault | MINimum | MAXimumAlternative query qualifiers for the delay value.
:SOURce2:BSIZeset / query

Sets the digital I/O port bit size for source 2. The manual states the bit size can be 3 or 4.

NameTypeValuesDescription
bit sizeinteger3 - 4Digital I/O port bit size.

Queries the bit size of digital I/O port 2 used for contact check operation.

integer - Returns the digital I/O port bit size.

:CALC2:LIM2:LOWset / query

Sets the lower value for Limit 2 in the limits test example. It is used with :CALC2:LIM2:UPP to define the pass/fail window.

NameTypeValuesDescription
valuerealVLower limit value for Limit 2.
:CALC2:LIM2:UPPset / query

Sets the upper limit value for Limit 2 in the limits test. The example shows it being assigned 0.85 V.

NameTypeValuesDescription
valuerealVUpper limit value for Limit 2.
:CALC2:LIM2:UPP 0.85
:CALC2:CLLM:BCONset / query

Sets the digital I/O pattern used for the contact check limit test. The default setting is 0-7 for 3-bit and alternately 0-15 for 4-bit.

NameTypeValuesDescription
patterninteger0 - 15Contact check digital I/O pattern.
:CALC2:CLIM:BCON IMM
:CALC2:LIMit4set / query

Configure the contact check limit test (Limit 4) in the hardware limits menu. The slice indicates this limit can be enabled or disabled and used with a contact check digital I/O pattern.

:OUTPutset / query

Sets or queries the output state. The command is used as the root for output control commands such as :OUTPut ON.

NameTypeValuesDescription
stateenumON | OFFTurns the output on or off.
:outp on <PMT>

Enables or disables the output enable for output 1. The optional state query returns the state of output enable.

NameTypeValuesDescription
stateoptboolean-Enable or disable output enable.
boolean - Returns the state of output enable.

Queries the present state of source output. Returns whether the source is on or off.

enum - Present state of source output.

Queries the state of the source output. Measurements cannot be made while the source is off.

enum - Returns the source output state.

Turns the source output on or off. Measurements cannot be made while the source is off, which places the SourceMeter in the idle state except when source auto clear is enabled.

NameTypeValuesDescription
benum0 | OFF | 1 | ONTurns the source off (standby) or on (operate).

This query command is used to determine whether the output enable has been tripped. The tripped condition ("1") means the source can be turned on; "0" means it cannot be turned on because the output enable line is at logic high.

enum - Returns whether the output enable line has tripped: "1" means the source can be turned on (logic low), and "0" means the source cannot be turned on (logic high).

Controls the output enable line. When enabled, the source cannot be turned on unless the output enable line is pulled low; when the line goes high, the source turns off.

NameTypeValuesDescription
benum0 | OFF | 1 | ONEnable or disable output enable.

Controls the output enable line state. When enabled, the source cannot be turned on unless the output enable line is pulled low; when the line goes high, the source turns off.

NameTypeValuesDescription
stateenum0 | OFF | 1 | ONSelects whether output enable is disabled or enabled.
enum - Returns the state of the output enable line.
:OUTPut[1]:SMODeset / query

Selects the output-off state of the SourceMeter. The HIMPedance setting opens the output relay when the source is turned off, disconnecting external circuitry from the SourceMeter input/output.

NameTypeValuesDescription
nameenumHIMPedance | NORMal | ZERO | GUARdSelects the output-off mode.

Selects the output-off mode of the SourceMeter. The manual says the available modes are HIMPedance, NORMal, ZERO, and GUARd, and describes how each mode behaves when the output is turned off.

NameTypeValuesDescription
nameenumHIMPedance | NORMal | ZERO | GUARdSelects the output-off state mode.
enum - Returns the current output-off mode.
:OUTput:STATeset / query

Turns the output on or off. The manual notes that output must be on before readings can be taken.

NameTypeValuesDescription
stateenumON | OFFOutput state.

*CLSset / query

Resets the event registers to 0 and clears all messages from the Error Queue. The Standard Event Enable Register is not reset by *CLS.

*STB?query

Reads the Status Byte Register. When used with *STB?, bit B6 is the Master Summary Status (MSS) bit.

integer - Returns the status byte register value; bit B6 is MSS when read with *STB?.
*SREset / query

Sets the Service Request Enable register. The register controls whether status summary bits can set bit B6 (RQS/MSS); sending 0 clears the register.

NameTypeValuesDescription
valueinteger0 - 255Service Request Enable register value.
*SRE 0
*SRE?query

Reads the Service Request Enable Register. The register is programmed with *SRE and clears when power is cycled or when *SRE 0 is sent.

integer - Returns the current Service Request Enable Register value.
*ESEset / query

Programs the Standard Event Enable Register.

NameTypeValuesDescription
<NRf>integer-Numeric value written to the Standard Event Enable Register.
*ESE?query

Reads the Standard Event Enable Register.

integer - Returns the value of the Standard Event Enable Register.
*IDN?query

Returns the manufacturer, model number, serial number, and firmware revision levels of the unit.

string - Returns the manufacturer, model number, serial number, and firmware revision levels of the unit.
*SAVset / query

Saves the present setup in nonvolatile memory. The stored setup can later be recalled with *RCL <NRf>.

NameTypeValuesDescription
NRfinteger0 - 4Memory location to save the setup in.
*TST?query

Performs a checksum self-test on ROM and places the coded result in the output queue. A returned value of 0 indicates the test passed; 1 indicates the test failed.

integer - Coded self-test result from the output queue: 0 means passed, 1 means failed.
*ESR?query

Reads the Standard Event Status Register.

integer - Returns the contents of the Standard Event Status Register.
*OPT?query

Queries installed options. The response message indicates the presence or absence of options, such as no options present or a contact check board present.

enum - Returns the installed option status as a keyword response.
*OPCset / query

Sets or queries the IEEE-488.1 operation complete status. It is still functional in 488.1 protocol mode, but is not needed there.

*OPC?query

Sets the operation complete query for GPIB 488.1 protocol use. In this manual slice it is mentioned as functional but not needed for the 488.1 protocol.

boolean - Returns 1 when all pending operations are complete and 0 otherwise.
*RCLset / query

Recalls a stored instrument state. It is listed as an initiate command in the 488.1 protocol notes and releases bus hold-off immediately during data acquisition.

*RSTset / query

Resets the instrument. In 488.1 protocol mode, it is one of the commands that immediately releases bus hold-off during data acquisition.

*TRGset / query

Triggers the interface bus; the manual notes it as an initiate command that releases bus hold-off immediately during data acquisition. It is supported during data acquisition along with GET, SDC, IFC, *RCL, *RST, SYSTem:PRESet, and ABORt.

*WAIset / query

Waits until all previously received commands have finished executing. In this protocol it is still functional but not needed because the bus is automatically held off until command execution completes.

Returns the active function(s).

enum - The active measurement function(s).

Selects the function to be configured for subsequent measurements or source-measure operations.

NameTypeValuesDescription
functionenum-Measurement or source-measure function to configure.

Specifies the data elements included in formatted data output for the Sense path. The optional ":SENSe[1]" node selects the Sense subsystem.

NameTypeValuesDescription
item listenumVOLTage | CURRent | RESistance | TIME | STATusData elements to include in the formatted output.
sense[1]opt--Optional Sense path selector.

Queries the selected data elements for format output. The optional :SENSe[1] node scopes the element list to sense data.

NameTypeValuesDescription
item listenumVOLTage | CURRent | RESistance | TIME | STATusSelects which data elements are included in the format output.
sense[1]opt--Optional sense path modifier for the elements query.
enum - Returns the selected data format elements.
:FORMat[:DATA]set / query

Selects the data format for transferring instrument readings over the bus. For the REAL format, <length> is optional and defaults to 32 if omitted; ASCII is the only format allowed over RS-232.

NameTypeValuesDescription
typeenumASCii | REAL,32 | SREalData format to use for transferred readings.
lengthoptinteger-Optional length for the REAL parameter; not used for ASCii or SREal.

Queries the current data format used for transferring instrument readings over the bus. The command affects output formatting for READ?, FETCh?, MEASure?, TRACe:DATA?, CALC1:DATA? and CALC2:DATA? over GPIB; ASCII is the only format allowed over RS-232.

enum - Returns the current data format setting.
:FORMat:ELEMentsset / query

Specifies which data elements are included in the data string. The selected elements are used in responses to :FETCh?, :READ?, :MEASure?, and :TRACe:DATA?.

NameTypeValuesDescription
item listenumVOLTage | CURRent | RESistance | TIME | STATusData elements to include in the data string.

Specifies which data elements are included in the data string returned by measurement and fetch queries. Up to five elements can be selected, separated by commas.

NameTypeValuesDescription
item listenumVOLTage | CURRent | RESistance | TIME | STATusData elements to include in the data string.
enum - Returns the currently selected data elements in the data string format.

Sets the CALC data element list used for CALC1:DATA? and CALC2:DATA? responses, and when TRACe:FEED is set to CALC1 or CALC2. The list can include calculation data, timestamp, and status information.

NameTypeValuesDescription
item listenumCALC | TIME | STATusSelects which CALC data elements to include in the response list.
enum - Returns the CALC data element list.
:FORMat:SOURce2set / query

Sets the response format for SOUR2 and TTL queries. The format applies to all CALC2:XXXX:SOUR2 and SOUR2:TTL queries in a manner similar to FORM:SREG.

NameTypeValuesDescription
nameenumASCii | HEXadecimal | OCTal | BINaryResponse format to use.

Sets the response format for all CALC2:XXXX:SOUR2 and SOUR2:TTL queries. The format can be ASCII, hexadecimal, octal, or binary.

NameTypeValuesDescription
nameenumASCii | HEXadecimal | OCTal | BINaryResponse format name.
enum - Returns the current response format.
:FORMat:SOURce2 <name>
:FORMat:BORDerset / query

Specifies the byte order used for IEEE-754 binary formats. The ASCII data format is always sent in normal byte order, and the SWAPped selection is ignored when ASCII is selected.

NameTypeValuesDescription
nameenumNORMal | SWAPpedSelects normal or reverse byte order for binary formats.

Queries the current binary byte order used for IEEE-754 binary formats. The command controls whether binary data is sent in normal or reversed byte order; ASCII data always uses normal order.

enum - Returns the current byte order setting.

Sets the response message format for reading status registers.

NameTypeValuesDescription
nameenumASCii | Hexadecimal | OCTal | BINarySelects the data format used when reading status registers.

Sets the response message format used when reading status event registers. The query form returns the current format.

NameTypeValuesDescription
nameenumASCii | Hexadecimal | OCTal | BINarySets the format for reading status registers.
enum - Returns the format for reading status registers.

Specifies the status data element for contact check. Bit 18 (262144) of the status word returned by various read, fetch, measure, and data commands signifies contact check status; if LIMIT 4 is enabled, the limit pass/fail status is also returned.

NameTypeValuesDescription
elementenumSTATusSelects the status data element.

Reads the measurement enable register. The response format depends on the currently selected status register format (ASCII, hexadecimal, octal, or binary).

numeric - The current measurement enable register value.

Reads the operation status enable register. The response format depends on the currently selected status register format.

numeric - The current operation status enable register value, returned in the selected status register format.

Reads the enable register for the questionable status registers.

numeric - Returns the enable register value in the currently selected register format (ASCII, hexadecimal, octal, or binary).

Reads the disabled messages in the status error queue. This command is part of the :STATus:QUEue subsystem.

string - Returns the disabled messages.

Read the enabled messages in the error queue.

string - Returns the enabled error/status messages.

Reads the contents of the Measurement condition register and returns a value indicating which bits are set.

integer - Returns the contents of the Measurement condition register as a bit value indicating which bits are set.

Programs the contents of the Measurement Event Enable Register.

NameTypeValuesDescription
NDN or NRfnumeric-Value written to the measurement event enable register.

Reads the contents of the Measurement Event Register. After the command is sent and the instrument is addressed to talk, it returns a value indicating which bits in the register are set.

integer - Value indicating which bits in the Measurement Event Register are set.

Reads the contents of the operation condition register.

integer - Returns a value indicating which bits in the Operation Condition register are set.

Programs the contents of the Operation event enable register.

NameTypeValuesDescription
NDNinteger0 - 65535Enable register value.

Reads the contents of the Operation event register and returns a value indicating which bits are set. After sending the command and addressing the SourceMeter to talk, the register contents are sent to the computer.

integer - Value indicating which bits in the Operation event register are set.

Reads the contents of the Questionable condition register. A value is sent to the controller indicating which bits in the register are set.

integer - The current contents of the Questionable condition register.

Programs the contents of the Questionable event enable register.

NameTypeValuesDescription
NDN or NRfnumeric-Value written to the Questionable event enable register.

Reads the contents of the Questionable Event Register. After the command is sent and the instrument is addressed to talk, a value is returned indicating which bits in the register are set.

integer - Value indicating which bits in the Questionable Event Register are set.
:STATus:PRESetset / query

Returns the event registers to default conditions. The Operation Event Enable Register, Event Enable Register, and Measurement Event Enable Register are cleared to zero; the Standard Event Register is not affected.

Specifies which error messages are enabled for entry into the Error Queue. Messages not specified are disabled and prevented from entering the queue.

NameTypeValuesDescription
listchannel_list-Specified list of messages to enable for the Error Queue.

Reads the next message from the Error Queue. It performs the same function as :SYSTem:ERRor? and is used to retrieve error and status messages as they occur.

string - Returns the next error or status message from the Error Queue.

Disables specified messages for the Error Queue. Disabled messages are prevented from going into the Error Queue.

NameTypeValuesDescription
listnumeric-Specified list of messages to disable for the Error Queue.

Specifies the CALC3 statistical format used for readings stored in the buffer. The selected statistic is used when CALC3 data is acquired.

NameTypeValuesDescription
nameenumMEAN | SDEViation | MAXimum | MINimum | PKPKStatistic to use for CALC3 buffer data.
:CALCulate3:FORMat <name>

Queries the programmed CALC3 math format. The selected statistic is used for readings stored in the buffer.

enum - Returns the programmed math format name.
:CALCulate3:FORMat?

:DISPlay:DIGitsset / query

Sets the display resolution. Rational numbers may also be used; the instrument rounds them to an integer.

NameTypeValuesDescription
ninteger4 - 7Display resolution setting: 4=3.5 digits, 5=4.5 digits, 6=5.5 digits, 7=6.5 digits.
qualifieroptenumDEFault | MINimum | MAXimumSelects the default, lowest, or largest allowable display resolution.

Sets or queries the display resolution for the SourceMeter display. Rational numbers may also be used for the setter; the instrument rounds them to an integer.

NameTypeValuesDescription
ninteger4 - 7Display resolution setting: 4=3.5 digit, 5=4.5 digit, 6=5.5 digit, 7=6.5 digit.
qualifieroptenumDEFault | MINimum | MAXimumOptional keyword form for the setter value.
integer - Returns the current display resolution; the query can also request the default, minimum, or maximum allowable resolution using a qualifier.
:DISPlay:DIGits <n>
:DISPlay:DIGits?

Returns the instrument to the source-measure display state, showing source, measure, and compliance readings. If a menu is currently displayed, this command exits the menu and restores the normal display.

:DISPlay:ENABleset / query

Enables or disables the front panel display circuitry. When disabled, the instrument operates at a higher speed and the display is frozen until re-enabled or the instrument is put into local.

NameTypeValuesDescription
benum0 | OFF | 1 | ONSelects whether the display circuitry is disabled or enabled.

Query state of the display circuitry. The command reports whether the front-panel display is enabled or disabled.

enum - Returns the display state.

Queries which characters on the specified display are blinking. The response message returns the blink status of each character position for the top display (20 characters) or bottom display (32 characters).

string - A string of 0s and 1s indicating the blink status of each character position for the specified display; 1 means blinking and 0 means not blinking.

Reads what is currently being displayed on the selected display window. After the command is sent and the SourceMeter is addressed to talk, the displayed data (message or reading) is returned to the computer.

returns - The displayed data currently shown on the selected display window (message or reading).

Defines a text message for the display. The message can be up to 20 characters for the top display and up to 32 characters for the bottom display; excess characters cause an error.

NameTypeValuesDescription
astring-ASCII characters for the message.
:DISPlay[:WINDow[1]]:TEXT:DATA <a>

Reads the text currently defined for the display message on the specified display window. After the command is sent and the instrument is addressed to talk, the message is returned to the computer.

string - The currently defined display text message.

Controls whether the text message mode is enabled for the specified display. The command is documented for the top display with the optional window node; the same function is shown for the bottom display on Window2.

NameTypeValuesDescription
stateenum0 | OFF | 1 | ONEnables or disables the text message for the specified display.

Queries the message mode state for the specified display. The query returns whether the text message is enabled or disabled.

enum - Returns the state of message mode for the specified display.

:STATe?query

Queries the state of the digital averaging filter. When enabled, voltage, current, and resistance readings are filtered according to the filter configuration.

enum - Returns the digital filter state.

:AUTO?query

Queries the state of auto output-off for the source. Auto output-off controls whether the source output turns on at the beginning of each SDM cycle and turns off after each measurement, or remains on until explicitly turned off.

enum - Returns the state of auto output-off.

:MODE?query

Queries the auto output-off mode for the source. For this command, the source will turn off after every SDM cycle with ALWAYS, or after the trigger count has expired with TCOunt.

enum - Returns the selected auto output-off mode.

:SHAPe?query

Selects the output mode for the Model 2430. DC selects DC Mode and PULSE selects Pulse Mode.

enum - Returns the selected output mode.

Queries the selected source mode. The command is documented alongside the source-mode setting command, but this query returns the current source mode.

enum - Returns the selected source mode.

Queries the triggered amplitude for the fixed source. It can also return the *RST default, minimum, or maximum allowable amplitude when used with the DEFault, MINimum, or MAXimum qualifier.

NameTypeValuesDescription
qualifieroptenumDEFault | MINimum | MAXimumSelects which amplitude value to query.
real (A) - Returns the triggered amplitude for the fixed source, or the *RST default, minimum, or maximum allowable amplitude when queried with DEFault, MINimum, or MAXimum.

Queries the V-Source over-voltage protection limit level. The query can also request the default, minimum, or maximum allowable limit using a qualifier.

NameTypeValuesDescription
qualifieroptenumDEFault | MINimum | MAXimumSelects the default limit, the lowest allowable limit, or the highest allowable limit to query.
enum (V) - Returns the selected limit level or the requested default, minimum, or maximum limit value.

:DELay?query

Returns the currently programmed source delay. The query also accepts the qualifiers DEFault, MINimum, and MAXimum to return the *RST default, lowest allowable, or highest allowable delay.

NameTypeValuesDescription
qualifieroptenumDEFault | MINimum | MAXimumSelects the default, minimum, or maximum delay value to query.
real (s) - Returns the delay in seconds.
:DELay?

Queries the source ranging mode used for sweeps.

enum - Returns the current source ranging mode.

Selects the scale for a sweep. LINear performs the sweep on a linear scale, and LOGarithmic performs it on a logarithmic scale.

enum - Returns the present sweep scale.

:STARt?query

Queries the start level for a sweep. Special query forms allow retrieving the default, minimum, or maximum allowable start level.

NameTypeValuesDescription
qualifieroptenumDEFault | MINimum | MAXimumSelects the default, minimum, or maximum start level.
real (A) - Returns the sweep start level, or the default, minimum, or maximum allowable start level when queried with DEFault, MINimum, or MAXimum.

:STOP?query

Queries the stop level for a sweep. The query also supports the qualifiers DEFault, MINimum, and MAXimum to return the *RST default, lowest allowable, or highest allowable level.

NameTypeValuesDescription
qualifieroptenumDEFault | MINimum | MAXimumSelects the default, minimum, or maximum stop level to query.
real (A or V) - Returns the stop level for the sweep, or the *RST default, lowest allowable, or highest allowable level when queried with DEFault, MINimum, or MAXimum respectively.
:STOP?

Queries the sweep center point. Optional qualifiers return the default, minimum, or maximum allowable level for the sweep center.

NameTypeValuesDescription
qualifieroptenumDEFault | MINimum | MAXimumSelects which center-point value to query.
real - Returns the sweep center point, or the *RST default level, lowest allowable level, or highest allowable level when queried with DEFault, MINimum, or MAXimum respectively.

:SPAN?query

Queries the sweep span. The command also supports the keywords DEFault, MINimum, and MAXimum to query the *RST default, lowest allowable, or highest allowable span.

NameTypeValuesDescription
qualifieroptenumDEFault | MINimum | MAXimumSelects the default span, the minimum allowable span, or the maximum allowable span.
real - Returns the span for sweep, or the default, minimum, or maximum allowable span when queried with DEFault, MINimum, or MAXimum respectively.

:STEPset / query

Specifies the step size for a linear sweep. The source level changes in equal steps from the start level to the stop level, with a measurement performed at each source step including the start and stop levels.

NameTypeValuesDescription
qualifieroptenumMINimum | MAXimum | DEFaultSelects the lowest allowable level, highest allowable level, or the *RST default level when used in the query form.
real - Returns the step size for the sweep, or the *RST default level, the lowest allowable level, or the highest allowable level when queried with the corresponding qualifier.
:STEP? DEFault
:STEP? MINimum
:STEP? MAXimum
:STEP?query

Queries the step size for a linear sweep. Optional qualifiers return the *RST default level, the lowest allowable level, or the highest allowable level.

NameTypeValuesDescription
qualifieroptenumDEFault | MINimum | MAXimumSelects the default, minimum, or maximum step size value to query.
real - Returns the step size for a linear sweep, or the requested default/minimum/maximum level when a qualifier is used.

Queries the number of source-measure points in a sweep. Optional qualifiers return the default, minimum, or maximum allowable number of sweep points.

NameTypeValuesDescription
qualifieroptenumDEFault | MINimum | MAXimumSelects the default, minimum, or maximum number of sweep points to query.
integer - Returns the requested sweep-point count; with DEFault returns the *RST default, with MINimum returns the lowest allowable value, and with MAXimum returns the highest allowable value.
:POINts?
:POINts? DEFault
:POINts? MINimum
:POINts? MAXimum

Controls whether a sweep aborts when compliance is detected. Three modes are available: NEVer, EARLy, and LATE.

enum - Returns the current abort-on-compliance mode.

Queries the current direction of the sweep. Use the corresponding setter to run the sweep from start to stop or from stop to start.

enum - Returns the present sweep direction.

Queries the I-Source list used for list sourcing mode. The instrument sequentially sources each current or voltage value in the list when the list is executed.

block - Returns the present I-Source list.

Returns the present V-Source list.

string - Returns the V-Source list.
:VOLTage?

Queries the current scaling factor for the triggered source level. The manual also shows related setter forms for current and voltage scaling factors.

real - Returns the scaling factor.

:ABORtset / query

Aborts operation and returns the SourceMeter to the idle state.

Reads the Measurement Event Register.

integer - Returns the Measurement Event Register value.

Reads the Operation Event Register.

integer - Returns the contents of the Operation Event Register.

Reads the Questionable Event Register.

integer - Returns the Questionable Event Register value.

Reads the Error Queue. The manual notes that error and status conditions are stored in the queue and that each message is preceded by a code number.

string - Error queue message(s), each preceded by a code number. Negative numbers indicate SCPI-defined messages and positive numbers indicate Keithley-defined messages.

Returns the present function used for sensing measurements.

enum - Returns the current sense function setting.

:FETCh?query

Reads data stored in the Sample Buffer. It returns all stored data arrays and does not affect the buffer contents, so subsequent executions return the same data until overwritten.

returns - Returns the data stored in the Sample Buffer.

:READ?query

Reads data from the Sample Buffer. It performs an INITiate followed by FETCh? to acquire the new data.

returns - Performs an INITiate and then a FETCh?; no separate return format is documented in this slice.

:GETset / query

Initiates a bus-level GET action in the 488.1 protocol. The manual notes that GET is supported during data acquisition and immediately releases bus hold-off.

:IFCset / query

Controls NDAC hold-off in the GPIB 488.1 protocol mode. The default is OFF, but NRFD hold-off remains enabled.

NameTypeValuesDescription
stateenumON | OFFEnables or disables NDAC hold-off.

:SDCset / query

Immediate bus-holdoff release during data acquisition; listed as an exception to the normal hold-off behavior for 488.1 protocol. Used to support bus-control actions during acquisition.

:GTLset / query

GTL is an IEEE-488 bus command that attempts to return the SourceMeter to local mode. If the unit is in remote, it may not put the instrument into local mode, but it will still disable local lockout (LLO).

:LLOset / query

Places the instrument in local lockout so the front panel LOCAL key is disabled from returning control to the local interface. This is a common command.

Returns the latest reading in the selected sense function. This is a signal-oriented measurement query under :SENSe[1]:DATA.

real - Returns the most recent measured value.