Sets the N value in N dB BW measurement and queries the current N value. The command is invalid for the PvT measurement mode in RTSA.
| Name | Type | Values | Description |
|---|
| <rel_ampl> | real | -140 dB - -0.01 dB | Consecutive real number used as the relative amplitude for the N dB bandwidth measurement. |
real - The query returns the N value in scientific notation.
:CALCulate:BANDwidth:NDB -4
:CALCulate:BWIDth:NDB -4
:CALCulate:BANDwidth:NDB?
:CALCulate:BWIDth:NDB?
Queries the measurement results of the N dB band. The returned bandwidth value is in Hz and is reported in scientific notation; if the two points at both sides of the current marker are not found, the query returns -1.000000000e+02.
real (Hz) - Bandwidth value of the N dB band in scientific notation. If the two points at both sides of the current marker with N dB fall or rise in peak amplitude are not found, returns -1.000000000e+02.
:CALCulate:BANDwidth|BWIDth:RESult?
Queries the frequency value of the frequency point located to the left of the current marker at the N dB fall or rise in peak amplitude. Invalid for PvT measurement mode in RTSA.
real (Hz) - Returns the frequency value in scientific notation, in Hz. If the frequency point is not found, returns -1.000000000e+02.
Queries the frequency value of the frequency point that is located at the right side of the current marker with N dB fall or rise in its peak amplitude. If the frequency point is not found, the query returns -1.000000000e+02.
real (Hz) - The frequency value in scientific notation, in Hz. If the point is not found, the query returns -1.000000000e+02.
Deletes all the limit line data.
Enables or disables the limit line test function. The query returns the setting state as 0 or 1.
| Name | Type | Values | Description |
|---|
| state | enum | OFF | ON | 0 | 1 | Enables or disables the limit line test function. |
integer - The query returns 0 or 1.
:CALCulate:LLINe:TEST OFF or :CALCulate:LLINe:TEST 0
:CALCulate:LLINe:TEST?
Enables or disables the coupling between the selected limit line data points and the reference level. The query returns the setting state as 0 or 1.
| Name | Type | Values | Description |
|---|
| state | enum | OFF | ON | 0 | 1 | Enables or disables the coupling function. |
enum - Returns the setting state of the coupling function as 0 or 1.
:CALCulate:LLINe<n>:AMPLitude:CMODe:RELative OFF|ON|0|1
:CALCulate:LLINe<n>:AMPLitude:CMODe:RELative?
Builds the selected limit line from the specified trace.
| Name | Type | Values | Description |
|---|
| nopt | integer | 1 - 6 | Selects the limit line number. |
| trace | enum | TRACE1 | TRACE2 | TRACE3 | Selects the trace used to build the limit line. |
:CALCulate:LLINe2:BUILd TRACE2
Copies the selected limit line to the current limit line.
| Name | Type | Values | Description |
|---|
| n | integer | 1 - 6 | Limit line number to copy to. |
| source | enum | LLINE1 | LLINE2 | LLINE3 | LLINE4 | LLINE5 | LLINE6 | Limit line to copy from. |
:CALCulate:LLINe2:COPY LLINE1
Edits one limit line and marks it with n. The query returns the limit line data currently being edited.
| Name | Type | Values | Description |
|---|
| n | integer | 1 - 6 | Limit line number to edit. |
| x | real | 0 Hz - 3 GHz | Frequency value for a data point. |
| ampl | real | -1,000 dBm - 1,000 dBm | Amplitude value for a data point. |
| connect | integer | 0 - 1 | Connection flag for the data point. |
string - The query returns the limit line data currently being edited.
:CALCulate:LLINe1:CONTrol:INTerpolate:TYPE LINear
:CALCulate:LLINe1:CONTrol:INTerpolate:TYPE?
Deletes the selected limit line.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | Selects the limit line to delete. |
::CALCulate:LLINe2:DELete
Turns on or off the selected limit line. Queries the setting state of the selected limit line.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | Selects the limit line number. |
| state | enum | OFF | ON | 0 | 1 | Turns the selected limit line off or on. |
integer - Returns 0 or 1 indicating the setting state of the selected limit line.
:CALCulate:LLINe2:DISPlay OFF|ON|0|1
:CALCulate:LLINe2:DISPlay?
Queries the measurement results of the selected limit line and its associated trace.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 6 | |
enum - Returns 0 for pass or 1 for fail.
Enables or disables the coupling function between the selected limit line data points and the center frequency. The query returns the current coupling state.
| Name | Type | Values | Description |
|---|
| n | integer | 1 - 6 | Selects the limit line number. |
| state | enum | OFF | ON | 0 | 1 | Enables or disables the coupling function. |
enum - Returns the current coupling setting: OFF/0 for disabled or ON/1 for enabled.
Sets the margin for the selected limit line. Queries the margin for the selected limit line.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 6 | Selects the limit line number. |
| rel_ampl | real | -40 dB - 0 dB | Sets the margin value for the selected limit line. |
real - The query returns the margin for the limit line in scientific notation.
:CALCulate:LLINe2:MARGin 1
:CALCulate:LLINe2:MARGin?
Enables or disables the margin for the selected limit line. Queries the setting state of the margin for the selected limit line.
| Name | Type | Values | Description |
|---|
| <n> | integer | 1 - 6 | Selects the limit line number. |
| state | enum | OFF | ON | 0 | 1 | Enables or disables the margin. |
enum - Returns the margin state as 0 or 1.
Sets the trace for the current limit line test. Queries the trace for the current limit line test.
| Name | Type | Values | Description |
|---|
| n | integer | 1 - 6 | Limit line number. |
| trace | integer | 1 - 3 | Trace number used for the current limit line test. |
integer - Returns the trace number for the current limit line test.
:CALCulate:LLINe3:TRACe 2
:CALCulate:LLINe3:TRACe?
Sets the type of the specified limit line. Queries the type of the specified limit line.
| Name | Type | Values | Description |
|---|
| <n> | integer | 1 - 6 | |
| type | enum | UPPer | LOWer | Selects whether the limit line is upper or lower type. |
enum - The query returns UPP or LOW.
Turns off all the enabled markers.
Enables or disables the couple marker function. When enabled, moving any marker makes other non-fixed, non-off markers move along with it.
| Name | Type | Values | Description |
|---|
| stateopt | enum | OFF | ON | 0 | 1 | Couple marker function state. |
:CALCulate:MARKer:COUPle:STATe OFF or :CALCulate:MARKer:COUPle:STATe 0
:CALCulate:MARKer:COUPle:STATe?
Sets the peak offset in dB. Queries the peak offset.
| Name | Type | Values | Description |
|---|
| rel_ampl | real | dB | Peak offset value. |
Enables or disables the peak offset function. Queries the status of the peak offset function.
| Name | Type | Values | Description |
|---|
| state | enum | OFF | ON | 0 | 1 | Enables or disables the peak offset function. |
enum - The query returns 0 or 1.
:CALCulate:MARKer:PEAK:EXCursion:STATe ON or :CALCulate:MARKer:PEAK:EXCursion:STATe 1
:CALCulate:MARKer:PEAK:EXCursion:STATe?
Sets the peak search mode. Queries the peak search mode.
| Name | Type | Values | Description |
|---|
| mode | enum | PARameter | MAXimum | Peak search mode. |
Sets or queries the sorting order of the data displayed in the peak table.
| Name | Type | Values | Description |
|---|
| mode | enum | FREQuency | AMPLitude | Sort order for the peak table data. |
enum - Returns the current peak table sorting order.
:CALCulate:MARKer:PEAK:SORT AMPLitude
:CALCulate:MARKer:PEAK:SORT?
Queries or sets the peak-table readout mode used to list peaks meeting the peak criteria. The available modes are ALL, GTDLine, and LTDLine.
| Name | Type | Values | Description |
|---|
| mode | enum | ALL | GTDLine | LTDLine | Selects which peaks are listed in the peak table readout. |
enum - Returns the current readout mode as ALL (NORM), GTDL (DLM), or LTDL (DLL).
:CALCulate:MARKer:PEAK:TABLe:READout GTDLine
:CALCulate:MARKer:PEAK:TABLe:READout?
Enables or disables the peak table. Queries the status of the peak table.
| Name | Type | Values | Description |
|---|
| state | enum | OFF | ON | 0 | 1 | Turns the peak table off or on. |
enum - Returns the peak table status.
:CALCulate:MARKer:PEAK:TABLe:STATe OFF|ON|0|1
:CALCulate:MARKer:PEAK:TABLe:STATe?
Sets the peak threshold, with a default unit of dBm. Queries the peak threshold.
| Name | Type | Values | Description |
|---|
| ampl | real | -200 dBm - 0 dBm | Peak threshold amplitude. |
real - The query returns the peak threshold in scientific notation.
:CALCulate:MARKer:PEAK:THReshold -100
:CALCulate:MARKer:PEAK:THReshold?
Enables or disables the peak threshold function.
| Name | Type | Values | Description |
|---|
| state | enum | OFF | ON | 0 | 1 | Enables or disables the peak threshold function. |
Enables or disables the marker table. Queries the status of the marker table.
| Name | Type | Values | Description |
|---|
| stateopt | enum | OFF | ON | 0 | 1 | Enables or disables the marker table. |
enum - The query returns 0 or 1.
Enables or disables the signal track. Queries the status of the signal track.
| Name | Type | Values | Description |
|---|
| stateopt | boolean | OFF | ON | 0 | 1 | Boolean state to enable or disable signal track. |
integer - The query returns 0 or 1.
:CALCulate:MARKer:TRCKing:STATe OFF or :CALCulate:MARKer:TRCKing:STATe 0
:CALCulate:MARKer:TRCKing:STATe?
Enables or disables continuous peak search function, and marks the peak value with n. Queries the status of the continuous peak search function.
| Name | Type | Values | Description |
|---|
| <n> | discrete | 1 - 8 | Marker number. |
| state | enum | OFF | ON | 0 | 1 | Turns continuous peak search on or off. |
enum - Returns the status of continuous peak search.
:CALCulate:MARKer1:CPSeatch:STATe ON or :CALCulate:MARKer1:CPSeatch:STATe 1
:CALCulate:MARKer1:CPSeatch:STATe?
Sets the gate time for Marker n. The unit is s. Queries the gate time for Marker n.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Marker number. |
| time | real | 1 μs - 500 ms | Gate time for Marker n. |
real - The query returns the gate time for Marker n in scientific notation.
:CALCulate:MARKer2:FCOunt:GATetime 0.01
:CALCulate:MARKer2:FCOunt:GATetime?
Turns on or off the auto gate time for Marker n. Queries the status of the auto gate time for Marker n.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | |
| state | boolean | OFF | ON | 0 | 1 | Enables or disables auto gate time. |
boolean - Returns 1 if auto gate time is enabled and 0 if it is disabled.
:CALCulate:MARKer2:FCOunt:GATetime:AUTO ON or :CALCulate:MARKer2:FCOunt:GATetime:AUTO 1
:CALCulate:MARKer2:FCOunt:GATetime:AUTO?
Enables or disables the frequency counter function of the specified marker. The query returns the status of that function as 0 or 1.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Marker number. |
| state | enum | OFF | ON | 0 | 1 | Turns the frequency counter function off or on. |
enum - Returns the frequency counter status as 0 or 1.
:CALCulate:MARKer2:FCOunt:STATe ON or :CALCulate:MARKer2:FCOunt:STATe 1
:CALCulate:MARKer2:FCOunt:STATe?
Queries the readout of the frequency counter of the specified marker. The readout is returned as an integer in Hz; when the frequency counter function is disabled, the command returns NAN.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Marker number. |
integer (Hz) - Readout of the frequency counter of the specified marker, returned in Hz.
:CALCulate:MARKer2:FCOunt:X?
Selects the special measurement type for the specified marker. Queries the special measurement type of the specified marker.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Marker number. |
| mode | enum | NOISe | BPOWer | BDENsity | OFF | Special measurement type for the specified marker. |
enum - The query returns NOIS, BPOW, BDEN, or OFF.
:CALCulate:MARKer1:FUNCtion NOISe
:CALCulate:MARKer1:FUNCtion?
Sets or queries the left edge frequency or time of the signal used in the band function calculation for the specified marker. It is only valid when the corresponding bandwidth function is enabled.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Marker number. |
| freq | real | 0 - band right | Left edge frequency or time of the signal involved in the band function calculation. |
real - Returns the left edge frequency or time of the signal in scientific notation.
:CALCulate:MARKer1:FUNCtion:BAND:LEFT <freq>
:CALCulate:MARKer1:FUNCtion:BAND:LEFT?
Sets or queries the right edge frequency or time of the signal involved in the calculation for the band function. It is only valid when the corresponding bandwidth function is enabled, and is used to set the right edge time in zero span or PvT view.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Marker number. |
| freq | real | band left - +∞ | Right edge frequency or time of the signal involved in the calculation. |
real - Returns the right edge frequency or time of the signal in scientific notation.
:CALCulate:MARKer1:FUNCtion:BAND:RIGHt 4000000000
:CALCulate:MARKer1:FUNCtion:BAND:RIGHt?
Sets the bandwidth or the time span of the signal involved in the calculation for the band function. Queries the bandwidth or the time span of the signal involved in the calculation for the band function.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | |
| freq | real | 0 - +∞ | Consecutive real number specifying the bandwidth or time span used in the band function calculation. |
:CALCulate:MARKer1:FUNCtion:BAND:SPAN 500000000
:CALCulate:MARKer1:FUNCtion:BAND:SPAN?
Enables or disables the band span auto function. The query returns the status of band span auto function.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Marker number. |
| state | enum | OFF | ON | 0 | 1 | Turns band span auto off or on. |
enum - Returns the status of band span auto function.
Enables or disables the marker line of the specified marker. Queries the status for the marker line of the specified marker.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Marker number. |
| state | enum | OFF | ON | 0 | 1 | Enables or disables the marker line. |
enum - Returns 0 or 1.
:CALCulate:MARKer1:LINes:STATe ON or :CALCulate:MARKer1:LINes:STATe 1
:CALCulate:MARKer1:LINes:STATe?
Searches for and marks the nearest peak located to the left of the current peak that meets the peak search condition.
Performs one peak search based on the search mode set by :CALCulate:MARKer:PEAK:SEARch:MODE and marks it with the specified marker. When no peak is found, the instrument displays "No peak found".
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Marker number to use for the peak search. |
:CALCulate:MARKer2:MAXimum:MAX
Searches for and marks the peak whose amplitude on the trace is next to that of the current peak and which meets the peak search condition.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Marker number to use. |
:CALCulate:MARKer2:MAXimum:NEXT
Searches for and marks the nearest peak located on the right side of the current peak that meets the peak search condition.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Marker number to use. |
:CALCulate:MARKer2:MAXimum:RIGHt
Searches for and marks the peak with the minimum amplitude on the trace.
| Name | Type | Values | Description |
|---|
| n | integer | 1 - 8 | |
:CALCulate:MARKer2:MINimum
Sets the type of the specified marker. The query returns the current marker type.
| Name | Type | Values | Description |
|---|
| n | integer | 1 - 8 | Specifies the marker number. |
| mode | enum | POSition | DELTa | FIXed | OFF | Selects the marker mode. |
enum - Returns the type of the specified marker as POS, DELT, FIX, or OFF.
:CALCulate:MARKer1:MODE POSition
:CALCulate:MARKer1:MODE?
Performs the peak-peak search and marks it with the specified marker. After execution, the marker mode of the specified marker is set to Delta automatically.
| Name | Type | Values | Description |
|---|
| n | integer | 1 - 8 | Marker number. |
:CALCulate:MARKer1:PTPeak
Sets the reference marker for the specified marker. Queries the reference marker for the specified marker.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Specifies the marker number. |
| integer | integer | 1 - 8 | Specifies the reference marker number. |
integer - Returns the reference marker number for the specified marker.
:CALCulate:MARKer1:REFerence 2
:CALCulate:MARKer1:REFerence?
Sets the center frequency of the analyzer to the frequency of the specified marker.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Specifies which marker to use. |
:CALCulate:MARKer1:SET:CENTer
Sets the center frequency of the analyzer to half of the frequency difference (absolute frequency) between the two Delta markers. It is only valid when the current marker mode is "Delta".
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Marker number. |
:CALCulate:MARKer1:SET:DELTa:CENTer
Sets the frequency difference of the specified Delta marker to the span of the analyzer.
| Name | Type | Values | Description |
|---|
| n | integer | 1 - 8 | Specifies the marker number. |
:CALCulate:MARKer1:SET:DELTa:SPAN
Sets the reference level of the analyzer to the amplitude of the specified marker.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Marker number to use. |
Sets the start frequency of the analyzer to the frequency of the specified marker. If the marker is in Delta mode, the start frequency is set to the Delta marker frequency.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Specifies which marker is used. |
:CALCulate:MARKer3:SET:STARt
Sets the center frequency step of the analyzer to the frequency of the specified marker. If the marker mode of the specified marker is Position or Fixed, the center frequency step is set to the frequency of the marker.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Marker number. |
:CALCulate:MARKer3:SET:STEP
Sets the stop frequency of the analyzer to the frequency of the specified marker. If the marker mode is Position or Fixed, the stop frequency is set to the marker frequency; if the marker mode is Delta, it is set to the Delta marker frequency.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Specifies the marker number. |
:CALCulate:MARKer2:SET:STOP
Enables or disables the specified marker. Queries the status of the specified marker.
| Name | Type | Values | Description |
|---|
| <n> | integer | 1 - 8 | Marker number to select. |
| state | enum | OFF | ON | 0 | 1 | Sets the marker state. |
enum - Returns 0 or 1 for the marker status.
:CALCulate:MARKer1:STATe ON
:CALCulate:MARKer1:STATe 1
:CALCulate:MARKer1:STATe?
Sets the marker trace for the specified marker. Queries the marker trace for the specified marker.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Marker number. |
| integer | integer | 1 - 6 | Marker trace number to select. |
discrete - Returns the marker trace for the specified marker.
:CALCulate:MARKer1:TRACe 1
:CALCulate:MARKer1:TRACe?
Enables or disables the auto trace marking of the specified marker. Queries the status of the auto trace marking of the specified marker.
| Name | Type | Values | Description |
|---|
| <n> | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | |
| state | boolean | OFF | ON | 0 | 1 | Turns auto trace marking off or on for the specified marker. |
enum - Returns the auto trace marking status as 0 or 1.
:CALCulate:MARKer1:TRACe:AUTO ON or :CALCulate:MARKer1:TRACe:AUTO 1
:CALCulate:MARKer1:TRACe:AUTO?
Sets the frequency of the specified marker. Queries the frequency of the specified marker.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Specifies which marker to access. |
| param | real | 30 MHz - 1 GHz | Sets the marker frequency. |
string - The query returns the frequency of the specified marker in scientific notation.
:CALCulate:MARKer1:X 515000000
:CALCulate:MARKer1:X?
Sets the readout mode of the X axis of the specified marker. The readout mode can be Frequency, Time, ITime, or Period.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | |
| mode | enum | FREQuency | TIME | ITIMe | PERiod | Selects the X-axis readout mode for the specified marker. |
Enables or disables the auto readout mode of the specified marker. When enabled, if the marker trace changes, the readout mode is re-determined based on the destination trace.
| Name | Type | Values | Description |
|---|
| <n> | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Marker number. |
| state | boolean | OFF | ON | 0 | 1 | Turns auto readout mode off or on. |
:CALCulate:MARKer1:X:READout:AUTO ON or :CALCulate:MARKer1:X:READout:AUTO 1
:CALCulate:MARKer1:X:READout:AUTO?
Sets the Y-axis value of the specified fixed marker. Queries the Y-axis value of the specified marker, with a default unit of dBm.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Specifies the marker number. |
| amp | real | -170 dBm - 30 dBm | Sets the marker Y-axis value. |
real - Returns the Y-axis value in scientific notation.
:CALCulate:MARKer1:Y -59.6
:CALCulate:MARKer1:Y?
Sets the trace number of the trace where the marker stays in the Spectrogram view. Queries the trace number of the trace where the marker stays in the Spectrogram view.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Marker number. |
| integer | integer | 1 - 8192 | Trace number of the trace where the marker stays in the Spectrogram view. |
integer - The query returns the trace number in integer.
:CALCulate:MARKer1:Z:POSition 1
:CALCulate:MARKer1:Z:POSition?
Sets mathematical operations between traces and, in some cases, user-defined offsets. Queries the math operation function of the specified trace.
| Name | Type | Values | Description |
|---|
| trace_destination | enum | TRACE1 | TRACE2 | TRACE3 | TRACE4 | TRACE5 | TRACE6 | |
| function_math | enum | PDIFference | PSUM | LOFFset | LMOFFset | LDIFference | OFF | |
| trace1_operation | enum | TRACE1 | TRACE2 | TRACE3 | TRACE4 | TRACE5 | TRACE6 | |
| trace2_operation | enum | TRACE1 | TRACE2 | TRACE3 | TRACE4 | TRACE5 | TRACE6 | |
| rel_offset | real | -100 dB - 100 dB | |
| rel_reference | real | -170 dBm - 30 dBm | |
enum - Returns the operation function of the specified trace.
Enables or disables the normalize function. The query reports the current normalization status.
| Name | Type | Values | Description |
|---|
| state | boolean | - | Turns normalization off or on. |
integer - Returns the normalization status as 0 or 1.
:CALCulate:NTData:STATe ON or :CALCulate:NTData:STATe 1
:CALCulate:NTData:STATe?
Deletes signals from the EMI signal table according to the selected scope. The scope can be all, current, marked, or unmarked signals.
| Name | Type | Values | Description |
|---|
| scope | enum | ALL | CURRent | MARKed | UNMarked | Selects which signals to delete from the signal table. |
:CALCulate:EMI:SLISt:DELete ALL
Adds the current measurement results to the signal table.
Turns off all the enabled markers.
:CALCulate:FSCan:MARKer:AOFF
Enables or disables the couple marker function. The query returns the state of the couple marker function.
| Name | Type | Values | Description |
|---|
| stateopt | enum | OFF | ON | 0 | 1 | Sets the couple marker function state. |
enum - State of the couple marker function.
:CALCulate:FSCan:MARKer:COUPle[:STATe] OFF|ON|0|1
:CALCulate:FSCan:MARKer:COUPle[:STATe]?
Sets the peak excursion. The unit is dB. Queries the peak excursion.
| Name | Type | Values | Description |
|---|
| rel_ampl | real | 0 dB - 100 dB | Consecutive real number specifying the peak excursion. |
real (dB) - The query returns the peak excursion in scientific notation.
:CALCulate:FSCan:MARKer:PEAK:EXCursion 12
:CALCulate:FSCan:MARKer:PEAK:EXCursion?
Enables or disables the peak excursion function. Queries the status of the peak excursion function.
| Name | Type | Values | Description |
|---|
| state | enum | OFF | ON | 0 | 1 | Turns the peak excursion function off or on. |
enum - Returns the status of the peak excursion function as 0 or 1.
Sets the peak search mode and queries the current peak search mode. The command is only valid for peak search executed by the :CALCulate:FSCan:MARKer<n>:MAXimum command.
| Name | Type | Values | Description |
|---|
| mode | enum | PARameter | MAXimum | Peak search mode selection. |
enum - Returns the current peak search mode.
:CALCulate:FSCan:MARKer:PEAK:SEARch:MODE PARameter
:CALCulate:FSCan:MARKer:PEAK:SEARch:MODE?
Sets the peak threshold and queries the peak threshold.
| Name | Type | Values | Description |
|---|
| ampl | real | -200 dBm - 0 dBm | Peak threshold amplitude. |
real - The query returns the peak threshold in scientific notation.
:CALCulate:FSCan:MARKer:PEAK:SEARch:THReshold -100
:CALCulate:FSCan:MARKer:PEAK:SEARch:THReshold?
Enables or disables the peak threshold function. Queries the status of the peak threshold function.
| Name | Type | Values | Description |
|---|
| state | enum | OFF | ON | 0 | 1 | Enables or disables the peak threshold function. |
Queries the result measured at the marker. The query returns the detector 1, detector 2, detector 3, and frequency values at the marker in scientific notation, separated by commas; if the detector is disabled or the measurement is not completed, it returns -1.700000000e+02.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Marker number to query. |
string - Returns the result measured at the marker as four comma-separated scientific-notation values: detector 1 amplitude, detector 2 amplitude, detector 3 amplitude, and frequency. If the detector is disabled or the measurement is not completed, returns -1.700000000e+02.
:CALCulate:FSCan:MARker1:FUNCtion:MAMarker?
Performs one peak search based on the search mode set by the :CALCulate:FSCan:MARKer:PEAK:SEARch:MODE command and marks it with the specified marker.
| Name | Type | Values | Description |
|---|
| n | integer | 1 - 8 | Marker number. |
:CALCulate:FSCan:MARKer2:MAXimum:MAX
Searches for and marks the nearest peak located to the left of the current peak that meets the peak search condition.
| Name | Type | Values | Description |
|---|
| n | integer | 1 - 8 | Marker number. |
Searches for and marks the peak whose amplitude on the trace is next to that of the current peak and which meets the peak search condition.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Marker number. |
:CALCulate:FSCan:MARKer2:MAXimum:NEXT
Searches for and marks the nearest peak located to the right of the current peak that meets the peak search condition.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Marker number. |
:CALCulate:FSCan:MARKer2:MAXimum:RIGHt
Searches for and marks the peak with the minimum amplitude on the trace.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Specifies the marker number. |
:CALCulate:FSCan:MARKer2:MINimum
Sets the type of the specified marker. Queries the type of the specified marker.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | |
| mode | enum | POSition | DELTa | OFF | Selects the marker type: position, delta, or off. |
enum - Returns the marker type as POS, DELT, or OFF.
:CALCulate:FSCan:MARKer1:MODE POSition
:CALCulate:FSCan:MARKer1:MODE?
Sets the reference marker for the specified marker. The query returns the current reference marker number for that marker.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Specifies the marker number. |
| integer | integer | 1 - 8 | Reference marker number to assign. |
integer - Returns the reference marker number for the specified marker.
:CALCulate:FSCan:MARKer1:REFerence 2
:CALCulate:FSCan:MARKer1:REFerence?
Adds the frequency at the specified marker to the signal table.
| Name | Type | Values | Description |
|---|
| n | integer | 1 - 8 | Marker number to use. |
:CALCulate:FSCan:MARKer1:SET:SLISt
Sets the frequency at the currently selected marker to the frequency of the meter.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Marker number. |
Sets the frequency of the meter to the frequency at the currently selected marker.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Marker number. |
:CALCulate:FSCan:MARKer1:TO:METer
Sets or queries the trace assigned to the specified marker. The trace index is from 1 to 3.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Marker number. |
| integer | discrete | 1 | 2 | 3 | Marker trace number. |
integer - Returns the marker trace number.
:CALCulate:FSCan:MARKer1:TRACe 2
:CALCulate:FSCan:MARKer1:TRACe?
Adds the frequency of the meter to the signal table.
Deletes the selected signal(s) from the signal table.
| Name | Type | Values | Description |
|---|
| integer | integer | 1 - 2,000 | |
:CALCulate:FSCan:SLISt:DELete:SIGNal 20
Marks all the signals in the signal table.
Clears all the marks in the signal table.
Clears the mark of the specified signal.
| Name | Type | Values | Description |
|---|
| integer | integer | 1 - 2,000 | |
Marks the specified signal.
| Name | Type | Values | Description |
|---|
| integer | integer | 1 - 2,000 | Specifies the signal number to mark in the signal table. |
The following command marks Signal 20 in the signal table. :CALCulate:FSCan:SLISt:MARK:SIGNal 20
Replaces the frequency of the currently selected signal with the currently set frequency of the meter.
| Name | Type | Values | Description |
|---|
| integer | integer | 1 - 2,000 | |
:CALCulate:FSCan:SLISt:REPLace:METer 20
Sets the frequency of the meter to the frequency of the specified signal.
| Name | Type | Values | Description |
|---|
| integer | integer | 1 - 2,000 | |
Enables or disables the auto sort function. Queries the status of the auto sort function.
| Name | Type | Values | Description |
|---|
| state | enum | OFF | ON | 0 | 1 | Turns auto sort off or on. |
enum - The query returns 0 or 1.
:CALCulate:FSCan:SLISt:SORT:AUTO OFF or :CALCulate:FSCan:SLISt:SORT:AUTO 0
:CALCulate:FSCan:SLISt:SORT:AUTO?
Sets the detector amplitude sorting mode for the signal. Queries the detector amplitude sorting mode for the signal.
| Name | Type | Values | Description |
|---|
| mode | enum | DET1 | DET2 | DET3 | Detector amplitude sorting mode for the signal. |
:CALCulate:FSCan:SLISt:SORT:DAMPlitude DET1
Sets the detector limit delta amplitude mode for the signal. Queries the detector limit delta amplitude mode for the signal.
| Name | Type | Values | Description |
|---|
| mode | enum | DET1 | DET2 | DET3 | Detector limit delta amplitude mode. |
:CALCulate:FSCan:SLISt:SORT:DLDelta DET1
:CALCulate:FSCan:SLISt:SORT:DLDelta?
Sets the sort order of the signal. Queries the sort order of the signal.
enum - The query returns ASC or DESC.
:CALCulate:FSCan:SLISt:SORT:ORDer ASCending
:CALCulate:FSCan:SLISt:SORT:ORDer?
Sets the sort mode of the signal. Queries the sort mode of the signal.
| Name | Type | Values | Description |
|---|
| mode | enum | FREQuency | DAMPlitude | DLDelta | Sort mode of the signal. |
enum - The query returns FREQ, DAMP, or DLD.
::CALCulate:FSCan:SLISt:SORT:TYPE FREQuency
::CALCulate:FSCan:SLISt:SORT:TYPE?
Takes the midspan frequency as the reference and zooms the coordinate display at the selected signal of the signal table to 10% of the current span.
| Name | Type | Values | Description |
|---|
| integer | integer | 1 - 2,000 | |
:CALCulate:FSCan:SLISt:ZOOM 3
Returns the coordinate display of the current signal to its previous display set before the Signal Zoom.
Deletes all the limit line data.
Enables or disables the limit line test function. The query returns the setting state as 0 or 1.
| Name | Type | Values | Description |
|---|
| state | enum | OFF | ON | 0 | 1 | Enables or disables the limit line test function. |
integer - The query returns 0 or 1.
:CALCulate:LLINe:TEST OFF or :CALCulate:LLINe:TEST 0
:CALCulate:LLINe:TEST?
Enables or disables the coupling between the selected limit line data points and the reference level. The query returns the setting state as 0 or 1.
| Name | Type | Values | Description |
|---|
| state | enum | OFF | ON | 0 | 1 | Enables or disables the coupling function. |
enum - Returns the setting state of the coupling function as 0 or 1.
:CALCulate:LLINe<n>:AMPLitude:CMODe:RELative OFF|ON|0|1
:CALCulate:LLINe<n>:AMPLitude:CMODe:RELative?
Builds the selected limit line from the specified trace.
| Name | Type | Values | Description |
|---|
| nopt | integer | 1 - 6 | Selects the limit line number. |
| trace | enum | TRACE1 | TRACE2 | TRACE3 | Selects the trace used to build the limit line. |
:CALCulate:LLINe2:BUILd TRACE2
Sets the frequency interpolation mode of the selected limit line. The query returns the current frequency interpolation mode.
| Name | Type | Values | Description |
|---|
| mode | enum | LOGarithmic | LINear | Frequency interpolation mode for the selected limit line. |
Copies the selected limit line to the current limit line.
| Name | Type | Values | Description |
|---|
| n | integer | 1 - 6 | Limit line number to copy to. |
| source | enum | LLINE1 | LLINE2 | LLINE3 | LLINE4 | LLINE5 | LLINE6 | Limit line to copy from. |
:CALCulate:LLINe2:COPY LLINE1
Edits one limit line and marks it with n. The query returns the limit line data currently being edited.
| Name | Type | Values | Description |
|---|
| n | integer | 1 - 6 | Limit line number to edit. |
| x | real | 0 Hz - 3 GHz | Frequency value for a data point. |
| ampl | real | -1,000 dBm - 1,000 dBm | Amplitude value for a data point. |
| connect | integer | 0 - 1 | Connection flag for the data point. |
string - The query returns the limit line data currently being edited.
:CALCulate:LLINe1:CONTrol:INTerpolate:TYPE LINear
:CALCulate:LLINe1:CONTrol:INTerpolate:TYPE?
Deletes the selected limit line.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | Selects the limit line to delete. |
::CALCulate:LLINe2:DELete
Turns on or off the selected limit line. Queries the setting state of the selected limit line.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | Selects the limit line number. |
| state | enum | OFF | ON | 0 | 1 | Turns the selected limit line off or on. |
integer - Returns 0 or 1 indicating the setting state of the selected limit line.
:CALCulate:LLINe2:DISPlay OFF|ON|0|1
:CALCulate:LLINe2:DISPlay?
Queries the measurement results of the selected limit line and its associated trace.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 6 | |
enum - Returns 0 for pass or 1 for fail.
Enables or disables the coupling function between the selected limit line data points and the center frequency. The query returns the current coupling state.
| Name | Type | Values | Description |
|---|
| n | integer | 1 - 6 | Selects the limit line number. |
| state | enum | OFF | ON | 0 | 1 | Enables or disables the coupling function. |
enum - Returns the current coupling setting: OFF/0 for disabled or ON/1 for enabled.
Sets the margin for the selected limit line. Queries the margin for the selected limit line.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 6 | Selects the limit line number. |
| rel_ampl | real | -40 dB - 0 dB | Sets the margin value for the selected limit line. |
real - The query returns the margin for the limit line in scientific notation.
:CALCulate:LLINe2:MARGin 1
:CALCulate:LLINe2:MARGin?
Enables or disables the margin for the selected limit line. Queries the setting state of the margin for the selected limit line.
| Name | Type | Values | Description |
|---|
| <n> | integer | 1 - 6 | Selects the limit line number. |
| state | enum | OFF | ON | 0 | 1 | Enables or disables the margin. |
enum - Returns the margin state as 0 or 1.
Sets the frequency offset of the specified limit line. Queries the frequency offset of the specified limit line.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 6 | Specifies the limit line number. |
| freq | real | -1 GHz - 1 GHz | Sets the frequency offset. |
string - The query returns the frequency offset in scientific notation.
:CALCulate:LLINe2:OFFSet:X 5000
:CALCulate:LLINe2:OFFSet:X?
Sets the amplitude offset of the specified limit line. Queries the amplitude offset of the specified limit line.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 6 | Limit line number to apply the offset to. |
| re_ampl | real | -50 - 50 | Amplitude offset value for the specified limit line. |
real - The query returns the amplitude offset in scientific notation.
Sets the trace for the current limit line test. Queries the trace for the current limit line test.
| Name | Type | Values | Description |
|---|
| n | integer | 1 - 6 | Limit line number. |
| trace | integer | 1 - 3 | Trace number used for the current limit line test. |
integer - Returns the trace number for the current limit line test.
:CALCulate:LLINe3:TRACe 2
:CALCulate:LLINe3:TRACe?
Enables or disables the peak threshold line. Queries the status of the peak threshold line.
| Name | Type | Values | Description |
|---|
| state | enum | OFF | ON | 0 | 1 | Enables or disables the peak threshold line. |
enum - Returns the status of the peak threshold line.
Enables or disables the marker line of the specified marker. Queries the status for the marker line of the specified marker.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Marker number. |
| state | enum | OFF | ON | 0 | 1 | Enables or disables the marker line. |
enum - Returns 0 or 1.
:CALCulate:MARKer1:LINes:STATe ON or :CALCulate:MARKer1:LINes:STATe 1
:CALCulate:MARKer1:LINes:STATe?
Enables or disables the specified marker. Queries the status of the specified marker.
| Name | Type | Values | Description |
|---|
| <n> | integer | 1 - 8 | Marker number to select. |
| state | enum | OFF | ON | 0 | 1 | Sets the marker state. |
enum - Returns 0 or 1 for the marker status.
:CALCulate:MARKer1:STATe ON
:CALCulate:MARKer1:STATe 1
:CALCulate:MARKer1:STATe?
Sets the frequency of the specified marker. Queries the frequency of the specified marker.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Specifies which marker to access. |
| param | real | 30 MHz - 1 GHz | Sets the marker frequency. |
string - The query returns the frequency of the specified marker in scientific notation.
:CALCulate:MARKer1:X 515000000
:CALCulate:MARKer1:X?
Queries the amplitude of the specified marker.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | |
Queries the measurement results of all the limit lines and their associated meters. If no limit line is set for the meter, the query returns 0; otherwise it returns 0 if all meters are within limit and 1 if the test fails.
integer - Returns 0 if all meters are within the limit and the test is successful, or 1 if the test is failed.
Queries the current power of the meter.
real - Returns the current power values of three meters in scientific notation, separated by commas. Returns -1.700000000e+02 for certain filter/detector combinations or when the meter is disabled.
Queries the peak power of the meter. If the meter is disabled, the query returns the minimum data value.
real - The peak power of the meter.
Sets or queries the limits of the specified meter. The query returns the limits of the specified meter in scientific notation.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 3 | Specifies the meter number. |
| ampl | real | dBuV | Current amplitude range limit. |
string - Returns the limits of the specified meter in scientific notation.
:CALCulate:METer1:LIMit:DATA 10
:CALCulate:METer1:LIMit:DATA?
Tests the measurement results of one limit line and its associated specified meter. The query returns 0 if the selected meter is within the limit and 1 if it exceeds the limit line.
| Name | Type | Values | Description |
|---|
| n | integer | 1 - 3 | Specifies the meter number. |
enum - Returns 0 when the selected meter is within the limit and the test is successful, or 1 when the selected meter exceeds the limit line and the test fails.
Enables or disables the limit line of the specified meter. The query reports the limit-line status of the specified meter.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 3 | Specifies the meter number. |
| state | boolean | OFF | ON | 0 | 1 | Turns the meter limit line off or on. |
integer - The query returns 0 or 1.
:CALCulate:METer1:LIMit:STATe ON or :CALCulate:METer1:LIMit:STATe 1
:CALCulate:METer1:LIMit:STATe?
Turns off all the enabled markers.
Enables or disables the couple marker function. The query returns the state of the couple marker function.
| Name | Type | Values | Description |
|---|
| stateopt | enum | OFF | ON | 0 | 1 | Turns the couple marker function on or off. |
enum - Returns the couple marker function state as 0 or 1.
:CALCulate:DDEMod:MARKer:COUPle:STATe OFF or :CALCulate:DDEMod:MARKer:COUPle:STATe 0
:CALCulate:DDEMod:MARKer:COUPle:STATe?
Enables or disables the marker table. Queries the state of the marker table.
| Name | Type | Values | Description |
|---|
| stateopt | enum | OFF | ON | 0 | 1 | Turns the marker table on or off. |
Enables or disables continuous peak search function, and marks the peak value with Marker N. Queries the status of continuous peak search function.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | |
| stateopt | boolean | OFF | ON | 0 | 1 | Enables or disables continuous peak search. |
integer - The query returns 0 or 1.
:CALCulate:DDEMod:MARKer1:CPSearch:STATe ON or :CALCulate:DDEMod:MARKer1:CPSearch:STATe 1
:CALCulate:DDEMod:MARKer1:CPSearch:STATe?
Selects the special measurement type for the specified marker. Queries the special measurement type of the specified marker.
| Name | Type | Values | Description |
|---|
| n | integer | 1 - 8 | Marker number. |
| function | enum | BPOWer | OFF | Special measurement type to use for the marker. |
enum - Returns the special measurement type of the specified marker.
:CALCulate:DDEMod:MARKer1:FUNCtion BPOWer
:CALCulate:DDEMod:MARKer1:FUNCtion?
Sets the left edge frequency or time of the signal involved in the calculation for the band function. The query returns the left edge frequency or time in scientific notation.
| Name | Type | Values | Description |
|---|
| n | integer | 1 - 8 | Marker number. |
| real | real | 0 - band right | Left edge frequency or time. |
real - Returns the left edge frequency or time of the signal involved in the calculation for the band function.
:CALCulate:DDEMod:MARKer1:FUNCtion:BAND:LEFT 2000000
:CALCulate:DDEMod:MARKer1:FUNCtion:BAND:LEFT?
Sets the right edge frequency or time of the signal involved in the calculation for the band function. Queries the right edge frequency or time of the signal involved in the calculation for the band function.
| Name | Type | Values | Description |
|---|
| <n> | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | |
| <real> | real | 0 - band right | Consecutive real number for the right edge frequency or time. |
returns
:CALCulate:DDEMod:MARKer1:FUNCtion:BAND:RIGHt 2000000
:CALCulate:DDEMod:MARKer1:FUNCtion:BAND:RIGHt?
Sets or queries the bandwidth or time span of the signal involved in the calculation for the band function. It is only valid when the marker is open on frequency-domain data and the corresponding bandwidth function is enabled.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Marker number. |
| freq | real | 0 - +∞ | Bandwidth or time span value. |
real - Returns the bandwidth of the signal involved in the calculation for the band function in scientific notation.
:CALCulate:DDEMod:MARKer1:FUNCtion:BAND:SPAN 500000000
:CALCulate:DDEMod:MARKer1:FUNCtion:BAND:SPAN?
Performs one peak search and marks the found peak with the specified marker. If no peak is found, the instrument displays the prompt message "No peak found".
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Marker number to use. |
:CALCulate:DDEMod:MARKer2:MAXimum
Searches for the nearest peak which is located at the left side of the current peak and marks it with the specified marker.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Marker number to use. |
:CALCulate:DDEMod:MARKer2:MAXimum:LEFT
Searches for the peak whose amplitude on the trace is next lower than that of the current peak and marks it with the specified marker.
| Name | Type | Values | Description |
|---|
| n | enum | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Marker number to use. |
:CALCulate:DDEMod:MARKer2:MAXimum:NEXT
Searches for the peak whose amplitude on the trace is next higher than that of the current peak and marks it with the specified marker.
| Name | Type | Values | Description |
|---|
| <n> | discrete | 1 - 8 | Marker number. |
:CALCulate:DDEMod:MARKer2:MAXimum:PREVious
Searches for the nearest peak located at the right side of the current peak and marks it with the specified marker.
| Name | Type | Values | Description |
|---|
| <n> | discrete | 1 - 8 | Marker number to use for the peak search. |
:CALCulate:DDEMod:MARKer2:MAXimum:RIGHt
Searches for the peak with the minimum amplitude on the trace and marks it with the specified marker.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Marker number to use. |
:CALCulate:DDEMod:MARKer2:MINimum
Sets the type of the specified marker, or queries the current marker type. The marker can be set to POSition, DELTa, FIXed, or OFF.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Marker number. |
| mode | enum | POSition | DELTa | FIXed | OFF | Marker type to set. |
enum - Returns the current marker type as POS, DELT, FIX, or OFF.
:CALCulate:DDEMod:MARKer1:MODE POSition
:CALCulate:DDEMod:MARKer1:MODE?
Sets or queries the reference marker for the specified marker. Each marker can have another marker as its reference marker, and a marker cannot reference itself.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Specifies the marker number. |
| integer | integer | 1 - 8 | Sets the reference marker number for the specified marker. |
integer - Returns the reference marker number for the specified marker.
:CALCulate:DDEMod:MARKer1:REFerence 2
:CALCulate:DDEMod:MARKer1:REFerence?
Sets the marker trace for the specified marker. Queries the marker trace for the specified marker.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | |
| integer | discrete | 1 | 2 | 3 | 4 | Marker trace selection, where 1 corresponds to Trace 1 and 4 corresponds to Trace 4. |
integer - Returns the marker trace as an integer ranging from 1 to 4.
:CALCulate:DDEMod:MARKer1:TRACe 2
Sets the X-axis value of the specified marker. The default unit is Hz; the query returns the marker's X-axis value in scientific notation.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Specifies which marker to access. |
| real | real | Hz | Sets the X-axis value of the marker within the available range of the current X axis. |
real - Returns the X-axis value of the marker in scientific notation.
The following command sets the X-axis value of Marker 1 to 150 MHz.
:CALCulate:DDEMod:MARKer1:X 150000000
The following query returns 1.500000000e+08.
:CALCulate:DDEMod:MARKer1:X?
Sets the selected marker's quadrature component (imaginary) of the Y value. Queries the selected marker's quadrature component (imaginary) of the Y value.
| Name | Type | Values | Description |
|---|
| n | integer | 1 - 8 | Marker number. |
| real | real | - | Quadrature component (imaginary) of the Y value to set. |
real - Returns the selected marker's quadrature component (imaginary) of the Y value.
Sets or queries the Y-axis value of the specified marker. When queried, it returns the marker's Y-axis value or the in-phase component of the Y value in scientific notation.
| Name | Type | Values | Description |
|---|
| n | integer | 1 - 8 | Marker number. |
| real | real | - | Y-axis value or in-phase component of the Y value. |
real - The Y-axis value of the marker or the in-phase component of the Y value in scientific notation.
:CALCulate:DDEMod:MARKer1:Y:REAL 0.325
:CALCulate:DDEMod:MARKer1:Y:REAL?
Sets the center frequency of the analyzer to the frequency of the specified marker.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Specifies which marker to use. |
:CALCulate:MARKer1:SET:CENTer
Sets the center frequency of the analyzer to half of the frequency difference (absolute frequency) between the two Delta markers. It is only valid when the current marker mode is "Delta".
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Marker number. |
:CALCulate:MARKer1:SET:DELTa:CENTer
Sets the reference level of the analyzer to the amplitude of the specified marker.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Marker number to use. |
Sets the start frequency of the analyzer to the frequency of the specified marker. If the marker is in Delta mode, the start frequency is set to the Delta marker frequency.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Specifies which marker is used. |
:CALCulate:MARKer3:SET:STARt
Sets the center frequency step of the analyzer to the frequency of the specified marker. If the marker mode of the specified marker is Position or Fixed, the center frequency step is set to the frequency of the marker.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 - 8 | Marker number. |
:CALCulate:MARKer3:SET:STEP
Sets the stop frequency of the analyzer to the frequency of the specified marker. If the marker mode is Position or Fixed, the stop frequency is set to the marker frequency; if the marker mode is Delta, it is set to the Delta marker frequency.
| Name | Type | Values | Description |
|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Specifies the marker number. |
:CALCulate:MARKer2:SET:STOP