Aborts the current operation and restarts the sweep.
SCPI COMMAND REFERENCE
The Rigol RSA5032 is Rigol's spectrum analyzer in the RSA5000 Series family. Its SCPI remote-control interface exposes 300 commands across 23 subsystems - 236 settable and 64 queries - and every command is documented here with its parameters, accepted values, return format, and a copy-paste PyVISA example for automating test and measurement.
300
Commands
236
Settable
64
Queries
23
Subsystems
SCPI
Verified
2026-07-16
All commands
Sets or queries the N value used in N dB bandwidth measurement.
| Name | Type | Values | Description |
|---|---|---|---|
| rel_ampl | real | -140 dB - -0.01 dB | Relative amplitude threshold for the N dB bandwidth measurement. |
:CALCulate:BANDwidth:NDB -4 :CALCulate:BWIDth:NDB -4 :CALCulate:BANDwidth:NDB? :CALCulate:BWIDth:NDB?
Queries the measurement results of the N dB band, with the result reported in Hz. If the two points on either side of the current marker at N dB are not found, the query returns -1.000000000e+02.
Queries the frequency value of the frequency point located at the left side of the current marker at N dB fall or rise in its peak amplitude.
:CALCulate:BANDwidth|BWIDth:RLEFt?
Queries the frequency value of the frequency point that is located at the right side of the current marker with N dB fall or rise in its peak amplitude. If the point is not found, the query returns -1.000000000e+02.
Deletes all the limit line data.
Enables or disables the limit line test function. Queries the status of the limit line test function.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | OFF | ON | 0 | 1 | Enables or disables the limit line test function. |
:CALCulate:LLINe:TEST OFF or :CALCulate:LLINe:TEST 0 :CALCulate:LLINe:TEST?
Enables or disables the coupling functions between the selected limit line data points and the reference level. Queries the status of the coupling functions between the selected limit line data points and the reference level.
| Name | Type | Values | Description |
|---|---|---|---|
| n | integer | 1 - 6 | Selects the limit line data point index. |
| state | enum | OFF | ON | 0 | 1 | Enables or disables the coupling function. |
:CALCulate:LLINe<n>:AMPLitude:CMODe:RELative OFF|ON|0|1 :CALCulate:LLINe<n>:AMPLitude:CMODe:RELative?
Builds the limit line from the selected trace.
| Name | Type | Values | Description |
|---|---|---|---|
| trace | enum | TRACE1 | TRACE2 | TRACE3 | TRACE4 | TRACE5 | TRACE6 | Selects which trace is used to build the limit line. |
:CALCulate:LLINe2:BUILd TRACE2
Copies the selected limit line to the current limit line.
| Name | Type | Values | Description |
|---|---|---|---|
| limit line | enum | LLINE1 | LLINE2 | LLINE3 | LLINE4 | LLINE5 | LLINE6 | Selects the limit line to copy from. |
:CALCulate:LLINe2:COPY LLINE1
Edits one limit line and marks it with n. The query form returns the limit line data that is currently being edited.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | Limit line number. |
| x | real | 0 Hz - 6.5 GHz | X-axis value; frequency when the X-axis indicates frequency. |
| x | real | 0 μs - 6.5 ks | X-axis value; time when the X-axis indicates time. |
| ampl | real | -1,000 dBm - 1,000 dBm | Amplitude value. |
| connect | discrete | 0 | 1 | Whether the current point connects with the previous point. |
:CALCulate:LLINe2:DATA 50,100,0,100,150,1,200,200,1 :CALCulate:LLINe2:DATA?
Deletes the selected limit line.
| Name | Type | Values | Description |
|---|---|---|---|
| n | integer | 1 - 6 | Selects which limit line to delete. |
:CALCulate:LLINe2:DELete
Turns on or off the selected limit line. Queries the status of the selected limit line.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 6 | Selects the limit line number. |
| state | boolean | OFF | ON | 0 | 1 | Turns the selected limit line display off or on. |
:CALCulate:LLINe2:DISPlay ON or :CALCulate:LLINe2:DISPlay 1 :CALCulate:LLINe2:DISPlay?
Queries whether the selected limit line and its associated trace pass the test. The query returns 0 for pass or 1 for fail.
| Name | Type | Values | Description |
|---|---|---|---|
| n | integer | 1 - 6 | Selects limit line 1 through 6. |
Enables or disables the coupling functions between the selected limit line data points and the center.
| Name | Type | Values | Description |
|---|---|---|---|
| <n> | integer | 1 - 6 | Selected limit line number. |
| state | enum | OFF | ON | 0 | 1 | Turns the coupling functions off or on. |
:CALCulate:LLINe2:FREQuency:CMODe:RELative ON :CALCulate:LLINe2:FREQuency:CMODe:RELative?
Sets the margin for the selected limit line. Queries the margin for the selected limit line.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | |
| rel_ampl | real | -40 - 0 | Margin relative amplitude for the selected limit line. |
:CALCulate:LLINe2:MARGin 1 dB :CALCulate:LLINe2:MARGin?
Enables or disables the margin for the selected limit line. Queries the status of the margin for the selected limit line.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 6 | Selects the limit line number. |
| state | enum | OFF | ON | 0 | 1 | Turns the margin off or on. |
:CALCulate:LLINe2:MARGin:STATe ON or :CALCulate:LLINe2:MARGin:STATe 1 :CALCulate:LLINe2:MARGin:STATe?
Sets the trace to be tested against the current limit line. Queries which trace is being tested against the current limit line.
| Name | Type | Values | Description |
|---|---|---|---|
| <n> | integer | 1 - 6 | Selects the limit line number. |
| trace | integer | 1 - 6 | Specifies the trace to be tested against the current limit line. |
:CALCulate:LLINe3:TRACe 2 :CALCulate:LLINe3:TRACe?
Sets the type of the specified limit line. Queries the type of the specified limit line.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 6 | |
| type | enum | UPPer | LOWer |
:CALCulate:LLINe2:TYPE UPPer :CALCulate:LLINe2:TYPE?
Turns off all the enabled markers.
Enables or disables the couple marker function. Queries the state of the couple marker function.
| Name | Type | Values | Description |
|---|---|---|---|
| stateopt | boolean | OFF | ON | 0 | 1 | Enables or disables the couple marker function. |
Sets the peak search mode. Queries the peak search mode.
| Name | Type | Values | Description |
|---|---|---|---|
| mode | enum | PARameter | MAXimum | Selects whether peak search uses the specified parameter or the maximum value on the trace. |
:CALCulate:MARKer:PEAK:SEARch:MODE PARameter :CALCulate:MARKer:PEAK:SEARch:MODE?
Sets the peak sort criterion. The query reads back the current peak sort setting.
| Name | Type | Values | Description |
|---|---|---|---|
| sort | enum | FREQuency | AMPLitude | Peak sort criterion. |
The following command sets the peak sort criterion to frequency. :CALCulate:MARKer:PEAK:SORT FREQuency The following query returns FREQuency. :CALCulate:MARKer:PEAK:SORT?
Sets the amplitude of the peak. Its unit is dB. Queries the peak amplitude.
| Name | Type | Values | Description |
|---|---|---|---|
| rel_ampl | real | 0 dB - 100 dB | Peak amplitude setting in dB. |
:CALCulate:MARKer:PEAK:EXCursion 12 :CALCulate:MARKer:PEAK:EXCursion?
Turns peak excursion state on or off. The command accepts the discrete values OFF, ON, 0, and 1.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | OFF | ON | 0 | 1 | Sets the peak excursion state. |
Sets the peak criteria that the displayed peak must meet. Queries the peak criteria that the displayed peak must meet.
| Name | Type | Values | Description |
|---|---|---|---|
| mode | enum | ALL | GTDLine | LTDLine | Peak criteria used by the peak table readout. |
:CALCulate:MARKer:PEAK:TABLe:READout GTDLine :CALCulate:MARKer:PEAK:TABLe:READout?
Enables or disables the peak table. Queries the status of the peak table.
| Name | Type | Values | Description |
|---|---|---|---|
| stateopt | enum | OFF | ON | 0 | 1 | Enables or disables the peak table. |
CALCulate:MARKer:PEAK:TABLe:STATe ON or :CALCulate:MARKer:PEAK:TABLe:STATe 1 The following query returns 1. CALCulate:MARKer:PEAK:TABLe:STATe?
Sets the peak threshold, with default unit dBm. Queries the peak threshold.
| Name | Type | Values | Description |
|---|---|---|---|
| ampl | real | -200 dBm - 0 dBm | Peak threshold value. |
:CALCulate:MARKer:PEAK:THReshold -100 :CALCulate:MARKer:PEAK:THReshold?
Enables or disables the peak threshold function. Queries the status of the peak threshold function.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | OFF | ON | 0 | 1 | Enables or disables the peak threshold function. |
:CALCulate:MARKer:PEAK:THReshold:STATe ON or :CALCulate:MARKer:PEAK:THReshold:STATe 1 :CALCulate:MARKer:PEAK:THReshold:STATe?
Enables or disables the marker table. Queries the status of the marker table.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | OFF | ON | 0 | 1 | Selects whether the marker table is enabled or disabled. |
The following command disables the marker table. :CALCulate:MARKer:TABLe:STATe OFF or :CALCulate:MARKer:TABLe:STATe 0 The following query returns 0. :CALCulate:MARKer:TABLe:STATe?
Enables or disables signal track. Queries the status of the signal track.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | OFF | ON | 0 | 1 | Enables or disables signal track. |
:CALCulate:MARKer:TRCKing:STATe ON or :CALCulate:MARKer:TRCKing:STATe 1 :CALCulate:MARKer:TRCKing:STATe?
Enables or disables continuous peak search function, and marks the peak value with n. Queries the status of the continuous peak search function.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 8 | Marker number used to select which marker is affected. |
| stateopt | enum | OFF | ON | 0 | 1 | Enables or disables continuous peak search. |
:CALCulate:MARKer1:CPSeatch:STATe ON or :CALCulate:MARKer1:CPSeatch:STATe 1 :CALCulate:MARKer1:CPSeatch:STATe?
Sets the gate time for Marker n. The unit is s. Queries the gate time for Marker n.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | |
| time | real | 1 μs - 500 ms | Gate time for Marker n. |
:CALCulate:MARKer2:FCOunt:GATetime 0.01 :CALCulate:MARKer2:FCOunt:GATetime?
Turns on or off the auto gate time for Marker n. Queries the status of the auto gate time for Marker n.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Marker number to control. |
| state | enum | OFF | ON | 0 | 1 | Enables or disables auto gate time. |
:CALCulate:MARKer2:FCOunt:GATetime:AUTO ON or :CALCulate:MARKer2:FCOunt:GATetime:AUTO 1 :CALCulate:MARKer2:FCOunt:GATetime:AUTO?
Enables or disables the frequency counter function of the specified marker. Queries the status of the frequency counter function of the specified marker.
| Name | Type | Values | Description |
|---|---|---|---|
| n | integer | 1 - 8 | Marker number. |
| stateopt | boolean | OFF | ON | 0 | 1 | Enables or disables the frequency counter function. |
:CALCulate:MARKer2:FCOunt:STATe ON or :CALCulate:MARKer2:FCOunt:STATe 1 :CALCulate:MARKer2:FCOunt:STATe?
Queries the readout of the frequency counter of the specified marker. This command is only available for GPSA mode.
| Name | Type | Values | Description |
|---|---|---|---|
| n | integer | 1 - 8 | Marker number. |
:CALCulate:MARKer2:FCOunt:X?
Selects the special measurement type for the specified marker. The query returns the current special measurement type of the specified marker.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 8 | Marker number. |
| mode | enum | NOISe | BPOWer | BDENsity | OFF | Special measurement type to select for the marker. |
:CALCulate:MARKer1:FUNCtion NOISe :CALCulate:MARKer1:FUNCtion?
Sets the left edge frequency of the signal involved in the calculation for the band function. Queries the left edge frequency of the signal involved in the calculation for the band function.
| Name | Type | Values | Description |
|---|---|---|---|
| <n> | discrete | 1 - 8 | |
| freq | real | 0 Hz - 3.4125 GHz | Left edge frequency of the signal involved in the calculation for the band function. |
:CALCulate:MARKer1:FUNCtion:BAND:LEFT <freq> :CALCulate:MARKer1:FUNCtion:BAND:LEFT?
Sets the right edge frequency of the signal involved in the calculation for the band function. Queries the right edge frequency of the signal involved in the calculation for the band function.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 8 | Marker number. |
| freq | real | 3.4125 GHz - 6.5 GHz | Right edge frequency. |
:CALCulate:MARKer1:FUNCtion:BAND:RIGHt 4000000000 :CALCulate:MARKer1:FUNCtion:BAND:RIGHt?
Sets the bandwidth of the signal involved in the calculation for the band function. Queries the bandwidth of the signal involved in the calculation for the band function.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 8 | Marker number. |
| freq | real | 0 Hz - 6.5 GHz | Bandwidth frequency value. |
:CALCulate:MARKer1:FUNCtion:BAND:SPAN 4000000000 :CALCulate:MARKer1:FUNCtion:BAND:SPAN?
Enables or disables the band span auto function. Queries the status of band span auto function.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 8 | Marker number. |
| state | enum | OFF | ON | 0 | 1 | Turns band span auto off or on. |
:CALCulate:MARKer1:FUNCtion:BAND:SPAN:AUTO ON :CALCulate:MARKer1:FUNCtion:BAND:SPAN:AUTO 1 :CALCulate:MARKer1:FUNCtion:BAND:SPAN:AUTO?
Enables or disables the marker line of the specified marker. Queries the status for the marker line of the specified marker.
| Name | Type | Values | Description |
|---|---|---|---|
| n | integer | 1 - 8 | Marker index. |
| stateopt | enum | OFF | ON | 0 | 1 | Turns the marker line off or on. |
Searches for and marks the nearest peak which is located at the left side of the current peak and meets the peak search condition.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 8 |
:CALCulate:MARKer2:MAXimum:LEFT
Performs one peak search based on the search mode set by the :CALCulate:MARKer:PEAK:SEARch:MODE command and marks it with the specified marker.
| Name | Type | Values | Description |
|---|---|---|---|
| n | integer | 1 - 8 | Marker number to use. |
:CALCulate:MARKer2:MAXimum:LEFT
Searches for and marks the peak whose amplitude on the trace is next to that of the current peak and meets the peak search condition.
| Name | Type | Values | Description |
|---|---|---|---|
| <n> | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Marker number. |
:CALCulate:MARKer2:MAXimum:NEXT
Searches for and marks the nearest peak located to the right of the current peak, subject to the peak search condition.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 8 | Marker number to use. |
Searches for and marks the peak with the minimum amplitude on the trace.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 8 | Specifies the marker number. |
:CALCulate:MARKer2:MINimum
Sets the type of the specified marker, or queries the current marker type.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 8 | Marker number. |
| mode | enum | POSition | DELTa | FIXed | OFF | Marker type to set. |
:CALCulate:MARKer1:MODE POSition :CALCulate:MARKer1:MODE?
Performs the peak-peak search and marks it with a specified marker.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 8 | Marker number to use for the peak-peak search. |
:CALCulate:MARKer1:PTPeak
Sets or queries the reference marker for the specified marker. Each marker can have another marker as its reference marker, but a marker cannot reference itself.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Specifies the marker number. |
| integer | integer | 1 - 8 | Specifies the reference marker number. |
Sets the frequency of the specified marker to the center frequency of the analyzer.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Marker number to set. |
:CALCulate:MARKer1:SET:CENTer
Sets the frequency difference of the specified Delta marker to the center frequency of the analyzer.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Specifies the marker number. |
:CALCulate:MARKer1:SET:DELTa:CENTer
Sets the frequency difference of the specified Delta marker to the span of the analyzer.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 8 | Marker number to apply the command to. |
Sets the amplitude of the specified marker to the reference level of the analyzer.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 8 | Specifies the marker number. |
:CALCulate:MARKer2:SET:RLEVel
Sets the frequency of the specified marker to the start frequency of the analyzer.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 8 | Marker number. |
Sets the frequency of the specified marker to the center frequency step of the analyzer.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | Specifies the marker number. |
:CALCulate:MARKer4:SET:STEP
Sets the frequency of the specified marker to the stop frequency of the analyzer.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 8 | Marker number to set. |
:CALCulate:MARKer2:SET:STOP
Enables or disables the specified marker. Queries the status for the specified marker.
| Name | Type | Values | Description |
|---|---|---|---|
| n | integer | 1 - 8 | Marker number. |
| state | enum | OFF | ON | 0 | 1 | Enables or disables the marker. |
:CALCulate:MARKer1:STATe ON or :CALCulate:MARKer1:STATe 1 :CALCulate:MARKer1:STATe?
Sets or queries the marker trace for the specified marker. In Swept SA and RTSA modes the marker trace can be Trace 1 to Trace 6; in ACP measurement mode it can be Trace 1 to Trace 3.
| Name | Type | Values | Description |
|---|---|---|---|
| n | integer | 1 - 8 | Marker number. |
| integer | integer | 1 - 3 | Marker trace number in ACP measurement mode. |
:CALCulate:ACP:MARKer<n>:TRACe <integer> :CALCulate:ACP:MARKer<n>:TRACe? :CALCulate:MARKer<n>:TRACe <integer> :CALCulate:MARKer<n>:TRACe?
Sets or queries the trace associated with the specified marker.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 8 | Marker number to target. |
| integer | discrete | 1 - 3 | Trace number for the marker trace setting. |
:CALCulate:MARKer1:TRACe 1 :CALCulate:MARKer1:TRACe?
Enables or disables the auto trace marking of the specified marker. Queries the status of the auto trace marking of the specified marker.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 8 | Marker number. |
| state | boolean | OFF | ON | 0 | 1 | Enables or disables auto trace marking. |
:CALCulate:MARKer1:TRACe:AUTO ON or :CALCulate:MARKer1:TRACe:AUTO 1 :CALCulate:MARKer1:TRACe:AUTO?
Sets the marker trace of the specified marker to Auto.
| Name | Type | Values | Description |
|---|---|---|---|
| n | integer | 1 - 8 | Marker number. |
:CALCulate:MARKer1:TRACe 2 :CALCulate:MARKer1:TRACe?
Sets the X-axis value of the specified marker, with default unit Hz. Queries the X-axis value of the specified marker.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | |
| param | real | - | Consecutive real number within the available range of the current X axis. |
:CALCulate:MARKer1:X 150000000 :CALCulate:MARKer1:X?
Sets or queries the readout mode of the X axis of the specified marker. The readout mode keywords are frequency, time, instantaneous time, and period.
| Name | Type | Values | Description |
|---|---|---|---|
| n | integer | 1 - 8 | Marker number. |
| mode | enum | FREQuency | TIME | ITIMe | PERiod | Readout mode of the X axis. |
Enables or disables the auto readout mode of the specified marker. When enabled, if the marker trace changes, the readout mode is re-determined based on the destination trace.
| Name | Type | Values | Description |
|---|---|---|---|
| n | integer | 1 - 8 | Marker number. |
| state | boolean | OFF | ON | 0 | 1 | Turns auto readout mode off or on. |
:CALCulate:MARKer1:X:READout:AUTO ON :CALCulate:MARKer1:X:READout:AUTO 1 :CALCulate:MARKer1:X:READout:AUTO?
Queries the Y-axis value of the specified marker, with a default unit of dBm. If the marker mode is Position or Fixed, it returns the marker Y value; if Delta, it returns the Y-axis difference between the reference and Delta markers.
:CALCulate:MARKer1:Y?
Sets mathematical operations between traces and, in some cases, user-defined offsets. Queries the math operation function of the specified trace.
| Name | Type | Values | Description |
|---|---|---|---|
| trace_destination | enum | TRACE1 | TRACE2 | TRACE3 | TRACE4 | TRACE5 | TRACE6 | |
| function_math | enum | PDIFference | PSUM | LOFFset | LMOFFset | LDIFference | OFF | |
| trace1_operation | enum | TRACE1 | TRACE2 | TRACE3 | TRACE4 | TRACE5 | TRACE6 | |
| trace2_operation | enum | TRACE1 | TRACE2 | TRACE3 | TRACE4 | TRACE5 | TRACE6 | |
| rel_offset | real | -100 - 100 | |
| rel_reference | real | -170 - 30 |
Enables or disables the normalize function. The query reports the current normalization status.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | OFF | ON | 0 | 1 | Enables or disables normalization. |
:CALCulate:NTData:STATe ON or :CALCulate:NTData:STATe 1 :CALCulate:NTData:STATe?
Executes self-calibration immediately.
:CALibration:ALL
Enables or disables auto calibration. The query reports the current auto-calibration setting.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | OFF | ON | 0 | 1 | Sets auto calibration state. |
Queries the current measurement function.
Sets the analyzer to be in the ACP measurement state.
Sets the analyzer to be in the CNR measurement state.
Sets the analyzer to be in the Density measurement state.
Sets the analyzer to be in the Density Spectrogram measurement state.
Sets the analyzer to be in the EBW measurement state.
Sets the analyzer to be in the harmonic distortion measurement state. This command is only available for GPSA mode and restores the specified measurement to the default factory state.
Sets the analyzer to be in the MCHP measurement state.
Sets the analyzer to be in the Normal measurement state. This command is only available for RTSA mode and restores the specified measurement to the default factory state.
Sets the analyzer to be in the OBW measurement state. This command is only available for GPSA mode and restores the instrument to the default factory state.
Sets the analyzer to be in the PvT Spectrogram measurement state.
Sets the analyzer to be in the PvT Spectrum measurement state.
Sets the analyzer to be in the PvT measurement state. This command is only available for RTSA mode and discontinuing the current measurement restores the specified measurement to the default factory state.
Available only in GPSA mode. Disables the currently activated measurement function and sets the analyzer to swept SA mode.
Sets the analyzer to be in the Spectrogram measurement state.
Sets the analyzer to be in the TOI measurement state.
Sets the analyzer to be in the time-domain power measurement state.
Sets all the manual/auto settings in the current measurement mode to Auto.
| Name | Type | Values | Description |
|---|---|---|---|
| mode | enum | ALL | Selects all manual/auto settings. |
:COUPle ALL
Sets the brightness of the LCD backlight. The query returns the current backlight brightness as an integer.
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 1 - 100 | Backlight brightness setting. |
:DISPlay:BACKlight 50 :DISPlay:BACKlight?
Turns on or off the LCD.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | OFF | ON | 0 | 1 | Controls whether the LCD is enabled or disabled. |
:DISPlay:ENABle ON or :DISPlay:ENABle 1 :DISPlay:ENABle?
Enables or disables the display of the graticule. Queries the display of the graticule.
| Name | Type | Values | Description |
|---|---|---|---|
| stateopt | enum | OFF | ON | 0 | 1 | Enables or disables the graticule display. |
:DISPlay:GRATicule:STATe ON or :DISPlay:GRATicule:STATe 1 :DISPlay:GRATicule:STATe?
Enables or disables HDMI. Queries the status of HDMI.
| Name | Type | Values | Description |
|---|---|---|---|
| stateopt | enum | OFF | ON | 0 | 1 | Enables or disables HDMI. |
:DISPlay:HDMI:STATe ON or :DISPlay:HDMI:STATe 1 The following query returns 1. :DISPlay:HDMI:STATe?
Enables or disables the auto scale function for the horizontal axis in the PvT view. The query returns the status of the auto scale function.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | OFF | ON | 0 | 1 | Enables or disables auto scale for the horizontal axis in the PvT view. |
:DISPlay:PVTime:WINDow:TRACe:X:SCALe:COUPle ON :DISPlay:PVTime:WINDow:TRACe:X:SCALe:COUPle 1 :DISPlay:PVTime:WINDow:TRACe:X:SCALe:COUPle?
Sets the unit per division in the horizontal axis of the PvT view. Queries the unit per division in the horizontal axis of the PvT view.
| Name | Type | Values | Description |
|---|---|---|---|
| time | real | 20 μs - 4 s | Consecutive real number specifying the horizontal axis scale per division. |
:DISPlay:PVTime:WINDow:TRACe:X:SCALe:PDIVision 0.00004 :DISPlay:PVTime:WINDow:TRACe:X:SCALe:PDIVision?
Sets the reference time for the horizontal axis in the PvT view. Queries the reference time for the horizontal axis in the PvT view.
| Name | Type | Values | Description |
|---|---|---|---|
| real | real | -1 s - 40 s | Consecutive real number specifying the reference time. |
:DISPlay:PVTime:WINDow:TRACe:X:SCALe:RLEVel 2 :DISPlay:PVTime:WINDow:TRACe:X:SCALe:RLEVel?
Sets the position of the reference time in the horizontal axis of the PvT view. Queries the position of the reference time in the horizontal axis of the PvT view.
| Name | Type | Values | Description |
|---|---|---|---|
| -- | enum | LEFT | CENTer | RIGHt | Position of the reference time in the horizontal axis of the PvT view. |
:DISPlay:PVTime:WINDow:TRACe:X:SCALe:RPOSition LEFT :DISPlay:PVTime:WINDow:TRACe:X:SCALe:RPOSition?
Sets the unit per division in the vertical axis of the PvT view. Queries the Y-axis scale value in scientific notation.
| Name | Type | Values | Description |
|---|---|---|---|
| rel_ampl | real | 0.1 - 20 | Unit per division in the vertical axis of the PvT view. |
:DISPlay:PVTime:WINDow:TRACe:Y:SCALe:PDIVision 15 :DISPlay:PVTime:WINDow:TRACe:Y:SCALe:PDIVision?
Sets the highest density hue to the highest density value found in the current bitmap, and the lowest density hue to the lowest non-zero density value found in the current bitmap.
Sets or queries the curve nonlinearity. Increasing the value compresses colors toward the higher end of the color bar, while decreasing it compresses colors toward the lower end.
| Name | Type | Values | Description |
|---|---|---|---|
| real | real | -100 - 100 | Consecutive real number specifying the curve nonlinearity. |
:DISPlay:VIEW:DENSity:CNONlinear 50 :DISPlay:VIEW:DENSity:CNONlinear?
Sets the color palette of Density. Queries the color palette of Density.
| Name | Type | Values | Description |
|---|---|---|---|
| palette | enum | COOL | WARM | RADar | FIRE | FROSt | Color palette for Density. |
:DISPlay:VIEW:DENSity:CPALettes COOL :DISPlay:VIEW:DENSity:CPALettes?
Sets the highest density hue. Queries the highest density hue.
| Name | Type | Values | Description |
|---|---|---|---|
| real | real | 0.1 - 100 | Consecutive real number for the highest density hue. |
:DISPlay:VIEW:DENSity:HDHue 60 :DISPlay:VIEW:DENSity:HDHue?
Sets the lowest density hue. Queries the lowest density hue.
| Name | Type | Values | Description |
|---|---|---|---|
| real | real | 0 - 99.9 | Consecutive real number specifying the lowest density hue. |
:DISPlay:VIEW:DENSity:LDHue 30 :DISPlay:VIEW:DENSity:LDHue?
Sets the persistence time. Queries the persistence time.
| Name | Type | Values | Description |
|---|---|---|---|
| time | real | 0 - 10 | Persistence time for Density display. |
The following command sets the persistence time to 5 s. :DISPlay:VIEW:DENSity:PERSistence 5 The following query returns 5.000000e+00. :DISPlay:VIEW:DENSity:PERSistence?
Enables or disables the infinite mode of the persistence time. Queries the status of the infinite mode of the persistence time.
| Name | Type | Values | Description |
|---|---|---|---|
| state | boolean | OFF | ON | 0 | 1 | Enables or disables the infinite mode of the persistence time. |
The following command enables the infinite mode of the persistence time. :DISPlay:VIEW:DENSity:PERSistence:INFinite ON or :DISPlay:VIEW:DENSity:PERSistence:INFinite 1 The following query returns 1. :DISPlay:VIEW:DENSity:PERSistence:INFinite?
Sets the current display view and queries the current display view.
| Name | Type | Values | Description |
|---|---|---|---|
| view | enum | NORMal | SPECtrogram | DENSity | DSPectrogram | PVT | PVTSpectrum | PSPectrogram | Current display view to select. |
:DISPlay:VIEW[:SELect] NORMal|SPECtrogram|DENSity|DSPectrogram|PVT|PVTSpectrum|PSPectrogram :DISPlay:VIEW[:SELect]?
Auto adjusts the reference hue position and bottom hue position based on the highest and lowest amplitude values found in the spectrogram, respectively. This command is only available for Spectrogram in RTSA mode or the view that contains the spectrogram.
Sets the bottom hue position displayed in the graticule, or queries the current bottom hue position. Any amplitudes lower than the bottom hue position are displayed as black.
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 0 - min (90%, with the reference position value -10%) | Bottom hue position displayed in the graticule. |
:DISPlay:VIEW:SPECtrogram:BOTTom 40 :DISPlay:VIEW:SPECtrogram:BOTTom?
Sets the reference hue. Queries the reference hue.
| Name | Type | Values | Description |
|---|---|---|---|
| real | real | 0 - 359.9 | Consecutive real number reference hue value. |
:DISPlay:VIEW:SPECtrogram:HUE 120 :DISPlay:VIEW:SPECtrogram:HUE?
Sets the trace to be displayed in the spectrogram. Queries the trace to be displayed in the spectrogram.
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 1 - 8,192 | Displayed trace number. |
:DISPlay:VIEW:SPECtrogram:POSition 146
Sets the position of the reference hue displayed in the graticule. The query returns the reference hue position in integer form.
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | Max (10%, bottom hue value +10%) - 100 | Reference hue position displayed in the graticule. |
:DISPlay:VIEW:SPECtrogram:REFerence 60 :DISPlay:VIEW:SPECtrogram:REFerence?
Enables or disables coupling the marker to the trace. The query reports whether the marker is coupled to the trace.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | ON | OFF | 1 | 0 | Enables or disables coupling the marker to the trace. |
Sets the selection method for the displayed trace. Queries the selection method for the displayed trace.
| Name | Type | Values | Description |
|---|---|---|---|
| mode | enum | TIME | TNUMber | Selection method for the displayed trace. |
::DISPlay:VIEW:SPECtrogram:TRACe:SELection TIME ::DISPlay:VIEW:SPECtrogram:TRACe:SELection?
Selects a window in the current view. Queries the currently selected window type.
| Name | Type | Values | Description |
|---|---|---|---|
| window | enum | SPECtrum | PVT | Window type to select. |
Sets the position of the display line. Its default unit is dBm. Queries the position of the display line.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 3 | |
| ampl | real | Current amplitude range | Position of the display line. |
:DISPlay:WINDow1:TRACe:Y:DLINe -10 :DISPlay:WINDow1:TRACe:Y:DLINe?
Turns on or off the display line. Queries the status of the display line.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | OFF | ON | 0 | 1 | Display line state. |
Example The following command sets the position of the display line to -10 dBm. :DISPlay:WINDow1:TRACe:Y:DLINe -10 The following query returns -1.000000e+01. :DISPlay:WINDow1:TRACe:Y:DLINe?
Sets the reference level of normalization. Queries the reference level of normalization.
| Name | Type | Values | Description |
|---|---|---|---|
| n | integer | 1 - 3 | |
| rel_ampl | real | -200 dB - 200 dB | Reference level of normalization. |
:DISPlay:WINDow1:TRACe:Y:SCALe:NRLevel -20 :DISPlay:WINDow1:TRACe:Y:SCALe:NRLevel?
Sets the reference position of normalization. Queries the reference position of normalization.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 3 | |
| integer | integer | - | Reference position of normalization. |
Sets the Y-axis scale type. Queries the Y-axis scale type.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 | 2 | 3 | |
| rel_ampl | real | 0.1 - 20 | Consecutive real number specifying the Y-axis scale value. |
:DISPlay:WINDow1:TRACe:Y:SCALe:PDIVision 15 :DISPlay:WINDow1:TRACe:Y:SCALe:PDIVision?
Sets the reference level. Queries the reference level.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 3 | |
| ampl | real | 0% - 100% | Reference level amplitude. |
Sets the reference level offset. Queries the reference level offset.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 3 | |
| rel_ampl | real | -300 - 300 |
:DISPlay:WINDow1:TRACe:Y:SCALe:RLEVel:OFFSet 10 :DISPlay:WINDow1:TRACe:Y:SCALe:RLEVel:OFFSet?
Sets the Y-axis scale type and queries the current Y-axis scale type.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 3 | Window number. |
| mode | enum | LINear | LOGarithmic | Y-axis scale type. |
:DISPlay:WINDow1:TRACe:Y:SCALe:SPACing LOGarithmic :DISPlay:WINDow1:TRACe:Y:SCALe:SPACing?
Queries the results of adjacent channel power measurement. It selects data from the latest measurement results and sends it to the output buffer.
:FETCh:ACPower? -5.150423000e+01,-5.173441000e+01,-2.301865000e-01,-5.142665000e+01,7.757568000e-02. :FETCh:ACPower?
Queries the lower channel power of the adjacent channel power measurement. The result is taken from the latest measurement and placed in the output buffer.
:FETCh:ACPower:LOWer?
Queries the main channel power of the adjacent channel power measurement. It selects data from the latest measurement results and transmits it to the output buffer.
:FETCh:ACPower:MAIN?
Queries the upper channel power of the adjacent channel power measurement.
:FETCh:ACPower:UPPer? The following query returns -5.173441000e+01.
Queries the results of C/N ratio measurement. It selects data from the latest measurement results and transmits the carrier power, noise power, and C/N ratio to the output buffer.
:FETCh:CNRatio?
Queries the carrier power. It is only valid when the C/N ratio measurement is enabled and returns data from the latest measurement results.
:FETCh:CNRatio:CARRier?
Queries the results of C/N ratio measurement. It selects the latest measurement results and transmits them to the output buffer.
:FETCh:CNRatio?
Queries the noise power from the latest C/N ratio measurement results. This command is only valid when the C/N ratio measurement is enabled, and the returned power uses the current Y-axis unit.
:FETCh:CNRatio:NOISe?
Queries the result of the emission bandwidth measurement. It selects data from the latest measurement results and transmits the data to the output buffer.
Queries the amplitudes of the first 10 harmonics. The first harmonic is the fundamental waveform.
:FETCh:HARMonics:AMPLitude:ALL?
Queries the percentage of the total harmonic distortion. It selects data from the latest measurement results and transmits it to the output buffer.
:FETCh:HARMonics:DISTortion?
Queries the frequencies of the first 10 harmonics. The first harmonic is the fundamental waveform.
:FETCh:HARMonics:FREQuency:ALL?
Queries the frequency of the fundamental waveform. It has the same function as :FETCh:HARMonics:FREQuency? 1 and returns data from the latest measurement results.
:FETCh:HARMonics:FUNDamental?
Queries the results of the occupied bandwidth measurement. It selects data from the latest measurement results and transmits them to the output buffer.
::FETCh:OBWidth?
Queries the occupied bandwidth. It is only valid when occupied bandwidth measurement is enabled, and it returns the latest measurement result to the output buffer.
Queries the transmit frequency error. This command is only valid when the occupied bandwidth measurement is enabled, and it returns data from the latest measurement results to the output buffer.
":FETCh:OBWidth:OBWidth:FERRor?"
Queries the measurement results or trace data in the cache area. The query returns the measurement data of Trace n in scientific notation as x,y pairs separated by commas, with units of Hz.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 | 2 | 3 | 4 | 5 | 6 | Trace number to query. |
:FETCh:SANalyzer1?
Queries the results of TOI measurement. The command is only valid when TOI measurement is enabled and returns data from the latest measurement results to the output buffer.
:FETCh:TOIntercept?
Queries the minor value of the intercepts of the Third Order Lower and Third Order Upper. The query is only valid when TOI measurement is enabled and returns the minor one in scientific notation.
*:FETCh:TOIntercept:IP3?
Queries the result of T-power measurement. It returns data from the latest measurement results in scientific notation.
:FETCh:TPOWer?
Selects the binary data byte order for data transmission. The query returns the current byte order.
| Name | Type | Values | Description |
|---|---|---|---|
| mode | enum | NORMal | SWAPped | Binary data byte order for data transmission. |
:FORMat:BORDer NORMal :FORMat:BORDer?
Sets the input/output format of the trace data. Queries the input/output format of the trace data.
| Name | Type | Values | Description |
|---|---|---|---|
| format | enum | ASCii | INTeger,32 | REAL,32 | REAL,64 | Input/output format of the trace data. |
:FORMat:TRACe:DATA REAL,32 :FORMat:TRACe:DATA?
Clears all the event registers and status byte registers.
Sets the enable register for the standard event status register, and the query reads back that register. The returned integer is the binary-weighted sum of the enabled bits.
| Name | Type | Values | Description |
|---|---|---|---|
| value | integer | 0 - 255 | Enable-register value to write. |
*ESE 16 *ESE?
Queries and clears the event register for the standard event status register.
*ESR?
Queries the ID string of the instrument.
Sets bit 0 (Operation Complete, OPC) in the standard event status register to 1 after the current operation is finished. Queries whether the current operation is finished.
Recalls the selected register.
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 1 - 128 | Selected register number to recall. |
*RCL 1
Restores the instrument to its factory default settings.
Saves the current instrument state to the selected register.
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 1 - 128 | Register number to save the current instrument state to. |
*SAV 1
Sets or queries the enable register for the status byte register.
| Name | Type | Values | Description |
|---|---|---|---|
| value | integer | 0 - 255 | Integer value of the enable register; bits 0 and 1 are always treated as 0. |
*SRE 16 *SRE?
Queries the event register for the status byte register.
*STB?
Triggers a sweep or measurement immediately.
Queries whether the self-check operation is finished.
Selects continuous (ON|1) or single (OFF|0) measurement mode. The query returns the current measurement mode.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | OFF | ON | 0 | 1 | Sets continuous or single measurement mode. |
:INITiate:Continuous ON or :INITiate:Continuous 1 :INITiate:Continuous?
Restarts the current measurement.
Restarts the current measurement.
Pauses the currently executed measurement.
Continues the currently paused measurement.
Turns on or off the global center frequency of the instrument. The query returns the current setting status.
| Name | Type | Values | Description |
|---|---|---|---|
| mode | enum | ALL | NONE | Selects whether the global center frequency is enabled or disabled. |
:INSTrument:COUPle:FREQuency:CENTer ALL :INSTrument:COUPle:FREQuency:CENTer?
Resets the parameters of the current mode to the factory default settings.
Selects the working mode of the instrument. The query returns the current working mode as 1 or SA.
| Name | Type | Values | Description |
|---|---|---|---|
| mode | enum | 1 | 2 | Selects the working mode; 1 corresponds to SA and 2 corresponds to RTSA. |
:INSTrument:NSELect 1 :INSTrument:NSELect?
Selects the working mode of the instrument. Queries the working mode of the instrument.
| Name | Type | Values | Description |
|---|---|---|---|
| mode | enum | SA | RTSA | Working mode to select. |
:INSTrument:SELect SA :INSTrument:SELect?
Deletes a specified file.
| Name | Type | Values | Description |
|---|---|---|---|
| file_name | string | - | File path and filename to delete. |
:MMEMory:DELete /gpsa/state/state1.sta
Loads the edited FMT file (.csv).
| Name | Type | Values | Description |
|---|---|---|---|
| label | enum | UPPer | LOWer | Selects the target mask label. |
| file_name | string | - | Specifies the file path and filename to load. |
:MMEMory:LOAD:FMT upper,/rtsa/mask/mask1.csv
Imports the edited limit line file (.csv).
| Name | Type | Values | Description |
|---|---|---|---|
| label | enum | LLINE1 | LLINE2 | LLINE3 | LLINE4 | LLINE5 | LLINE6 | Selects the limit line to load. |
| file_name | string | - | Specifies the path and filename of the limit line file. |
:MMEMory:LOAD:LIMit LLINE1,/gpsa/limit/upp1.csv
Imports the specified state file (.sta).
| Name | Type | Values | Description |
|---|---|---|---|
| file_name | string | - | State file name or path. |
Imports the specified trace file (.trs) and the current instrument state.
| Name | Type | Values | Description |
|---|---|---|---|
| label | enum | TRACE1 | TRACE2 | TRACE3 | TRACE4 | TRACE5 | TRACE6 | Trace destination label. |
| file_name | string | - | Path and filename of the trace file to load. |
:MMEMory:LOAD:TRACe TRACE2,/gpsa/tracestate/t1.trs
Imports the specified measurement data file (.csv).
| Name | Type | Values | Description |
|---|---|---|---|
| label | enum | TRACE1 | TRACE2 | TRACE3 | TRACE4 | TRACE5 | TRACE6 | Selects the trace to import the measurement data into. |
| file_name | string | - | Specifies the path and filename of the measurement data file to load. |
Renames the specified file from <file_name1> to <file_name2>. The paths and filenames must be included, and the operation fails if the target file does not exist.
| Name | Type | Values | Description |
|---|---|---|---|
| file_name1 | string | - | Source file path and filename. |
| file_name2 | string | - | New file path and filename. |
EXAMPLE UNKNOWN
Saves the currently edited FMT file with the specified filename, adding ".csv" by default, to the default mask path. If the file already exists, it is overwritten.
| Name | Type | Values | Description |
|---|---|---|---|
| label | enum | UPPer | LOWer | Selects whether to save the upper or lower mask FMT file. |
| file_name | string | - | Specifies the output filename. |
:MMEMory:STORe:FMT UPPer,mask1
Saves the currently edited limit line to a default limit folder with the specified filename, with ".csv" added by default. If the file already exists, it is overwritten.
| Name | Type | Values | Description |
|---|---|---|---|
| label | enum | LLINE1 | LLINE2 | LLINE3 | LLINE4 | LLINE5 | LLINE6 | ALL | Selects which limit line to save. |
| file_name | string | - | Specifies the filename to save. |
:MMEMory:STORe:LIMit LLINE1,low
Saves the current measurement results with a specified filename suffixed with ".csv" by default to the default path ("/mode name"/measdata).
| Name | Type | Values | Description |
|---|---|---|---|
| file_name | string | - | Filename used to save the current measurement results. |
:MMEMory:STORe:RESults data
Saves the marker table to the default path with a filename suffixed with ".csv" by default. If the specified file already exists, it is overwritten.
| Name | Type | Values | Description |
|---|---|---|---|
| file_name | string | - | Filename to use when saving the marker table. |
:MMEMory:STORe:MTABle MAK1
Saves the peak table with a specified filename suffixed with ".csv" by default to the default path ("/mode name"/measdata). If the specified file already exists, it is overwritten.
| Name | Type | Values | Description |
|---|---|---|---|
| file_name | string | - | Filename used to save the peak table. |
:MMEMory:STORe:PTABle PT1
Saves the current screen image with the specified filename, using the suffix to determine the image format (.jpg, .png, or .bmp). The file is stored in the default screen path, and any existing file with the same name is overwritten.
| Name | Type | Values | Description |
|---|---|---|---|
| file_name | string | - | Filename for the saved screen image. |
MMEMory:STORe:SCReen screen.jpg
Saves the current instrument state with the specified filename suffixed with ".sta" to the default path (/{mode name}/state). If the file already exists, it is overwritten.
| Name | Type | Values | Description |
|---|---|---|---|
| <file_name> | string | - | ASCII string filename for the saved instrument state. |
:MMEMory:STORe:STATe state
Saves the specified trace+state file with filename suffixed with ".trs" to the default path (/"mode name"/tracestate). If the specified file already exists, it is overwritten.
| Name | Type | Values | Description |
|---|---|---|---|
| label | enum | TRACE1 | TRACE2 | TRACE3 | TRACE4 | TRACE5 | TRACE6 | ALL | Selects which trace or traces to save. |
| file_name | string | - | Filename to save, without the ".trs" suffix. |
:MMEMory:STORe:TRACe TRACE1,mystate
Saves the trace measurement results to a file with the specified filename suffixed with ".csv" in the default "/mode name/measdata" path. If the file already exists, it is overwritten.
| Name | Type | Values | Description |
|---|---|---|---|
| label | enum | TRACE1 | TRACE2 | TRACE3 | TRACE4 | TRACE5 | TRACE6 | ALL | Selects which trace or traces to save. |
| file_name | string | - | Specifies the output filename without the ".csv" suffix. |
:MMEMory:STORe:TRACe:DATA TRACE1,mydata
Enables or disables the output of the tracking generator. Queries the on/off status of the tracking generator.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | OFF | ON | 0 | 1 | Turns the tracking generator output off or on. |
:OUTPut:EXTernal:STATe ON or :OUTPut:EXTernal:STATe 1 :OUTPut:EXTernal:STATe?
Executes one adjacent channel power measurement and returns the measurement results.
:READ:ACPower? -5.150423000e+01,-5.173441000e+01,-2.301865000e-01,-5.142665000e+01,7.757568000e-02
Executes one adjacent channel power measurement and returns the lower channel power. The returned power unit is consistent with the current Y-axis unit and is taken from the latest measurement results.
:READ:ACPower:LOWer?
Executes one adjacent channel power measurement and returns the main channel power. The returned power unit is consistent with the current Y-axis unit.
Executes one adjacent channel power measurement and returns the upper channel power.
Executes one C/N ratio measurement and returns the measurement results.
Executes one C/N ratio measurement and returns the carrier power.
:READ:CNRatio:CARRier?
Executes one C/N ratio measurement and returns the C/N ratio.
:READ:CNRatio:CNRatio?
Executes one C/N ratio measurement and returns the noise power. The query selects data from the latest measurement results and transmits it to the output buffer.
Executes one emission bandwidth measurement and returns the measurement results.
:READ:EBWidth?
Executes one harmonic distortion measurement and returns the amplitudes of the first 10 harmonics. The amplitudes are taken from the latest measurement results and transmitted to the output buffer.
-1.692102000e+01,-6.458423000e+01,-7.509421000e+01,-7.924328000e+01,-7.847027000e+01,-7.885457000e+01,-7.882358000e+01,-7.921457000e+01,-7.923057000e+01,-7.915358000e+01
Executes one harmonic distortion measurement and returns the percentage of the total harmonic distortion. It selects data from the latest measurement results and transmits the data to the output buffer.
Executes one harmonic distortion measurement and returns the frequencies of the first 10 harmonics. If fewer than 10 harmonics are measured, unmeasured harmonics have no return value.
Executes one harmonic distortion measurement and returns the frequency of the fundamental waveform. It selects data from the latest measurement results and transmits the data to the output buffer.
Executes one occupied bandwidth measurement and returns the measurement results. It selects data from the latest measurement results and transmits the data to the output buffer.
Executes one occupied bandwidth measurement and returns the occupied bandwidth. It selects data from the latest measurement results and transmits the data to the output buffer.
Executes one occupied bandwidth measurement and returns the transmit frequency error. It selects data from the latest measurement results and transmits the data to the output buffer.
Queries the measurement results or trace data in the buffer. It returns the measurement data of Trace n in scientific notation as comma-separated x,y pairs.
| Name | Type | Values | Description |
|---|---|---|---|
| n | enum | 1 | 2 | 3 | 4 | 5 | 6 | Selects Trace n. |
:FETCh:SANalyzer1?
Executes one TOI measurement and returns the measurement results. The return value includes base lower and upper frequencies, 3rd-order lower and upper frequencies, amplitudes, and intercepts; amplitude units follow the current Y-axis unit.
Executes one TOI measurement and returns the smaller of the Third Order Lower or Third Order Upper intercept values.
Executes one T-power measurement and returns the measurement results. The return value is in scientific notation, and its power unit is consistent with the current Y-axis unit.
:READ:TPOWer?
Sets the input impedance for voltage-to-power conversion. Queries the input impedance for voltage-to-power conversion.
| Name | Type | Values | Description |
|---|---|---|---|
| impedance | enum | 50 | 75 | Input impedance value. |
:INPut:IMPedance 75 or :SENSe:CORRection:IMPedance:INPut:MAGNitude 75 :INPut:IMPedance? or :SENSe:CORRection:IMPedance:INPut:MAGNitude?
Sets the offset of the tracking generator output amplitude. The query returns the offset in scientific notation; this affects only the readout, not the actual output power.
| Name | Type | Values | Description |
|---|---|---|---|
| rel_ampl | real | -200 - 200 | Offset of the output amplitude. |
:SOURce:CORRection:OFFSet 10 :SOURce:CORRection:OFFSet?
Sets the output amplitude of the tracking generator in fixed power output mode. Queries the output amplitude of the tracking generator in fixed power output mode.
| Name | Type | Values | Description |
|---|---|---|---|
| ampl | real | -40 - 0 | Output amplitude setting for fixed power output mode. |
Sets whether to display the reference trace of normalization. Queries whether to display the reference trace of normalization.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | OFF | ON | 0 | 1 | Whether to display the reference trace of normalization. |
:SOURce:TRACe:REF:STATe ON or :SOURce:TRACe:REF:STATe 1 :SOURce:TRACe:REF:STATe?
Saves the reference trace of normalization.
Queries the condition register of the Operation Status Register.
Sets the enable register of the Operation Status Register. The enable register value is a decimal integer corresponding to the binary-weighted sum of status bits 0 through 15, with some bits reserved and bit 15 always treated as 0.
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 0 - 32,767 | Enable register value as a decimal integer corresponding to the status bit mask. |
Queries the event register of the Operation Status Register.
:STATus:OPERation[:EVENt]?
Sets the bit enabled in the Operation Event Register when the operation condition register bit has a negative transition (1 to 0). Queries the enable bit of the Operation Event Register.
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 0 - 32,767 | Binary-weighted sum of the bits to enable; reserved bits may be set but do not affect the system. |
*None*
Sets the bit enabled in the Operation Event Register when the condition register bit has a positive transition (0 to 1). Queries the bit enabled in the Operation Event Register when the condition register bit has a positive transition (0 to 1).
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 0 - 32,767 | Bit-enabled value for the Operation Event Register; any decimal value corresponding to the binary-weighted sum of the allowed bits. |
:STATus:OPERation:PTRansition 100 :STATus:OPERation:PTRansition?
Clears the enable register of the Operation Status Register and Questionable Status Register.
Queries the condition register of the Questionable Status Register.
Sets or queries the enable register of the Questionable Status Register.
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 0 - 32,767 | Decimal value corresponding to the binary-weighted sum of the enabled status bits. |
:STATus:QUEStionable:ENABle 100 :STATus:QUEStionable:ENABle?
Queries the event register of the Questionable Status Register.
:STATus:QUEStionable[:EVENt]?
Sets the bit enabled in the Questionable Event Register when the questionable condition register bit has a negative transition (1 to 0). The query form returns the enable bit of the Questionable Event Register.
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 0 - 32,767 | Binary-weighted sum of the bits in the Questionable Status Register. |
:STATus:QUEStionable:NTRansition 100 :STATus:QUEStionable:NTRansition?
Sets the bit enabled in the Questionable Event Register when the questionable condition register bit has a positive transition (0 to 1). Queries the enable bit in the Questionable Event Register for that transition.
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 0 - 32,767 | Bit-encoded enable mask for the Questionable Status Register; reserved bits may be set but do not affect the system. |
:STATus:QUEStionable:PTRansition 100 :STATus:QUEStionable:PTRansition?
Queries the condition register of the Questionable Calibration Status Register.
Sets or queries the enable register of the Questionable Calibration Status Register.
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 0 - 32,767 | Decimal value corresponding to the binary-weighted sum of the enabled bits in the Questionable Calibration Status Register. |
Queries the event register of the Questionable Calibration Status Register.
:STATus:QUEStionable:CALibration[:EVENt]?
Sets the enable bit in the Questionable Calibration Event Register when the questionable calibration condition register bit has a negative transition (1 to 0). Queries the enable bit in the same register.
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 0 - 32,767 | Binary-weighted decimal value for the enable bit mask; bits 3-8, 12-13, and 15 are always treated as 0. |
:STATus:QUEStionable:CALibration:NTRansition 100 :STATus:QUEStionable:CALibration:NTRansition?
Sets the enable bit in the Questionable Calibration Event Register when the questionable calibration condition register bit has a positive transition (0 to 1). Queries the enable bit in the same register.
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 0 - 32,767 | Binary-weighted sum of the enabled bits; bit 3-8 and bit 12-13 are always treated as 0, and bit 15 is always treated as 0. |
:STATus:QUEStionable:CALibration:PTRansition 100 :STATus:QUEStionable:CALibration:PTRansition?
Queries the condition register of the Questionable Frequency Status Register.
Sets the enable register of the Questionable Frequency Status Register. Queries the enable register of the Questionable Frequency Status Register.
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 0 - 32,767 | Enable-bit mask for the Questionable Frequency Status Register; bits 3-9 and 11-14 are unused and treated as 0, and bit 15 is always 0. |
:STATus:QUEStionable:FREQuency:ENABle 100
Queries the event register of the Questionable Frequency Status Register.
Sets the enable bit in the Questionable Frequency Event Register when the questionable frequency condition register bit has a negative transition (1 to 0). The query form returns that enable bit as an integer.
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 0 - 32767 | Binary-weighted sum of the enabled bits; bits 3-9 and 11-14 are always treated as 0, and bit 15 is always treated as 0. |
:STATus:QUEStionable:FREQuency:NTRansition 100 :STATus:QUEStionable:FREQuency:NTRansition?
Sets the enable bit in the Questionable Frequency Event Register when the questionable frequency condition register bit has a positive transition (0 to 1). Queries the enable bit in the Questionable Frequency Event Register when the questionable frequency condition register bit has a positive transition (0 to 1).
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 0 - 32,767 | Enable-bit value for the Questionable Frequency Event Register; bit 3-9, bit 11-14, and bit 15 are treated as 0. |
:STATus:QUEStionable:FREQuency:PTRansition 100 :STATus:QUEStionable:FREQuency:PTRansition?
Queries the condition register of the Questionable Integrity Status Register.
Sets or queries the enable register of the Questionable Integrity Status Register.
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 0 - 32,767 | Decimal value corresponding to the binary-weighted sum of the enabled bits; bits 2-14 and bit 15 are always treated as 0. |
Queries the event register of the Questionable Integrity Status Register.
:STATus:QUEStionable:INTegrity[:EVENt]?
Sets the enable bit in the Questionable Integrity Event Register when the questionable integrity condition register bit has a negative transition (1 to 0). Queries the same enable bit for its current integer value.
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 0 - 32,767 | Value corresponding to the binary-weighted sum of the enabled bits; bits 2-14 and bit 15 are always treated as 0. |
:STATus:QUEStionable:INTegrity:NTRansition 100 :STATus:QUEStionable:INTegrity:NTRansition?
Sets the enable bit in the Questionable Integrity Event Register when the questionable integrity condition register bit has a positive transition (0 to 1). The query returns that enable bit as an integer.
| Name | Type | Values | Description |
|---|---|---|---|
| <integer> | integer | 0 - 32767 | Decimal value corresponding to the binary-weighted sum of the bits; bits 2-14 and bit 15 are always treated as 0. |
:STATus:QUEStionable:INTegrity:PTRansition 100 :STATus:QUEStionable:INTegrity:PTRansition?
Queries the condition register of the Questionable Power Status Register.
:STATus:QUEStionable:POWer:CONDition?
Sets or queries the enable register of the Questionable Power Status Register.
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 0 - 32767 | Decimal value corresponding to the enable bits; bit 7 to 14 are always treated as 0 and bit 15 is always treated as 0. |
:STATus:QUEStionable:POWer:ENABle 100 :STATus:QUEStionable:POWer:ENABle?
Queries the event register of the Questionable Power Status Register.
Sets the enable bit in the Questionable Power Event Register when the questionable power condition register bit has a negative transition (1 to 0). Queries the same enable bit for the Questionable Power Event Register.
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 0 - 32,767 | Decimal value corresponding to the binary-weighted sum of bits 0 to 14, with bits 7 to 14 treated as 0. |
Sets the enable bit in the Questionable Power Event Register when the questionable power condition register bit has a positive transition (0 to 1). The query returns the enable bit of the Questionable Power Event Register.
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 0 - 32,767 | Decimal value corresponding to the binary-weighted sum of the enabled bits; bits 7 to 14 are always 0 and bits 1 to 5 are reserved. |
:STATus:QUEStionable:POWer:PTRansition 100 :STATus:QUEStionable:POWer:PTRansition?
Queries the condition register of the Questionable Temperature Status Register.
Sets or queries the enable register of the Questionable Temperature Status Register.
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 0 - 1 | Decimal value corresponding to the enable bits of the Questionable Temperature Status Register. |
:STATus:QUEStionable:TEMPerature:ENABle <integer>
Queries the event register of the Questionable Temperature Status Register.
:STATus:QUEStionable:TEMPerature:ENABle 100 :STATus:QUEStionable:TEMPerature:ENABle? :STATus:QUEStionable:TEMPerature:NTRansition 100 :STATus:QUEStionable:TEMPerature:NTRansition?
Sets or queries the enable bit in the Questionable Temperature Event Register when the questionable temperature condition register bit has a negative transition (1 to 0).
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 0 - 1 | Decimal value corresponding to the binary-weighted sum of the bits; bit 1 to 14 and bit 15 are always treated as 0. |
:STATus:QUEStionable:TEMPerature:NTRansition 100 :STATus:QUEStionable:TEMPerature:NTRansition?
Sets the enable bit in the Questionable Temperature Event Register when the questionable temperature condition register bit has a positive transition (0 to 1). Queries that enable bit.
| Name | Type | Values | Description |
|---|---|---|---|
| integer | integer | 0 - 1 | Decimal value corresponding to the binary-weighted sum of the enabled bits in the Questionable Temperature Status Register. |
Turns on or off the beeper. Queries the status of the beeper.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | OFF | ON | 0 | 1 | Beeper on/off state. |
.:SYSTem:BEEPer:STATe ON or :SYSTem:BEEPer:STATe 1 .:SYSTem:BEEPer:STATe?
Applies the LAN interface parameter settings.
Enables or disables the auto IP setting mode. Queries the status of the auto IP setting mode.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | OFF | ON | 0 | 1 | Enables or disables the auto IP mode. |
:SYSTem:COMMunicate:LAN[:SELF]:AUToip:STATe OFF|ON|0|1 :SYSTem:COMMunicate:LAN[:SELF]:AUToip:STATe?
Enables or disables the DHCP configuration mode. The query returns the current DHCP configuration status.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | OFF | ON | 0 | 1 | Enables or disables DHCP configuration mode. |
SYSTem:COMMunicate:LAN:SELF:DHCP:STATe ON or :SYSTem:COMMunicate:LAN:SELF:DHCP:STATe 1 SYSTem:COMMunicate:LAN:SELF:DHCP:STATe?
Sets the IP address. Queries the IP address.
| Name | Type | Values | Description |
|---|---|---|---|
| ip_address | string | - | IP address to set. |
Sets or queries the DNS auto mode for the LAN self IP configuration. ON/1 enables auto DNS mode; OFF/0 disables it and uses manual mode.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | OFF | ON | 0 | 1 | Selects whether DNS is obtained automatically or manually. |
SYSTem:COMMunicate:LAN:SELF:IP:DNS:AUTO:STATe ON SYSTem:COMMunicate:LAN:SELF:IP:DNS:AUTO:STATe 1 SYSTem:COMMunicate:LAN:SELF:IP:DNS:AUTO:STATe?
Sets the backup DNS server address. This command is only valid in manual IP setting mode.
| Name | Type | Values | Description |
|---|---|---|---|
| ip_address | string | - | Backup DNS server IP address in the format nnn.nnn.nnn.nnn. |
Sets the preferred address for the DNS, or queries the currently preferred DNS address. This command is only valid in manual DNS setting mode.
| Name | Type | Values | Description |
|---|---|---|---|
| ip_address | string | - | Preferred DNS IP address in the form nnn.nnn.nnn.nnn. |
Sets the default gateway. Queries the current default gateway.
| Name | Type | Values | Description |
|---|---|---|---|
| ip_address | string | - | Default gateway address in nnn.nnn.nnn.nnn format. |
:SYSTem:COMMunicate:LAN:SELF:IP:GATeway 172.16.3.1 :SYSTem:COMMunicate:LAN:SELF:IP:GATeway?
Sets the subnet mask. Queries the subnet mask.
| Name | Type | Values | Description |
|---|---|---|---|
| ip_address | string | - | Subnet mask in the format nnn.nnn.nnn.nnn. |
:SYSTem:COMMunicate:LAN:SELF:IP:SUBMask 255.255.255.0
Enables or disables the manual IP setting mode. Queries the status of the manual IP setting mode.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | OFF | ON | 0 | 1 | Enables or disables the manual IP setting mode. |
:SYSTem:COMMunicate:LAN:SELF:MANuip:STATe ON or :SYSTem:COMMunicate:LAN:SELF:MANuip:STATe :SYSTem:COMMunicate:LAN:SELF:MANuip:STATe?
Enables or disables the state of the network information sending. Queries the on/off status of the network information sending.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | OFF | ON | 0 | 1 | Controls the state of the network information sending. |
:SYSTem:COMMunicate:LAN:SELF:MDNS:STATe OFF|ON|0|1 :SYSTem:COMMunicate:LAN:SELF:MDNS:STATe?
Resets the LAN setting by enabling DHCP and disabling Auto IP and Manual IP.
Sets or queries the instrument date.
| Name | Type | Values | Description |
|---|---|---|---|
| year | string | 2000 - 2099 | Year value. |
| month | string | 01 - 12 | Month value. |
| day | string | 01 - 31 | Day value. |
SYSTem:DATE 2017,11,16 SYSTem:DATE?
Sets or queries the front-panel switch state. The command accepts OFF/ON or 0/1 to control the state, and the query returns the current state as 0 or 1.
| Name | Type | Values | Description |
|---|---|---|---|
| stateopt | enum | OFF | ON | 0 | 1 | Front-panel switch state. |
Sets the language of the instrument. Queries the language of the instrument.
| Name | Type | Values | Description |
|---|---|---|---|
| language | enum | ENGLish | CHINese | Language to set for the instrument. |
:SYSTem:LANGuage ENGLish :SYSTem:LANGuage?
Installs and activates the specified option.
| Name | Type | Values | Description |
|---|---|---|---|
| option info | - | - | |
| license info | - | - |
Queries whether an option is activated or not.
| Name | Type | Values | Description |
|---|---|---|---|
| option name | string | - | Option name to check. |
:SYSTem:OPTion:STATe? RSA5000-PA
Selects the setting type the instrument recalls at power-on. Queries what setting type the instrument recalls at power-on.
| Name | Type | Values | Description |
|---|---|---|---|
| mode | enum | MODE | PRESet | USER | LAST | Setting type recalled at power-on. |
:SYSTem:PON:TYPE MODE|PRESet|USER|LAST :SYSTem:PON:TYPE?
Resets the current mode to the factory mode.
Saves the specified user setting.
| Name | Type | Values | Description |
|---|---|---|---|
| setting | enum | USER1 | USER2 | USER3 | USER4 | USER5 | USER6 | Specifies which user setting to save. |
Selects the preset type of the system to be the factory mode or one of the user preset items. FACTory may be used in place of MODE to select the factory mode.
| Name | Type | Values | Description |
|---|---|---|---|
| type | enum | MODE | USER1 | USER2 | USER3 | USER4 | USER5 | USER6 | Selects the system preset type. |
:SYSTem:PRESet:TYPe?
Saves the current user setting. If the current preset type is Mode, the setting is saved to User1; if the current preset type is User1 through User6, it is saved to the specified user.
Clears all the data that you set safely. It deletes all saved files and configuration files, and performs a factory reset.
Displays or queries the system-related information. OFF turns off the system information display.
| Name | Type | Values | Description |
|---|---|---|---|
| mode | enum | OFF | SYSTem | OPTion | LICense | Selects which system-related information to display. |
Sets or queries the system time of the instrument.
| Name | Type | Values | Description |
|---|---|---|---|
| hour | integer | 00 - 23 | Hour value of the system time. |
| minute | integer | 00 - 59 | Minute value of the system time. |
| second | integer | 00 - 59 | Second value of the system time. |
:SYSTem:TIME 15,10,30 :SYSTem:TIME?
Clears all traces by resetting all points on the trace to the minimum trace value.
Resets all the traces. It enables Trace 1 and sets the other traces to blank state without affecting the trace type or detector type.
Enables or disables the display of the specified trace. The query reports the status for the specified trace.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 6 | Specifies the trace number. |
| state | boolean | OFF | ON | 0 | 1 | Turns the trace display off or on. |
Sets the type of the specified trace. Queries the type of the specified trace.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 6 | |
| type | enum | WRITe | AVERage | MAXHold | MINHold | Trace type to select. |
:TRACe1:TYPE MAXHold :TRACe1:TYPE?
Enables or disables update of the specified trace. The query returns the update status.
| Name | Type | Values | Description |
|---|---|---|---|
| n | integer | 1 - 6 | Trace number. |
| state | enum | OFF | ON | 0 | 1 | Enables or disables trace update. |
:TRACe1:UPDate:STATe ON or :TRACe1:UPDate:STATe 1 :TRACe1:UPDate:STATe?
Sets the time that the analyzer will wait for the trigger to be initiated automatically. Queries the same time value.
| Name | Type | Values | Description |
|---|---|---|---|
| time | real | 1 ms - 100 s | Time the analyzer waits before automatic trigger initiation. |
:TRIGger:SEQuence:ATRigger 0.01 :TRIGger:SEQuence:ATRigger?
Enables or disables the auto trigger function. Queries the setting status of the auto trigger function.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | OFF | ON | 0 | 1 | Enables or disables the auto trigger function. |
:TRIGger:SEQuence:ATRigger:STATe ON :TRIGger:SEQuence:ATRigger:STATe 1 :TRIGger:SEQuence:ATRigger:STATe?
Sets the delay time for the external trigger. Queries the delay time for the external trigger.
| Name | Type | Values | Description |
|---|---|---|---|
| n | integer | 1 - 2 | Selects External Trigger 1 or External Trigger 2. |
| time | real | 0 μs - 500 ms | Sets the external trigger delay time. |
:TRIGger:SEQuence:EXTernal1:DELay 0.1 :TRIGger:SEQuence:EXTernal1:DELay?
Enables or disables the external trigger delay function. The parameter n selects External Trigger 1 or 2.
| Name | Type | Values | Description |
|---|---|---|---|
| n | integer | 1 - 2 | Selects which external trigger to configure: 1 for External Trigger 1, 2 for External Trigger 2. |
| state | enum | OFF | ON | 0 | 1 | Enables or disables the external trigger delay function. |
:TRIGger:SEQuence:EXTernal1:DELay:STATe OFF|ON|0|1
Sets the trigger edge for the external trigger and queries the current trigger edge.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 2 | Selects External Trigger 1 or 2. |
| mode | enum | POSitive | NEGative | Selects the trigger edge polarity. |
:TRIGger:SEQuence:EXTernal1:SLOPe POSitive :TRIGger:SEQuence:EXTernal1:SLOPe?
Sets the number of times for acquisition after each effective trigger is completed. Queries the number of times for acquisition after each effective trigger is completed.
| Name | Type | Values | Description |
|---|---|---|---|
| number | integer | 1 | Number of acquisitions after each effective trigger is completed. |
:TRIGger[:SEQuence]:FMT:APTRigger <number> :TRIGger[:SEQuence]:FMT:APTRigger?
Sets the trigger criteria for FMT. Queries the trigger criteria for FMT.
| Name | Type | Values | Description |
|---|---|---|---|
| mode | enum | ENTer | LEAVe | INSide | OUTSide | ELEave | LENTer | Trigger criterion for FMT. |
:TRIGger:SEQuence:FMT:CRITeria ENTer :TRIGger:SEQuence:FMT:CRITeriar?
Sets the delay time for FMT. Queries the delay time for FMT.
| Name | Type | Values | Description |
|---|---|---|---|
| time | real | - | Delay time for FMT. |
Enables or disables the FMT trigger delay function. Queries the status of the FMT trigger delay function.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | OFF | ON | 0 | 1 | Bool setting for the FMT trigger delay function. |
:TRIGger:SEQuence:FMT:DELay:STATe ON :TRIGger:SEQuence:FMT:DELay:STATe 1 :TRIGger:SEQuence:FMT:DELay:STATe?
Sets the mask to be used for the current trigger. The query returns the current trigger mask.
| Name | Type | Values | Description |
|---|---|---|---|
| mask | enum | UPPer | LOWer | BOTH | Selects the mask to use for the current trigger. |
:TRIGger:SEQuence:FMT:MASK UPPer :TRIGger:SEQuence:FMT:MASK?
Deletes all the masks.
Sets the mask type that is currently viewed/edited. Queries the mask type that is currently viewed/edited.
| Name | Type | Values | Description |
|---|---|---|---|
| mode | enum | UPPer | LOWer | Mask type to view or edit. |
Sets whether the amplitudes of the mask points are coupled to the instrument reference level. Queries whether that coupling is enabled.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | ON | OFF | 1 | 0 | Enables or disables coupling of mask-point amplitudes to the reference level. |
:TRIGger:SEQuence:FMT:MASK:RELative:AMPLitude ON :TRIGger:SEQuence:FMT:MASK:RELative:AMPLitude 1 :TRIGger:SEQuence:FMT:MASK:RELative:AMPLitude?
Sets whether the frequencies of the mask points are coupled to the center frequency of the instrument. This command is only valid when FMT is selected as the trigger source in RTSA mode.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | ON | OFF | 1 | 0 | Enables or disables coupling of mask-point frequencies to the center frequency. |
:TRIGger:SEQuence:FMT:MASK:RELative:FREQuency ON :TRIGger:SEQuence:FMT:MASK:RELative:FREQuency 1 :TRIGger:SEQuence:FMT:MASK:RELative:FREQuency?
Creates a mask from a trace. The command is only valid when FMT is selected as the trigger source in RTSA mode.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 2 | Selects which mask is being edited: 1 for the upper mask, 2 for the lower mask. |
Edits the mask parameters for the specified mask. The query returns the mask parameters currently being edited.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 2 | Selects which mask is edited: upper mask for 1, lower mask for 2. |
| freq | real | 0 Hz - 6.5 GHz | Frequency point of the mask. |
| ampl | real | -1,000 dBm - 1,000 dBm | Amplitude point of the mask. |
:TRIGger:SEQuence:FMT:MASK1:DATA –80000000, –50, 80000000, –50 :TRIGger:SEQuence:FMT:MASK1:DATA?
Deletes all the masks for the selected mask index. It is only valid when FMT is selected as the trigger source in RTSA mode.
| Name | Type | Values | Description |
|---|---|---|---|
| n | integer | 1 - 2 | Selects which mask to delete: 1 for the upper mask or 2 for the lower mask. |
Clears the currently activated mask and creates a default new mask. <n> selects whether the upper mask (1) or lower mask (2) is edited.
| Name | Type | Values | Description |
|---|---|---|---|
| n | discrete | 1 - 2 | Selects the mask to edit: 1 for the upper mask, 2 for the lower mask. |
Sets the trigger holdoff time, or queries the current trigger holdoff time. This command is only valid when the trigger holdoff function is enabled.
| Name | Type | Values | Description |
|---|---|---|---|
| time | real | 100 us - 500 ms | Trigger holdoff time. |
:TRIGger:SEQuence:HOLDoff 0.1 :TRIGger:SEQuence:HOLDoff?
Enables or disables the trigger holdoff function. Queries the status of the trigger holdoff function.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | OFF | ON | 0 | 1 | Enables or disables the trigger holdoff function. |
:TRIGger:SEQuence:HOLDoff:STATe ON or :TRIGger:SEQuence:HOLDoff:STATe 1 :TRIGger:SEQuence:HOLDoff:STATe?
Sets the trigger source. Queries the trigger source.
| Name | Type | Values | Description |
|---|---|---|---|
| source | enum | EXTernal1 | EXTernal2 | IMMediate | VIDeo | FMT | POWer | Trigger source selection. |
:TRIGger:SEQuence:SOURce IMMediate :TRIGger:SEQuence:SOURce?
Sets the delay time for the video trigger. Queries the delay time for the video trigger.
| Name | Type | Values | Description |
|---|---|---|---|
| time | real | 0 μs - 500 ms | Consecutive real number specifying the video trigger delay time. |
:TRIGger:SEQuence:VIDeo:DELay 0.1 :TRIGger:SEQuence:VIDeo:DELay?
Enables or disables the video trigger delay function. Queries the status of the video trigger delay function.
| Name | Type | Values | Description |
|---|---|---|---|
| state | enum | OFF | ON | 0 | 1 | Enables or disables the video trigger delay function. |
:TRIGger:SEQuence:VIDeo:DELay:STATe ON or :TRIGger:SEQuence:VIDeo:DELay:STATe 1 :TRIGger:SEQuence:VIDeo:DELay:STATe?
Sets the trigger level of the video trigger. Queries the trigger level of the video trigger.
| Name | Type | Values | Description |
|---|---|---|---|
| ampl | real | -140 - 30 | Trigger level amplitude. |
:TRIGger:SEQuence:VIDeo:LEVel 10 :TRIGger:SEQuence:VIDeo:LEVel?
Sets the polarity of the video trigger. Queries the polarity of the video trigger.
| Name | Type | Values | Description |
|---|---|---|---|
| slope | enum | POSitive | NEGative | Polarity of the video trigger. |
:TRIGger:SEQuence:VIDeo:SLOPe POSitive :TRIGger:SEQuence:VIDeo:SLOPe?
Sets the interface type of External Trigger 2. Queries the interface type of External Trigger 2.
| Name | Type | Values | Description |
|---|---|---|---|
| mode | enum | IN | OUT | 0 | 1 | Interface type for External Trigger 2. |
:TRIGger2:MODE IN :TRIGger2:MODE?
Sets or queries the unit of the Y axis.
| Name | Type | Values | Description |
|---|---|---|---|
| power unit | enum | DBM | DBMV | DBUV | V | W | Selects the Y-axis amplitude unit. |
:UNIT:POWer W :UNIT:POWer?