Rohde & Schwarz

/

Rohde & Schwarz ROHDE-SCHWARZ-MXO34

SCPI COMMAND REFERENCE

Rohde & Schwarz ROHDE-SCHWARZ-MXO34

Rohde & Schwarz ROHDE-SCHWARZ-MXO34

The Rohde & Schwarz ROHDE-SCHWARZ-MXO34 is Rohde & Schwarz's oscilloscope in the MXO 3 family. Its SCPI remote-control interface exposes 3381 commands across 49 subsystems - 2353 settable and 1028 queries - and every command is documented here with its parameters, accepted values, return format, and a copy-paste PyVISA example for automating test and measurement.

3,381

Commands

2,353

Settable

1,028

Queries

49

Subsystems

SCPI

Verified

2026-07-16

All commands

Enables or disables the efficiency display result for the selected output waveform. The manual points to related power-efficiency display commands for output power, total efficiency, and total output power.

NameTypeValuesDescription
enableoptenumENABleTurns the efficiency display result on or off.

Enables or disables the total efficiency result display for channel <n> of power measurement <m>. This command is cross-referenced as part of the power efficiency display result commands.

NameTypeValuesDescription
stateoptenumENABleEnables the total efficiency result display.

Deprecated legacy command for setting the harmonic display frequency start value in the power analysis display. It is replaced by `POWer<m>:HARMonics:DISPlay:HARMonics:STARt[:VALue]` since FW 2.9.

NameTypeValuesDescription
valueoptreal-Start frequency value.

Sets the display frequency stop value for harmonics in the power analysis display. This legacy command was replaced by POWer<m>:HARMonics:DISPlay:HARMonics:STOP[:VALue] on page 1164.

NameTypeValuesDescription
valuereal-Stop frequency value.

Executes the power skew function for the specified power measurement. The optional EXECute node may be used to explicitly trigger the operation.

Controls whether a displayed result output efficiency trace is enabled. The command belongs to the Power Efficiency display result output node.

NameTypeValuesDescription
stateoptenumENABle | DISableEnables or disables the efficiency display result output.

Enables or disables display of the total efficiency result for a power measurement. This command is part of the efficiency result display subtree for total results.

NameTypeValuesDescription
stateenumON | OFFTurns the total efficiency result display on or off.

Turns the total output power display result on or off for the specified total result item. The command is part of the power efficiency display result subtree.

NameTypeValuesDescription
ninteger-Total result index.
stateoptenumENABleEnables display of the result.

Enables or disables display of the total waveform output power for the efficiency display. The command belongs to the Power Efficiency waveform total output power display settings.

NameTypeValuesDescription
stateoptenumON | OFFEnables or disables the waveform total output power display.

Sets the power efficiency gate. The command is listed in the power efficiency subsystem command index.

Sets the current source used for efficiency input channel <n> of power measurement <m>.

NameTypeValuesDescription
sourceenum-Current source selection for the efficiency input.

Returns the power data values for the specified efficiency input. The command is listed as a query and appears with an optional [:VALues] node.

block - Power data values for the specified efficiency input.

Sets the vertical offset for the input power waveform used in the power efficiency measurement. This command belongs to the Power Efficiency input power configuration subtree.

NameTypeValuesDescription
offsetreal-Vertical offset for the input power waveform.

Sets the vertical scale value for the specified power efficiency input waveform. The scale is used for the input power trace under POWer<m>:EFFiciency:INPut<n>:POWer:VERTical:SCALe.

NameTypeValuesDescription
valueoptreal-Vertical scale value.

Sets the mode used for the relative reference level of the selected power efficiency input. This command is part of the Relative reference-level subtree under :POWer<m>:EFFiciency:INPut<n>:REFLevel<rl>.

NameTypeValuesDescription
modeenum-Relative reference-level mode.

Sets the voltage source for an input efficiency measurement. The command is part of the input configuration for power efficiency measurements.

NameTypeValuesDescription
sourceoptenum-Voltage source selection.

Sets or queries the type of power-efficiency input for the selected module and input channel.

NameTypeValuesDescription
minteger1 - 4Module index.
ninteger1 - 4Input index.
typeoptenumAC | DCSpecifies the input type.

Returns the number of output channels for the selected power efficiency measurement. The command is a query and yields the ONUMber value for the current measurement setup.

integer - The number of output channels used by the power efficiency measurement.

Returns the power data values for the selected output power efficiency result. The optional [:VALues] node is part of the query syntax as listed in the command index.

string - Returns the data values for the selected output power efficiency result.

Sets the lower limit level for an absolute reference level in the power efficiency output measurement path. This command belongs to the output reference-level absolute settings.

NameTypeValuesDescription
levelreal-Lower limit level for the absolute reference level.

Sets the relative hysteresis for the selected output power efficiency reference level. This command belongs to the relative reference-level controls under :POWer<m>:EFFiciency:OUTPut<n>:REFLevel<rl>.:

NameTypeValuesDescription
hysteresisreal-Relative hysteresis value.

Sets the output type for the specified power efficiency output. The command is part of the power efficiency output configuration subsystem.

Returns the average input power result for the selected power efficiency measurement. The command is a query and yields the measured value for the input power average.

real - Average input power result.

Returns the peak negative input power result for the selected power efficiency measurement. The command is a query and yields the value for the current input power result.

real - Peak negative input power result.

Returns the standard deviation of the input power result for the selected power efficiency measurement. This is a query and yields a single numeric value.

real - Standard deviation of the input power result.

Returns the peak positive efficiency value for the total efficiency result. This is a query command and yields the current peak value for the selected measurement group.

real - Peak positive efficiency value of the total efficiency result.

Returns the actual or average total efficiency result for the selected power-efficiency measurement. The optional [:ACTual] node is shown in the syntax and does not change the return type in the slice.

real - Returns the total efficiency result as an actual or averaged value.

Returns the average total output power efficiency result for the selected power measurement. The command is listed as a query and has no additional syntax details on this page.

real - Average total output power efficiency result.

Returns the peak output power result for the total efficiency measurement. This command is part of the efficiency results queries.

real - Peak output power value for the total efficiency result.

Returns the peak-to-peak total output power efficiency result for the selected measurement channel. This is a read-only query result under the POWer<m>:EFFiciency:RESult:TOTal:OPOWer subtree.

real - The peak-to-peak total output power efficiency value.

Returns the standard deviation of the total output power efficiency result for the selected power efficiency measurement. This is a query command.

real - The standard deviation of the total OPOWer efficiency result.

Returns the waveform count for the total output power efficiency result. This command is part of the POWer<m>:EFFiciency:RESult:TOTal:OPOWer result family.

integer - The number of waveforms used for the total output power efficiency calculation.

Returns the total output power result for the selected power efficiency measurement. The optional :ACTual node is documented in the syntax line.

real - The total output power result for the selected measurement.

Returns the waveform count for the efficiency statistics subsystem. It is a query on the POWer<m>:EFFiciency:STATistics subsystem.

integer - The number of waveforms counted in the efficiency statistics measurement.

Enables or disables power efficiency statistics for the selected measurement channel. The command is part of the power efficiency subsystem.

NameTypeValuesDescription
stateenumON | OFFTurns statistics enable on or off.

Returns whether harmonics are available for the selected power measurement. This is a query command.

boolean - Indicates whether harmonics are available.

Sets the harmonic stop value used for the displayed harmonics range. The value may be omitted in the short form as indicated by the optional ":VALue" node.

NameTypeValuesDescription
valueoptreal-Stop value for the harmonics display range.

Enables the harmonics frequency function for the selected power measurement. This command is listed under the power harmonics subsystem.

Selects MIL frequency handling for the harmonics measurement. The command is listed as a setter with no query form shown in this slice.

Sets the harmonic frequency RTCA mode for the selected power measurement. The command is listed as a setter and has no documented query form on this page slice.

Sets or queries the power harmonics power factor mode. The optional mode node is part of the command mnemonic.

NameTypeValuesDescription
modeoptenum-Power factor mode.

Returns the header information for harmonic power data. This is the query form of the harmonics power data header command.

string - Header information for harmonic power data.

Sets the vertical offset for the harmonics power display. The command applies to the selected power measurement instance <m>.

NameTypeValuesDescription
offsetreal-Vertical offset value for the power harmonics display.

Sets the vertical scale for the harmonic power display. The optional value form is documented as the same command with optional `[:VALue]` suffix.

NameTypeValuesDescription
scalereal-Vertical scale value for the harmonic power display.

Sets the lower limit level for the absolute reference level of the harmonics measurement. This command belongs to the harmonics reference level settings under the power measurements subsystem.

Sets the absolute maximum level for the selected harmonics reference level. The reference level is addressed by the <rl> suffix.

NameTypeValuesDescription
levelreal-Absolute maximum level for the selected harmonics reference level.

Sets the lower relative reference level for harmonic measurements for the selected math channel. The command is part of the harmonic reference-level settings.

NameTypeValuesDescription
rlreal-Relative reference level value.

Sets the middle reference level for relative harmonic measurements. The level is specified as a value under the harmonic reference level path.

NameTypeValuesDescription
rlreal-Reference level index or identifier.

Sets the relative reference mode for harmonics reference levels. The command is part of the harmonics reference level subsystem.

NameTypeValuesDescription
modeenum-Relative reference mode.

Sets the upper relative reference level for the harmonics measurement in the selected power-measurement group.

NameTypeValuesDescription
rlreal-Relative reference level index or value placeholder as used in the command mnemonic.

Returns the maximum RMS magnitude of the specified harmonic result. This is a query command for harmonic result magnitude in RMS units.

NameTypeValuesDescription
moptinteger1Harmonic measurement instance.
noptinteger1Harmonic result index.
real - Maximum RMS magnitude of the selected harmonic result.

Returns the minimum RMS magnitude result for the selected harmonic result. This is a query form of the harmonic result magnitude command.

real - The minimum RMS magnitude result for the selected harmonic number.

Returns the frequency of harmonic result <n>. The optional [:FREQuency] suffix is shown in the command listing, but no additional parameter syntax is documented in this slice.

real - The harmonic result frequency for the selected result <n>.

Returns the revision of the harmonics power measurement subsystem.

string - The harmonics power measurement revision.

Sets or queries the reactive power calculation mode for the harmonics measurement. The optional mode node selects the mode used by the harmonics power routine.

NameTypeValuesDescription
modeoptenum-Selects the harmonics reactive power mode.
enum - Returns the current harmonics reactive power mode.

Selects the harmonics standard used for the power harmonics measurement setup. The command is part of the power harmonics subsystem for channel <m>.

Sets the absolute value for the AC input of the power on/off measurement. The command belongs to the power measurement subsystem for source m.

NameTypeValuesDescription
valuereal-Absolute AC input value.

Sets the DC absolute input power measurement value for the selected power measurement. The optional value node is part of the command syntax.

NameTypeValuesDescription
valueoptreal-Absolute DC input power value.

Sets or queries the input type for the power measurement on the selected meter. The command belongs to the ON/OFF power measurement subsystem.

Sets or queries the ON/OFF number setting for the power on/off measurement subsystem. The command is part of the POWer<m>:ONOFf group.

Controls whether the display result for a specific output is enabled.

NameTypeValuesDescription
moptinteger-Power measurement index.
noptinteger-Output number.
enableoptboolean-Enables or disables display of the result.

Enables or disables the display of the calculated real-line for the selected power output result. This command belongs to the power ON/OFF output display subsystem.

NameTypeValuesDescription
noptinteger-Output index.
stateoptenumENABle | DISableTurns the real-line display on or off.

Sets or queries the output source for the selected power-on/off function output. The optional source node is part of the command syntax.

NameTypeValuesDescription
moptinteger1Measurement index.
noptinteger1Output index.
sourceopt--Source selector.

Returns the current time result for a selected power on/off measurement result. The optional ACTual suffix is supported in the query form.

real - Returns the time result value for the selected result item.

Selects the on/off function type for a power measurement. The command belongs to the POWer<m>:ONOFf subsystem and supports an optional type qualifier.

NameTypeValuesDescription
typeoptenum-Optional function type selector.

Controls whether the active power display is shown in the power quality display group. The command is listed as a settable display command with no additional syntax shown on this page.

Turns on display of the power phase measurement for the specified measurement index. This command is part of the Power Quality display selection group.

Enables or disables display of the power waveform under the quality display branch. The command is listed in the command index only; no syntax or parameter details are shown in this slice.

Sets or queries the voltage crest factor display setting for power quality measurements. The page slice only lists the command in the command index and does not provide syntax details or a return description.

Sets or queries the display of voltage frequency for the selected power quality measurement. The command is listed without additional syntax or return details in this slice.

Displays the peak voltage measurement in the power quality display. This command is part of the power quality display measurement set for channel <m>.

Sets or queries the user-defined power-quality frequency. The command is part of the POWer<m>:QUALity subsystem and is related to POWer<m>:QUALity:FREQuency[:VALue].

Sets or queries the power quality frequency value for measurement m. The optional value node is shown in the syntax; this page slice only lists the command in the command index and does not further describe it.

Sets the power quality gate setting for the selected power measurement channel.

Sets or queries the vertical scale value for the power quality measurement. The value is part of the power quality vertical scaling controls.

NameTypeValuesDescription
valueoptreal-Vertical scale value.

Sets the absolute hysteresis for the reference level of a power quality measurement. The command is part of the reference-level absolute settings under POWer<m>:QUALity:REFLevel<rl>:ABSolute.

Sets the absolute middle level for the selected power quality reference level. The command belongs to the `POWer<m>:QUALity:REFLevel<rl>:ABSolute` group.

NameTypeValuesDescription
levelreal-Absolute middle level value.

Sets the mode used for the reference level of a power quality measurement. The command belongs to the reference level settings for the selected measurement  m  and reference level  rl .

Returns the present N-peak crest factor result for the current power quality measurement. This command is part of the power quality result tree under current measurements.

real - The current N-peak crest factor result.

Returns the present peak-to-peak value for the current crest factor result in the power quality measurement. The command is a query and has no parameters.

real - Present peak-to-peak value of the current crest factor result.

Returns the current crest factor standard deviation result for the selected power measurement. The slice only shows the command in the list of commands and does not provide further detail.

returns - Returns the current crest factor standard deviation result for the selected power measurement.

Returns the actual current crest factor result for the selected power quality measurement. The manual lists this query as an optional ":ACTual" form under the current crest-factor result family.

real - Returns the actual crest factor value.

Returns the average value of the current peak result in the power quality measurement. This is a query command for the current measurement result subgroup.

real - Returns the average value of the current peak result.

Returns the event count for the current peak power quality result. This is a query in the CURRent:PEAK result subtree.

integer - The event count for the current peak power quality result.

Returns the negative peak current result for the current power-quality measurement. This is a query form in the PEAK results group.

real - The negative peak current result.

Returns the current peak RMS result for the power quality measurement on the selected module. The command is a query and reports the present value of the peak RMS result.

real - Current peak RMS result.

Returns the standard deviation of the present current peak result for power quality measurements. This is a query command.

real - Standard deviation of the current peak result.

Returns the waveform count for the current peak power quality result. The command is part of the POWer<m>:QUALity:RESult:CURRent:PEAK result group.

integer - Waveform count for the current peak power quality result.

Returns the average RMS current result for the current power quality measurement. This is a query command in the power quality results tree.

real - Returns the average RMS current result.

Returns the event count for the current RMS power quality result. This is a query under the current power-quality result hierarchy.

integer - Returns the event count for the current RMS result.

Returns the present RMS negative peak result for the current power quality measurement. This is a query under the current-result tree.

real - Present RMS negative peak result.

Returns the current RMS standard deviation for the power quality result. The command is part of the POWer<m>:QUALity:RESult:CURRent:RMS result group.

real - Current RMS standard deviation.

Returns the present current RMS value for the selected measurement. The optional ACTual node is shown in the command listing but no separate behavior is described on this page slice.

real - Present current RMS value.

Returns the average active power result for the selected measurement m. The slice only shows the command listing; no further return details are provided here.

returns - Returns the average active power result for the selected measurement m.

Returns the current peak negative power result for the active power measurement. This query is part of the Power Quality result group.

real - Current peak negative power result.

Returns the waveform count for the active power result. The slice only lists the query in the command index and does not provide additional syntax or explanation.

integer - Waveform count for the active power result.

Returns the actual active power result for the selected measurement channel. The command is documented as an optional "[:ACTual]" form under the active power result subtree.

real - Actual active power result.

Returns the average apparent power result for the selected power measurement. This is a query for the POWer<m>:QUALity:RESult:POWer:APParent function.

returns - Returns the average apparent power result.

Returns the event count for the apparent power result measurement. This command is part of the Power Quality Result readout tree.

integer - Event count for the apparent power result measurement.

Returns the peak apparent power result for the selected measurement. The command is listed under the power quality results tree for apparent power peak values.

real - Returns the peak apparent power result.

Returns the standard deviation of the apparent power result for the selected measurement memory. The command is part of the power quality results queries.

real - Standard deviation of the apparent power result.

Returns the apparent power result for the selected measurement channel, with an optional Actual form in the command tree. The slice only lists the command in the index and does not provide further details.

returns

Returns the event count for the power factor result. The command is part of the POWer< m >:QUALity:RESult:POWer:PFACtor measurement group.

integer - The event count for the power factor result.

Returns the RMS value of the power factor result for the selected measurement. This is one of the power quality result queries under :POWer<m>:QUALity:RESult:POWer:PFACtor.

real - The RMS value of the power factor result.

Returns the waveform count for the power factor result. The command is part of the Power Quality Results hierarchy for the selected measurement channel <m>.

integer - Returns the waveform count for the power factor result.

Returns the actual power factor result for the selected measurement path. The command is a query for the POWer<m>:QUALity:RESult:POWer:PFACtor branch.

real - Returns the actual power factor result.

Returns the event count for the power phase result. The slice only lists the query in the command index and does not provide a fuller functional description.

integer - Event count for the power phase result.

Returns the negative peak of the power phase result for the selected measurement. This is a query under the power quality results tree.

real - Negative peak value of the power phase result.

Returns the event count for the reactive power result. This query is part of the POWer<m>:QUALity:RESult:POWer:REACtive measurement results tree.

integer - Event count for the reactive power result.

Returns the positive peak result for the reactive power measurement. This is a query in the power quality results subsystem.

real - Positive peak value of the reactive power result.

Returns the actual or current reactive power quality result for the selected power measurement. The optional [:ACTual] node is part of the query syntax.

real - Returns the reactive power result value.

Returns the negative peak value of the voltage crest factor result for the selected measurement. This is a query command and yields the current result value.

real - Returns the negative peak voltage crest factor result value.

Returns the peak value of the voltage crest factor result for the selected measurement channel. This is a query command in the Power quality results group.

real - Returns the peak voltage crest factor result.

Returns the RMS value of the voltage crest factor result for the selected power measurement. This is a query in the POWer qualification results branch.

real - Returns the RMS voltage crest factor result.

Returns the RMS value of the voltage frequency result for the selected power measurement. This is a query command in the POWer<m>:QUALity:RESult:VOLTage:FREQuency subtree.

real - The RMS value of the voltage frequency result.

Returns the waveform count for the voltage frequency quality result. This is a query command and yields the number of waveforms used in the measurement.

integer - Number of waveforms used for the voltage frequency quality result.

Returns the actual frequency result for the specified voltage quality measurement. The optional :ACTual node is shown in the syntax line.

real - The actual voltage frequency result for the selected measurement.

Returns the average value of the voltage peak result for the selected power quality measurement. The command is part of the POWER<m>:QUALity:RESult:VOLTage:PEAK subtree.

real - Average voltage peak result value.

Returns the event count for the peak voltage quality result. This is a query command for the POWer<m>:QUALity:RESult:VOLTage:PEAK measurement.

integer - Event count for the voltage peak quality result.

Returns the peak-negative peak voltage result for the selected power quality measurement. This is a query command for the voltage peak measurement branch.

real - Returns the peak-negative peak voltage result.

Returns the present positive peak value of the voltage result for the selected measurement. The command belongs to the Power Quality results group.

real - Present positive peak voltage result value.

Returns the standard deviation of the peak voltage results for the selected power quality measurement. This is a query command for the voltage peak result statistics.

real - Returns the standard deviation of the peak voltage result.

Returns the waveform count for the peak voltage result of the selected power quality measurement. This is a query under the power quality result tree.

integer - The waveform count for the peak voltage result.

Returns the present actual peak voltage result for the selected power quality measurement. The optional actual-node suffix is shown in the syntax.

real - Returns the present peak voltage result.

Returns the average RMS voltage result for the selected power quality measurement. This is a query command and does not take parameters.

real - Returns the average RMS voltage result.

Returns the standard deviation of the voltage RMS result for the selected power quality measurement. This is a query command.

real - Standard deviation of the voltage RMS result.

Returns the present actual RMS voltage result for the selected power quality measurement. The optional `ACTual` node is part of the documented query form.

real - Present actual RMS voltage result.

Sets the current source value for the power quality measurement. The command belongs to the POWer<m>:QUALity source settings group.

NameTypeValuesDescription
currentreal-Current source value.

Sets the source voltage used for power quality measurements. The command belongs to the power quality source subtree and the voltage node is optional in the syntax list.

Returns the number of waveforms included in the power quality statistics. This is a query form of the statistics waveform count command.

integer - The number of waveforms included in the power quality statistics.

Enables or disables the power quality statistics function for the selected measurement. The corresponding reset command is :POWer<m>:QUALity:STATistics:RESet.

NameTypeValuesDescription
stateoptenumENABleEnables power quality statistics.

Applies the current SOA limit settings. This is a setter command with no documented return value.

Sets the maximum current limit used by the SOA limit function for the specified power source.

NameTypeValuesDescription
moptinteger-Power source index.
currentreal-Maximum current limit.

Sets the maximum power limit for the selected safe operating area (SOA) limit. This command belongs to the SOA limit group for the selected power source instance.

NameTypeValuesDescription
moptinteger1 - 4Selects the power source instance.
pmaxreal-Maximum power limit.

Sets the maximum voltage limit for the SOA limit definition of the selected power measurement. The command belongs to the `POWer<m>:SOA:LIMit` group.

Sets the name used for an imported SOA multitest. The command is part of the `POWer<m>:SOA:MTESt<n>:IMEXport` subsystem.

NameTypeValuesDescription
namestring-Name to assign for the imported multitest.

Adds a segment to the SOA test segment list for the selected power measurement.

NameTypeValuesDescription
segmentinteger1Segment index to add.

Returns the number of points in the specified segment of the SOA test waveform. This is a query command.

NameTypeValuesDescription
minteger-Waveform channel or module selector.
ninteger-SOA test index.
ointeger-Segment index.
pinteger-Point index.
integer - Number of points in the specified segment.

Sets the X coordinate of a point within a segment of a SOA test waveform. The command belongs to the power meter SOA test segment point configuration commands.

NameTypeValuesDescription
minteger-Power meter index.
ninteger-SOA test index.
ointeger-Segment index.
pinteger-Point index.
xreal-X coordinate value for the point.

Returns the number of power waveforms in the specified SOA result set. The result set is selected by the <n> index.

integer - Number of power waveforms in the selected SOA result set.

Returns the switching power result frequency rate for the selected measurement and result index.

real - Returns the frequency rate for the selected SOA result.

Returns the result data for the specified SOA result entry, with an optional additional result selector. The slice does not provide details of the returned format beyond the command listing.

returns

Sets the SOA source current for the selected power channel. The command is listed as part of the SOA source settings; no parameter details are shown in this slice.

Sets the voltage source used for SOA measurements. The command belongs to the POWer<m>:SOA:SOURce subsystem.

Sets the display color used for conduction-related switching measurements. The slice lists the command name only and does not provide syntax, parameter details, or return information.

Enables or disables the display of conduction information for switching measurements. This command is part of the switching display subsystem.

NameTypeValuesDescription
stateoptenumENABle | DISableTurns the conduction display on or off.

Sets or enables the label display for switching power measurements on channel <m>.

NameTypeValuesDescription
stateoptenumENABle | DISableEnables or disables the label display.

Enables or disables the display of the switching nonconduction trace. The command belongs to the switching display subsystem for power measurements.

NameTypeValuesDescription
stateoptenumON | OFFTurns the nonconduction display on or off.

Controls whether the power waveform is displayed in the switching display. The optional ENABle form is listed in the command index for this function.

NameTypeValuesDescription
stateenum-Enable or disable the power waveform display.

Enables or disables the TOFF display item for the selected power measurement channel. The optional ":ENABle" node is part of the command mnemonic in the manual listing.

Sets the display color for the switching turn-on waveform. This command belongs to the switching display subsystem for power measurement.

Controls whether the TON display for switching power measurements is enabled. This command is part of the switching display settings for power measurements.

NameTypeValuesDescription
enableoptenumON | OFFEnables or disables the TON display.

Sets the switching display type for the selected power measurement. The command is part of the power switching display subsystem.

Returns the values of the power switching data. The optional ":VALues" node is part of the command mnemonic.

string - Returns the power switching data values.

Sets the start value for the conduction region of the switching-power measurement. The optional node indicates this command may also be used without the final START keyword.

Sets the stop time for the N-conduction region in the selected power measurement. This command belongs to the switching region timing commands.

NameTypeValuesDescription
stopreal-Stop time for the N-conduction region.

Sets or queries the start point of the non-conduction region for the selected power measurement channel. The optional ":STARt" node is part of the command name in the manual listing.

Sets or queries the start time of the TON region for the selected power switch measurement. The optional start node may be omitted in the command syntax.

NameTypeValuesDescription
moptinteger1 - 4Measurement index selector.

Sets the start of the total switching region for the selected power source. The start point is optional in the syntax line.

NameTypeValuesDescription
startoptreal-Start of the total switching region.

Returns the number of peak energy values in the conduction result. The query is part of the switching result tree for power measurements.

integer - Returns the number of peak energy values in the conduction result.

Returns the actual conduction energy result for the selected measurement. The command applies to power switching results and uses the optional ACTual node in the mnemonic.

real - The actual conduction energy value.

Returns the event count for the conduction power result in the switching analysis. This is a query command with no parameters documented in the slice.

integer - Returns the event count for the conduction power result.

Returns the RMS power result for the conduction interval. This is a query under the switching result measurements for the selected channel <m>.

real - RMS power result for the conduction interval.

Returns the standard deviation of the conduction power result for the selected switching event measurement. The command is a query and yields the value for the specified measurement channel <m>.

real - Standard deviation of the conduction power result.

Returns the average negative-conduction energy result for the selected switching measurement memory. The command is a query and yields the value for :POWer<m>:SWITching:RESult:NCONduction:ENERgy:AVERage?.

real - Average negative-conduction energy result.

Returns the number of negative-conduction energy peaks for the selected measurement. This is a result query under the switching analysis result tree.

integer - Returns the count of negative-conduction energy peaks.

Returns the RMS value of the nonconduction energy result for switching measurements. This is a query command with no parameters documented in the slice.

real - The RMS value of the nonconduction energy result.

Returns the number of negative-conduction power peaks for the selected measurement memory. The command is a query and yields the count for the NCONduction power result.

integer - Number of negative-conduction power peaks.

Returns the waveform count for the nonconduction power result of the switching measurement. The slice lists this as a query only and does not provide any additional format details.

integer - Waveform count for the nonconduction power result.

Returns the peak negative energy for the turn-off interval. This is a read-only measurement query in the switching result subsystem.

real - Peak negative energy for the turn-off interval.

Returns the RMS value for the TOFF energy result of the selected power measurement. This is a query command and takes no parameters.

real - Returns the RMS value for the TOFF energy result.

Returns the waveform count for the TOFF energy result of the switching analysis. This is a query command and yields the count value for that result item.

integer - Returns the waveform count for the TOFF energy result.

Returns the event count for the TOFF power result. This is a query command under the switching result measurement tree.

integer - Returns the number of events for the TOFF power result.

Returns the peak power result for the TOFF switching interval of measurement m. This command is listed in the switching results command index.

real - Returns the peak power result for the TOFF interval.

Returns the RMS power result for the TOFF switching result measurement. No parameters are documented on this page slice.

real - Returns the RMS power value for the TOFF switching result.

Returns the actual turn-off power result for the selected measurement. The command is listed as part of the switching result queries for TOFF power.

real - Returns the actual turn-off power result.

Returns the number of negative peaks for the turning-on energy result. This is a query command and does not take parameters.

integer - Number of negative peaks for the turning-on energy result.

Returns the RMS result for the selected switching energy measurement. The command is a query for the energy RMS measurement under the POWer<m>:SWITching:RESult:TON hierarchy.

real - Returns the RMS value of the selected switching energy measurement.

Returns the standard deviation of the TON energy result for the selected power switching measurement. This is a query command.

real - Standard deviation of the TON energy result.

Returns the number of waveforms counted for the selected power-switching energy result. This is a query command for the TON:ENERgy result statistics path.

integer - Returns the waveform count for the selected TON:ENERgy result.

Returns the actual energy result for the selected source in the switching results tree. The optional ":ACTual" node is part of the query mnemonic shown in the list of commands.

real - Returns the actual energy value.

Returns the event count for switching power results in the TON:POWer branch. The command is a query and returns the current EVTCount value.

integer - Returns the event count for switching power results.

Returns the peak positive power measured over the switching result TON power interval.

real - Peak positive power measured over the switching result TON power interval.

Returns the standard deviation of the power result for the selected switching measurement window. This is a query command and returns a numeric value.

real - Standard deviation of the power result for the selected switching measurement window.

Returns the WFMCount result for the power switching TON power measurement. This is a query command under :POWer<m>:SWITching:RESult:TON:POWer.

integer - The WFMCount result for the selected power switching TON measurement.

Returns the RMS value of the total energy result for the selected power measurement. This is a query for the :POWer<m>:SWITching:RESult:TOTal:ENERgy subsystem.

real - RMS value of the total energy result.

Returns the standard deviation of the total energy measurement for the switching result. This is a query command.

real - Returns the standard deviation of the total energy measurement.

Returns the actual or total energy result for the selected switching measurement. The optional ":ACTual" node is part of the query mnemonic in the manual listing.

real - Returns the actual or total energy result value.

Returns the total negative peak value for the selected power switching result. This is a query command for the total-power branch of the switching results hierarchy.

real - The total negative peak power value.

Returns the total peak power result for the specified switching measurement. The measurement index is the optional channel-like placeholder <m>.

real - The total peak power result.

Returns the RMS total power switching result. This is a query command and yields the current value of the total power RMS measurement for the selected source <m>.

real - The RMS total power switching result.

Returns the total power actual result for the switching measurement. The optional :ACTual node may be included in the query form.

real - The total power actual value for the switching result.

Sets the source selection for power switching measurements. The optional voltage suffix indicates the command also applies to voltage source selection.

Returns the number of waveforms used in the switching statistics measurement.

integer - The number of waveforms used in the switching statistics measurement.
:POWer<m>:TYPEset / query

Sets the type of the selected power measurement. This command is part of the power measurement subsystem for measurement m.

NameTypeValuesDescription
typeenum-Power measurement type.

Enables or disables power switching measurement for the specified math channel <m>.

NameTypeValuesDescription
stateenumON | OFFTurns the power switching measurement on or off.

Sets the comparison operator used for the data setup of the selected Manchester frame field. The data setup consists of a comparison condition and one or two data patterns.

NameTypeValuesDescription
frinteger-Frame index.
flinteger-Field index.
operatorenum-Comparison operator for the selected field.

Sets the trigger level for the specified event and source (channel).

NameTypeValuesDescription
levelreal-Trigger level for the specified event and source (channel).
valueoptreal-Optional value suffix for the trigger level.

Selects the trigger type for the specified trigger event. In a trigger sequence, the type is set for each condition.

NameTypeValuesDescription
typeenumEDGE | GLITch | WIDTh | RUNT | WINDow | TIMeout | INTerval | SLEWrate | ANEDge | SETHold | STATe | PATTernTrigger type to use for the event.

Selects whether triggering is based on a single event or on a series of events.

NameTypeValuesDescription
ClassenumSINGle | SEQuenceTrigger mode class.

Selects the source of the trigger signal for the selected trigger event. The available sources depend on the trigger sequence setting; for a single event, all inputs can be used, while in a trigger sequence only analog channels can be used for A, B, and R events.

NameTypeValuesDescription
evinteger1 - 3Selects the trigger event: 1 = A-trigger, 2 = B-trigger, 3 = reset event.
sourceenumC1 | C2 | ... | C8 | EXTernanalog | LINE | D0 | D1 | D2 | ... | D14 | D15Sets the trigger source for the selected event.

Sets the trigger level automatically to 0.5 × (MaxPeak − MinPeak). In a trigger sequence, it affects all active events of the sequence (A, B, and R event).

Selects the type of the trigger sequence.

NameTypeValuesDescription
TypeenumAONLy | ABR | AZ | ASBSelects the trigger sequence type.

Sets the number of B-trigger conditions to be fulfilled after an A-trigger. The last B-trigger causes the trigger event, and the waiting time for B-triggers can be restricted with a reset condition.

NameTypeValuesDescription
<Events>integer1 - 3Number of B-trigger conditions to be fulfilled after an A-trigger.

Sets the time that the instrument waits after an A-trigger until it recognizes B-triggers.

NameTypeValuesDescription
Delayreal0 - 50Delay time after an A-trigger before B-triggers are recognized.

Enables or disables reset-event handling for the trigger sequence. If enabled, the sequence is restarted by the R-trigger condition if the specified number of B-triggers does not occur before the R-trigger conditions are fulfilled.

NameTypeValuesDescription
stateenumOFF | ONTurns reset-event handling off or on.

Sets the time the instrument waits for the specified number of B-events before the sequence is restarted with the A-trigger.

NameTypeValuesDescription
ResetTimeoutreal0 - 50Time to wait before the sequence is restarted.

Sets the edge direction for the trigger.

NameTypeValuesDescription
SlopeenumPOSitive | NEGative | EITHerEdge direction for the trigger.

Enables or disables reset-by-time behavior for the specified sequence event. If set to ON, the instrument waits for the time defined by TRIGger:MEVents:SEQuence<se>:RESet:TIMeout:TIME and restarts the sequence with the A-event if no trigger occurs during that time.

NameTypeValuesDescription
StateenumOFF | ONTurns the reset-by-time enable state off or on.

Sets the connection of the external trigger signal, including input impedance and termination. The coupling determines what part of the signal is used for triggering.

NameTypeValuesDescription
CouplingenumDC | DCLimit | ACSelects the coupling mode for the external trigger signal.

Frequencies below the cutoff frequency are rejected, while higher frequencies pass the filter.

NameTypeValuesDescription
cutoff frequencyreal-Cutoff frequency for the highpass filter.

Selects the filter mode for the external trigger signal.

NameTypeValuesDescription
FilterenumOFF | LFReject | RFRejectFilter mode for the external trigger signal.

Sets the trigger level for the external trigger source.

NameTypeValuesDescription
ExtTrigLevreal-5 - 5Trigger level for the external trigger source.

Sets the low-pass trigger cutoff frequency for asynchronous trigger filtering. Frequencies higher than the cutoff frequency are rejected, and lower frequencies pass the filter.

NameTypeValuesDescription
cutoff frequencyenumKHZ50 | MHZ50Selects the low-pass cutoff frequency.

Enables automatic hysteresis on the trigger level to avoid unwanted trigger events caused by noise.

NameTypeValuesDescription
NoiseRejectenumOFF | ONTurns noise reject off or on.

Sets the polarity of a pulse, which is the direction of the first pulse slope.

NameTypeValuesDescription
polarityenum-Polarity of the pulse.

Selects whether glitches shorter or longer than the specified width are identified. The width is set by TRIGger:EVENt<ev>:GLITch:WIDTh.

NameTypeValuesDescription
rangeModeenumSHORter | LONGerSelects glitchs shorter or longer than the specified width.

Sets the length of a glitch. The instrument triggers on pulses shorter or longer than this value, depending on the range mode set with TRIGger:EVENt<ev>:GLITch:RANGe.

NameTypeValuesDescription
widthreal1E-10 - 10000Length of the glitch to detect.

Defines a range around the width value.

NameTypeValuesDescription
WidthDeltareal0 - 432Range around the width value.

Sets the polarity of a pulse, which is the direction of the first pulse slope.

NameTypeValuesDescription
evinteger1 - 3Event selector: 1 = A-trigger, 2 = B-trigger, 3 = reset event.
PolarityenumPOSitive | NEGative | EITHerPulse polarity setting.

Selects how the range of a pulse width is defined.

NameTypeValuesDescription
RangeModeenumWITHin | OUTSide | SHORter | LONGerSelects how the range of a pulse width is defined.

Sets the pulse width used by trigger event width measurements. For the ranges Within and Outside, the width defines the center of a time range limited by ±Delta.

NameTypeValuesDescription
evinteger1 - 3Trigger event selector: 1 = A-trigger, 2 = B-trigger, 3 = reset event.
widthreal-Width value used by the selected trigger event width range.

Sets the lower voltage limit for the runt trigger on the specified trigger event and analog channel.

NameTypeValuesDescription
evenum1 | 2 | 3Trigger event selector: 1 = A-trigger, 2 = B-trigger, 3 = reset event.
ninteger1 - 8Index of the analog channel.
Levelreal-10 - 10Lower voltage limit.

Sets the upper voltage limit for runt triggering on the selected analog channel. The command is asynchronous.

NameTypeValuesDescription
evenum1 | 2 | 3Selects the trigger event: A-trigger, B-trigger, or reset event.
ninteger1 - 8Index of the analog channel.
levelreal-10 - 10Upper voltage limit.

Defines a range around the runt width specified using TRIGger:EVENt<ev>:RUNT:WIDTh. It is available if TRIGger:EVENt<ev>:RUNT:RANGe is set to WITHin or OutSide.

NameTypeValuesDescription
widthdeltareal1E-10 - 864Range around the runt width.

Sets the polarity of a pulse, which is the direction of the first pulse slope.

NameTypeValuesDescription
polarityenumPOSitive | NEGative | EITHerSelects the pulse polarity.

Defines the time limit of the runt pulse in relation to the runt width and delta settings. It is available for the specified event trigger.

NameTypeValuesDescription
modeenum-Defines the time limit of the runt pulse in relation to the runt width and delta settings.

Defines the upper or lower voltage threshold for runt-width triggering. It is not available if TRIGger:EVENt<ev>:RUNT:RANGe is set to ANY.

NameTypeValuesDescription
Widthreal1E-10 - 10000Range width threshold.

Sets the lower voltage limit.

NameTypeValuesDescription
evinteger1 - 3Selects the trigger event: 1 = A-trigger, 2 = B-trigger, 3 = reset event.
ninteger1 - 8Selects the analog channel index.
Levelreal-10 - 10Sets the lower voltage limit.

Sets the upper voltage limit.

NameTypeValuesDescription
evinteger1 - 3Event selector: 1 = A-trigger, 2 = B-trigger, 3 = reset event.
ninteger1 - 8Index of the analog channel.
levelreal0 - 1000Upper voltage limit.

Defines a range around the width value.

NameTypeValuesDescription
evenum1 | 2 | 3Selects the event: 1 = A-trigger, 2 = B-trigger, 3 = reset event.
widthdeltareal0 - 432Defines the delta around the width value.

Selects how the signal run is compared with the window.

NameTypeValuesDescription
rangemodeenumENTer | EXIT | WITHin | OUTSideSelects the window comparison mode.

Selects how the time limit of the window is defined. It is available when :TRIGger:EVENt<ev>:WINDow:RANGe is WITHin or OUTSide.

NameTypeValuesDescription
time range modeenumWITHin | OUTSide | SHORter | LONGerSelects the definition of the window time limit.

Sets the relation of the signal level to the trigger level for the timeout trigger.

NameTypeValuesDescription
evinteger1 - 3Event selector: 1 = A-trigger, 2 = B-trigger, 3 = reset event.
TimeoutModeenumWITHin | OUTSide | SHORter | LONGerRelation of the signal level to the trigger level for the timeout trigger.

Sets the trigger window width. For ranges Within and Outside, the width defines the center of a time range; for Shorter and Longer, it defines the maximum and minimum time lapse, respectively.

NameTypeValuesDescription
evinteger1 - 3Selects the trigger event: 1 = A-trigger, 2 = B-trigger, 3 = reset event.
widthreal1E-10 - 10000Sets the window width.

Sets a range around the interval width value specified with TRIGger:EVENt<ev>:INTerval:WIDTh.

NameTypeValuesDescription
WidthDeltareal0 - 10Range around the interval width value.

Sets the time limit for the timeout at which the instrument triggers. The event suffix selects A-trigger, B-trigger, or the reset event.

NameTypeValuesDescription
evinteger1 - 3Event number: 1 = A-trigger, 2 = B-trigger, 3 = reset event.
Timereal1E-10 - 10000Time limit for the timeout.

Defines the range of an interval in relation to the interval width specified using TRIGger:EVENt<ev>:INTerval:WIDTh and TRIGger:EVENt<ev>:INTerval:DELTa.

NameTypeValuesDescription
RangeModeenumWITHin | OUTSide | SHORter | LONGerSelects whether the trigger responds inside, outside, shorter than, or longer than the defined interval range.

Sets the edge for the interval trigger. You can analyze the interval between positive edges or between negative edges.

NameTypeValuesDescription
evinteger1 - 3Trigger event suffix: 1 = A-trigger, 2 = B-trigger, 3 = reset event.
SlopeenumPOSitive | NEGative | EITHerSelects the trigger edge polarity.

Sets the time between two pulses for the interval trigger.

NameTypeValuesDescription
widthrealsTime between two pulses for the interval trigger.

Sets the lower voltage threshold for slew rate triggering. When the signal crosses this level, the slew rate measurement starts or stops depending on the selected slope.

NameTypeValuesDescription
evinteger1 - 3Trigger event index: 1 = A-trigger, 2 = B-trigger, 3 = reset event.
ninteger1 - 8Analog channel index.
Levelreal-10 - 10Lower voltage threshold.

Sets the upper voltage threshold for slew rate triggering. When the signal crosses this level, the slew rate measurement starts or stops depending on the selected slope.

NameTypeValuesDescription
evinteger1 - 3Trigger event selector: 1 = A-trigger, 2 = B-trigger, 3 = reset event.
ninteger1 - 8Index of the analog channel.
levelreal-10 - 10Upper voltage threshold.

Defines a time range around the given slew rate.

NameTypeValuesDescription
TimeDeltareal0 - 10Time range around the given slew rate.

Selects how the time limit for the slew rate is defined. The measurement starts when the signal crosses the first trigger level and stops when it crosses the second level.

NameTypeValuesDescription
RangeModeenumINSRange | OUTRange | LTHan | GTHanDefines how the slew-rate time limit is interpreted.

Sets the edge of the clock signal. Edge and level define the time reference point.

NameTypeValuesDescription
ClockEdgeenum-Edge of the clock signal.

Sets the slew rate for trigger event <ev>. For TRIGger:EVENt<ev>:SLEW:RANGe = INSRange and OUTRange, the slew rate defines the center of a range defined by the limits ±Delta; for LTHan and GTHan, it defines the maximum and minimum slew rate limits, respectively.

NameTypeValuesDescription
<Time>real1E-10 - 10000Slew rate value.

Sets the edge direction for the trigger.

NameTypeValuesDescription
slopeenumPOSitive | NEGative | EITHerEdge direction for the trigger.

Sets the voltage level for the clock signal used to determine the starting point for setup and hold time calculation.

NameTypeValuesDescription
ClockLevelreal-10 - 10Voltage level of the clock signal.

Selects the input channel of the clock signal used for setup and hold time calculation.

NameTypeValuesDescription
ClockSourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8Input channel of the clock signal.

Sets the minimum time after the clock edge while the data signal must stay steady above or below the data level.

NameTypeValuesDescription
hold timereal-Minimum time after the clock edge that the data signal must remain steady.

Sets the minimum setup time before the clock edge while the data signal must remain steady above or below the data level.

NameTypeValuesDescription
SetupTimereal-9.9999E-08 - 1E-07Minimum time before the clock edge while the data signal must stay steady.

Sets the state for each analog channel in a state trigger qualification. The clock source is indicated and does not get a state.

NameTypeValuesDescription
evinteger1 - 3Event selector: 1 = A-trigger, 2 = B-trigger, 3 = reset event.
ninteger1 - 8Index of the analog channel.
HLXenumHIGH | LOW | DONTcareState of the individual channels.

Defines the logic combination of the channels and their states for the specified trigger event.

NameTypeValuesDescription
evinteger1 - 3Selects the trigger event: 1 = A-trigger, 2 = B-trigger, 3 = reset event.
logicenumAND | ORSets the logic operator used to combine the channel states.

Defines the logic combination of the channels and their states.

NameTypeValuesDescription
evinteger1 - 3Trigger event index: 1 = A-trigger, 2 = B-trigger, 3 = reset event.
StateOperatorenumAND | ORLogic operator used to combine the channel states.

Sets the edge direction for the trigger. The event suffix <ev> selects A-trigger, B-trigger, or reset event.

NameTypeValuesDescription
evinteger1 - 3Event selector: 1 = A-trigger, 2 = B-trigger, 3 = reset event.
slopeenumPOSitive | NEGative | EITHerEdge direction for the trigger.

Sets the required state for each analog channel used for pattern triggering. For the state trigger, the clock source is indicated and does not get a state.

NameTypeValuesDescription
<n>integer1 - 8Index of the analog channel.
<HLX>enumHIGH | LOW | DONTcareRequired state of the individual channel.
TRIG:EVEN1:PATT:QUAL:ANAL:CHAN1:HLX LOW
TRIG:EVEN1:PATT:QUAL:ANAL:CHAN2:HLX HIGH
TRIG:EVEN1:PATT:QUAL:ANAL:CHAN3:HLX LOW
TRIG:EVEN1:PATT:QUAL:ANAL:CHAN4:HLX HIGH

Sets the required state for each digital channel that is used for triggering.

NameTypeValuesDescription
<HLX>enumHIGH | LOW | DONTcareRequired state for the digital channel.
TRIG:EVEN1:PATT:QUAL:DIG:CHAN<n>:HLX <HLX>

Sets the required state for each digital channel that is used for triggering.

NameTypeValuesDescription
evinteger1 - 3Trigger event index: 1 = A-trigger, 2 = B-trigger, 3 = reset event.
ninteger0 - 15Index of the logic channel.
HLXenumHIGH | LOW | DONTcareRequired state for the digital channel.

Defines the logic combination of the channels and their states for the digital pattern qualifier.

NameTypeValuesDescription
LogicenumAND | ORLogic combination to apply to the channels and their states.

Selects whether the source of the pattern trigger is an analog or digital channel.

NameTypeValuesDescription
stateSourcesenumANALog | DIGitalSelects the trigger source type for the pattern trigger.
:TRIGger:FORCeset / query

Provokes an immediate single acquisition. It forces a trigger in normal mode if no valid trigger occurs, allowing the signal to be confirmed and used for trigger setup.

:TRIGger:MODEset / query

Sets the trigger mode, which determines the behavior of the instrument with and without a trigger event.

NameTypeValuesDescription
TriggerModeenumAUTO | NORMal | FREerunTrigger mode.

Returns the resulting holdoff time when TRIGger:HOLDoff:MODE is set to AUTO. Auto time is calculated as auto time scaling times horizontal scale.

real (s) - Returns the resulting holdoff time in seconds.

Selects the method used to define the trigger holdoff condition. The holdoff mode affects when the next trigger after the current one is recognized.

NameTypeValuesDescription
ModeenumTIME | EVENts | RANDom | AUTO | OFFHoldoff mode selection.

Defines the number of triggers to be skipped when Holdoff mode is set to Events. The next trigger occurs only when the specified number of events is reached.

NameTypeValuesDescription
eventsinteger1 - 2147483647Number of triggers to skip.

Defines the upper limit for the random time holdoff when holdoff mode is set to RANDom.

NameTypeValuesDescription
RandomMaxTimereal1E-07 - 10Upper limit for the random time holdoff.

Sets the auto time scaling factor used when trigger holdoff mode is AUTO. The next trigger occurs only after the resulting auto time has passed.

NameTypeValuesDescription
AutoTimeSclreal0.001 - 1000Auto time scaling factor multiplied with the horizontal scale.

Defines the lower limit for the random time holdoff when trigger holdoff mode is set to RANDom.

NameTypeValuesDescription
RandomMinTimereal1E-07 - 5Lower limit for the random holdoff time.

Defines the holdoff time period when TRIGger:HOLDoff:MODE is set to TIME. The next trigger occurs only after this time has passed.

NameTypeValuesDescription
timereal1E-07 - 10Holdoff time period.

Defines a range in absolute values around the trigger level. If the signal oscillates inside this range and crosses the trigger level, no trigger event occurs.

NameTypeValuesDescription
Absolutereal0 - 10000000Absolute trigger hysteresis range around the trigger level.

Returns the hysteresis that is set by the instrument in automatic hysteresis mode.

NameTypeValuesDescription
minteger1 - 8Index of the analog channel.
real (V) - Numeric effective hysteresis value in volts.

Selects whether the hysteresis is defined in absolute or relative values. The setting is available only in manual hysteresis mode.

NameTypeValuesDescription
ModeenumABS | RELSelects the hysteresis mode as absolute or relative.

Defines a range in divisions around the trigger level in division units. If the signal oscillates inside this range and crosses the trigger level thereby, no trigger event occurs.

NameTypeValuesDescription
minteger1 - 8Index of the analog channel.
inDivisionreal0 - 5Range around the trigger level, in division units.

Selects the filter mode for the trigger channel.

NameTypeValuesDescription
Modeenum-Filter mode for the trigger channel.

Defines a relative hysteresis range around the trigger level as a percentage. If the signal oscillates inside this range and crosses the trigger level, no trigger event occurs.

NameTypeValuesDescription
minteger1 - 8Index of the analog channel.
relativereal0 - 50Relative hysteresis range.

Selects how the hysteresis is set for the trigger noise function. AUTO lets the instrument set the hysteresis automatically to reject internal noise, while MANual uses the value defined by the absolute or relative hysteresis command.

NameTypeValuesDescription
modeenumAUTO | MANualSelects the hysteresis setting mode.

Generates a beep sound if the command is set to TRIGger.

NameTypeValuesDescription
beep--Beep action setting.

Sets the trigger filter mode. The mode can be OFF, LFReject, or RFReject.

NameTypeValuesDescription
modeenumOFF | LFReject | RFRejectTrigger filter mode.

Sets the lowpass filter reject frequency for the trigger signal. Frequencies higher than the specified bandwidth are rejected, lower frequencies pass the filter.

NameTypeValuesDescription
bandwidthreal1E+3 - 500E+6Limit frequency for the trigger signal lowpass filter reject setting.

Defines the delay of the first pulse edge relative to the trigger point. The minimum delay is 600 ns.

NameTypeValuesDescription
delayreal8E-07 - 1Delay of the first pulse edge relative to the trigger point.

Defines when the trigger out signal is initiated: at the trigger point, when waiting for the trigger, or when the post-trigger time is finished.

NameTypeValuesDescription
SignalSourceenumTRIG | TRIGGER | POST | POSTTRIGGER | WAIT | WAITTRIGGERSelects when the trigger out signal is initiated.

Activates the outgoing pulse on the rear-panel [Trigger Out] connector. If ON, a pulse is sent out each time a trigger occurs.

NameTypeValuesDescription
stateenumOFF | ONSets whether the outgoing trigger pulse is active.

Sets the length of the trigger out pulse.

NameTypeValuesDescription
PulseLengthreal1.6E-08 - 0.05Length of the trigger out pulse.

Sets the polarity of the trigger out pulse, which is the direction of the first pulse edge.

NameTypeValuesDescription
polarityenumPOSitive | NEGativePolarity of the trigger out pulse.

Saves a screenshot at each trigger when set to TRIGger. Screenshot settings are configured with the commands in the Screenshots section.

NameTypeValuesDescription
SaveScreenshotenumNOACtion | TRIGgerSelects whether a screenshot is saved on each trigger.

Stops the running acquisition if the command is set to TRIGger.

NameTypeValuesDescription
StopAcqenumNOACtion | TRIGgerControls whether the command stops the running acquisition when set to TRIGger.

Saves the waveform data to file at each trigger when set to TRIGger. Use the EXPort:WAVeform:AUTonaming commands to define the path and file names.

NameTypeValuesDescription
SaveWfmenumNOACtion | TRIGgerSelects whether waveform data is saved on each trigger.

Defines the zone trigger using a logical expression. The available operators for combining zones are AND, NOT, OR, and XOR.

NameTypeValuesDescription
LogicExpressionstring-String with the logical expression.
TRIGger:ZONE:EXPRession 'Zone1 and Zone2'
TRIGger:ZONE:EXPRession?

Applies the zone trigger condition to acquisitions stored in history memory, allowing history waveforms to be filtered by zone conditions.

NameTypeValuesDescription
ApplyToHistoryenumOFF | ONTurns applying the zone trigger condition to history acquisitions off or on.

Defines how the specified data pattern is searched.

NameTypeValuesDescription
DataAlignmentenumWORD | BIT

Selects the trigger type for SPI analysis.

NameTypeValuesDescription
typeenumFRSTart | FRENd | MOSI | MISOSelects the SPI trigger type.

Specifies a data bit pattern, or sets the start value of a pattern range.

NameTypeValuesDescription
DataPattern--

Sets the number of bits or words to be ignored before the first bit or word of interest. The effect is defined by TRIGger:SBHW:SPI:PALignment.

NameTypeValuesDescription
DataPositioninteger1 - 4096Number of bits or words to be ignored before the first bit or word of interest.

Selects the operator for the MISO and MOSI pattern.

NameTypeValuesDescription
DataOperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHanOperator used for the MISO and MOSI pattern.

Sets the bit state of a one-bit field for the software trigger.

NameTypeValuesDescription
bitstateenumONE | ZEROBit state to apply to the selected field.

Sets the bit state of a one-bit field for the software trigger. The manual also shows the related frame/field form and a query form that takes a bit index.

NameTypeValuesDescription
bitinteger-Bit index for the selected field.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
enableenumOFF | ONEnables or disables the selected frame checking condition.

Sets the bit state of a field that only consists of one bit for the software trigger.

NameTypeValuesDescription
BitStateenumONE | ZEROBit state to apply to the selected field.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
EnableenumOFF | ONEnables or disables the checking condition for the selected frame.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
enablerenumOFF | ONEnables or disables the checking condition for the selected frame.
enum - Returns the enable state for the selected frame.

Inverts the current state of the frame and error types for the software trigger, so all enabled frames and error types become disabled and vice versa.

Presets the state of the selected frames and error types for the software trigger.

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe. The operator is set with TRIGger:SBSW:SPI:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
Frameinteger-Index of the frame.
Fieldinteger-Index of the field.
Datastring-

Returns the end value of a data pattern range for the software trigger.

string - Returns the end value of a data pattern range for the software trigger.

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
Data_Min--

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
Data--

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
Operator--Operator for the data pattern.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
operatorenum-

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe. The operator is set with TRIGger:SBSW:SPI:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
Data_Maxstring-

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger. This command applies to the selected frame and field.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
Data_Min--

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Data_Operator--Operator for the data pattern.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
Frame--Frame index.
Field--Field index.
EnablerenumOFF | ONEnables or disables the checking condition.
enum - Returns the enabler state.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
frame--Index of the frame.
field--Index of the field.
enablerenumOFF | ONEnables or disables the checking condition.
returns

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
<CondEnabler>enumOFF | ONEnables or disables the checking condition.

Sets the end value of an index range for the software trigger when the operator is set to INRange. Use with the selected frame and field of the SPI software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
index_maxinteger1 - 65535End value of an index range.

Sets the end value of an index range for the software trigger when the operator is set to INRange. The operator is set with TRIGger:SBSW:SPI:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
datainteger1 - 65535End value of the index range.

Sets the end value of an index range for the software trigger when the operator is set to INRange. The operator is set with TRIGger:SBSW:SPI:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
datainteger1 - 65535End value of an index range for the software trigger.
integer - Returns the end value of an index range for the software trigger.

Defines the error type for the software trigger.

NameTypeValuesDescription
enablerenumOFF | ONEnables or disables the error type.

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
Index_Mininteger1 - 65535Start value of the index range.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
frame--Index of the frame.
field--Index of the field.
operatorenumEQUal | INRange | RANGeOperator for the index in the selected field.

Specifies the start value of an index range for the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
Data--

Returns the start value of an index range for the software trigger. The query form takes a data value and is associated with the selected frame and field.

NameTypeValuesDescription
data--Data value for the query.
returns - Returns the start value of an index range for the software trigger.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
operatorenumEQUal | INRange | RANGeOperator used for the index.

Sets or queries the operator for the index in the selected field of the selected frame for the software trigger.

enum - Returns the current index operator.

Sets the checking condition for all QSPI frames in the software trigger settings. The page slice only lists the command name in the QSPI software trigger section and does not provide syntax or further parameter details for this command.

Clears the software trigger QUAD SPI settings. The command is listed under Software trigger and cross-referenced on the same page.

Inverts the current software trigger state for all frames and error types, disabling previously enabled items and enabling previously disabled ones.

Sets the bit state of a field that only consists of one bit for the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
BitenumONE | ZEROBit state to set for the software trigger.

Sets the bit state of a field that only consists of one bit for the software trigger.

NameTypeValuesDescription
bitenumONE | ZEROBit state to set.

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe.

NameTypeValuesDescription
data_maxreal-End value of the data pattern range.

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe.

NameTypeValuesDescription
data_max--End value of the data pattern range.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
enableenumOFF | ONEnables or disables the checking condition for the selected frame.

Sets the bit state of a one-bit field for the software trigger.

NameTypeValuesDescription
bit stateenumONE | ZEROBit state to set for the selected field.

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
data_maxreal-End value of the data pattern range.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
EnablerenumOFF | ONEnables or disables the checking condition.

Returns the frame enable state for the selected frame in the software trigger setup.

enum - Returns the enable state of the selected frame.

Presets the state of the selected frames and error types for the software trigger.

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
frame--Index of the frame.
field--Index of the field.
data--

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
Datareal-
real - Returns the current minimum data pattern value.
TRIGger:SBSW:QSPI:DMIN? <Data>

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GEThan | INRange | OORangeOperator for the data pattern.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

enum - Returns the data operator.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator used for the data pattern.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
EnablerenumOFF | ONEnables or disables the checking condition.
enum - Returns the enabler state of the selected field of the selected frame.

Returns the enabler state for the selected field of the selected software trigger frame. The response is the current condition enabler value.

enum - Returns the current checking-condition enabler for the selected field.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
<CondEnabler>enumOFF | ONEnables or disables the checking condition.

Sets the end value of an index range for the selected field of the selected frame of the software trigger when the operator is set to INRange.

NameTypeValuesDescription
index_maxinteger1 - 65535End value of the index range.

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
index_mininteger1 - 65535Start value of the index range for the software trigger.

Sets the end value of an index range for the software trigger if the operator is set to INRange. The operator is set with the corresponding IOperator command.

NameTypeValuesDescription
Index_Maxinteger1 - 65535End value of the index range.

Sets the end value of an index range for the software trigger when the operator is set to INRange. The corresponding query returns the current maximum index value.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
datainteger1 - 65535End value of the index range.
integer - Returns the current end value of the index range.

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
Frameinteger-Index of the frame.
Fieldinteger-Index of the field.
Datainteger1 - 65535Index value or range start value for the software trigger.

Specifies the index, or sets the start value of an index range for the software trigger.

returns

Defines the error type for the software trigger.

NameTypeValuesDescription
enableenumOFF | ONError type enable state.

Queries the software trigger error-enable state. If no software trigger event has occurred, the command returns Invalid.

enum - Returns the enabled state of the selected error type, or Invalid if no software trigger event has occurred.

Defines whether the selected software-trigger error is enabled. The error index is given by <m>.

NameTypeValuesDescription
enableenumOFF | ONEnables or disables the error type.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
Frame--
Field--
OperatorenumEQUal | INRange | RANGe

Queries the software trigger index. If no software trigger event has occurred, the command returns Invalid.

enum - Returns the software trigger index, or Invalid if no software trigger event has occurred.

Toggles the trigger condition between read and write access of the primary. Select "Either" if the transfer direction is not relevant for the trigger condition.

NameTypeValuesDescription
RWBitAddressenumREAD | WRITe | EITHerRead/write access condition for the primary trigger.

Selects the trigger type for I²C analysis.

NameTypeValuesDescription
typeenumSTARt | REPStart | STOP | NACK | ADDRess | DATA | ADATTrigger type for I²C analysis.

Sets the operator used to define a specific address or an address range. The address values are set with `TRIGger:SBHW:I2C:ADDRess` and `TRIGger:SBHW:I2C:ADDTo`.

NameTypeValuesDescription
AddrOptorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeSelects the address comparison operator.

Triggers on the specified address, or sets the start value of an address range depending on the condition set with TRIGger:SBHW:I2C:ACONdition.

NameTypeValuesDescription
Address--

Sets the end value of an address range if the condition is set to an address range with TRIGger:SBHW:I2C:ACONdition.

NameTypeValuesDescription
AddressTo--

Triggers if the address acknowledge bit is missing, meaning no target recognizes the address.

NameTypeValuesDescription
AddressNackenumOFF | ONEnables or disables triggering on missing address acknowledge.

Sets the address length to be triggered on: 7 bit or 10 bit.

NameTypeValuesDescription
AddressTypeenumBIT7 | BIT7RW | BIT7_RW | BIT10Selects whether to enter only the seven or ten address bits, or seven address bits plus the read/write bit.

Sets the operator used to specify a specific data value or a data range.

NameTypeValuesDescription
DataOperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHanSelects the comparison operator for the I2C data condition.

Specifies the data bit pattern, or sets the start value of a data pattern range. Enter the bytes in MSB first bit order; the maximum pattern length is 64 bit and waveform data is compared with the pattern byte-by-byte.

NameTypeValuesDescription
data--Data bit pattern or start value of a data pattern range.

Sets the number of data bytes to be skipped after the address.

NameTypeValuesDescription
DataPositioninteger1 - 4096Number of data bytes to skip after the address.

Triggers on the end of the read process when the controller reads data from the target. The NACK is sent according to the protocol definition and is not an error.

NameTypeValuesDescription
DataReadNackenumOFF | ONEnables or disables triggering on the read-process NACK condition.

Triggers if a data acknowledge bit is missing, meaning the addressed target does not accept the data.

NameTypeValuesDescription
DataWriteNackenumOFF | ONEnables or disables triggering on a missing data acknowledge bit.

Inverts the current state of the frame and error types for the software trigger; enabled frames and error types are disabled and vice versa.

Presets the state of the selected frames and error types for the software trigger.

Sets the bit state of a one-bit field for the software trigger.

NameTypeValuesDescription
frame--Index of the frame.
field--Index of the field.
bitenumONE | ZEROBit state to set for the field.

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe. The operator is set with TRIGger:SBSW:I2C:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
data_maxreal-

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
EnableenumOFF | ON

Sets the bit state of a field that only consists of one bit for the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
BitStateenumONE | ZERO

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe. The operator can be set with TRIGger:SBSW:I2C:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
data_max--End value of the data pattern range.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
frameinteger-Index of the frame.
enablerenumOFF | ONEnables or disables the checking condition.

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
Data--Data pattern or range start value.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator used for the data pattern.
enum - Returns the current data operator.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
enablerenum-
enum

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator for the data pattern.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
CondEnablerenum-

Sets the end value of an index range for the software trigger if the operator is set to INRange. The operator is set with TRIGger:SBSW:I2C:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
Index_Maxinteger1 - 65535End value of an index range.

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
fr--Index of the frame.
fl--Index of the field.
Index_Mininteger1 - 65535Start value of an index range for the software trigger.

Sets the end value of an index range for the software trigger if the operator is set to INRange. You can set the operator with TRIGger:SBSW:I2C:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
frameinteger1 - 65535Index of the frame.
fieldinteger1 - 65535Index of the field.
datastring-

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
Index_Mininteger1 - 65535Start value of the index range.

Defines the error type for the software trigger.

NameTypeValuesDescription
ErrorName--
EnablerenumOFF | ON

Defines the error type for the software trigger. The error enable state is set with the indexed error suffix.

NameTypeValuesDescription
minteger-Index of the error.
enableenumOFF | ONEnables or disables the error type.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
Index_OperatorenumEQUal | INRange | RANGeOperator for the index in the selected field of the selected frame.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
Frame--
Field--
OperatorenumEQUal | INRange | RANGeSelects the comparison operator used for the index.

Returns the end value of a data pattern range for the software trigger. This applies when the operator is set to INRange or OORANGe.

string - The end value of the data pattern range for the software trigger.

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

returns - Returns the start value of the data pattern range for the software trigger, or the data pattern itself.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
Operator--Operator for the data pattern.
returns - Returns the operator for the data pattern of the software trigger in the selected field of the selected frame.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
EnablerenumOFF | ONTurns the checking condition off or on.
enum - Returns the enable state of the selected field condition.

Sets the end value of an index range for the software trigger when the operator is set to INRange. The operator is set with TRIGger:SBSW:I3C:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
datainteger1 - 65535End value of the index range.

Defines the error type for the software trigger.

NameTypeValuesDescription
ErrorName--
EnablerenumOFF | ON
enum - Returns the enable state.

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
data--
returns

Queries the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
operatorenum-Operator of the index in the selected field of the selected frame.

Specifies the data pattern to be found on the specified trigger source. Enter the words in MSB-first bit order.

NameTypeValuesDescription
DataPattern--Data pattern to be found on the specified trigger source.

Sets the number of words before the first word of interest. These offset words are ignored.

NameTypeValuesDescription
DataPositioninteger1 - 4096Number of words before the first word of interest.

Selects the operator for the "Data" pattern.

NameTypeValuesDescription
DataOperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHanOperator for the Data pattern.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
DataOperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHanOperator used for the data pattern.

Selects the transmitter or receiver line as the trigger source.

NameTypeValuesDescription
sourceenum-Trigger source line.

Selects the UART trigger type condition.

NameTypeValuesDescription
TypeenumSTBT | PCKS | DATA | PRER | BRKC | STPerrorTrigger type condition.

Inverts the current state of the frame and error types for the software trigger, disabling enabled items and enabling disabled ones.

Presets the state of the selected frames and error types for the software trigger.

Sets the bit state of a field that only consists of one bit for the software trigger.

NameTypeValuesDescription
Frame--Index of the frame
Field--Index of the field
BitenumONE | ZERO

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe. The operator is set with TRIGger:SBSW:UART:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
data--

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
enableenumOFF | ONEnables or disables the checking condition.

Sets the bit state of a one-bit field for the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
BitStateenumONE | ZEROBit state to set for the field.

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe. The operator is set with TRIGger:SBSW:UART:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
data_max--

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
EnablerenumOFF | ONEnables or disables the checking condition for the selected frame.

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
Frameinteger-Index of the frame.
Fieldinteger-Index of the field.
Data--

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator for the data pattern.
enum - Returns the current data pattern operator.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
frame--Index of the frame.
field--Index of the field.
enablerenum-
enum

Sets the start value of a data pattern range for the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Data_Min--Start value of the data pattern range.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator used for the data pattern.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
CondEnablerenumON | OFFEnables or disables the checking condition.

Sets the end value of an index range for the software trigger when the operator is set to INRange. The frame and field are selected by the suffixes <fr> and <fl>.

NameTypeValuesDescription
Index_Maxinteger1 - 65535End value of the index range.

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
frinteger1 - 65535Index of the frame.
flinteger1 - 65535Index of the field.
Index_Mininteger1 - 65535Start value of an index range for the software trigger.

Sets the end value of an index range for the software trigger if the operator is set to INRange. You can set the operator with TRIGger:SBSW:UART:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
frame--Frame index.
field--Field index.
data--Data value.

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
Index_Mininteger1 - 65535Start value of an index range for the software trigger.

Defines the error type for the software trigger.

NameTypeValuesDescription
enablerenumOFF | ONEnables or disables the error type for the software trigger.

Defines the error type for the software trigger.

NameTypeValuesDescription
enableenumOFF | ONEnables or disables the error type.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame
flinteger-Index of the field
index_operatorenumEQUal | INRange | RANGeOperator for the index in the selected field.

Sets the operator for the selected field of the selected frame for the software trigger. A query form is also documented for the operator setting.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
OperatorenumEQUal | INRange | RANGeOperator for the selected field.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
EnableenumOFF | ONEnables or disables the checking condition for the selected frame.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
Frame--Frame index selector.
EnablerenumOFF | ONEnables or disables the checking condition.

Inverts the current state of the frame and error types for the software trigger; all frames and error types that were enabled are disabled and vice versa.

Presets the state of the selected frames and error types for the software trigger.

Sets the bit state of a field that only consists of one bit for the software trigger.

NameTypeValuesDescription
Frame--
Field--
Bit--

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe. You can set the operator with TRIGger:SBSW:NRZC:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
data--

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
Data_Min--

Sets the end value of a data pattern range for the software trigger when the operator is set to INRange or OORANGe. The operator is set with TRIGger:SBSW:NRZC:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
data_max--

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Frameinteger-Index of the frame.
Fieldinteger-Index of the field.
OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator used for the data pattern comparison.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
enablerenumOFF | ONEnables or disables the checking condition for the selected field of the selected frame of the software trigger.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator used for the data pattern.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
CondEnablerenumOFF | ONEnables or disables the checking condition for the selected field of the selected frame.

Sets the end value of an index range for the software trigger if the operator is set to INRange. The operator can be set with the corresponding IOPerator command.

NameTypeValuesDescription
Index_Maxinteger-End value of the index range.

Sets the end value of an index range for the software trigger if the operator is set to INRange.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
datainteger-End value of an index range.

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
index_mininteger1 - 65535Start value of the index range for the selected frame and field.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
Index_OperatorenumEQUal | INRange | RANGeOperator for the index in the selected field of the selected frame.

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
Frameinteger1 - 65535Index of the frame.
Fieldinteger1 - 65535Index of the field.
Datainteger-Index value to be specified or set as the start value of the range.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
operatorenumEQUal | INRange | RANGeOperator for the index in the selected field of the selected frame.

Defines the error type for the software trigger.

NameTypeValuesDescription
ErrorName--
EnablerenumOFF | ON

Defines the error type for the software trigger for error index <m>.

NameTypeValuesDescription
minteger-Index of the error.
EnableenumOFF | ONEnables or disables the error type.

Sets the bit state of a one-bit field for the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Field index.
Bit--Bit state for the selected field.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
enableenumOFF | ONEnables or disables the checking condition.

Sets the bit state of a one-bit field for the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
bitinteger-Bit state.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
Frame--Selected frame.
EnablerenumOFF | ONEnables or disables the selected frame's checking condition.

Inverts the current state of the frame and error types for the software trigger; all enabled frames and error types are disabled and vice versa.

Presets the state of the selected frames and error types for the software trigger.

Sets the end value of a data pattern range for the software trigger when the operator is set to INRange or OORANGe. The operator is configured with TRIGger:SBSW:NRZU:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
data--End value of the data pattern range.

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
Data_Minstring-

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe.

NameTypeValuesDescription
frinteger-Index of the frame
flinteger-Index of the field
Data_Max--

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
data_minnull-Start value of a data pattern range.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator for the data pattern.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
EnablerenumOFF | ONEnables or disables the checking condition.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
Data_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator for the data pattern.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
condenablerenumOFF | ONEnables or disables the checking condition for the selected field of the selected frame.

Sets the end value of an index range for the software trigger if the operator is set to INRange.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
datainteger-End value of the index range.

Defines the error type for the software trigger.

NameTypeValuesDescription
enableenum-Enable setting for the selected error type.

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
frinteger1 - 65535Index of the frame.
flinteger1 - 65535Index of the field.
Index_Mininteger1 - 65535Start value of an index range for the software trigger.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
frinteger1 - 65535Index of the frame.
flinteger1 - 65535Index of the field.
index_operatorenumEQUal | INRange | RANGeOperator for the index.

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
Frameinteger1 - 65535Index of the frame.
Fieldinteger1 - 65535Index of the field.
Datainteger1 - 65535Data index value.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
OperatorenumEQUal | INRange | RANGeIndex operator to apply.

Disables the software trigger for all available frames and error types. It is a setting-only asynchronous command.

Inverts the current state of the frame and error types for the software trigger, so enabled items become disabled and vice versa.

Sets the bit state of a field that only consists of one bit for the software trigger.

NameTypeValuesDescription
BitStateenumONE | ZEROBit state to set for the selected field.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
EnableenumOFF | ONEnables or disables the checking condition for the selected frame.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
Frame--Selected frame.
EnablerenumOFF | ONEnables or disables the selected frame check condition.

Presets the state of the selected frames and error types for the software trigger.

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe.

NameTypeValuesDescription
Frameinteger-Index of the frame.
Fieldinteger-Index of the field.
Datanumeric-

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
data--

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Frameinteger-Index of the frame.
Fieldinteger-Index of the field.
Operatorenum-Operator for the data pattern.
enum - Returns the current data pattern operator for the selected field of the selected frame.

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe. Use TRIGger:SBSW:MANCh:FRAMe<fr>:FLD<fl>:DOPerator to set the operator.

NameTypeValuesDescription
Data_Max--

Specifies the start value of a data pattern range for the software trigger, or sets the data pattern itself. It applies to the selected frame and field.

NameTypeValuesDescription
Data_Minreal-Start value of the data pattern range.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
EnablerenumOFF | ONEnables or disables the checking condition.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
CondEnablerenumOFF | ONEnables or disables the checking condition.

Sets the end value of an index range for the software trigger if the operator is set to INRange. The operator is set with TRIGger:SBSW:MANCh:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
index_maxinteger1 - 65535End value of the index range.

Sets the end value of an index range for the software trigger when the operator is set to INRange. The operator is set with TRIGger:SBSW:MANCh:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
datainteger1 - 65535End value of the index range.

Defines the error type for the software trigger.

NameTypeValuesDescription
errorname--
enablerenumOFF | ON

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
Index_Mininteger1 - 65535Index minimum value for the selected frame and field.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
Index_OperatorenumEQUal | INRange | RANGeOperator for the index in the selected frame and field.

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
frame--Index of the frame.
field--Index of the field.
datainteger1 - 65535Index_Min value.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
operatorenumEQUal | INRange | RANGeOperator for the index in the selected field of the selected frame.

Triggers when the transmitter does not receive an acknowledgment, i.e. a dominant bit during the ACK Slot. Available only if TRIGger:SBHW:CAN:TYPE is set to ERRC.

NameTypeValuesDescription
AckErrorenumOFF | ONEnables or disables the ACK error trigger condition.

Triggers if a stuff error occurs when the 6th consecutive equal bit level in the mentioned fields is detected. Available only if TRIGger:SBHW:CAN:TYPE is set to ERRC.

NameTypeValuesDescription
BitStuffErrorenumOFF | ONEnables or disables bit stuff error triggering.

Sets the byte order (endianness) of the data transfer. Only for CAN protocol.

NameTypeValuesDescription
endiannessenumBENDian | LENDianByte order used for CAN data transfer.

Triggers on CRC errors. A CRC error occurs when the CRC calculated by the receiver differs from the received value in the CRC sequence.

NameTypeValuesDescription
ChecksumErrorenumOFF | ONEnables or disables CRC-error triggering.

Sets the operator used to define a specific CAN data pattern or a data pattern range.

NameTypeValuesDescription
DataOperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHanSelects the comparison used for the CAN data pattern.

Sets the data length code, the number of data bytes to be found. For complete definition, set also the operator with TRIGger:SBHW:CAN:DLCCondition.

NameTypeValuesDescription
DLCinteger1 - 8Data length code to be found.

Sets the data length code operator. The number of data bytes to be found is set with TRIGger:SBHW:CAN:DLC.

NameTypeValuesDescription
DLCOperatorenumEQUal | GETHanSelects how the data length code is evaluated.

Sets a data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
DataPatternstring-List of comma separated values.

Sets the bit rate switch bit.

NameTypeValuesDescription
BRS_BitenumONE | ZERO | DCBit rate switch bit value.

Defines the number of the first data byte at which the data pattern may start.

NameTypeValuesDescription
DataPositioninteger1 - 2048Number of the first data byte at which the data pattern may start.

Sets the error state indicator bit.

NameTypeValuesDescription
ESI_BitenumONE | ZERO | DCError state indicator bit; DC means do not care, bit is not relevant.

Triggers on stuff count errors. A stuff bit error occurs if more than five consecutive bits of the same level occur on the bus.

NameTypeValuesDescription
StuffCntErrenumOFF | ON

Triggers when a fixed-form bit field contains one or more illegal bits. Available if `TRIGger:SBHW:CAN:TYPE` is set to `ERRC`.

NameTypeValuesDescription
FormErrorenumOFF | ONEnables or disables triggering on form errors.

Sets the CAN frame type.

NameTypeValuesDescription
FrameTypeenumCBFF | CBFR | CEFF | CEFR | FBFF | FEFF | XLFF | ERRor | OVERloadCAN frame type to set.

Sets the operator used to match a specific CAN identifier or an identifier range. The selected condition determines whether one identifier pattern or a min/max range must be set with TRIGger:SBHW:CAN:IMIN and/or TRIGger:SBHW:CAN:IMAX.

NameTypeValuesDescription
IdOperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeSelects how the CAN identifier is compared.

Sets the end value of an identifier range if TRIGger:SBHW:CAN:ICONdition is set to INRange or OORange.

NameTypeValuesDescription
IdPatternstring-List of comma separated values.

Specifies a message identifier pattern, or sets the start value of an identifier range.

NameTypeValuesDescription
IdPatternstring-Message identifier pattern or the start value of an identifier range.

Selects the format of data and remote frames. Remote frames are not available in the CAN FD protocol.

NameTypeValuesDescription
IdentifierTypeenumB11 | B29Selects the CAN identifier format.

Selects the trigger type for CAN analysis.

NameTypeValuesDescription
typeenumSTOF | EDOF | FTYP | ID | IDDT | ERRCTrigger type for CAN analysis.

Sets the comparison condition for the acceptance field to a specific value or a range.

NameTypeValuesDescription
AfOperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeComparison condition for the acceptance field; single-value comparison modes require one pattern set with TRIGger:SBHW:CAN:XDATa:AF:MIN, while range modes use AF:MIN and AF:MAX.

Sets the end value of an acceptance field when the acceptance-field condition is INRange or OORange.

NameTypeValuesDescription
AfPatternToblock-List of comma separated values.

Specifies an acceptance field pattern, or sets the start value of a range.

NameTypeValuesDescription
AfPatterndiscrete-Acceptance field pattern, or start value of a range.

Sets the comparison condition for the service data unit type to a specific value or a range.

NameTypeValuesDescription
SdtOperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeComparison condition for the service data unit type.

Sets the end value of a service data unit type range when TRIGger:SBHW:CAN:XDATa:SDT:CONDition is set to INRange or OORange.

NameTypeValuesDescription
sdtpatterntostring-List of comma separated values.

Specifies a service data unit type pattern, or sets the start value of a range.

NameTypeValuesDescription
<SdtPattern>list-List of comma separated values.

Sets a value for the simple extended content (SEC) field. It indicates whether the CAN XL data frame uses the CADsec protocol.

NameTypeValuesDescription
SEC_BitenumONE | ZERO | DCSEC field value.

Sets the comparison condition for the VCID to a specific value or a range. Equality and inequality conditions require one pattern set with VCID:MIN; INRange and OORange require VCID:MIN and VCID:MAX.

NameTypeValuesDescription
VcidOperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeComparison condition for the VCID.

Sets the end value of a VCID range when the VCID comparison condition is INRange or OORange.

NameTypeValuesDescription
vcidpatterntostring-List of comma separated values.

Specifies a VCID pattern, or sets the start value of a range.

NameTypeValuesDescription
VcidPatterndiscrete-List of comma separated values.

Sets the bit state of a field that only consists of one bit for the software trigger.

NameTypeValuesDescription
Frame--Frame index.
Field--Field index.
Bit--Bit state.

Sets the bit state of a one-bit field for the software trigger.

NameTypeValuesDescription
bitenum-Bit state to apply to the selected field.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
enableenumOFF | ONEnables or disables the checking condition for the selected frame.

Sets the bit state of a one-bit field for the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
BitStateenum-Bit state to set.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
EnableenumOFF | ONEnables or disables the checking condition for the selected frame.
enum - Returns the current enable state of the selected frame.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
enablerenumOFF | ONEnables or disables the checking condition for the selected frame.
enum - Returns the current enable state for the selected frame.

Inverts the current state of the frame and error types for the software trigger, so enabled frames and error types become disabled and vice versa.

Presets the state of the selected frames and error types for the software trigger.

Sets the end value of a data pattern range for the software trigger when the operator is set to INRange or OORANGe.

NameTypeValuesDescription
data_max--End value of the data pattern range.

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe.

NameTypeValuesDescription
data_max--
returns

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
Data_Minstring-

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
Data_Min--

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe. The operator is set with TRIGger:SBSW:CAN:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
Data_Max--

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Frameinteger-Index of the frame.
Fieldinteger-Index of the field.
OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeData operator to use for the software trigger pattern.

Returns the operator for the data pattern of the software trigger in the selected field of the selected frame.

enum - Returns the current data operator for the selected field of the selected frame.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
enablerenumOFF | ONEnables or disables the checking condition for the selected field of the selected frame of the software trigger.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
enablerenumOFF | ONEnables or disables the checking condition.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeSelects the comparison operator for the data pattern.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
CondEnablerenumOFF | ONEnables or disables the checking condition.

Sets the end value of an index range for the software trigger if the operator is set to INRange.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
Index_Maxinteger-End value of the index range.

Sets the end value of an index range for the software trigger if the operator is set to INRange.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
datainteger-End value of the index range.

Sets the end value of an index range for the software trigger if the operator is set to INRange.

NameTypeValuesDescription
datareal-End value of the index range.
real - Returns the end value of an index range for the software trigger.

Defines the error type for the software trigger.

NameTypeValuesDescription
errorname--Error type to define for the software trigger.
enabler--Enables the selected error type.

Defines the error type for the software trigger.

NameTypeValuesDescription
enableenumEnable state for the selected CAN error.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
Index_OperatorenumEQUal | INRange | RANGeOperator used for the index comparison.

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
frame--Index of the frame.
field--Index of the field.
datainteger1 - 65535Start value of the index range.

Returns the start value of an index range for the software trigger.

NameTypeValuesDescription
data--
returns - Returns the start value of an index range for the software trigger.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
operatorenumEQUal | INRange | RANGeOperator used for the index comparison.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
operatorenumEQUal | INRange | RANGeOperator used for the index comparison.

Triggers if a synchronization error occurs. Available only when TRIGger:SBHW:LIN:TYPE is set to ERRC.

Selects the trigger type for LIN analysis.

NameTypeValuesDescription
TypeenumSTARtframe | ID | IDDT | WKFR | ERRCTrigger type for LIN analysis.

Sets the end value of an identifier range when TRIGger:SBHW:LIN:ICONdition is set to INRange or OORange.

NameTypeValuesDescription
IdPattern--

Specifies a secondary identifier pattern, or sets the start value of an identifier range.

NameTypeValuesDescription
IdPattern--

Triggers if an error occurs in the identifier parity bits. The parity bits are bits 6 and 7 of the identifier. Available if TRIGger:SBHW:LIN:TYPE is set to ERRC.

NameTypeValuesDescription
IdParityErrorenumOFF | ONEnables or disables triggering on identifier parity errors.

Sets the operator to set a specific data pattern or a data pattern range.

NameTypeValuesDescription
DataOperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeSelects the comparison operator for a data pattern or pattern range.

Specifies a data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
DataPattern--

Sets the number of data events that are ignored before trigger condition checking of the data starts.

NameTypeValuesDescription
data positioninteger1 - 4096Number of data events to ignore before trigger condition checking begins.

Sets the operator to select a specific identifier or an identifier range. The conditions EQUal, NEQual, LTHan, LETHan, GTHan, and GETHan require an identifier pattern set with TRIGger:SBHW:LIN:IMIN; INRange and OORange use TRIGger:SBHW:LIN:IMIN and TRIGger:SBHW:LIN:IMAX to define the range.

NameTypeValuesDescription
IdOperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeSelects the comparison operator for the identifier condition.

Triggers on checksum errors. Available only if `TRIGger:SBHW:LIN:TYPE` is set to `ERRC`.

NameTypeValuesDescription
ChecksumErrorenumOFF | ONEnables or disables triggering on checksum errors.

Inverts the current state of the frame and error types for the software trigger: all frames and error types that were enabled are disabled and vice versa.

Presets the state of the selected frames and error types for the software trigger.

Sets the bit state of a one-bit field for the software trigger.

NameTypeValuesDescription
frame--Index of the frame.
field--Index of the field.
bitenumONE | ZEROBit state to set for the field.

Sets the end value of a data pattern range for the software trigger. The setting applies when the operator is set to INRange or OORANGe.

NameTypeValuesDescription
frame--Index of the frame.
field--Index of the field.
data--

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
enableenumOFF | ONEnables or disables the frame checking condition.

Sets the bit state of a one-bit field for the software trigger. The command is addressed by frame and field suffix indices.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
BitStateenumONE | ZEROBit state to set for the field.

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe. The operator is set with TRIGger:SBSW:LIN:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
Data_Max--

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
frame--Index of the frame.
enablerenumOFF | ONEnables or disables the checking condition.

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
Data--

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator for the data pattern.
enum - Returns the current data pattern operator.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
enabler--

Sets the start value of a data pattern range for the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
Data_Min--

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator for the data pattern.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
CondEnablerenumON | OFFEnables or disables the checking condition.

Sets the end value of an index range for the software trigger when the operator is set to INRange.

NameTypeValuesDescription
Index_Maxinteger1 - 65535End value of the index range.

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
frinteger1 - 65535Index of the frame
flinteger1 - 65535Index of the field
Index_Mininteger1 - 65535Start value of an index range for the software trigger

Sets the end value of an index range for the software trigger if the operator is set to INRange. The operator is set with TRIGger:SBSW:LIN:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
datainteger1 - 65535End value of an index range.

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
Frame--
Field--
Data--

Defines the error type for the software trigger. The command uses an error index and an enable state.

NameTypeValuesDescription
ErrorName--
EnablerenumOFF | ON

Defines the enable state for a selected software-trigger error type. The error is selected by index <m>.

NameTypeValuesDescription
enableenumOFF | ONEnables or disables the selected error type.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
Index_OperatorenumEQUal | INRange | RANGeSets the operator for the selected index.

Sets the operator for the selected frame and field for the software trigger.

NameTypeValuesDescription
Frame--Frame index.
Field--Field index.
OperatorenumEQUal | INRange | RANGeOperator for the selected field index.

Enables the software trigger for a selected frame. The command is documented with the setting form `TRIGger:SBSW:SENT:FRAMe<fr>:ENABle <Enable>` and related aliases.

NameTypeValuesDescription
fr--Frame selector for the software trigger.
Enable--Enables or disables the selected frame.

Enables a software-trigger frame selection with an enabler setting. The page shows both the frame-specific form and the abbreviated FRENable form.

NameTypeValuesDescription
Frame--
Enabler--

Inverts the current state of the frame and error types for the software trigger. All frames and error types that were enabled are disabled, and vice versa.

Presets the state of the selected frames and error types for the software trigger.

Sets the bit state of a single-bit field for the software trigger. The frame and field are specified by index.

NameTypeValuesDescription
Frameinteger-Index of the frame.
Fieldinteger-Index of the field.
Bitinteger-Bit state to set.

Sets the end value of a data pattern range for the software trigger when the operator is set to INRange or OORANGe. The operator is set with TRIGger:SBSW:SENT:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
datareal-End value of the data pattern range.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
enablerenumOFF | ONEnables or disables the checking condition.

Sets the bit state of a field that only consists of one bit for the software trigger.

NameTypeValuesDescription
BitStateenumONE | ZEROBit state to set for the selected field bit.

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe. The operator is set with TRIGger:SBSW:SENT:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
data_maxreal-End value of a data pattern range.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
condenablerenumOFF | ONEnables or disables the checking condition for the selected field.

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
Data--

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator used for the data pattern comparison.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator for the data pattern.

Sets the end value of an index range for the software trigger if the operator is set to INRange. The frame and field indices are specified by the <fr> and <fl> suffixes.

NameTypeValuesDescription
Index_Maxinteger1 - 65535End value of the index range.

Sets the end value of an index range for the software trigger if the operator is set to INRange. The index range operator can be set with TRIGger:SBSW:SENT:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
data--

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
Frameinteger1 - 65535Index of the frame.
Fieldinteger1 - 65535Index of the field.
Datainteger1 - 65535Index value for the software trigger.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
Operator--Operator value.

Defines the error type for the software trigger.

NameTypeValuesDescription
<ErrorName>--
<Enabler>enumOFF | ONEnables or disables the error type.

Defines the error type for the software trigger.

NameTypeValuesDescription
enableenumOFF | ONEnables or disables the selected error type.

Inverts the current state of the frame and error types for the software trigger. All frames and error types that were enabled are disabled and vice versa.

Sets the bit state of a one-bit field for the software trigger.

NameTypeValuesDescription
BitStateenumONE | ZEROBit state to set for the field.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
EnableenumOFF | ONEnables or disables the checking condition for the selected frame.

Sets the bit state of a field that only consists of one bit for the software trigger.

NameTypeValuesDescription
bitstateenumONE | ZEROBit state to set for the field.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
EnableenumOFF | ONEnables or disables the checking condition for the selected frame.

Presets the state of the selected frames and error types for the software trigger.

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
datanumeric-

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
Data--

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Frameinteger-Index of the frame.
Fieldinteger-Index of the field.
Operatorenum-Operator for the data pattern.
enum - Returns the current data pattern operator for the selected field of the selected frame.

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe. Use this command together with the corresponding data pattern start value command.

NameTypeValuesDescription
Data_Maxreal-End value of the data pattern range.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
frinteger-Index of the frame
flinteger-Index of the field
Data_Operator--

Sets the end value of a data pattern range for the software trigger in the selected frame and field, if the operator is set to INRange or OORANGe.

NameTypeValuesDescription
Data_Maxreal-End value of the data pattern range.

Sets the start value of a data pattern range for the software trigger. Use with the selected frame and field.

NameTypeValuesDescription
Data_Minreal-Start value of the data pattern range.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Data_Operatorenum-Data pattern operator for the software trigger.

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe. You can set the operator with TRIGger:SBSW:ARINc:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
data--

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
Data--

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
operatorenum-Operator for the data pattern.

Defines the error type for the software trigger.

NameTypeValuesDescription
minteger1Index of the error.
enableenumOFF | ONEnables or disables the selected error type.

Enables or disables the error type for the selected error index of the software trigger.

NameTypeValuesDescription
EnableenumOFF | ONEnables or disables the error type.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
EnablerenumOFF | ONEnables or disables the checking condition.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
frinteger1Index of the frame.
flinteger1Index of the field.
CondEnablerenumOFF | ONEnables or disables the checking condition.

Sets the bit state of a field that only consists of one bit for the software trigger.

NameTypeValuesDescription
BitStateenumONE | ZEROBit state to set for the selected one-bit field.

Sets the bit state of a one-bit field for the software trigger.

NameTypeValuesDescription
BitStateenumONE | ZEROBit state to set for the selected field.
enum - Returns the bit state of the selected field.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
enableenumOFF | ONEnables or disables the checking condition for the selected frame.

Sets the bit state of a field that only consists of one bit for the software trigger.

NameTypeValuesDescription
bitenumONE | ZEROBit state to set for the selected field.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
<Frame>--Index of the frame.
<Enabler>enumOFF | ONTurns the checking condition off or on.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
enablerenumOFF | ONEnables or disables the checking condition for the selected frame.
enum - Returns the enable state for the selected frame.

Inverts the current state of the frame and error types for the software trigger; all enabled frames and error types become disabled and vice versa.

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe.

NameTypeValuesDescription
Frameinteger-Index of the frame.
Fieldinteger-Index of the field.
Data--

Returns the end value of a data pattern range for the software trigger. The command is used when the operator is set to INRange or OORANGe.

NameTypeValuesDescription
data--
returns - Returns the end value of the data pattern range for the software trigger.

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
Frameinteger-Index of the frame.
Fieldinteger-Index of the field.
Data--

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
datareal-Data pattern or range start value.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Frameinteger-Index of the frame.
Fieldinteger-Index of the field.
Operatorenum-Operator for the data pattern.
enum - Returns the data pattern operator for the selected field of the selected frame.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Frameinteger-Index of the frame.
Fieldinteger-Index of the field.
Operatorenum-Data pattern operator.
enum - Returns the operator for the data pattern of the software trigger.

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
Data_Maxnumeric-End value of the data pattern range.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
Data_Operator--

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
frame--Index of the frame.
field--Index of the field.
enablerenumOFF | ONEnables or disables the checking condition.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
frame--Index of the frame.
field--Index of the field.
enablerenumOFF | ONEnables or disables the checking condition.
enum - Returns the enabler state.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
CondEnablerenumOFF | ON

Sets the end value of an index range for the software trigger if the operator is set to INRange. You can set the operator with TRIGger:SBSW:MILStd:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
index_maxinteger1 - 65535End value of the index range.

Sets the end value of an index range for the software trigger if the operator is set to INRange. You can set the operator with TRIGger:SBSW:MILStd:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
frame--Index of the frame
field--Index of the field
datainteger1 - 65535End value of the index range.

Sets the end value of an index range for the software trigger when the operator is set to INRange. The operator is set with TRIGger:SBSW:MILStd:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
datainteger1 - 65535End value of the index range.
integer - Returns the end value of the index range when queried.

Defines the error type for the software trigger.

NameTypeValuesDescription
enableenumOFF | ONSets the error enable state.

Defines the error type for the software trigger.

NameTypeValuesDescription
EnablerenumOFF | ONEnables or disables the error type.
enum - Returns the enable state.

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
index_mininteger1 - 65535Start value of the index range.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
Index_OperatorenumEQUal | INRange | RANGeOperator for the index.

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
Data--

Returns the data value for the selected MIL-STD software trigger index setting.

NameTypeValuesDescription
data--Data value for the selected frame and field.
returns - Returns the current data value for the selected MIL-STD software trigger index setting.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
Frame--Frame index placeholder used in the command syntax.
Field--Field index placeholder used in the command syntax.
OperatorenumEQUal | INRange | RANGeOperator to use for the index.
enum - Returns the operator for the index in the selected field of the selected frame.

Sets or queries the operator for the index in the selected field of the selected frame for the software trigger.

enum - Returns the index operator.

Inverts the current state of the frame and error types for the software trigger; all frames and error types that were enabled are disabled and vice versa.

Presets the state of the selected frames and error types for the software trigger.

Sets the bit state of a field that only consists of one bit for the software trigger.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
bitenumONE | ZEROBit state to set for the selected one-bit field.

Returns the bit state of a field that only consists of one bit for the software trigger.

enum - Returns the bit state of the selected field.

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe. Use it together with the field operator command TRIGger:SBSW:SWIRe:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
data_max--

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe. The operator can be set with the corresponding DOPerator command.

NameTypeValuesDescription
data_max--

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
enableenumOFF | ONEnables or disables the checking condition.

Sets the bit state of a field that only consists of one bit for the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame
flinteger-Index of the field
BitStateenumONE | ZERO

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
data_max--End value of the data pattern range.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
frameinteger-Index of the frame.
enablerenumOFF | ONEnables or disables the checking condition.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
enablerenumOFF | ONEnables or disables the checking condition for the selected frame.
enum - Returns the enable state of the selected frame for the software trigger.

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
Data--

Queries the start value of a data pattern range for the software trigger. The value applies to the specified frame and field.

returns - Returns the data pattern range start value for the specified frame and field.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
operatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator used for the data pattern comparison.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator used for the data pattern comparison.
enum - Returns the current data-pattern operator.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
enablerenum-

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
enabler--
returns

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator used for the data pattern of the software trigger.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
CondEnabler--Condition enable state for the selected frame field.

Defines the error type for the software trigger.

NameTypeValuesDescription
<ErrorName>--
<Enabler>enumOFF | ON

Queries the software trigger error enable state for the specified error. The response returns the enable setting.

enum - Returns the enable state.

Inverts the current state of the frame and error types for the software trigger. All frames and error types that were enabled are disabled and vice versa.

Sets the bit state of a field that only consists of one bit for the software trigger.

NameTypeValuesDescription
BitStateenumONE | ZEROBit state to set for the selected field.

Sets the bit state of a field that only consists of one bit for the software trigger.

NameTypeValuesDescription
bitenumONE | ZEROBit state to set.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
EnableenumOFF | ONEnables or disables the checking condition for the selected frame.

Sets the bit state of a field that only consists of one bit for the software trigger.

NameTypeValuesDescription
BitStateenumONE | ZEROBit state to set for the selected field.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
EnablerenumOFF | ONEnables or disables the checking condition.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
enablerenumOFF | ONEnables or disables the checking condition for the selected frame.
enum - Returns the enabled state for the selected frame.

Presets the state of the selected frames and error types for the software trigger.

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe.

NameTypeValuesDescription
Frameinteger-Index of the frame.
Fieldinteger-Index of the field.
Data--Data value for the pattern range end.

Returns the end value of a data pattern range for the software trigger. Use it when the operator is set to INRange or OORANGe.

NameTypeValuesDescription
datareal-End value of the data pattern range.
real - Returns the end value of a data pattern range for the software trigger.

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
data--Data pattern or start value of the range.

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Frameinteger-Index of the frame.
Fieldinteger-Index of the field.
OperatorenumINRange | OORANGeOperator for the data pattern.
enum - Returns the current operator for the data pattern of the software trigger in the selected field of the selected frame.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
Operator--Data operator to use for the software trigger pattern.
returns - Returns the current data operator for the selected field of the selected frame.

Specifies the start value of a data pattern range for the software trigger. It applies to the selected frame and field.

NameTypeValuesDescription
Data_Minreal-Start value of a data pattern range.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
EnablerenumOFF | ONTurns checking for the selected field on or off.

Returns the current field-enable state for the selected software trigger field.

enum - Returns the field enable state.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
frinteger1 - 65535Index of the frame.
flinteger1 - 65535Index of the field.
CondEnablerenumOFF | ONEnables or disables the checking condition.

Sets the end value of an index range for the software trigger if the operator is set to INRange. Use TRIGger:SBSW:SPMI:FRAMe<fr>:FLD<fl>:IOPerator to set the operator.

NameTypeValuesDescription
index_maxinteger1 - 65535End value of the index range.

Sets the end value of an index range for the software trigger when the operator is set to INRange. The operator is set with TRIGger:SBSW:SPMI:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
datainteger1 - 65535End value of the index range.

Sets the end value of an index range for the software trigger when the operator is set to INRange. The operator is set with TRIGger:SBSW:SPMI:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
datainteger1 - 65535End value of the index range.
integer - Returns the end value of the index range for the software trigger.

Defines the error type for the software trigger.

NameTypeValuesDescription
EnablerenumOFF | ONSets the error enable state.
enum - Returns the error enable state.

Defines the error type for the software trigger.

enum - Returns the current enable state for the error type.

Specifies the index, or sets the start value of an index range, for the software trigger.

NameTypeValuesDescription
Index_Mininteger1 - 65535Index minimum value for the selected frame and field.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
Index_OperatorenumEQUal | INRange | RANGeOperator used for the index.

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
frame--
field--
data--
TRIGger:SBSW:SPMI:IMIN <Frame>,<Field>,<Data>
TRIGger:SBSW:SPMI:IMIN? <Data>
TRIGger:SBSW:SPMI:FRAMe<fr>:FLD<fl>:IMIN <Index_Min>

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
Data--
returns

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
operatorenumEQUal | INRange | RANGeOperator used for the index.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
operatorenumEQUal | INRange | RANGeOperator for the index in the selected field of the selected frame.
enum - Returns the index operator.

Inverts the current state of the frame and error types for the software trigger. All frames and error types that were enabled are disabled, and vice versa.

Sets the bit state of a one-bit field for the software trigger.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
bitenumONE | ZEROBit state to set.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
EnableenumOFF | ON

Sets the bit state of a field that only consists of one bit for the software trigger.

NameTypeValuesDescription
BitStateenumONE | ZEROBit state to set.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
EnablerenumOFF | ONEnables or disables the selected frame checking condition.

Presets the state of the selected frames and error types for the software trigger.

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe. You can set the operator with TRIGger:SBSW:RFFE:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
Frameinteger-Index of the frame.
Fieldinteger-Index of the field.
Datanumeric-End value of the data pattern range.

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
frame--Index of the frame.
field--Index of the field.
data--

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
operatorenum-
enum - Returns the current operator for the data pattern of the software trigger in the selected field of the selected frame.

Sets the end value of a data pattern range for the software trigger when the operator is set to INRange or OORANGe. The frame and field suffixes select the target frame and field.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
Data_Max--End value of the data pattern range.

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
Data_Min--

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
operator--Operator for the data pattern.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
EnablerenumOFF | ONEnables or disables the checking condition.
enum - Returns the enable state of the selected field of the selected frame.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
CondEnablerenumOFF | ONEnables or disables the checking condition.

Sets the end value of an index range for the software trigger when the operator is set to INRange. The operator is set with the corresponding IOPerator command.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
Index_Maxinteger1 - 65535End value of an index range.

Sets the end value of an index range for the software trigger if the operator is set to INRange. The operator is set with the corresponding IOPerator command.

NameTypeValuesDescription
frameinteger1 - 65535Index of the frame.
fieldinteger1 - 65535Index of the field.
datainteger1 - 65535End value of the index range.

Defines the error type for the software trigger.

NameTypeValuesDescription
enablerenumOFF | ONSelects whether the error type is enabled.

Defines the error type for the software trigger.

NameTypeValuesDescription
enableenumOFF | ONEnables or disables the error type for the selected error index.

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
Index_Mininteger1 - 65535Sets the start value of an index range for the software trigger.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
operatorenumEQUal | INRange | RANGeOperator for the index.

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
datainteger1 - 65535Index value or start value of an index range.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
operatorenumEQUal | INRange | RANGeOperator used for the index comparison.
enum - Returns the current index operator.

Inverts the current state of the frame and error types for the software trigger. All frames and error types that were enabled are disabled and vice versa.

Sets the bit state of a field that only consists of one bit for the software trigger.

NameTypeValuesDescription
bitstateenumONE | ZEROBit state for the selected field.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
enableenumOFF | ONEnables or disables the checking condition for the selected frame.

Sets the bit state of a field that only consists of one bit for the software trigger.

NameTypeValuesDescription
BitStateenumONE | ZEROBit state to assign to the selected field.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
EnablerenumOFF | ONEnables or disables the frame checking condition.

Presets the state of the selected frames and error types for the software trigger.

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe.

NameTypeValuesDescription
Frameinteger-Index of the frame.
Fieldinteger-Index of the field.
Data--

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
Data_Min--

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
operatorenum-Data pattern operator to apply.

Sets the start value of a data pattern range for the software trigger in the selected frame and field. The command is used for the data pattern trigger and applies to the field identified by the frame and field suffixes.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
data_minnumeric-Start value of the data pattern range.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
EnablerenumOFF | ONEnables or disables the checking condition.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
CondEnablerenumOFF | ONEnables or disables the checking condition.

Sets the end value of an index range for the software trigger when the operator is set to INRange. The operator is set with TRIGger:SBSW:TNOS:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
index_maxinteger1 - 65535End value of an index range.

Sets the end value of an index range for the software trigger when the operator is set to INRange. The operator is set with TRIGger:SBSW:TNOS:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
frame--Index of the frame.
field--Index of the field.
datainteger1 - 65535End value of the index range.

Defines the error type for the software trigger.

NameTypeValuesDescription
ErrorName--
EnablerenumOFF | ON

Specifies the start value of an index range for the software trigger for the selected frame and field. The command uses the frame and field suffixes to address a specific index element.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
Index_Mininteger1 - 65535Start value of the index range.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
Index_OperatorenumEQUal | INRange | RANGe

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
Frameinteger1 - 65535Index of the frame.
Fieldinteger1 - 65535Index of the field.
Datainteger1 - 65535Index minimum value.
integer - Returns the start value of an index range for the software trigger.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
OperatorenumEQUal | INRange | RANGeOperator used for the index comparison.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
EnableenumOFF | ONSets whether the selected frame is enabled or disabled for the software trigger.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
EnablerenumOFF | ONEnables or disables the checking condition for the selected frame.

Inverts the current state of the frame and error types for the software trigger, switching enabled items to disabled and vice versa.

Sets the bit state of a one-bit field for the software trigger.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
bitinteger-

Sets the end value of a data pattern range for the software trigger when the operator is set to INRange or OORANGe. The operator can be set with TRIGger:SBSW:USBone:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
Frameinteger-Index of the frame.
Fieldinteger-Index of the field.
Datastring-List of comma-separated values.

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
Data_Minstring-List of comma-separated values.

Sets the bit state of a one-bit field for the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
BitStateenumONE | ZEROBit state to set for the field.

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe.

NameTypeValuesDescription
data_Maxblock-List of comma-separated values.

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
Data_Minstring-List of comma-separated values.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
frame--Index of the frame.
field--Index of the field.
operatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeSets the data operator for the selected field.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
EnablerenumOFF | ONEnables or disables the checking condition for the selected field of the selected frame of the software trigger.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
data_operatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator for the data pattern.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
flinteger-Index of the field.
enablerenumOFF | ONEnables or disables the checking condition.

Sets the end value of an index range for the software trigger if the operator is set to INRange.

NameTypeValuesDescription
datainteger-End value of the index range.

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
Index_Mininteger1 - 65535Start value of the index range.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
index_operatorenumEQUal | INRange | RANGeOperator for the index in the selected field of the selected frame.

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
Frameinteger1 - 65535Frame index.
Fieldinteger1 - 65535Field index.
Data--

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
Index_OperatorenumEQUal | INRange | RANGeOperator for the index in the selected field of the selected frame.

Defines the error type for the software trigger.

NameTypeValuesDescription
<ErrorName>--
<Enabler>enumOFF | ON

Defines the enable state for the specified software-trigger error type. The suffix <m> identifies the error index.

NameTypeValuesDescription
enableenumOFF | ONEnables or disables the error type.

Sets the bit state of a field that only consists of one bit for the software trigger.

NameTypeValuesDescription
BitStateenumONE | ZEROBit state to set for the field.

Sets the bit state of a one-bit field for the software trigger. The command is asynchronous.

NameTypeValuesDescription
BitStateenumONE | ZEROBit state to set for the selected field.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
enableenumOFF | ONEnables or disables the checking condition for the selected frame.

Sets the bit state of a one-bit field for the software trigger. The command applies to the indexed frame and field.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
BitStateenumONE | ZEROBit value to set.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
EnableenumOFF | ONEnables or disables the checking condition for the selected frame.

Enables or disables the checking condition for the selected frame for the software trigger.

NameTypeValuesDescription
enablerenumOFF | ONEnables or disables the checking condition for the selected frame.
enum - Returns the enable state for the selected frame.

Inverts the current state of the frame and error types for the software trigger, disabling all previously enabled frames and error types and enabling the others.

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
datastring-List of comma-separated values.

Returns the end value of a data pattern range for the software trigger. The command is used when the operator is set to INRange or OORANGe.

returns - Returns the end value of a data pattern range for the software trigger.

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
Data_Minblock-List of comma-separated values.

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

string - Returns the data pattern start value for the software trigger.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
operatorenum-

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

enum - Returns the data pattern operator for the software trigger in the selected field of the selected frame.

Sets the end value of a data pattern range for the software trigger, if the operator is set to INRange or OORANGe. The operator is set with the corresponding DOPerator command.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
Data_Maxnumeric-List of comma-separated values.

Specifies the data pattern, or sets the start value of a data pattern range for the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
Data_Minnumeric-List of comma-separated values.

Sets the operator for the data pattern of the software trigger in the selected field of the selected frame.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
operatorenum-Operator for the data pattern.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
Frame--Frame index.
Field--Field index.
EnablerenumOFF | ONEnables or disables the checking condition.
enum - Returns the current enabler state.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

enum - Returns the enable state of the checking condition for the selected frame field.

Enables or disables the checking condition for the selected field of the selected frame of the software trigger.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
condenablerenumOFF | ONEnables or disables the checking condition for the selected field.

Sets the end value of an index range for the software trigger when the operator is set to INRange. The frame and field are selected by the frame and field suffixes.

NameTypeValuesDescription
frinteger1Index of the frame.
flinteger1Index of the field.
index_maxinteger1 - 65535End value of the index range.

Sets the end value of an index range for the software trigger if the operator is set to INRange. The operator is set with TRIGger:SBSW:USBone:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
index_maxinteger1 - 65535End value of the index range.

Sets the end value of an index range for the software trigger if the operator is set to INRange. The operator is set with TRIGger:SBSW:USBone:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
index_maxinteger1 - 65535End value of the index range.
integer - Returns the end value of an index range.

Defines the error type for the software trigger.

NameTypeValuesDescription
EnablerenumOFF | ON
enum - Returns the current enable state of the error type.

Defines the error type for the software trigger.

NameTypeValuesDescription
EnableenumOFF | ONEnables or disables the error type.
enum - Returns the current enable state for the selected error type.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
Index_OperatorenumEQUal | INRange | RANGeOperator for the index in the selected frame/field.

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
Frameinteger1 - 65535Index of the frame.
Fieldinteger1 - 65535Index of the field.
Data--

Specifies the index, or sets the start value of an index range for the software trigger.

NameTypeValuesDescription
Data--

Sets the operator for the index in the selected field of the selected frame for the software trigger. The manual shows the operator choices as EQUal, INRange, and RANGe.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
OperatorenumEQUal | INRange | RANGeOperator used for the index in the selected field of the selected frame.

Sets the operator for the index in the selected field of the selected frame for the software trigger.

NameTypeValuesDescription
OperatorenumEQUal | INRange | RANGeOperator for the index in the selected frame and field.

Returns the CONDition part of the trigger analog edge overload status. This is a status-reporting query under the questionable status group.

integer - Returns the condition part of the TRIGger:ANEDge:OVERload:STATus status.

Returns or sets the contents of the EVENT part of the overload status register for trigger actions output. Reading an EVENT register deletes its contents.

NameTypeValuesDescription
Value--

Returns the contents of the EVENt part of the overload status register. Reading this register deletes its contents.

string - Bit-dependent, comma-separated values indicating event status; see the source values for the related status commands for the list of possible return values.

Sets the ENABle part that allows true conditions in the EVENt part to be reported for the summary bit in the status byte. The enabled events can be used for a service request and cause a positive transition in the summary bit when the corresponding event bit is true.

NameTypeValuesDescription
Valuestring-Bit-dependent, comma-separated values.

Sets the negative transition part for the trigger action output overload status. This allows true conditions in the EVENt part to be reported for the summary bit in the status byte.

NameTypeValuesDescription
Value--Bit dependent, comma-separated values.

Sets the ENABle part that allows true conditions in the EVENt part to be reported for the summary bit in the status byte. If a bit in the ENABle part is 1 and the corresponding EVENt bit is true, a positive transition occurs in the summary bit.

NameTypeValuesDescription
Value--Bit dependent, comma-separated values.

Sets the negative-transition part for the trigger analog edge overload status reporting. It controls which true conditions in the EVENt part are reported upward through the status byte summary bit.

NameTypeValuesDescription
<Value>--Bit dependent, comma-separated values.

Sets the positive transition filter for the selected status register path. A bit set causes a transition from 0 to 1 in the condition part to be entered into the corresponding bit of the event part of the register.

NameTypeValuesDescription
valuestring-Source dependent, comma-separated values

Sets the positive transition filter for the trigger analog edge overload status register. If a bit is set, a transition from 0 to 1 in the condition part causes an entry in the corresponding bit of the EVENT part of the register.

NameTypeValuesDescription
Valuestring-Source dependent, comma-separated values.

*SRE?query

Returns the Service Request Enable (SRE) register. The SRE can be set with *SRE and read with *SRE?.

integer - Returns the current value of the SRE register.
*ESEset / query

Sets the event status enable register (ESE). The register can also be read back with *ESE?.

NameTypeValuesDescription
enable registerinteger0 - 255Bit pattern written to the event status enable register.
*ESE?query

Reads the event status enable register (ESE). The register controls which event status bits are enabled to set the ESB bit in the status byte.

integer - Returns the current ESE register value as a bit mask.
*ESR?query

Reads out the event status register (ESR). The ESR is defined in IEEE 488.2 and corresponds to the event part of a SCPI register.

integer - Returns the event status register value.
*SRQ?query

Generates a service request indication. The example shows that issuing the command results in a service request being sent.

boolean - Indicates whether a service request was sent.
*SRQ?
*CAL?query

Queries the calibration status of the instrument.

enum - Returns the calibration status.
*CLSset / query

Clears the status data structures. This command does not return a response.

*IDN?query

Returns the instrument identification string. The response includes the manufacturer, model, serial number, and firmware revision.

string - Instrument identification string containing the manufacturer, model, serial number, and firmware revision.
*OPCset / query

Sets the operation complete bit in the Standard Event Status Register when all preceding overlapped commands have finished. It is used to synchronize command processing.

*OPT?query

Returns the option code of the installed options. The response is a comma-separated list of option codes.

string - A comma-separated list of option codes identifying the installed options.
*RCLset / query

Recalls a previously saved instrument state or setup from nonvolatile memory.

NameTypeValuesDescription
setupinteger0 - NMemory location to recall.
*RSTset / query

Resets the instrument to its default state.

*SAVset / query

Saves the current instrument state to nonvolatile memory in the specified save area. The saved setup can later be recalled with *RCL.

NameTypeValuesDescription
registerinteger0 - 9Save register number.
*SREset / query

Sets the event status register enable mask. This command controls which event status bits are summarized in the status byte.

NameTypeValuesDescription
enable_maskinteger0 - 255Bit mask enabling event status register bits.
*STB?query

Returns the contents of the Status Byte register. The register summarizes the state of the instrument’s event and status reporting system.

integer - Status Byte register value.
*TRGset / query

Triggers the selected subsystem or initiation event. This is a common IEEE 488.2 common command with no parameters.

*TST?query

Performs a self-test and returns the result.

integer - Returns the self-test result code.
*WAIset / query

Waits until all previously sent commands have been executed. It is used to synchronize command execution before sending subsequent commands.

Reads the condition register of the STATus:QUEStionable status subsystem. The register contains information on instrument operation and signal reliability.

integer - Returns the current contents of the STATus:QUEStionable condition register as a bit-coded value.

Returns the contents of the STATus:QUEStionable register as a single decimal number. The register contains specific information on instrument operation and signal reliability.

integer - The contents of the questionable status register, with each set bit contributing its weighted value to the returned decimal number.
:STATus:PRESetset / query

Resets the status registers. PTRansition bits are set to 1, NTRansition bits to 0, EVENt bits to 0, and the ENABle bits of the status registers to 0.

Returns the bit of the action the instrument is currently executing. The contents of the STATus:OPERation register is retained.

integer - Returns the bit of the action the instrument is currently executing.

Controls the ENABle part of the STATus:OPERation register. It defines which events in the EVENt part of the status register are forwarded to the OPERation summary bit of the status byte.

NameTypeValuesDescription
Enableinteger1 - 65535Sets which operation status events are enabled for forwarding to the summary bit.
STATus:OPERation:ENABle 5

Returns information on actions the instrument has executed since the last reading. Reading the EVENt register deletes its contents.

integer - Returns the event register contents as bit-coded information about executed actions since the last reading. The register contents are cleared after reading.

Sets the bits of the NTRansition part of the STATus:OPERation register. A bit set in the NTRansition register causes a bit transition from 1 to 0 in the CONDition register to produce an entry in the EVENt register; reading the information does not clear the register.

NameTypeValuesDescription
SumBitinteger0 - 65535Bit mask to set in the NTRansition register.

Sets the bits of the PTRansition part of the STATus:OPERation register. A bit set in the PTRansition register causes a bit transition from 0 to 1 in the CONDition register to produce an entry in the EVENt register.

NameTypeValuesDescription
SumBitinteger0 - 65535Bit mask for the PTRansition register.

Reads out the condition part of the Questionable status register for generator overload. The command is listed among the Questionable status condition queries.

integer - Returns the condition part of the Questionable status register for generator overload.

Reads out the condition part of the specified questionable status register bit. It is listed among the status-reporting condition queries.

integer - Returns the condition part of the questionable status register bit.

Returns the condition part of the STATus:QUEStionable:LIMit event register. This is a query-only command.

integer - Returns the condition part of the questionable limit status register.

Returns the condition part of the questionable status register mask. This is a readout-only query.

integer - Returns the condition register value.

Reads out the CONDition part of the questionable status register for PLL. It is listed among the status condition queries in the questionable status section.

integer - Returns the current condition register value for STATus:QUEStionable:PLL.

Returns the contents of the EVENt part of the questionable status register to check if an event has occurred since the last reading. Reading an EVENt register deletes its contents.

string - Bit-dependent, comma-separated values; see the source values for STATus:QUEStionable:...:[:EVENt] and STATus:QUEStionable:...:[:ENABLe] for the list of return values.

Returns the contents of the EVENt part of the questionable status register to check if an event has occurred since the last reading. Reading an EVENt register deletes its contents.

enum - Bit-dependent, comma-separated status/event values as documented in the source values tables for STATus:QUEStionable:...:[:EVENt] and STATus:QUEStionable:...:[:ENABLe].

Returns the contents of the EVENt part of the questionable status register to check whether an event has occurred since the last reading. Reading an EVENt register deletes its contents.

enum - Bit-dependent, comma-separated source values for the questionable status event register.

Returns the contents of the EVENt part of the questionable status register to check if an event has occurred since the last reading. Reading an EVENt register deletes its contents.

bit dependent, comma-separated values - Bit-dependent, comma-separated values; see the Source values for STATus:QUEStionable:...:[:EVENt] and STATus:QUEStionable:...:[:ENABLe] for the list of return values.

Returns the contents of the EVENt part of the questionable status register to check if an event has occurred since the last reading. Reading an EVENt register deletes its contents.

bit dependent, comma-separated values - Returns the contents of the EVENt part of the questionable status register. See the Source values for STATus:QUEStionable:...:[:EVENt] and STATus:QUEStionable:...:[:ENABLe] for the list of return values.

Returns the contents of the EVENt part of the questionable status register to check whether an event has occurred since the last reading. Reading an EVENt register deletes its contents.

string - Bit-dependent, comma-separated values; see the source values for STATus:QUEStionable:...:[:EVENt] and STATus:QUEStionable:...:[:ENABLe] for the list of return values.

Returns the contents of the EVENt part of the questionable status register to check whether an event has occurred since the last reading. Reading an EVENt register deletes its contents.

enum - Bit-dependent, comma-separated values; see the Source values for STATus:QUEStionable:...:[:EVENt] and STATus:QUEStionable:...:[:ENABLe] for the list of return values.

Returns the contents of the CONDition part of the questionable status register for the power supply status group. Reading the CONDition register does not delete its contents, since it reflects the current hardware status.

string - Bit-dependent, comma-separated values from the CONDition part of the status register.

Returns the contents of the EVENt part of the questionable power supply status register to check if an event has occurred since the last reading. Reading an EVENt register deletes its contents.

bit dependent, comma-separated values - Bit-dependent, comma-separated event values from the status register. See the Source values for STATus:QUEStionable:...:[:EVENt] and STATus:QUEStionable:...:[:ENABLe] listings for the possible return values.

Returns the contents of the CONDition part of the questionable status register for temperature states. Reading the CONDition register does not delete its contents because it indicates current hardware status.

string - Bit dependent, comma-separated values from the CONDition part of the questionable status register.

Returns the contents of the EVENt part of the questionable status register to check if an event has occurred since the last reading. Reading an EVENt register deletes its contents.

discrete - Bit-dependent, comma-separated values; see the source values for STATus:QUEStionable:...:[:EVENt] and STATus:QUEStionable:...:[:ENABLe] for the list of return values.

Sets the ENABle part that allows true conditions in the EVENt part to be reported for the summary bit in the status byte. If a bit in the ENABle part is 1 and the corresponding EVENt bit is true, a positive transition occurs in the summary bit.

NameTypeValuesDescription
Valuediscrete-Bit dependent, comma-separated values.

Sets the ENABle part that allows true conditions in the EVENt part to be reported for the summary bit in the status byte. These events can be used for a service request.

NameTypeValuesDescription
Valuenumeric-Bit-dependent, comma-separated values.

Sets the ENABle part that allows true conditions in the EVENt part to be reported for the summary bit in the status byte. If a bit in the ENABle part is 1 and the corresponding EVENt bit is true, a positive transition occurs in the summary bit.

NameTypeValuesDescription
Valuenumeric-Bit dependent, comma-separated values.

Sets the ENABle part that allows true conditions in the EVENt part to be reported for the summary bit in the status byte. If a bit in the ENABle part is 1 and the corresponding EVENt bit is true, a positive transition occurs in the summary bit.

NameTypeValuesDescription
Valuenumeric-Bit-dependent, comma-separated values.

Sets the ENABle part that allows true conditions in the EVENt part to be reported for the summary bit in the status byte. These events can be used for a service request; if a bit in the ENABle part is 1 and the corresponding EVENt bit is true, a positive transition occurs in the summary bit.

NameTypeValuesDescription
Value--Bit dependent, comma-separated values.

Sets the enable part for the questionable power supply status reporting. If the corresponding event bit is true and the enable bit is 1, a positive transition occurs in the summary bit and can be used for service request reporting.

NameTypeValuesDescription
valueblock-Bit-dependent, comma-separated values.

Sets the ENABle part that allows true conditions in the EVENt part to be reported in the status byte summary bit. If the ENABle bit is 1 and the corresponding EVENt bit is true, a positive transition is reported to the next higher level.

NameTypeValuesDescription
valueenumbit dependent | comma-separated valuesBit dependent, comma-separated values.

Sets the negative transition filter for the questionable ADC state status register. If a bit is set, a transition from 1 to 0 in the condition part causes an entry in the corresponding bit of the event part of the register.

NameTypeValuesDescription
Value--Source dependent, comma-separated values.

Sets the positive transition filter for the questionable ADC state status register. If a bit is set, a transition from 0 to 1 in the condition part causes an entry to be made in the corresponding bit of the event part of the register.

NameTypeValuesDescription
valuenumeric-Source-dependent, comma-separated values.

Sets the negative transition filter for the questionable overload status register. If a bit is set, a transition from 1 to 0 in the condition part causes an entry in the corresponding bit of the event part of the register.

NameTypeValuesDescription
Value--Source dependent, comma-separated values.

Sets the positive transition filter for the questionable COVerload status register. If a bit is set, a 0-to-1 transition in the condition part causes an entry in the corresponding bit of the EVENt part of the register.

NameTypeValuesDescription
Valuestring-Source dependent, comma-separated values

Sets the negative transition filter for the questionable status register. If a bit is set, a transition from 1 to 0 in the condition part causes an entry to be made in the corresponding bit of the event part of the register.

NameTypeValuesDescription
value--Source dependent, comma-separated values.

Sets the positive transition filter for the questionable status register. If a bit is set, a transition from 0 to 1 in the condition part causes an entry to be made in the corresponding bit of the event part of the register.

NameTypeValuesDescription
Value--Source dependent, comma-separated values.

Sets the negative transition filter for the questionable status register. If a bit is set, a transition from 1 to 0 in the condition part causes an entry to be made in the corresponding bit of the EVENt part of the register.

NameTypeValuesDescription
Value--Source dependent, comma-separated values

Sets the positive transition filter for the questionable status register. If a bit is set, a transition from 0 to 1 in the condition part causes an entry in the corresponding bit of the event part of the register.

NameTypeValuesDescription
value--Source dependent, comma-separated values.

Sets the negative transition filter for the questionable status register. If a bit is set, a transition from 1 to 0 in the condition part causes an entry in the corresponding bit of the EVENT part of the register.

NameTypeValuesDescription
Value--Source dependent, comma-separated values.

Sets the positive transition filter for the questionable limit status register. If a bit is set, a transition from 0 to 1 in the condition part causes an entry to be made in the corresponding bit of the event part of the register.

NameTypeValuesDescription
Value--Source dependent, comma-separated values.

Sets the negative transition filter for the Questionable status register subtree. If a bit is set, a 1-to-0 transition in the condition part causes an entry in the corresponding event bit.

NameTypeValuesDescription
Valuediscrete-Source dependent, comma-separated values

Sets the positive transition filter for the STATus:QUEStionable:PLL register. If a bit is set, a transition from 0 to 1 in the condition part causes an entry to be made in the corresponding bit of the EVENt part of the register.

NameTypeValuesDescription
valuediscrete-Source dependent, comma-separated values.

Sets the negative transition filter for the questionable power supply status register. If a bit is set, a transition from 1 to 0 in the condition part causes an entry in the corresponding bit of the event part of the register.

NameTypeValuesDescription
Valuestring-Source dependent, comma-separated values

Sets the positive transition filter for the questionable status register. If a bit is set, a transition from 0 to 1 in the condition part causes an entry in the corresponding event bit of the register.

NameTypeValuesDescription
valuestring-Source dependent, comma-separated values.

Sets the negative transition filter for the questionable temperature status register. If a bit is set, a transition from 1 to 0 in the condition part causes an entry to be made in the corresponding bit of the event part of the register.

NameTypeValuesDescription
value--Source dependent, comma-separated values

Sets the positive transition filter for the Questionable Temperature status register. If a bit is set, a transition from 0 to 1 in the condition part causes an entry to be made in the corresponding bit of the EVENT part of the register.

NameTypeValuesDescription
Valuestring-Source dependent, comma-separated values

Resets the condition to allow operation to continue after a PLL unlock of the external reference.

NameTypeValuesDescription
ValueenumPLL100Reset condition value.

Enables bits in the STATus:QUEStionable register. In the example, it is used to enable the LIMIt bit.

NameTypeValuesDescription
enableinteger0 - 512Bits to enable in the STATus:QUEStionable register.
STATus:QUEStionable:ENABle 512

Enables the limit check for the active measurements in the STATus:QUEStionable register. The example shows it can be set to ALL, and the corresponding query returns which measurements were active.

NameTypeValuesDescription
stateenumALLSelects which measurements have the limit check enabled.
STATus:QUEStionable:LIMit:ENABle ALL
STATus:QUEStionable:LIMit:ENABle?

Queries which measurements were active at the time the limit-enable command was sent. The query returns the active measurement list.

string - Active measurements at the time STATus:QUEStionable:LIMit:ENABle was sent.
STATus:QUEStionable:LIMit:ENABle?

Returns the questionable margin condition status. This deprecated command is provided for compatibility with previous oscilloscope versions only; use STATus:QUEStionable:LIMit: commands instead.

enum - Returns the status condition as a bit-coded status register value.

Deprecated status-reporting command for the questionable margin enable setting. The manual lists it as a legacy command and indicates it was replaced by STATus:QUEStionable:LIMit: commands (obsolete since FW 2.9).

Deprecated command provided for compatibility with previous oscilloscope versions only. For new remote control programs, use the corresponding STATus:QUEStionable:LIMit: commands.

Deprecated status-reporting command for the questionable margin transition status. It is provided for compatibility with previous oscilloscope versions only; use the corresponding :STATus:QUEStionable:LIMit command instead.

Returns the event status register bit for the questionable margin status group. This deprecated query is provided for compatibility with previous oscilloscope versions; use the corresponding limit status commands instead.

enum - Returns the questionable margin event status bit.

Enables or disables display of an additional text in remote control. To define the text, use SYSTem:DISPlay:MESSage[:TEXT].

NameTypeValuesDescription
stateenumOFF | ONEnables or disables the display message.

Defines whether the display is updated while the instrument is in the remote state. If the display is switched off, the normal GUI is replaced by a static image; switching it off can speed up measurement, and OFF is recommended.

NameTypeValuesDescription
DisplayUpdateenumOFF | ONControls whether the display is shown and updated during remote control.
SYSTem:DISPlay:UPDate 1

Sets the host name of the instrument, which is required when configuring a network. After changing the host name, the instrument must be rebooted.

NameTypeValuesDescription
Hostnamestring-String parameter specifying the host name to set.

Defines additional text displayed during remote control operation. Enable the text display with SYSTem:DISPlay:MESSage:STATe.

NameTypeValuesDescription
DisplayMessagestring-Additional text to display during remote control operation.
:SYSTem:DATEset / query

Sets the date of the internal calendar.

NameTypeValuesDescription
Yearinteger2012 - 2099Year, to be entered as a four-digit number including century and millennium information.
Monthinteger1 - 12Month of the year.
Dayinteger1 - maximum number of days in the specified monthDay of the month.
SYSTem:DATE?

Sets the path and filename of the firmware installation file.

NameTypeValuesDescription
FilePathstring-String with path and filename.
:SYSTem:PRESetset / query

Resets the instrument to the default state, with the same effect as *RST.

:SYSTem:FW:STARtset / query

Starts the firmware update. Before starting, make sure that the correct path is set with `SYSTem:FW:FILepath`.

:SYSTem:SHUTdownset / query

Starts the shutdown of the instrument, including firmware and operating system shutdown.

:SYSTem:TIMEset / query

Returns or sets the current time of the clock using hours, minutes, and seconds.

NameTypeValuesDescription
Hoursinteger0 - 24Hour value of the clock.
Minutesinteger0 - 59Minute value of the clock.
Secondsinteger0 - 59Second value of the clock.
:SYSTem:APUPset / query

Enables or disables automatic power-up when the instrument is connected to mains voltage.

NameTypeValuesDescription
AutoPowerUpenumOFF | ONControls whether the instrument powers up automatically on mains connection.

Assigns a color table to the source waveform instead of a dedicated color.

NameTypeValuesDescription
Signalstring-Signal name as returned by DISPlay:COLor:SIGNal:CATalog?.
ColorTableenumFalseColors | Spectrum | SingleEvent | TemperatureName of the color table to assign.

Returns the color of the selected waveform for the specified signal. The signal name must be one returned by DISPlay:COLor:SIGNal:CATalog?.

NameTypeValuesDescription
Signal--Signal name as returned by DISPlay:COLor:SIGNal:CATalog?.

Sets the height of an annotation arrow. The command applies to the specified arrow number <m>.

NameTypeValuesDescription
minteger0Arrow annotation index.

Returns the catalog of annotation colors. This is a query for the color catalog associated with a display annotation color index.

string - Returns the catalog of annotation colors.

Sets the color of a display annotation polyline. The command applies to the specified polyline instance.

NameTypeValuesDescription
minteger1Polyline number to be colored.
colorenum-Color selection for the polyline.

Removes the specified annotation polyline from the display.

NameTypeValuesDescription
minteger1 - MAXimumAnnotation polyline index to remove.

Sets or returns the value of a polyline annotation. The command belongs to the display annotation polyline group.

NameTypeValuesDescription
minteger1 - MAXimumAnnotation index.
valueoptstring-Polyline annotation value.

Sets the width of a rectangle annotation. The command belongs to a specific rectangle annotation identified by <m>.

NameTypeValuesDescription
widthreal0Width of the rectangle annotation.

Sets the color of the specified text annotation. The command belongs to the DISPlay:ANNotation:TEXT<m> subsystem.

NameTypeValuesDescription
colorenum-Color of the text annotation.

Sets the font size for a text annotation. The command belongs to the display annotation text subsystem for annotation <m>.

NameTypeValuesDescription
sizeinteger-Font size to apply to the text annotation.

Sets the horizontal position of a text annotation. The command belongs to the display annotation text subsystem.

NameTypeValuesDescription
positionreal-Horizontal position of the text annotation.

Sets the transparency of display annotations.

NameTypeValuesDescription
transparencyreal0 - 100Transparency level for annotations.

Sets the display backlight brightness. The command is listed alongside DISPlay:BACKlight:DIMMing on the same page.

Assigns a signal color to a signal source in the display color setup. The command is part of the display color signal configuration group.

Returns the catalog of signal colors available for display color assignment.

string - The catalog of signal colors available for display color assignment.

Sets the color used for signals in the display color scheme. The command is part of the display color/signal color subsystem.

Sets whether the diagram fine grid is displayed.

NameTypeValuesDescription
stateenumON | OFFTurns the fine grid display on or off.

Sets the X-fixed mode for the diagram display. The command is listed in the display diagram command index, but this slice does not provide syntax details or parameter values.

Sets the vertical fixed-grid position for the display diagram. The command is listed in the display diagram command index on the referenced page.

Sets the font size used in dialog displays.

NameTypeValuesDescription
fontsizeinteger-Font size for dialog text.

Sets the transparency of dialog boxes displayed on the screen.

NameTypeValuesDescription
transparencyinteger0 - 100Transparency level of the dialog boxes.

Sets the display intensity. The command adjusts how bright waveform and other display elements appear.

NameTypeValuesDescription
intensityreal0 - 100Display intensity level.

Turns infinite persistence display on or off.

NameTypeValuesDescription
stateenumON | OFFEnables or disables infinite persistence.

Sets the persistence time for the display. This command belongs to the display persistence subsystem.

NameTypeValuesDescription
timereal0 - INFPersistence time setting.

Sets or queries the display persistence state. Persistence controls whether previously acquired waveform data remain visible on the display.

NameTypeValuesDescription
stateoptenumON | OFFEnables or disables display persistence.

Returns the number of toolbars currently displayed.

integer - The number of toolbars currently displayed.

Sets the path, filename, and file format of the preset file. The name must be a string with path and filename ending in .set.

NameTypeValuesDescription
Namestring-String with path and filename with extension .set.
USERdefined:PRESet:NAME '/home/storage/userData/settings/Preset_SPI.set'

Saves the current settings as a preset file. The storage location and filename are defined with USERdefined:PRESet:NAME.

Enables or disables use of a user-defined preset. If enabled, the settings from the selected saveset are restored when the Preset key is pressed; if disabled, Preset restores the factory defaults.

NameTypeValuesDescription
NameenumOFF | ONEnables or disables the user-defined preset.

Returns the date of the last self-alignment.

string - The date of the last self-alignment.

Starts the self-alignment process without returning status information. The query form starts self-alignment and returns information on the state of the self-alignment, same as *CAL?.

numeric - Numeric status indicator for the self-alignment state; return values not equal to 0 indicate an error.

Returns the result of the last self-alignment and the current alignment status. In remote mode, *CAL? provides more detailed information.

enum - Returns the last self-alignment result and current alignment status.

Returns the time of the last self-alignment.

returns - Time of the last self-alignment.

:SERVice:REPortset / query

Creates a service report. The report is saved on a connected USB flash drive, or otherwise in /home/storage/userData.

:SINGleset / query

Starts a defined number of acquisition cycles. The number of cycles is set with ACQuire:COUNt.

:STOPset / query

Stops the running acquisition.

:TIMebase:RANGeset / query

Sets the time of one acquisition, i.e. the time across the 10 divisions of the diagram. Acquisition time equals time scale times 10 divisions.

NameTypeValuesDescription
TimebaseRangereal2E-9 - 100E+3Time of one acquisition.
:TIMebase:SCALeset / query

Sets the horizontal scale, the time per division, for all waveforms in the time domain, including channel and math waveforms.

NameTypeValuesDescription
TimebaseScalereal200E-12 - 10E+3Horizontal scale value.

Returns the number of horizontal divisions on the screen. The number cannot be changed.

integer - Number of horizontal divisions on the screen.

Defines the time distance between the reference point and the trigger point, which is the zero point of the diagram. The horizontal position is also known as trigger offset.

NameTypeValuesDescription
Positionreal-159.99E-3 - 1E+26Time distance between the reference point and the trigger point.

Sets the position of the reference point in percent of the screen. It defines which part of the waveform is shown.

NameTypeValuesDescription
RescaleCtrPosreal0 - 100Position of the reference point on the screen.

Selects whether roll mode is started automatically by the instrument or turned off.

NameTypeValuesDescription
modeenum-Selects whether roll mode is started automatically by the instrument or turned off.

Sets the minimum acquisition time for the automatic start of the roll mode.

NameTypeValuesDescription
MinAcquTimereal0.5 - 100000Minimum acquisition time for automatic start of roll mode.

Sets the channel signal coupling, which determines the input impedance/termination and coupling filter used for waveform analysis and triggering.

NameTypeValuesDescription
chinteger1 - 8Index of the analog channel.
usercouplingenumDC | DCLimit | ACChannel coupling mode.

Sets the offset voltage of the selected analog channel, shifting the vertical center of the channel by the specified amount. The offset corrects an offset-affected signal and repositions it within the diagram.

NameTypeValuesDescription
chinteger1 - 8Index of the analog channel.
Offsetreal-1 - 1Offset voltage to apply.

Moves the selected signal up or down in the diagram. The position is a graphical setting given in divisions; the visual effect is the same as for offset.

NameTypeValuesDescription
chinteger1 - 8Index of the analog channel.
positionreal-5 - 5Vertical position of the selected signal in divisions; positive values move the waveform up and negative values move it down.

Turns the inversion of the signal amplitude on or off. Inversion reflects the voltage values of all signal components against the ground level.

NameTypeValuesDescription
chinteger1 - 8Index of the analog channel.
InvertChannelenumOFF | ONSelects whether the waveform is normal or inverted.

Sets the bandwidth limit for the selected analog channel. The query returns a numeric result in Hz.

NameTypeValuesDescription
chinteger1 - 8Index of the analog channel.
BandwidthLimitenumFULL | B20 | B50 | B100 | B200 | B350 | B500 | B700Bandwidth limit setting; numeric values in Hz are also accepted.
numeric (Hz) - Numeric result in Hz.

Sets the attenuation scale for an external divider. The manual states the scale can be linear or logarithmic.

NameTypeValuesDescription
ExtAttSclenumLINear | LOGarithmicAttenuation scale for the external divider.

Sets the external attenuation for the selected analog channel. The instrument uses this value to account for a voltage divider in the DUT before the measuring point.

NameTypeValuesDescription
<ExtAtt>real0.005 - 1000000External attenuation value for the channel; limits shown are for linear scale.

Sets the impedance of the connected probe for power calculations and measurements.

NameTypeValuesDescription
<Impedance>real0.1 - 100000Impedance value for the connected probe.

Returns the header of channel waveform data, describing the waveform attributes. Query only.

NameTypeValuesDescription
chinteger1 - 8Index of the analog channel.
numeric - Returns four comma-separated values: XStart (time of the first sample in s), XStop (time of the last sample in s), RecordLength (number of samples), and ValuesPerSample (number of values per sample).
CHAN1:DATA:HEAD? -1E-07,9.980000000000001E-08,1000,1

Returns the data of the specified channel waveform points for transmission to the controlling computer. Without parameters, the complete waveform is retrieved; with one parameter it is treated as the offset, and with two parameters it retrieves a smaller portion of the waveform.

NameTypeValuesDescription
chinteger1 - 8Index of the analog channel.
Offsetoptinteger0 - mNumber of waveform samples to skip before retrieving data.
Lengthoptinteger1 - nNumber of waveform points to retrieve.
block - List of values according to the format and content settings.
CHANnel:DATA:VALues? 0,10
CHANnel:DATA:VALues? 5,5

Reads out the condition part of the channel ADC state negative clipping status. The command is listed under the status reporting commands.

integer - Returns the condition part of the status register for the specified channel ADC state negative clipping event.

Returns the condition part of the ADC state status questionable register for the selected channel's positive clipping status.

integer - Returns the condition bit field value for the channel ADCState positive clipping status.

Reads out the condition part of the questionable status for the specified channel overload status. It is listed under the status reporting commands for the CONDition part.

integer - Returns the condition part of the specified status register.

Returns the contents of the EVENt part of the status register to check if an event has occurred since the last reading. Reading an EVENt register deletes its contents.

enum - Bit-dependent, comma-separated values; see the Source values for STATus:QUEStionable:...:[:EVENt] and STATus:QUEStionable:...:[:ENABLe] for the list of return values.

Returns the contents of the EVENt part of the status register for the positive clipping state to check whether an event has occurred since the last reading. Reading an EVENt register deletes its contents.

bit dependent, comma-separated values - Returns the contents of the EVENt part of the status register. See the Source values for the corresponding status/event register for the list of return values.

Returns or reads the contents of the EVENt part of the channel overload status register. Reading an EVENt register deletes its contents.

bit dependent - Bit-dependent, comma-separated values indicating the channel overload status event contents; see the source values referenced in the manual.

Sets the ENABle part that allows true conditions in the EVENt part to be reported for the summary bit in the status byte. If a bit in the ENABle part is 1 and the corresponding EVENt bit is true, a positive transition occurs in the summary bit.

NameTypeValuesDescription
Value--Bit dependent, comma-separated values.

Sets the negative transition part for the channel ADC state clipping status register. This command is part of the status reporting control commands.

NameTypeValuesDescription
<Value>--Bit-dependent, comma-separated values.

Sets the ENABle part that allows true conditions in the EVENt part to be reported for the summary bit in the status byte. If a bit in the ENABle part is 1 and the corresponding EVENt bit is true, a positive transition occurs in the summary bit.

NameTypeValuesDescription
Valuenumeric-Bit-dependent, comma-separated values.

Sets the negative transition part for the specified status reporting source. The command is listed among status-questionable controls and uses a bit-dependent comma-separated value.

NameTypeValuesDescription
Valuestring-Bit dependent, comma-separated values.

Sets the ENABle part that allows true conditions in the EVENt part to be reported for the summary bit in the status byte. If the corresponding EVENt bit is true and the ENABle bit is 1, a positive transition is reported to the next higher level.

NameTypeValuesDescription
Valuenumeric-Bit-dependent, comma-separated values.

Sets the negative transition part for the channel overload status reporting. It controls the transition bit used in the questionable status register hierarchy.

NameTypeValuesDescription
Value--Bit-dependent, comma-separated values.

Sets the positive transition filter for the specified channel ADC state clipping status register. If a bit is set, a transition from 0 to 1 in the condition part causes an entry in the corresponding event bit.

NameTypeValuesDescription
Valuestring-Source dependent, comma-separated values

Sets the positive transition filter for the channel ADC state status register. If a bit is set, a transition from 0 to 1 in the condition part causes an entry in the corresponding bit of the event part of the register.

NameTypeValuesDescription
valuestring-Source dependent, comma-separated values.

Sets the positive transition filter. If a bit is set, a transition from 0 to 1 in the condition part causes an entry to be made in the corresponding bit of the EVENt part of the register.

NameTypeValuesDescription
<Value>numeric-Source dependent, comma-separated values

Returns the condition register for the channel power supply status. The value reflects the present condition of the status source.

integer - Condition register value for the channel power supply status.

Enables or disables the event status enable register for the selected channel's power-supply status system.

NameTypeValuesDescription
stateenum0 | 1 | OFF | ONTurns the enable register off or on.

Sets or queries the status event register for a channel power supply status structure. The optional [:EVENt] node is the event register subcommand.

Sets the input range for the specified channel.

NameTypeValuesDescription
chinteger-Channel number.

Sets the vertical scale of the specified channel. The command is listed alongside the channel range and state commands.

NameTypeValuesDescription
scalerealnull - nullVertical scale for the channel.

Adjusts the horizontal skew time for the specified channel.

NameTypeValuesDescription
chinteger-Channel number to which the skew time is applied.

Turns the specified channel on or off. This command is listed in the channel command set.

NameTypeValuesDescription
stateenumON | OFFChannel state.

Returns the overload warning status condition for the specified channel. This is a query command.

integer - Returns the condition register for CHANnel<ch>:WARNoverload:STATus.

Enables or disables the overload-status enable register for the specified channel. This command is listed under channel warning overload status control.

NameTypeValuesDescription
chinteger-Channel number.

Returns the event status of the overload warning for the specified channel. The event register indicates whether an overload warning event has occurred.

enum - Event status bit; indicates whether an overload warning event has occurred.

Returns the unit of the probe connected to the external trigger input.

enum - Returns the probe unit used on the external trigger input.

Returns the attenuation of the probe that is connected to the external trigger input.

real (V/V) - The attenuation value of the probe connected to the external trigger input.

Returns the name of the probe connected to the external trigger input.

string - The probe name.

Returns the state of the probe that is connected to the external trigger input.

enum - Returns the probe state connected to the external trigger input.

Returns the type of the probe that is connected to the external trigger input.

string - The probe type.

:ZONE<m>:ADDset / query

Adds a new trigger zone.

NameTypeValuesDescription
minteger1 - 4Index of the zone.

Sets the logic combination that applies to all areas in the indicated zone.

NameTypeValuesDescription
minteger1 - 4Index of the zone.
AreaCombinationenumAND | ORLogic combination for the areas in the zone.

Adds a new area to the trigger zone.

NameTypeValuesDescription
minteger1 - 4Index of the zone.
ninteger1 - 8Index of the zone area.

Returns the number of defined areas in the specified zone.

integer - The number of defined areas in the zone.

Defines whether the signal must intersect the selected zone area to allow triggering, or must not intersect the zone area.

NameTypeValuesDescription
IntersectionenumMUST | NOTSpecifies whether the signal must intersect the zone area to trigger.

Checks the validity of the indicated zone area. The area is invalid if one of its points is invalid.

boolean - Returns whether the indicated zone area is valid.

Adds a new point to the area at the indicated point index. Use ZONE<m>:AREA<n>:POINt<o>:X and ZONE<m>:AREA<n>:POINt<o>:Y to define the point coordinates.

NameTypeValuesDescription
minteger1 - 4Index of the zone.
ninteger1 - 8Index of the zone area.
ointeger1 - 16Index of the zone area point.

Queries the number of defined points in the selected zone area. `:POINt:COUNt? MAX` returns the maximum number of points that can be created.

NameTypeValuesDescription
qualifieroptenumMAXSelects the maximum number of points that can be created.
integer - Number of defined points in the area, or the maximum number of points when queried with `MAX`.

Removes the selected point from the area.

NameTypeValuesDescription
minteger1 - 4Index of the zone.
ninteger1 - 8Index of the zone area.
ointeger1 - 16Index of the zone area point.

Sets the horizontal X coordinates for the selected point of the area.

NameTypeValuesDescription
minteger1 - 4Index of the zone.
ninteger1 - 8Index of the zone area.
ointeger1 - 16Index of the zone area point.
Xreal-1E+26 - 1E+26Horizontal X coordinate for the selected point.

Sets the vertical Y coordinates for the selected point of the area.

NameTypeValuesDescription
Yreal-1E+26 - 1E+26Vertical Y coordinate for the selected point of the area.

Removes the selected area from the trigger zone.

NameTypeValuesDescription
minteger1 - 4Index of the zone.
ninteger1 - 8Index of the zone area.

Enables the selected area.

NameTypeValuesDescription
minteger1 - 4Index of the zone.
ninteger1 - 8Index of the zone area.
StateenumOFF | ONEnables or disables the selected area.

Returns the number of zones. With the MAX qualifier, returns the maximum number of zones that can be created.

NameTypeValuesDescription
qualifieroptenumMAXOptional query qualifier; MAX returns the maximum number of zones that can be created.
integer - Number of defined zones.
:ZONE<m>:DIAGramset / query

Selects the diagram on which the zone trigger is applied, for example layoutset1 diagram1 (L1_D1).

NameTypeValuesDescription
minteger1 - 4Index of the zone.
SignDiagKeystring-String that indicates the layout set and the diagram, e.g. "L1_D1".
:ZONE<m>:REMoveset / query

Removes the selected trigger zone.

NameTypeValuesDescription
minteger1 - 4Index of the zone.
:ZONE<m>:SOURceset / query

Sets the source of the zone trigger.

NameTypeValuesDescription
sourceenumNONE | C1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | XY1 | XY2 | XY3 | XY4 | SPECNORM1 | SPECNORM2 | SPECNORM3 | SPECNORM4Source used for the zone trigger.

Enables the display of the specified zone on the screen. The zone index is specified by suffix <m>; the display state can be turned OFF or ON.

NameTypeValuesDescription
minteger1 - 4Index of the zone.
DisplayStateenumOFF | ONTurns the zone display off or on.

Sets the software algorithm used for software clock data recovery.

NameTypeValuesDescription
algorithmenumCFRequency | PLL | FF | PLLRlockSelects the software algorithm for clock data recovery.

Sets the PLL bandwidth for the selected timing reference. It defines the part of the spectrum that the PLL can follow during synchronization.

NameTypeValuesDescription
minteger1 - 4Index of the timing reference.
Bandwidthreal20000 - Device specific: Double-Val(uso::db::wfa::IDeviceConstDataWFA::DOUBLE_ITEM::MaxDeviceBandwidth)PLL bandwidth.

Enables continuous synchronization of the clock with the data signal.

NameTypeValuesDescription
minteger1 - 4Index of the timing reference.
ClockResyncenumOFF | ONTurns continuous clock resynchronization off or on.

Sets the damping factor, which is only relevant for second order PLL.

NameTypeValuesDescription
minteger1 - 4Index of the timing reference.
dampingreal0.5 - 2Damping factor for the PLL.

Sets the order of the PLL for the selected timing reference. Higher order can compensate for more complex jitter behavior.

NameTypeValuesDescription
PLLOrderenumFIRSt | SECondPLL order to set.

Defines the phase reference for the first clock edge.

NameTypeValuesDescription
InitialPhaseenumSAMPle | DATaedgePhase reference for the first clock edge.

Sets the relative bandwidth, defined as the ratio of the nominal bit rate to the PLL bandwidth.

NameTypeValuesDescription
minteger1 - 4Index of the timing reference.
RelBwinteger10 - 6000Relative bandwidth value.

Selects when the CDR algorithm returns clock edges.

NameTypeValuesDescription
ResultsenumALL | LOCKed | AISYncSelects which clock edges are used for the CDR results.

Sets a value for the clock multiplier if `TREFerence<m>:TYPE CLOCk` is set. The multiplier is the ratio of an internal clock rate to the externally supplied clock and defines the number of samples per clock interval.

NameTypeValuesDescription
Multiplierinteger1 - 100Clock multiplier value.

Sets the frequency of the clock signal when TREFerence<m>:TYPE is set to CLOCk.

NameTypeValuesDescription
minteger1 - 4Index of the timing reference.
frequencyreal1000 - Device specific: Double-Val(uso::db::wfa::IDeviceConstDataWFA::DOUBLE_ITEM::MaxDeviceBandwidth)Clock signal frequency.

Sets the offset between the clock edge and the data edge if TREFerence<m>:TYPE CLOCk is set.

NameTypeValuesDescription
minteger1 - 4Index of the timing reference.
offsetreal-1000 - 1000Offset between the clock edge and the data edge.

Selects the reference level used for the timing reference when the indicated timing reference uses an explicit clock signal.

NameTypeValuesDescription
minteger1 - 4Index of the timing reference.
ReferenceLevelenumUPPer | MIDDle | LOWerReference level to use.

Sets the clock edges used for measurements when TREFerence<m>:TYPE is set to CLOCk, or sets the data edges for clock data recovery when TREFerence<m>:TYPE is set to SCDR.

NameTypeValuesDescription
edgeenumPOSitive | NEGative | EITHerSelects the clock or data edge used by the timing reference.

Adds the indicated timing reference when set to ON, or removes it when set to OFF.

NameTypeValuesDescription
timeRefenumOFF | ONAdds the indicated timing reference (ON) or removes it (OFF).

Sets the gate for the timing reference. Enable and configure a gate before assigning it.

NameTypeValuesDescription
gateinteger0Number of the gate to be used.

Selects the set of reference levels used for timing reference measurements. Define the reference level set before using it.

NameTypeValuesDescription
RefLevelSetinteger1 - 8Index of the reference level set.

Selects the clock source if TREFerence<m>:TYPE CLOCk is set, and sets the data source for clock data recovery if TREFerence<m>:TYPE SCDR is set for the indicated measurement.

NameTypeValuesDescription
minteger1 - 4Index of the timing reference.
sourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8Clock or data source selection.

Returns whether the specified timing reference is used by an application, such as jitter measurements.

NameTypeValuesDescription
minteger1 - 4Index of the timing reference.
boolean - Indicates whether the specified timing reference is used by an application.

Sets the symbol rate of the data signal for the indicated timing reference.

NameTypeValuesDescription
symbolratereal100000 - 5000000000Symbol rate for the timing reference.

Defines the origin of the clock signal for the indicated timing reference. The source can be an explicit clock signal or software CDR.

NameTypeValuesDescription
typeenumCLOCk | SCDROrigin of the clock signal.

Deprecated command for signal configuration. It is provided only for compatibility with previous oscilloscope versions; new remote control programs should use TREFerence<m>:CLK:REFLevel.

Returns the number of defined timing references.

integer - Number of defined timing references.

Sets file attributes for the specified file or files. The command applies to files only; wildcards (* and ?) are allowed in the file name.

NameTypeValuesDescription
FileNamestring-Name and path of the file whose attributes are queried.
string - String containing the file name and file attributes in the form "<file_name>,<file_attributes>".

Returns the data from the specified file. The file name is given as a parameter, followed by 488.2 block data; the delimiter EOI must be selected for correct transfer.

NameTypeValuesDescription
FileNamestring-Name of the file the data is stored to.
block - Returns the data from file abc.txt.
MMEM:DATA? 'abc.txt'
:MMEMory:NAMEset / query

Defines the filename for a screenshot that is stored to a file.

NameTypeValuesDescription
FileNamestring-String parameter specifying path and filename of the screenshot.
MMEMory:NAME <FileName>

Sets the auto-save interval for memory auto-save operations.

NameTypeValuesDescription
intervalreal0 - MAXimumAuto-save interval.

Sets the default path used by the automatic file naming feature. The manual slice only lists the command name and page reference, with no syntax or parameter details.

Sets the time stamp format used for autogenerated memory names.

NameTypeValuesDescription
timeenum-Not documented in the provided slice.

Returns the number of files in the current memory catalog.

integer - The number of files in the current memory catalog.

Returns the contents of the memory catalog. The command is listed in the memory subsystem command index.

string - Contents of the memory catalog.

Sets the current default directory used by mass storage commands.

:MMEMory:COPYset / query

Copies a file or directory in mass memory.

NameTypeValuesDescription
sourcestring-Source file or directory to copy.
destinationstring-Destination file or directory.
:MMEMory:DATAset / query

Saves data to a file in the mass memory. The command is listed in the mass-memory command tree, but this page slice does not provide syntax or parameter details.

Returns the names of files and directories in the specified directory on the memory device.

string - Returns a list of file and directory names in the specified memory directory.

Returns the list of available memory drives.

string - A list of available memory drives.

Saves the current generator settings to a file in memory. The command is part of the memory generator save/recall functions.

Creates or changes the current working directory used by mass memory commands.

:MMEMory:MOVEset / query

Moves a file or directory in mass memory.

Connects the selected memory drive to the instrument. The command applies to memory drive index m.

Sets the path for an external drive under the specified memory drive index. The command is part of the memory drive configuration group.

NameTypeValuesDescription
minteger-Memory drive index.
pathstring-Path to assign to the memory drive.

Changes the current working directory used by mass-memory commands.

:MMEMory:SAVset / query

Saves the current setup to a file. The manual slice only lists the command name in the index and does not provide syntax or parameter details.

Starts a fast export of the selected waveforms for use in automated analysis systems. The data stream contains the pure data of the exported sources; offset and length parameters are not available.

block - Returns the exported waveform data as a binary byte stream or as an ASCII comma-separated value list, depending on the selected format.

Sets the start time value of the waveform section for export when EXPort:WAVeform:SCOPe is set to Manual.

NameTypeValuesDescription
Startreal-1E+26 - 1E+26Start time value of the waveform section for export.

Sets the end time value of the waveform section for export, if EXPort:WAVeform:SCOPe is set to Manual.

NameTypeValuesDescription
stopreal-1E+26 - 1E+26End time value of the waveform section for export.

Sets the path, filename, and file format of the histogram export file. The command affects all histogram exports.

NameTypeValuesDescription
SaveAsPathstring-Path, filename, and file format of the histogram export file.

Sets histogram export normalization. When normalization is off, bin counts are exported as integer values; when on, each bin value is exported as a float ratio and the sum of all normalized values is 1. The command affects all histogram exports.

NameTypeValuesDescription
NormalizationenumOFF | ONTurns histogram export normalization off or on.

Saves the specified histogram to file. The target file is set using EXPort:HISTogram<m>:NAME.

NameTypeValuesDescription
minteger1 - 8Index of the histogram to save.

Sets the path, filename, and file format of the export file. The file name must include the .csv extension.

NameTypeValuesDescription
FilePathstring-String with path and file name of the export file, including extension .csv.

Saves the results to a file. The target file is set with EXPort:RESult:NAME, and the results included in the export are selected with EXPort:RESult:SELect:CURSor and EXPort:RESult:SELect:MEASurement.

Includes the current cursor results in the export file.

NameTypeValuesDescription
cursorResultenumOFF | ONSelects whether current cursor results are included in the export file.

Includes the current automatic measurement results in the export file.

NameTypeValuesDescription
MeasResultenumOFF | ONSelects whether the current automatic measurement results are included in the export file.

Returns the histogram data values for export. The command is listed as a query with an optional :VALues node.

definite_length_block - Histogram data values.

Selects the P3 phase result for export for a specified measurement. The command is indexed by measurement number <m>.

NameTypeValuesDescription
minteger1Measurement number to select for export.

Selects the power result export item for the specified measurement. The command is indexed by measurement number <m>.

NameTypeValuesDescription
minteger1 - MAXimumMeasurement number to select.

Sets the auto-generated waveform export name. The command belongs to the waveform export auto-naming subsystem.

NameTypeValuesDescription
namestring-Name used for automatic waveform export naming.

Sets the directory path used for automatic waveform file naming during export.

NameTypeValuesDescription
pathstring-Directory path used for automatic waveform file naming.

Sets the waveform export gate. This command is listed in the waveform export command group.

Sets the file name used for waveform export. The command is part of the EXPort:WAVeform subsystem.

Loads a session file. To reload user-specific display settings included in the session file, set SESSion:USERpref ON.

NameTypeValuesDescription
filepathstring-Path and filename of the session file.

Includes the channel waveform data in the session file.

NameTypeValuesDescription
ChWfmsenumOFF | ONControls whether channel waveform data is included in the session file.

Includes the reference waveform data in the session file.

NameTypeValuesDescription
RefWfmsenumOFF | ONControls whether reference waveform data is included in the session file.

Saves the current session with selected content to the specified file.

NameTypeValuesDescription
filepathstring-Path and filename of the target file.

If ON, the user-specific display settings for the toolbar, waveform colors, and diagram presentation are included in the session file. The setting affects both saving and recall actions.

NameTypeValuesDescription
UserSettingsenumOFF | ONEnables inclusion of user-specific display settings in the session file.

Selects the output medium for screenshots. The output configuration is selected by suffix <m>.

NameTypeValuesDescription
mediumenumMMEM | CLIPBOARDSelects the output medium.

Defines the file format for screenshot output to a file. To save to file, set the destination with HCOPy:DESTination<m> and parameter MMEM.

NameTypeValuesDescription
FileFormatenum-File format for output of the screenshot to file.

Inverts the colors of the output, so a dark waveform is shown on a white background.

NameTypeValuesDescription
InverseColorenumOFF | ONSelects whether the output colors are inverted.

Starts the output of the next display image, depending on the HCOPy:DESTination<m> destination setting. If the screenshot is saved to a file, the last used file name is automatically incremented to the next unused name.

NameTypeValuesDescription
moptinteger1 - 2Selects the output configuration.

Selects the output configuration for screenshot saving. The suffix <m> can be 1..2.

NameTypeValuesDescription
minteger1 - 2Selects the output configuration.
:HCOPy:ISBAset / query

If enabled, the screenshot shows the signal bar below the diagram area.

NameTypeValuesDescription
IncludeSignBarScpienumOFF | ONEnables or disables including the signal bar in the screenshot.
:HCOPy:SSDset / query

If enabled, the currently open dialog box is included in the screenshot.

NameTypeValuesDescription
ShowSetupDialogenumOFF | ONControls whether the currently open dialog box is included in the screenshot.
:HCOPy:WBKGset / query

Inverts the screenshot background color so waveforms can be shown on a white background. If both this command and HCOPy:DEVice<m>:INVerse are ON, the background is inverted twice and appears black.

NameTypeValuesDescription
WhiteBackgroundenumOFF | ONEnables or disables white-background screenshot rendering.

Creates a PNG screenshot and returns the image file data as a binary data stream. Write the received data into a PNG file to open it later.

block - Binary data stream containing the PNG screenshot.

Activates or deactivates the indicated measurement.

NameTypeValuesDescription
FirstenumOFF | ONSelects whether the indicated measurement is active.

Returns the maximum number of measurements, which is the maximum value for the <mg> suffix.

integer - Maximum number of measurements.

Sets the source of the measurement. The source must fit the selected measurement type.

NameTypeValuesDescription
SignalSourceenum-
SignalSource2optenum-

Defines the first measurement source. This command is an alternative to MEASurement<mg>:SOURce.

NameTypeValuesDescription
sourceenumNONE | C1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | TRK1 | TRK2 | TRK3 | TRK4 | TRK5 | TRK6 | TRK7 | TRK8 | TRK9 | TRK10 | TRK11 | TRK12 | TRK13 | TRK14 | TRK15 | TRK16 | HWAVeform1 | HWAVeform2 | HWAVeform3 | HWAVeform4 | HWAVeform5 | HWAVeform6 | HWAVeform7 | HWAVeform8 | PA1QPOWER | PA2QPOWER | PA3QPOWER | PA1HPOWER1 | PA2HPOWER1 | PA3HPOWER1 | PA1SPOWER | PA2SPOWER | PA3SPOWER | PA1SOA | PA2SOA | PA3SOA | PA1IPOWER | PA2IPOWER | PA3IPOWER | PA1OPOWER1 | PA2OPOWER1 | PA3OPOWER1 | PA1OPOWER2 | PA2OPOWER2 | PA3OPOWER2 | PA1OPOWER3 | PA2OPOWER3 | PA3OPOWER3 | PTHREE1HPOWER1 | PTHREE2HPOWER1 | PTHREE3HPOWER1 | PTHREE1HPOWER2 | PTHREE2HPOWER2 | PTHREE3HPOWER2 | PTHREE1HPOWER3 | PTHREE2HPOWER3 | PTHREE3HPOWER3 | PTHREE1HTPOWER | PTHREE2HTPOWER | PTHREE3HTPOWER | PTHREE1QPOWER1 | PTHREE2QPOWER1 | PTHREE3QPOWER1 | PTHREE1QPOWER2 | PTHREE2QPOWER2 | PTHREE3QPOWER2 | PTHREE1QPOWER3 | PTHREE2QPOWER3 | PTHREE3QPOWER3 | PTHREE1QTPOWER | PTHREE2QTPOWER | PTHREE3QTPOWER | PA1TOPOWER | PA2TOPOWER | PA3TOPOWERSelects the first measurement source.

Defines the second measurement source. The command is an alternative to MEASurement<mg>:SOURce.

NameTypeValuesDescription
source2enumNONE | C1 | C2 | ... | C8 | D0 | D1 | D2 | ... | D14 | D15 | M1 | M2 | ... | M8 | TRK1 | TRK2 | TRK3 | ... | TRK16 | HWAVeform1 | HWAVeform2 | ... | HWAVeform8 | PA1QPOWER | PA2QPOWER | PA3QPOWER | PA1HPOWER1 | PA2HPOWER1 | PA3HPOWER1 | PA1SPOWER | PA2SPOWER | PA3SPOWER | PA1SOA | PA2SOA | PA3SOA | PA1IPOWER | PA2IPOWER | PA3IPOWER | PA1OPOWER1 | PA2OPOWER1 | PA3OPOWER1 | PA1OPOWER2 | PA2OPOWER2 | PA3OPOWER2 | PA1OPOWER3 | PA2OPOWER3 | PA3OPOWER3 | PTHREE1HPOWER1 | PTHREE2HPOWER1 | PTHREE3HPOWER1 | PTHREE1HPOWER2 | PTHREE2HPOWER2 | PTHREE3HPOWER2 | PTHREE1HPOWER3 | PTHREE2HPOWER3 | PTHREE3HPOWER3 | PTHREE1HTPOWER | PTHREE2HTPOWER | PTHREE3HTPOWER | PTHREE1QPOWER1 | PTHREE2QPOWER1 | PTHREE3QPOWER1 | PTHREE1QPOWER2 | PTHREE2QPOWER2 | PTHREE3QPOWER2 | PTHREE1QPOWER3 | PTHREE2QPOWER3 | PTHREE3QPOWER3 | PTHREE1QTPOWER | PTHREE2QTPOWER | PTHREE3QTPOWER | PA1TOPOWER | PA2TOPOWER | PA3TOPOWERSecond measurement source.

Sets the gate of the indicated measurement. Enable a gate before assigning a measurement to it.

NameTypeValuesDescription
gateinteger0Number of the gate to be used.

Defines the measurement type assigned to the indicated measurement. To query the assigned measurement result, use MEASurement<mg>:RESult[:ACTual]?.

NameTypeValuesDescription
mginteger1 - 16Index of the measurement.
MeasTypeenumHIGH | LOW | AMPLitude | MAXimum | MINimum | PDELta | MEAN | RMS | STDDev | CRESt | POVershoot | NOVershoot | AREA | RTIMe | FTIMe | PPULse | NPULse | PERiod | FRE-Quency | PDCYcle | NDCYcle | CYCarea | CYCMean | CYCRms | CYCStddev | CYCCrest | CAMPlitude | CMAXimum | CMINimum | CPDelta | PULCnt | DELay | PHASe | BWIDth | EDGecount | SETup | HOLD | SHT | SHR | DTOTrigger | SLERising | SLEFalling | CCJitter | NCJitter | CCWidth | CCDutycycle | TIE | UINTerval | DRATe | SKWDelay | SKWPhase | WCOunt | HSAMples | HPEak | PEAK | UPEakvalue | LPEakvalue | HMAXimum | HMINimum | MEDian | MAXMin | HMEan | HSTDdev | M1STddev | M2STddev | M3STddev | MKPositive | MKNegative | F2F | T2F | F2T | FLDValue | MBITrate | SBITrate | BIDLe | GAP | FCNT | FEC | FER | CFERMeasurement type to assign to the indicated measurement.

Enables the measurement annotations for the selected measurement. These annotations can include periods, maximum and minimum values, relevant reference levels, and more.

NameTypeValuesDescription
ResultLinesenumOFF | ONControls whether measurement result lines are displayed.

Selects which envelope curve is used for the measurement. Relevant only for measurements on envelope waveforms, and requires ACQuire:TYPE to be set to ENVElope.

NameTypeValuesDescription
mginteger1 - 16Index of the measurement.
EnvelopeCurveenumMIN | MAX | BOTHEnvelope curve used for measurement.

Sets the edge of the clock from which the setup and hold times are measured.

NameTypeValuesDescription
<SetHoldClkSlp>enumPOSitive | NEGative | EITHerSelects which clock edge is used for the setup and hold time measurement.

Sets the edge direction to be counted for edge count measurement, selecting rising edges, falling edges, or both. The setting is only relevant when MEASurement<mg>:MAIN is set to EDGecount.

NameTypeValuesDescription
EdgesSlopeenumPOSitive | NEGative | EITHerSelects the edge direction to be counted.

Selects the direction for counting slopes for each source, from the beginning of the waveform or from the end.

NameTypeValuesDescription
EdgeCountDirenumFROMBeg | FROMEndDirection for counting slopes for each source.

Sets the first slope of the pulses to be counted. The setting is only relevant for pulse count and pulse train measurements.

NameTypeValuesDescription
PulsesSlopeenumPOSitive | NEGative | EITHerSelects the first slope of the pulses to be counted.

Sets the number of positive pulses for the pulse train measurement. It measures the duration of N positive pulses from the rising edge of the first pulse to the falling edge of the N-th pulse.

NameTypeValuesDescription
PulseCountinteger1 - 2147483647Number of positive pulses used for the pulse train measurement.

Sets the edge of each source, between which the delay is measured.

NameTypeValuesDescription
mginteger1 - 16Index of the measurement.
ninteger1 - 2Selects the source number.
slopeenumPOSitive | NEGative | EITHerSets the edge direction used for the delay measurement.

Sets the edge direction to be used for delay measurement.

NameTypeValuesDescription
DlyTrigSlpenumPOSitive | NEGative | EITHerEdge direction used for delay measurement.

Selects the unit for the phase measurement, in degrees or radians.

NameTypeValuesDescription
phase modeenumDEGrees | RADiansSelects the phase unit.

Selects the set of reference levels used for the measurement and the indicated measurement source. Each source of the measurement can have its own reference level set.

NameTypeValuesDescription
mginteger1 - 16Index of the measurement.
rlinteger1 - 2Measurement source to which the reference level set is applied.
RefLevSetinteger-Number of the reference level set.
MEASurement1:MAIN DELay
MEASurement1:SOURce C1,C2
MEASurement1:REFLevel1:REFerence 1
MEASurement1:REFLevel2:REFerence 2

Sets the number of periods (cycles) accumulated to measure the N-cycle jitter.

NameTypeValuesDescription
NumberCyclesinteger1 - 4294967295Number of cycles accumulated for the N-cycle jitter measurement.

Sets the polarity of pulses used for cycle-cycle width and cycle-cycle duty cycle measurements, or selects the starting edge for skew delay and skew phase measurements. The setting depends on the current MEASurement<mg>:MAIN measurement type.

NameTypeValuesDescription
PolarityenumPOSitive | NEGative | EITHerPolarity of pulses or edge used by the measurement.

Selects the reference level used for the jitter measurement and the indicated measurement source. Each measurement source can have its own reference level.

NameTypeValuesDescription
ReferenceLevel--

Sets the edge of the second waveform relative to the first waveform.

NameTypeValuesDescription
RelPolarityenumMATChing | INVerseSelects whether the measurement uses like polarity edges or opposite polarity edges.

Selects the slope used for jitter measurements. For cycle-cycle jitter, N-cycle jitter, and cycle-cycle duty cycle measurements it selects the period slope; for time-interval error measurements it selects the data-signal edges used for measurement.

NameTypeValuesDescription
slopeenumFIRSt | POSitive | NEGative | EITHerSlope or edge selection used for the measurement.

Selects the timing reference, which is one of the available clock configurations. The timing reference must be defined before it can be used.

NameTypeValuesDescription
TimingReferenceinteger1 - 4Timing reference index.

Sets the unit for data rate measurements.

NameTypeValuesDescription
UnitenumBIT_S | HZMeasurement unit for data rate measurements.

Sets or queries the name of the frame or the frame type at which the oscilloscope executes or starts the measurement. This command is asynchronous.

NameTypeValuesDescription
FrameNamestring-Name of the frame or frame type used for the measurement.

Sets or queries the name of the frame or the frame type at which the oscilloscope ends the measurement in a "From"-"To" condition.

NameTypeValuesDescription
Frame2Name--Name of the frame or frame type used as the measurement end point.

Sets or queries the name of the field or the field type at which the oscilloscope ends the measurement in a "From"-"To" condition.

NameTypeValuesDescription
Field2Name--

Sets or queries the one or more values of the field at which the oscilloscope ends the measurement in a "From"-"To" condition.

NameTypeValuesDescription
<Field2Value>string-List of comma separated values.

Returns the arithmetic mean of the measurement results for the specified measurement. The measurement type is defined with MEASurement<mg>:MAIN.

NameTypeValuesDescription
mginteger1 - 16Index of the measurement.
numeric - Numeric result value.

Returns the count of measurement events for the specified measurement. The measurement type is defined with MEASurement<mg>:MAIN.

numeric - Numeric result value.

Returns the negative peak statistic result of the specified measurement. The measurement type is defined with `MEASurement<mg>:MAIN`.

NameTypeValuesDescription
mginteger1 - 16Index of the measurement.
numeric - Numeric result value.

Returns the positive peak value of the specified measurement statistic. The measurement type is defined with MEASurement<mg>:MAIN.

NameTypeValuesDescription
mginteger1 - 16Index of the measurement.
numeric - Numeric result value.

Returns the RMS value of the statistic results for the specified measurement. The measurement type is defined with MEASurement<mg>:MAIN.

real - Numeric result value.

Returns the statistic result of the specified measurement. RSTDdev is the relative standard deviation (STDDev / AVG).

NameTypeValuesDescription
mginteger1 - 16Index of the measurement.
numeric - Numeric result value.

Returns the start time of the specified measurement. If no parameter is specified, the result of the main measurement is returned.

numeric - Start time of the specified measurement, or the main measurement if no parameter is specified.

Returns the standard deviation result of the specified measurement. The measurement type is defined with MEASurement<mg>:MAIN.

NameTypeValuesDescription
mginteger1 - 16Index of the measurement.
numeric - Numeric result value.

Returns the statistic result count for the specified measurement. The measurement type is defined with `MEASurement<mg>:MAIN` and `<mg>` selects measurement 1 to 16.

NameTypeValuesDescription
mginteger1 - 16Index of the measurement.
numeric - Numeric result value.

Returns the statistic results of the specified measurement. The measurement type is defined with `MEASurement<mg>:MAIN` and the query is asynchronous.

NameTypeValuesDescription
mginteger1 - 16Index of the measurement.
numeric - Numeric result value.

Returns the number of measured events in one acquisition. The command is relevant for measurements of all events, see MEASurement<mg>:MULTiple.

integer - Number of events measured in one acquisition.

Returns the start time of the indicated measured event. The command is relevant for measurements of all events.

NameTypeValuesDescription
EventIndexinteger-Index number of the measured event.
real - Start time of the indicated measured event.

Returns the end time of the indicated measured event. The command is relevant for measurements of all events.

NameTypeValuesDescription
EventIndexinteger1 - 16Index number of the measured event.
real - End time of the indicated measured event.

Returns the stop time of the indicated measured event. The command is relevant for measurements of all events.

NameTypeValuesDescription
EventIndexinteger-Index number of the measured event.
real - Stop time of the indicated measured event.

Returns the measured value of the indicated measured event. The command is relevant for measurements of all events, see MEASurement<mg>:MULTiple.

NameTypeValuesDescription
MeasResEvtIdxinteger-Index number of the measured event.
real - Measured value of the indicated measured event.

Returns the condition status register for measurement margin status. This deprecated command is provided for compatibility with previous oscilloscope versions only; use :MEASurement<mg>:LIMit:STATus: commands instead.

integer - Returns the content of the condition status register.

Deprecated status enable command for measurement margin status reporting. It is provided for compatibility with previous oscilloscope versions only; new programs should use the corresponding MEASurement<mg>:LIMit:STATus commands.

Deprecated status-reporting command provided for compatibility with previous oscilloscope versions only. Use the MEASurement<mg>:LIMit:STATus commands instead.

Deprecated status-reporting command provided for compatibility with previous oscilloscope versions only. Use the corresponding MEASurement<mg>:LIMit:STATus command instead.

Deprecated status reporting command provided for compatibility with previous oscilloscope versions only. The manual lists it under obsolete margin status commands and recommends using the corresponding LIMit:STATus commands instead.

Sets or queries the present measurement status transition condition for the impedance measurement status register. The slice lists this command in the status-register section but does not show the detailed syntax or return format.

Returns the event register for the measurement imprecise status group. The optional :EVENt node is shown in the syntax but no separate parameter details are given in this slice.

integer - Event register value for MEASurement<mg>:IMPRecise:STATus.

Queries the status of the measurement limit event condition. The optional :EVENt node is part of the documented query mnemonic.

enum - Returns the current measurement limit status/event condition.

Returns the lower result value for the selected reference level. The command belongs to the measurement reference level result group.

real - The lower reference level result value.

Returns the upper reference level result for the selected measurement and reference level. The command is a query and provides the upper value from :MEASurement<mg>:REFLevel<rl>:RESult.

real - Upper reference level result for the selected measurement and reference level.

Enables or disables statistics for the selected measurement. This command is the optional STATistics subtree under the measurement group.

NameTypeValuesDescription
stateenumON | OFFTurns measurement statistics on or off.

Enables or disables continuous tracking for the selected measurement group. The command belongs to the MEASurement<mg>:TRACk subsystem.

NameTypeValuesDescription
stateenumON | OFFTurns continuous tracking on or off.

Sets the measurement track offset. This command is part of the measurement track configuration commands.

NameTypeValuesDescription
offsetreal-Track offset value.

Enables or disables tracking for the selected measurement. The command is part of the measurement tracking subsystem.

NameTypeValuesDescription
stateenumON | OFFTurns measurement tracking on or off.

Returns the header information for the selected math waveform data. The command is a query for math trace <m> data headers.

string - Header information for the selected math waveform data.

Returns the data style used by the selected math channel. This is a query for the style setting under CALCulate:MATH<m>:DATA.

enum - Returns the current data style for the selected math channel.

Returns the selected values from the specified math data table. The optional `:VALues` node is part of the query form shown in the manual.

string - Returns the selected math data values.

Selects the environment for the specified math waveform. The command belongs to the MATH<m> subsystem and is used to set the environment selection.

Sets the label for the selected math trace.

NameTypeValuesDescription
labelstring-Label text for the math trace.

Sets or queries the state of a math channel calculation. The command enables or disables the selected math trace <m>.

NameTypeValuesDescription
stateenumON | OFFEnables or disables the math calculation.

Sets the unit used for math calculation results for the selected math waveform. The unit applies to the specified math memory slot.

NameTypeValuesDescription
minteger-Math waveform index.

Sets the vertical offset for a math waveform. This command applies to math traces identified by <m>.

NameTypeValuesDescription
offsetreal-Vertical offset value for the math waveform.

Sets the vertical scale value for the selected math trace. The optional value node is documented with this command.

NameTypeValuesDescription
valueoptreal-Vertical scale value for the math trace.

Defines the expression used for a math waveform. You can also use the command to access or modify the defined math expression for math channel m.

NameTypeValuesDescription
expressionoptstring-Expression to define for the math waveform.

Sets the resolution bandwidth value for the selected spectrum frequency bandwidth. The command supports an optional resolution node and an optional value node in the syntax.

NameTypeValuesDescription
resolutionopt--Optional resolution node in the spectrum frequency bandwidth branch.
valueoptreal-Resolution bandwidth value.

Sets the center frequency of the selected spectrum calculation. The command is part of the spectrum frequency setup group.

NameTypeValuesDescription
frequencyrealHzCenter frequency to set.

Sets the gate position for the spectrum calculation. The command belongs to the spectrum gate settings and is documented as a setter only.

NameTypeValuesDescription
positionreal-Gate position.

Sets the stop frequency of the spectrum gate. The gate stop can be defined for spectrum analysis on the specified spectrum bar.

Sets the spectrum gate width for the selected spectrum measurement. The command is part of the CALCulate:SPECtrum<sp>:GATE subsystem.

Sets the spectrum magnitude level. The command belongs to the spectrum calculation subsystem.

NameTypeValuesDescription
levelreal-Spectrum magnitude level.

Sets the magnitude range for the specified spectrum trace. The command is part of the Spectrum calculation subsystem.

NameTypeValuesDescription
rangereal-Magnitude range for the spectrum trace.

Enables or disables spectrum overlap for the selected spectrum analysis object.

NameTypeValuesDescription
stateenumON | OFFTurns spectrum overlap on or off.

Sets the overlap factor count value for the selected spectrum measurement. This command is part of the spectrum overlap settings.

NameTypeValuesDescription
valueinteger-Overlap factor count value.

Sets or queries the peak excursion setting for the selected spectrum trace. The command is part of the Spectrum calculator subsystem.

Controls whether the frequency label is shown for the spectrum peak list. The command supports an optional state argument.

NameTypeValuesDescription
stateoptenumON | OFFTurns the frequency label state on or off.

Sets the maximum number of labels that can be shown for the selected spectrum peak list.

NameTypeValuesDescription
countinteger0 - 1000Maximum number of labels to display.

Sets the maximum number of peaks in the peak list. The command belongs to the spectrum peak list subsystem.

NameTypeValuesDescription
countinteger1 - 1000Maximum number of peaks to include in the peak list.

Sets the spectrum power-list source or power list data associated with the selected spectrum calculation. The slice only lists the command in the index and does not provide syntax or parameter details.

Prepares the spectrum calculation for use with the selected source and settings. This command is listed as a setter.

Sets the spectrum source. The command belongs to the CALCulate:SPECtrum subsystem.

NameTypeValuesDescription
sourceenum-Spectrum source selection.

Turns spectrum calculation on or off.

NameTypeValuesDescription
stateenumON | OFFEnables or disables the spectrum calculation.

Enables or disables the maximum spectrum waveform display. The command belongs to the CALCulate:SPECtrum<sp>:WAVeform subtree.

NameTypeValuesDescription
stateenumON | OFFTurns maximum waveform display on or off.

Enables or disables the minimum spectrum waveform display. This command belongs to the CALCulate:SPECtrum waveform family.

NameTypeValuesDescription
stateenumON | OFFTurns the minimum waveform display on or off.

Returns waveform data header information for the normal spectrum waveform. This is the query form of the normal waveform data header command.

string - Returns the waveform data header information for the normal spectrum waveform.

Enables the amplitude signal for the frequency response analysis. You can then define the amplitude profile of the signal.

NameTypeValuesDescription
stateenumOFF | ONTurns the amplitude signal off or on.

Sets the amplitude offset for FRA analysis.

NameTypeValuesDescription
offsetreal-Amplitude offset value.

Sets the number of profile points used in amplitude profile analysis.

NameTypeValuesDescription
countinteger1 - 1000Number of points in the amplitude profile.

Automatically sets the display scaling for the frequency response analysis. The command is used to adjust the analysis display to the current data.

Returns the calibration result for the frequency response analysis calibration.

string - Returns the calibration result for FRANalysis calibration.

Enables or disables frequency response analysis. The command belongs to the FRANalysis subsystem.

NameTypeValuesDescription
stateenum-Turns frequency response analysis on or off.

Sets the stop frequency for the frequency analysis range.

NameTypeValuesDescription
frequencyreal-Stop frequency for the analysis range.

Enables or disables gain analysis. The command is part of the FRANalysis subsystem.

NameTypeValuesDescription
stateenumON | OFFTurns gain analysis on or off.

Sets the gain analysis offset value.

NameTypeValuesDescription
offsetreal-Gain analysis offset value.

Sets the gain scale for frequency-response analysis. This command is part of the FRANalysis:GAIN subsystem.

Sets or queries the channel used by the function-range analysis generator. The optional channel node selects a specific channel.

NameTypeValuesDescription
channeloptinteger1Channel number to use for the generator.

Defines or enables the harmonic distortion definition used by FRANalysis. The optional ENABle suffix indicates the command can be used to turn this definition on or off.

NameTypeValuesDescription
enableoptenum-Enables the harmonic definition.

Selects the input source used for frequency response analysis.

NameTypeValuesDescription
sourceoptenum-Input source for the FRANalysis function.

Returns the phase margin value from the FRANalysis margin measurement. This is a query command.

real - Phase margin value.

Turns the margin display on or off for frequency response analysis.

NameTypeValuesDescription
stateenumON | OFFEnables or disables margin display.

Sets the frequency of analysis marker m.

NameTypeValuesDescription
minteger1Marker number to address.
frequencyrealHzFrequency value for the marker.

Returns the gain value for the specified marker in frequency analysis.

NameTypeValuesDescription
minteger1Marker index.
real - Gain value for the specified marker.

Places a marker on the screen for the frequency analysis display. The manual lists this command in the FRANalysis:MARKer subsystem without further syntax or parameter details on this page.

Sets the state of the delay measurement function in FRANalysis. It is listed as part of the FRANalysis:MEASurement:DELay command group.

Sets the delay measurement time. The optional node form is shown as `:TIME` in the syntax listing.

NameTypeValuesDescription
timeoptreal-Delay measurement time.

Selects the output source for the FRANalysis function. The source is given as an optional parameter in the command syntax.

NameTypeValuesDescription
sourceoptenum-Output source selection.

Returns phase data for the selected analysis function.

definite_length_block - Phase data for the selected analysis function.

Enables or disables the phase analysis function.

NameTypeValuesDescription
stateenumON | OFFTurns phase analysis on or off.

Sets the maximum phase analysis value used by FRANalysis:PHASe. The command is listed in the FRANalysis phase command group index.

Sets the phase analysis offset.

NameTypeValuesDescription
offsetreal-Phase offset value.

Sets the phase scale for the FRANalysis display.

NameTypeValuesDescription
scalereal-Phase scale value.

Sets the logarithmic point spacing for FRANalysis measurements. This command is listed in the FRANalysis command group on the command index page.

Sets the mode used for FRANalysis points. The command is listed with related FRANalysis points commands, but this slice does not provide syntax details or allowed values.

Sets the total number of frequency analysis points.

NameTypeValuesDescription
totalinteger1 - 65536Total number of points to use for FRANalysis.

Sets the name of the referenced analysis curve. The curve is identified by the <rc> index.

NameTypeValuesDescription
namestring-Name assigned to the reference curve.

Sets the offset for the specified reference curve. The command is part of the FRAnalysis reference-curve subsystem.

NameTypeValuesDescription
rcinteger-Reference curve index.
offsetreal-Offset value for the reference curve.

Opens the specified reference curve. The reference curve is identified by the <rc> index.

NameTypeValuesDescription
rcinteger1Reference curve index.

Saves the current reference curve. The curve index is specified by <rc>.

NameTypeValuesDescription
rcinteger1 - 4Reference curve index.

Sets the source curve for a reference curve.

NameTypeValuesDescription
rcinteger1 - nReference curve index.
sourceenumCURVe | REFCurveSource curve selection.

Restores the selected reference curve to its original data. This command applies to the specified reference curve index <rc>.

NameTypeValuesDescription
rcinteger1 - nReference curve index.

Sets or queries the state of the FRAnalysis result display. The command controls whether the result state is enabled or disabled.

enum - Returns the current FRAnalysis result state.

Enables or disables the frequency analysis state.

NameTypeValuesDescription
stateenumON | OFFTurns frequency analysis on or off.

:XY<m>:SWAPset / query

Replaces the source of the x-axis with the source of the y-axis and vice versa.

NameTypeValuesDescription
minteger1 - 4Index of the XY-plot.
:XY<m>[:STATe]set / query

Activates an XY-waveform.

NameTypeValuesDescription
minteger1 - 4Index of the XY-plot.
stateopt--State of the XY-waveform.
:XY<m>:XSOurceset / query

Defines the signal source that supplies the x-values of the XY-diagram.

NameTypeValuesDescription
<XSource>enumNONE | C1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8 | TRK1 | TRK2 | TRK3 | TRK4 | TRK5 | TRK6 | TRK7 | TRK8 | TRK9 | TRK10 | TRK11 | TRK12 | TRK13 | TRK14 | TRK15 | TRK16Selects the signal source for the x-axis of the XY-diagram.
:XY<m>:YSOurceset / query

Defines the signal source that supplies the y-values of the XY-diagram.

NameTypeValuesDescription
YSourceenumNONE | C1 | C2 | ... | C8 | M1 | M2 | ... | M8 | R1 | R2 | ... | R8 | TRK1 | TRK2 | TRK3 | ... | TRK16Signal source for the y-values of the XY diagram.

:SBUS<sb>:TYPEset / query

Selects the bus type for analysis. The available bus types depend on the installed options.

NameTypeValuesDescription
typeenumI2C | I3C | SPI | QSPI | UART | CAN | LIN | SENT | TNOS | MILS1553 | MILStd | ARIN429 | ARINc429 | SWIR | SWIRe | RFFE | SPMI | MANC | NRZC | NRZU | USB1 | USB2 | USBHspeed | USBLow | USBFullProtocol type to use for the serial bus analysis.

Enables decoding of the specified serial bus. The suffix selects bus 1 to 4.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
ProtocolStateenumOFF | ONTurns decoding of the selected bus off or on.
:SBUS<sb>:FORMatset / query

Sets the number format for decoded data values of the indicated serial bus. It defines the format in the decode table and in the combs of the decoded signal on the screen.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
DataFormatenumHEX | OCT | BIN | ASCII | ASCii | SIGN | USIGNumber format for decoded data values.
:SBUS<sb>:RESultset / query

Enables a table with decoded data of the serial signal. The function requires the option for the analyzed protocol.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
ShwResTblenumOFF | ONEnables or disables the decoded data table.
:SBUS<sb>:RMSBusset / query

Deletes the specified serial bus.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

If enabled, includes the detailed results for all frames in the export result file.

NameTypeValuesDescription
IncludeDetailsenumOFF | ONEnables or disables inclusion of detailed results in the export result file.

Sets whether threshold lines are displayed in the diagram for the selected serial bus. The label of the threshold line is set according to the signal it refers to.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
ShwThresLinesenumOFF | ONEnables or disables display of threshold lines.

Synchronizes the protocol decode zoom and the result table when enabled.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
ZoomCouplingenumOFF | ONEnables or disables zoom coupling.

Selects the file format for exported results.

NameTypeValuesDescription
FileTypeenumHTML | CSV | XML | PYFile format to use for the export result file.

Saves the selected results to the indicated file.

NameTypeValuesDescription
FileName--
SBUS1:EXPResult:SAVE
'/home/storage/userData/export_10base-t1s.csv'

If enabled, includes the frame timing in the export result file.

NameTypeValuesDescription
IncludeTimingenumOFF | ON

Sets the path where the protocol export files are stored. On the instrument, the default path is /home/storage/userData/Protocol.

NameTypeValuesDescription
Pathstring-Path where the protocol export files are stored.

Selects the bit order, which determines whether SPI message data starts with MSB or LSB.

NameTypeValuesDescription
BitOrderenumLSBF | MSBFSelects the SPI bit order.

Defines how the start of an SPI frame is determined. A frame can begin and end either by the active CS signal or by a clock timeout on SCLK.

NameTypeValuesDescription
FrameConditionenumCS | CLKTimeoutSelects the condition used to define the start of a frame.

Sets a value for the hysteresis for the CS channel.

NameTypeValuesDescription
hysteresisreal-Hysteresis value for the CS channel.

Selects whether the transmitted signal for the respective line is active high or active low.

NameTypeValuesDescription
CSPolarityenumACTLow | ACTHighSelects the polarity of the transmitted signal for the CS line.

Sets the input channel of the CS line.

NameTypeValuesDescription
CSSourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8Selects the input channel of the CS line.

Sets the hysteresis for the MISO channel.

NameTypeValuesDescription
Hysteresisreal-Hysteresis value for the MISO channel.

Selects whether the transmitted signal for the MISO line is active high or active low.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
MISOPolarityenumACTLow | ACTHighPolarity of the transmitted signal for the MISO line.

Sets the input channel of the MISO line.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
MISOSourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8Selects the input channel for the MISO line.

Sets a user-defined threshold value for the line.

NameTypeValuesDescription
thresholdreal-User-defined threshold value for the line.

Selects whether the transmitted MOSI signal is active high or active low.

NameTypeValuesDescription
MOSIPolarityenumACTLow | ACTHighSelects the polarity of the MOSI line.

Sets the input channel of the MOSI line.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
MOSISource--

Sets the input channel of the SCLK line.

NameTypeValuesDescription
SCLKSourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8

Sets the vertical position of the MISO signal.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
positionreal-Vertical position of the MISO signal.

Sets a user-defined threshold value for the line.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
thresholdreal-User-defined threshold value for the line.

Sets the minimum clock idle time used as a timeout limiter on the SCLK line between two frames.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
clocktimeoutreal5E-08 - 10Minimum clock idle time. The default value is 0.001 s and the increment is 1E-06.

Sets the word length, or symbol size, for SPI messages. The maximum word length is 32 bits.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
WordLengthinteger4 - 32Number of bits in a message.

Enables the filter for all available frames and error types.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Frameinteger1 - 4Index of the frame.
Fieldinteger-Index of the field.
BitStateenumONE | ZEROSets the bit state operator.

Queries the bit setting for the selected field of the selected frame.

NameTypeValuesDescription
Bitinteger-Bit index.

Sets the end value of a data pattern range when the operator is set to INRange or OORANGe.

NameTypeValuesDescription
Frameinteger1 - 4Index of the frame.
Fieldinteger-Index of the field.
Data--End value of the data pattern range.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
BitStateenumONE | ZEROSelects the bit state operator.

Sets the end value of a data pattern range when the operator is set to INRange or OORANGe.

NameTypeValuesDescription
Data_Maxnull-End value of the data pattern range.

Inverts the current state of the frame and error types, so enabled frame and error types become disabled and vice versa.

Specifies the start value of a data pattern range for the selected SPI field in a serial bus frame. The command is asynchronous.

NameTypeValuesDescription
Frameinteger1 - 4Index of the serial bus frame.
Fieldinteger-Index of the field.
Data_Min--Start value of the data pattern range.

Queries the start value of the data pattern range for the selected SPI filter. The command is the query form of the DMIN setting.

numeric - Returns the start value of the data pattern range.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frame--Index of the frame.
field--Index of the field.
operatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator for the data pattern.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator for the data pattern.
enum - Returns the current data pattern operator.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator used for the data pattern in the selected field.

Defines the error type to be filtered on.

NameTypeValuesDescription
ErrorName--
EnablerenumOFF | ONEnables or disables the selected error filter.

Returns the enable state for the SPI filter error enable setting on the selected serial bus. The command is an asynchronous query.

enum - Returns the current enabler state.

Enables or disables the checking condition for the selected error type. The command uses the bus index and error index suffixes to identify the filter entry.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
ninteger-Index of the error.
enableenumOFF | ONEnables or disables the error filter.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
condenablerenumOFF | ONEnables or disables the checking condition for the selected field.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
condEnablerenumOFF | ONEnables or disables the checking condition.

Enables or disables the checking condition for the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
enableenumOFF | ONEnables or disables the checking condition for the selected frame.

Enables or disables the checking condition for the selected frame.

NameTypeValuesDescription
frinteger-Index of the frame.
enableenumOFF | ONTurns frame checking condition off or on.

Enables or disables the checking condition for the selected frame.

NameTypeValuesDescription
frameinteger1 - 4Index of the frame.
enableenumOFF | ONEnables or disables the checking condition.

Enables or disables the checking condition for the selected frame.

NameTypeValuesDescription
frameinteger1 - 4Index of the frame.
enablerenumOFF | ONEnables or disables the checking condition for the selected frame.
enum - Returns the current enable state.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
index_operatorenumINRangeOperator for the index in the selected field of the selected frame.

Sets the end value of an index range if the operator is set to INRange.

NameTypeValuesDescription
index_maxinteger1 - 65535End value of an index range.

Sets the end value of an index range if the operator is set to INRange.

NameTypeValuesDescription
Frame--Frame selector.
Field--Field selector.
Data--Data payload.

Specifies the index, or sets the start value of an index range.

NameTypeValuesDescription
Index_Mininteger1 - 65535Start value of the index range or index value.

Specifies the start value of an index range, or sets the index when the operator is INRange.

NameTypeValuesDescription
Index_Mininteger1 - 65535Index minimum value to set.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
Frame--
Field--
Operator--

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
Frame--
Field--
Operator--

Queries the software trigger time in seconds. If no software trigger event has occurred, the command returns Invalid.

real (s) - Software trigger time in seconds, or Invalid if no software trigger event has occurred.

Queries the software trigger index. If no software trigger event has occurred, the command returns Invalid.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
integer - Index of the software trigger event; returns Invalid if no software trigger event has occurred.

Returns the number of decoded frames.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
integer - Count of decoded frames.

Returns the bit rate of the specified SPI frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (bps) - Primary bit rate of the specified frame.

Returns the data words of the specified frame in comma-separated values.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - Comma-separated sequence of integer values (N, L1, R1,…, LN, RN). N is the number of word pairs in the frame; Lx and Rx are the MOSI and MISO channel values, respectively, and an empty value is returned if a channel is disabled.

Returns the start time of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
returns

Returns the overall state of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
enum - Overall state of the specified frame: OK for a valid frame, VOID for an empty frame, INComplete if the word is cut off, or LENGth if the frame length is unexpected.

Returns the end time of the specified frame. The command is query only and asynchronous.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (s) - End time of the specified frame.

Returns the number of words in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
integer - Number of words in the specified frame.

Returns the formatted value of the specified word on the MISO line.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
w--Index of the word.
returns - Returns the formatted value of the specified word on the MISO line.

Returns the formatted value of the specified word on the MOSI line.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
winteger-Index of the word.
string - Formatted MOSI value of the specified word.

Returns the data value of the specified word on the MISO line.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
winteger-Index of the word.
returns - Returns the data value of the specified word on the MISO line.

Returns the data value of the specified word on the MOSI line.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
winteger-Index of the word.
integer - The data value of the specified word on the MOSI line.

Returns the start time of the specified data word.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
winteger-Index of the word.
real (s) - Returns the start time of the specified data word.

Returns the end time of the specified data word.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
winteger-Index of the word.
real (s) - End time of the specified data word.

Sets a value for the hysteresis for the CS channel.

NameTypeValuesDescription
Hysteresisreal-Hysteresis value for the CS channel.

Selects whether the transmitted signal for the CS line is active high or active low.

NameTypeValuesDescription
CSPolarityenumACTLow | ACTHighSelects the polarity of the CS line signal.

Sets the input channel of the CS line.

NameTypeValuesDescription
CSSource--Input channel of the CS line.

Selects the instruction mode that defines how many lanes are used to transmit data.

NameTypeValuesDescription
InstructionenumSINGle | DUAL | QUADInstruction mode that defines how many lanes are used to transmit data.

Selects whether the transmitted signal for the respective line is active high or active low.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
IO1PolarityenumACTLow | ACTHighSets the polarity of the IO1 signal.

Sets the vertical position of the IO1 signal.

NameTypeValuesDescription
QUADSPI IO1 position--Vertical position of the IO1 signal.

Sets the vertical scale of the IO1 signal.

NameTypeValuesDescription
QUADSPI IO1 scale--Vertical scale of the IO1 signal.

Sets the input channel of the IO 1 line.

NameTypeValuesDescription
<IO1Source>enumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8Selects the input channel for IO1.

Selects whether the transmitted signal for the IO3 line is active high or active low.

NameTypeValuesDescription
IO3PolarityenumACTLow | ACTHighSets the polarity of the IO3 signal.

Sets the vertical position of the IO3 signal.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
QUADSPI IO3 position--Vertical position of the IO3 signal.

Sets the input channel of the IO 3 line.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
IO3SourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8

Sets a user-defined threshold value for the line.

NameTypeValuesDescription
Thresholdreal-User-defined threshold value for the line.

Sets a value for the hysteresis for the IO2 line.

NameTypeValuesDescription
Hysteresis--Value for the hysteresis for the IO2 line.

Selects whether the transmitted signal for the respective line is active high or active low.

NameTypeValuesDescription
<IO2Polarity>enumACTLow | ACTHighSets the polarity of the transmitted signal for the respective line.

Sets the vertical position of the IO2 signal.

NameTypeValuesDescription
<QUADSPI IO2 position>real-Vertical position of the IO2 signal.

Sets the input channel of the IO 2 line.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
IO2SourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8

Sets a value for the hysteresis for the IO2 line.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
Threshold--

Sets the input channel of the IO 0 line.

NameTypeValuesDescription
IO0SourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8Input channel assigned to the IO 0 line.

Sets a user-defined threshold value for the line.

NameTypeValuesDescription
thresholdreal-User-defined threshold value for the line.

Sets the polarity for the clock line.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
SCLKPolarityenumFALLing | RISingPolarity of the SCLK clock line.

Sets the input channel of the SCLK clock line.

NameTypeValuesDescription
SCLKSourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8Input channel of the SCLK clock line.

Deletes all currently defined opcodes from the list.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Sets a user-defined threshold value for the line.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
Threshold--

Loads an opcode file from the selected file.

NameTypeValuesDescription
FileNamestring-

Deletes the opcode with the selected index.

NameTypeValuesDescription
Indexinteger1 - 4Selected opcode index to delete.

Sets the size of the opcode, hence the number of opcodes currently defined on the table in the "Format" tab of QUAD-SPI.

Sets the address byte for the opcode item. It specifies the location in flash memory where the operation is performed.

NameTypeValuesDescription
AddressBytesinteger0 - 4Number of address bytes.

Selects how many lines or lanes are used to send the address bytes to the flash memory.

NameTypeValuesDescription
AddressLanesenumSINGle | DUAL | QUADNumber of lanes used for the address bytes.

Enables an alternative field, if one is available.

NameTypeValuesDescription
AltenumOFF | ONEnables or disables the alternative field.

Sets the opcode value for the specified serial bus and opcode item.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
ninteger-Index of the opcode.
Opcodeinteger0 - 255Opcode value.

Enables data transfer in the opcode operation if data is being transferred.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
ninteger-Index of the opcode.
DataenumOFF | ONEnable or disable data transfer in the opcode operation.

Enables or disables double-data-rate transfer for the selected opcode item. When enabled, data is transferred on both the rising and falling clock edges, effectively doubling the transfer rate compared to SDR.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
ninteger-Index of the opcode.
DDRenumOFF | ONEnables or disables double-data-rate transfer.

Sets the number of dummy cycles for the selected opcode item. Dummy cycles are inserted after the address or other command sequences but before data transfer begins.

NameTypeValuesDescription
DummyCyclesinteger0 - 255Number of dummy cycles to insert.

Selects how many lanes are used for the data transfer. QUADSPI can utilize multiple data lines to increase the speed and efficiency of data transfer.

NameTypeValuesDescription
DataLanesenumSINGle | DUAL | QUADNumber of data lanes used for the transfer.

Saves the opcode file for the selected serial bus. The command is asynchronous and applies to a specified bus index.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
FileNamestring-Name of the file to save.

Inverts the current state of the frame and error types for the selected serial bus; enabled items become disabled and vice versa.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Frame--
Field--
BitenumONE | ZEROBit state to set for the selected field.

Returns the bit value for the selected field of the selected frame in the serial bus QSPI filter.

NameTypeValuesDescription
Bitinteger-
integer - Returns the bit value for the selected field of the selected frame.

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe.

NameTypeValuesDescription
frameinteger-
fieldinteger-
datareal-

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe.

NameTypeValuesDescription
datastring-
string

Specifies the data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
Frame--
Field--
Data_Min--

Specifies the data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
Data--
returns

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
BitStateenumONE | ZERO

Specifies the data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
Data_Min--

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
<Frame>--
<Field>--
<Operator>enumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORange

Returns the operator for the data pattern in the selected field of the selected frame.

enum - Returns the data operator for the selected field and frame.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
operatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator used for the data pattern in the selected field of the selected frame.

Sets the end value of an index range if the operator is set to INRange.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
Index_Maxinteger1 - 65535End value of the index range.

Specifies the index, or sets the start value of an index range.

NameTypeValuesDescription
Index_Mininteger1 - 65535Index minimum value for the selected field; used as the start value of an index range.

Sets the end value of an index range if the operator is set to INRange.

NameTypeValuesDescription
index_maxinteger1 - 65535End value of the index range.

Sets the end value of an index range if the operator is set to INRange.

returns - Returns the end value of the index range.

Specifies the index, or sets the start value of an index range.

NameTypeValuesDescription
Frameinteger1 - 4Index of the frame.
Fieldinteger-Index of the field.
Datainteger1 - 65535Start value of an index range.

Defines the error type to be filtered on.

NameTypeValuesDescription
enableenumOFF | ONEnables or disables the error filter.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
Frameinteger1 - 65535Index of the frame.
Fieldinteger1 - 65535Index of the field.
OperatorenumEQUal | INRange | RANGeOperator to use for the selected index.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
operatorenumEQUal | INRange | RANGeOperator for the index in the selected field of the selected frame.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Frameinteger1 - 4Index of the frame.
Fieldinteger-Index of the field.
EnablerenumOFF | ONEnables or disables the operator for the selected field pattern.

Returns the enabled/disabled state of the selected filter-in enable condition for the selected frame and field.

enum - Returns the enabler state.

Enables or disables the checking condition for the selected frame.

NameTypeValuesDescription
enableenumOFF | ONEnables or disables the checking condition for the selected frame.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
<CondEnabler>enumOFF | ONEnables or disables the checking condition.

Enables or disables the checking condition for the selected frame.

NameTypeValuesDescription
Frame--
EnablerenumOFF | ON

Enables or disables the checking condition for the selected frame.

NameTypeValuesDescription
Frame--Selected frame.
EnablerenumOFF | ONEnables or disables the checking condition.
enum - Returns the checking-condition state of the selected frame.

Queries the software trigger index. If no software trigger event has occurred, the command returns Invalid.

enum - Returns the software trigger index, or Invalid if no software trigger event has occurred.

Queries the software trigger time in seconds. If no software trigger event has occurred, the command returns Invalid.

real (s) - Returns the software trigger time in seconds, or Invalid if no software trigger event has occurred.

Returns the address of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - Returns the frame address.

Returns the value of the alternate byte for the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - Alternate byte value for the specified frame.

Returns the bit rate of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (bps) - The primary bit rate of the specified frame.

Returns the data of the specified QSPI frame in comma-separated values.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
returns - Returns the data of the specified frame as comma-separated values.

Returns the formatted data value of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
returns - Returns the formatted data value of the specified frame.

Returns the field count for the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - Count of fields in the specified frame.

Returns the label of the selected field in the specified frame. Query only; asynchronous command.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
string - The label of the selected field in the specified frame.

Returns the state of the selected field in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
integer - Returns the state of the selected field in the specified frame.

Returns the value of the specified field in the specified frame of the selected serial bus. Query only and asynchronous.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
integer - Returns the value of the specified field. Increment is 1; *RST value is 0.

Returns the start time of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (s) - Returns the frame start time.

Returns the overall state of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
enum - Overall state of the specified frame.

Returns the end time of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (s) - End time of the specified frame.

Sets a user-defined threshold value for the clock line.

NameTypeValuesDescription
Thresholdreal-User-defined threshold value for the clock line.

Sets a hysteresis value for the data line.

NameTypeValuesDescription
hysteresisreal-Hysteresis value for the data line.

Sets the source channel to which the I2C data line is connected.

NameTypeValuesDescription
SDASourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8Source channel for the SDA line.

Sets the vertical position of the I2C signal.

NameTypeValuesDescription
Position--

Sets the vertical scale of the I2C signal.

NameTypeValuesDescription
Scale--

Sets a user-defined threshold value for the I2C data line.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
Thresholdreal-User-defined threshold value for the data line.

Activates the symbol list to be used for decoding.

NameTypeValuesDescription
ShowSymbolsenumOFF | ON

Loads a symbol list file for the selected serial bus. The command is setting-only and asynchronous.

NameTypeValuesDescription
FileNamestring-String parameter with path and file name.
SBUS1:I2C:NEWList
'/home/storage/userData/Protocol/I2C.csv'
SBUS1:I2C:SYMBols ON

Enables the filter for all available frames and error types.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Disables the filter for all available frames and error types on the selected I2C serial bus. This is a setting-only asynchronous command.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Inverts the current state of the frame and error types, so enabled items become disabled and vice versa.

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe. The operator is set with SBUS<sb>:I2C:FILTer:DOPerator.

NameTypeValuesDescription
frame--Frame index.
field--Field index.
data_max--End value of the data pattern range.

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe. You can set the operator with SBUS<sb>:I2C:FILTer:DOPerator.

NameTypeValuesDescription
datastring-End value of the data pattern range.

Specifies the data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
frameinteger-Frame index.
fieldinteger-Field index.
data_min--

Specifies the data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
Data_Min--

Enables or disables the specific frame to be filtered on.

NameTypeValuesDescription
enableenumOFF | ONEnables or disables the specific frame to be filtered on.

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe. The operator is set with SBUS<sb>:I2C:FILTer:DOPerator.

NameTypeValuesDescription
Data_Maxstring-End value of the data pattern range.

Specifies the data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
Data_Min--

Enables or disables the specific frame to be filtered on. The command is asynchronous and applies to the selected serial bus frame.

NameTypeValuesDescription
frame--
enablerenumOFF | ONEnables or disables filtering for the specified frame.

Enables or disables the specific frame to be filtered on.

NameTypeValuesDescription
enablerenumOFF | ONEnables or disables the specific frame to be filtered on.
enum - Returns the filter enable state.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Frame--
Field--
OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeSelects the comparison operator used for the data pattern.

Returns the operator for the data pattern in the selected field of the selected frame.

enum - Returns the current data-pattern operator.

Defines the error type to be filtered on for the selected serial bus. The query returns whether the error filter is enabled.

enum - Returns the enable state of the error filter.

Enables or disables filtering for the selected I2C error type in the serial bus filter setup.

NameTypeValuesDescription
EnableenumOFF | ONTurns error filtering off or on.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORange

Sets the end value of an index range if the operator is set to INRange. You can set the operator with SBUS<sb>:I2C:FILTer:IOPerator.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
fl--Index of the field.
index_max--End value of an index range.

Sets the end value of an index range if the operator is set to INRange. You can set the operator with SBUS<sb>:I2C:FILTer:IOPerator.

NameTypeValuesDescription
frame--
field--
data--

Sets the end value of an index range if the operator is set to INRange. You can set the operator with SBUS<sb>:I2C:FILTer:IOPerator.

real - Returns the end value of an index range.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
index_operatorenumEQUal | INRange | RANGeOperator for the index in the selected field of the selected frame.

Specifies the index, or sets the start value of an index range.

NameTypeValuesDescription
Frame--
Field--
Data--

Specifies the minimum index for a selected field in a selected frame, or returns the current minimum index value. This command is part of the serial bus I2C filter setup.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
Data--Data associated with the minimum-index setting or query.
returns - Returns the minimum index value for the selected field of the selected frame.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
Frame--Frame index.
Field--Field index.
Index_OperatorenumEQUal | INRange | RANGeOperator for the index in the selected field.

Sets the operator for the index in the selected field of the selected frame. The operator selects how the index is interpreted.

enum - Returns the selected index operator.

Sets the bit state of a field that only consists of one bit.

NameTypeValuesDescription
Frame--
Field--
Bit--

Returns the bit state of the selected bit in a field. The query form takes the bit index as its parameter.

NameTypeValuesDescription
Bitinteger1 - ?Index of the bit to query.
enum - Returns the bit state.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
Frame--
Field--
EnablerenumOFF | ON

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
enablerenumOFF | ONEnables or disables the checking condition for the selected field of the selected frame.

Queries the software trigger index. If no software trigger event has occurred, the command returns Invalid.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
integer - Returns the software trigger index, or Invalid if no software trigger event has occurred.

Returns the number of decoded frames.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
integer - Total number of decoded frames.

Queries the software trigger time in seconds. If no software trigger event has occurred, the command returns Invalid.

real (s) - Returns the software trigger time in seconds, or Invalid if no software trigger event has occurred.

Returns the address acknowledge bit value for the indicated frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
enum - Address acknowledge bit value for the indicated frame.

Returns the value of the R/W bit of the indicated frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
enum - R/W bit value for the indicated frame.

Returns whether the address is completely contained in the acquisition for the indicated frame.

enum - Indicates whether the address is completely contained in the acquisition.

Returns the start time of the address acknowledge bit.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (s) - Start time of the address acknowledge bit.

Returns the device address value of the indicated I2C frame, as shown in the decoded cells and decode results table.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - Device address value of the indicated frame.

Returns the pure device address of the indicated frame without the R/W bit.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - The pure device address of the indicated frame without the R/W bit.

Returns the address length for the indicated I2C frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
enum - Address length of the indicated frame.

Returns the start time of the address for the indicated frame.

real (s) - The start time of the address for the indicated frame.

Returns the primary bit rate for the indicated frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (bps) - Primary bit rate.

Returns the formatted value of the specified field.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
returns - The formatted value of the specified field.

Returns the name (label) of the specified field in the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
returns - Returns the label of the specified field in the selected frame.

Returns the start time of the specified field.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frnull-Index of the frame.
flnull-Index of the field.
null

Returns the number of fields in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - Number of fields in the specified frame.

Returns the overall state of the specified field in the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
fl--Index of the field.
enum - Returns the overall state of the specified field in the selected frame.

Returns the data value of the specified field.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
fl--Index of the field.
returns

Returns the data words of the specified I2C frame as comma-separated values.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
returns - Returns the data words of the specified frame as comma-separated values.
BUS:I2C:FRAMe4:DATA?

Returns the start time of the R/W bit for the specified I2C frame. This is a query-only asynchronous command.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (s) - Start time of the R/W bit.

Returns the start time of the specified frame. Query only and asynchronous.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (s) - Returns the start time of the specified frame.

Returns the overall state of the specified I2C frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
enum - Overall state of the frame.

Returns the end time of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (s) - End time of the specified frame.

Returns the symbolic label of the specified frame if the label list is enabled.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
string - String with the symbolic name of the frame.

Specifies whether a minimum gap before the first decoded frame is required.

NameTypeValuesDescription
MingapSelectenumOFF | ONSelects whether a minimum gap before the first decoded frame is required.

Specifies the minimum duration of the gap. It is available only when `SBUS<sb>:I3C:MINGap:SELect` is set to ON.

NameTypeValuesDescription
MingapWidthreal1E-11 - 1Minimum duration of the gap.
SBUS<sb>:I3C:MINGap:WIDTh <MingapWidth>

Returns the value of the acknowledge bit for the selected frame. The ACK bit is transmitted right after the address field and acknowledges receiving the address.

enum - Acknowledgement bit value for the selected frame.

Returns the start time of the acknowledge bit for the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (s) - Start time of the acknowledge bit for the selected frame.

Returns the cyclic redundancy check (CRC) checksum of the selected frame.

integer - CRC checksum of the selected frame.

Returns the formatted command from the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
string - Formatted command from the selected frame.

Returns the formatted CRC value for the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
returns - Returns the formatted CRC value for the selected frame.

Returns the value of the R/W bit of the selected frame. The value "0" denotes "Write", and the value "1" denotes "Read".

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
enum - The R/W bit value for the selected frame.

Sets the number of transmitted bits per second.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
Bitrateinteger300 - 20000000Transmit bit rate.

Selects the bit order for UART messages, determining whether data starts with MSB or LSB.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
BitOrderenumLSBF | MSBFSelects least-significant-bit-first or most-significant-bit-first bit order.

Defines the method of packet separation for UART data. A packet is a number of subsequent words in a data stream.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
FrmSeparationenumNONE | TOUTSelects how packets are separated.

Defines the optional parity bit used for error detection.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
ParityenumNONE | ODD | EVEN | MARK | SPC | DCParity setting for the UART bus.

Defines the logic levels of the bus. Idle high (high=1) is used for control signals, while idle low (low=1) is defined for data lines (RS-232).

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
polarityenumIDLLow | IDLHighSelects the bus logic-level polarity.

Sets the hysteresis for the Tx line.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
Rx hysteresis--

Selects the input channel for the receiver signal.

NameTypeValuesDescription
RxSourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8

Sets the number of stop bits for the serial bus UART. The possible stop bit settings are 1, 1.5, or 2 stop bits.

NameTypeValuesDescription
StopBitsenumB1 | B15 | B2Selects the number of stop bits.

Sets the number of data bits of a word in the range from 5 bits to 8 bits. If no parity bit is used, 9 data bits are possible.

NameTypeValuesDescription
DataBitsinteger5 - 8Number of data bits of a word.

Sets the timeout between packets in a UART data stream. A new packet starts with the first start bit after the timeout.

NameTypeValuesDescription
Timeoutreal1E-06 - 1Timeout between packets in a UART data stream.

Sets the hysteresis for the TX line.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
Tx hysteresis--

Selects the input channel for the transmitter signal.

NameTypeValuesDescription
TxSource--

Sets the vertical position of the RX UART signal.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
Rx position--

Sets the vertical scale of the RX UART signal.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
Rx scale--

Sets the vertical position of the TX UART signal.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
Tx position--

Sets a user-defined threshold value for the Tx line.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
Tx thresholdreal-User-defined clock threshold.

Sets the vertical scale of the TX UART signal.

NameTypeValuesDescription
Tx scale--

Inverts the current state of the frame and error types for the selected serial bus filter; all enabled frames and error types become disabled and vice versa.

Presets the state of the selected frames and error types.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Sets the end value of an index range if the operator is set to INRange.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
Data--

Specifies the index, or sets the start value of an index range.

NameTypeValuesDescription
Index_Mininteger1 - 65535Start value of the index range.

Returns the start value of an index range for the selected serial bus frame field. The command belongs to the UART filter index settings.

integer - Returns the current minimum index value for the selected field.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
Operator--Index operator to set.

Specifies the data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
frameinteger1 - 4Index of the frame.
fieldinteger-Index of the field.
datastring-Data pattern value.

Specifies the data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
frameinteger1 - 4Index of the frame.
fieldinteger-Index of the field.
datanull-Data pattern or start value for a data pattern range.
returns - Returns the data pattern or the start value of a data pattern range.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
frameinteger1 - 4Index of the frame.
fieldinteger-Index of the field.
operatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator for the data pattern.

Sets the operator for the data pattern in the selected field of the selected frame.

enum - Returns the selected data pattern operator.

Defines the error type to be filtered on.

NameTypeValuesDescription
enablerenumOFF | ON
enum - Returns the filter enable state.

Specifies the data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
data_minreal-Data pattern value or the start value of a data pattern range.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
frameinteger1 - 4Index of the frame.
fieldinteger-Index of the field.
operatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator for the data pattern.

Sets the bit state of a field that only consists of one bit.

NameTypeValuesDescription
bitstateenumONE | ZEROBit state to set for the field.

Sets the bit state of a field that only consists of one bit.

NameTypeValuesDescription
BitStateenumONE | ZEROBit state to set for a one-bit field.
enum - Returns the bit state of the selected field.

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe.

NameTypeValuesDescription
frame--
field--
data--

Queries the end value of a data pattern range for the selected serial bus UART filter. The value is used when the operator is set to INRange or OORANGe.

real - Returns the end value of a data pattern range.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
<Frame>--Frame index.
<Field>--Field index.
<Enabler>enumOFF | ONEnables or disables the checking condition.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
fieldinteger1 - 4Index of the field.
enablerenumOFF | ONEnables or disables the checking condition.
enum - Returns the field enabler state.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
CondEnablerenumOFF | ONEnables or disables the checking condition for the selected field.

Enables or disables the checking condition for the selected frame.

NameTypeValuesDescription
Frame--
EnablerenumOFF | ON

Enables or disables the checking condition for the selected frame.

NameTypeValuesDescription
EnablerenumOFF | ONEnables or disables the checking condition for the selected frame.
enum - Returns the current enabling state of the selected frame checking condition.

Queries the software trigger index for the selected serial bus. If no software trigger event has occurred, the command returns Invalid.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
integer - Returns the software trigger index; if no software trigger event has occurred, returns Invalid.

Queries the software trigger time in seconds. If no software trigger event has occurred, the command returns Invalid.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
real (s) - Software trigger time in seconds; returns Invalid if no software trigger event has occurred.

Returns the number of words in the acquisition.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
integer - Total number of words.

Returns the primary bit rate.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
winteger-Index of the word.
returns

Returns the value of the specified word on the Rx line.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
winteger-Index of the word.
integer - The value of the specified received word.

Returns the line on which the specified word was transferred.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
winteger-Index of the word.
enum - Indicates whether the word was transferred on the TX or RX line.

Returns the start time of the specified word.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
winteger-Index of the word.

Returns the stop time of the specified word.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
winteger-Index of the word.
real (s) - Returns the stop time of the specified word.

Returns the value of the specified word on the TX line.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
winteger-Index of the word.
integer - The value of the specified word on the TX line.

Sets the polarity for the data line.

NameTypeValuesDescription
<CLKPolarity>enumRISE | FALL | BOTHSelects when data is sampled relative to the clock signal edge.

Selects the source for the clock signal.

NameTypeValuesDescription
SourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8Channel or reference source to use for the clock signal.

Selects the channel for the data signal source.

NameTypeValuesDescription
SourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8Source channel for the data signal.

Sets the polarity for the data line.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
DATAPolarityenumACTLow | ACTHighSelects whether the transmitted data-line signal is active high or active low.

Sets the polarity for the enable line.

NameTypeValuesDescription
ENABPolarityenumACTLow | ACTHighSelects the polarity for the enable line.

Selects the channel for the enable signal source.

NameTypeValuesDescription
SourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8Channel used as the enable signal source.

Sets the threshold value for digitizing the clock signal.

NameTypeValuesDescription
CLKThreshold--Threshold value for the digitization of the clock signal.

Enables checking the minimum idle time between two frames during decoding. To set the idle time, use the command SBUS<sb>:NRZC:MINGap:WIDTh.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
MingapSelectenumOFF | ONEnables or disables checking the minimum idle time between two frames during decoding.

Sets the minimum duration of the idle time between frames during decoding. Any inactivity greater than this time is interpreted as a gap and causes resynchronization to the signal.

NameTypeValuesDescription
mingapwidthreal1E-09 - 1Minimum duration of the idle time.

Sets the threshold for the data channel.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
SDAThreshold--

Sets the threshold for the enable channel.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
ENABThreshold--

Sets the vertical position of the NRZ clocked signal.

NameTypeValuesDescription
Position--

Sets the vertical scale of the NRZ clocked signal.

NameTypeValuesDescription
Scale--

Disables the filter for all available frames and error types.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Inverts the current state of the frame and error types, so enabled frames and error types become disabled and vice versa.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Sets the bit state of a field that only consists of one bit. This is an asynchronous setting command for the selected serial bus and frame/field location.

NameTypeValuesDescription
BitStateenumONE | ZEROBit state to set for the one-bit field.

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe.

NameTypeValuesDescription
Frameinteger1 - 4Index of the serial bus.
Fieldinteger-Index of the field.
Data--

Specifies the data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
Data_Min--

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Frame--
Field--
Operatornull-

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
Frameinteger-Index of the frame.
Fieldinteger-Index of the field.
EnablerenumOFF | ONEnables or disables the checking condition for the selected field.

Enables the filtering on NRZ clocked frames. Only the frames that match the selected filter conditions are displayed.

NameTypeValuesDescription
enableenumOFF | ONTurns frame filtering on or off.

Enables or disables the filtering on the selected frame. Only frames that match the selected filter conditions are displayed.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
condEnablerenumOFF | ONEnables or disables the checking condition for the selected field of the selected frame.

Enables filtering on NRZ clocked frames. Only frames that match the selected filter conditions are displayed.

NameTypeValuesDescription
Frame--Index of the frame.
EnablerenumOFF | ONTurns filtering on or off.

Sets the end value of an index range if the operator is set to INRange. The operator is set with SBUS<sb>:NRZC:FILTer:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
Index_Maxinteger1 - 65535End value of an index range.

Specifies the start value of an index range, or sets the index for the selected frame field. It is part of the serial bus NRZC filter settings.

NameTypeValuesDescription
Index_Mininteger1 - 65535Start value of an index range or index value.

Sets the operator for the index in the selected field of the selected frame. The operator determines how the index range is applied.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
operatorenum-Operator to use for the index.

Sets the end value of an index range if the operator is set to INRange. The start value is set with :SBUS<sb>:NRZC:FILTer:IMIN, and the operator is set with :SBUS<sb>:NRZC:FILTer:IOPerator.

NameTypeValuesDescription
Frame--
Field--
Data--

Specifies the index, or sets the start value of an index range.

NameTypeValuesDescription
Index_Mininteger1 - 65535Start value of the index range.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
Operator--Index operator to use.
returns - Returns the current index operator for the selected field of the selected frame.

Loads the specified XML file containing a list of frame descriptions.

NameTypeValuesDescription
FileNamestring-Name of the XML file to load.

Erases from volatile memory all field and frame format descriptions created for a specific custom protocol.

Returns the number of frame format descriptions available for a specific custom protocol.

integer - The count of available frame format descriptions.

Saves the current list of frame descriptions to an XML file with the specified name.

NameTypeValuesDescription
FileNamestring-Name of the XML file to save the frame descriptions to.

Returns the number of field descriptions available for the selected frame.

NameTypeValuesDescription
<sb>integer1 - 4Index of the serial bus.
<fr>integer-Index of the frame.
integer - The number of field descriptions available for the selected frame.

Specifies the color for the selected frame. The color uses 32-bit RGB encoding in decimal format.

NameTypeValuesDescription
Colorinteger0 - 4294967295Use 32-bit RGB encoding in decimal format.

Specifies a user-defined condition operator for the selected field of the selected frame. The condition value format is set by the corresponding :FORMat command.

NameTypeValuesDescription
Conditionenumequal | arrayImplemented condition operator for the selected field.

Specifies the numerical format for the condition value of the selected field in the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
fl--Index of the field.
NumericFormatenumHEX | OCT | BIN | ASCII | ASCii | SIGN | USIG | DEC | LIST | FLINNumeric format for the condition value.

Specifies in which order the algorithm evaluates the bits of the condition value of the selected field in the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
BitOrderenumLSBF | MSBFBit evaluation order.

Specifies in which result column of the decode table to display the selected field of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
ColumnenumNONE | COL1 | COL2 | COL3Result column selection.

Queries the software trigger time in seconds. If no software trigger event has occurred, the command returns Invalid.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
real (s) - Time in seconds. Returns Invalid if no software trigger event has occurred.

Queries the software trigger index. If no software trigger event has occurred, the command returns Invalid.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
integer - Software trigger index; if no software trigger event has occurred, returns Invalid.

Returns the number of decoded frames for the selected serial bus. This is a query-only asynchronous command.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
integer - Number of decoded frames.

Returns the label of the specified field in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
returns

Returns the overall state of the specified field in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
enum - Overall state of the specified field in the specified frame.

Returns the formatted value of the specified field in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
fl--Index of the field.
returns - Returns the formatted value of the specified field in the specified frame.

Returns the value of the specified field in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
fl--Index of the field.
returns - Returns the value of the specified field in the specified frame.

Returns the overall state of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
enum - The frame state of the specified frame.

Returns the start time of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (s) - Start time of the specified frame.

Returns the end time of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (s) - End time of the specified frame.

Returns the content of the 1st result column of the specified decoded frame.

returns - The content of the 1st result column of the specified decoded frame.

Returns the content of the 3rd result column of the specified decoded frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
returns - The content of the 3rd result column of the specified decoded frame.

Returns the content of the 2nd result column of the specified decoded frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
string - The content of the 2nd result column of the specified decoded frame.

Returns the type of the specified decoded frame.

returns - The frame type is undefined, because it is custom specific.

Selects the channel used as the data signal source for NRZ unclocked decoding.

NameTypeValuesDescription
SourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8Channel to use as the data signal source.

Sets the polarity for the data line of the selected serial bus.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
DATAPolarityenumACTLow | ACTHighSelects whether the transmitted data signal is active high or active low.

Sets the polarity for the enable line.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
ENABPolarityenumACTLow | ACTHighSelects the enable-line signal polarity.

Selects the channel for the enable signal source.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
SourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8Channel to use as the enable signal source.

Enables setting the bit rate, which is required to trigger and decode unclocked NRZ signals. The bit rate itself is set with SBUS<sb>:NRZU:BITRate:WIDTh.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
stateenumOFF | ONEnables or disables bit rate selection.

Enables checking the minimum idle time between two frames during decoding. To set the idle time, use SBUS<sb>:NRZU:MINGap:WIDTh.

NameTypeValuesDescription
MingapSelectenumOFF | ON

Sets the minimum duration of the idle time. Any inactivity greater than this time is interpreted as a gap and causes resynchronization to the signal.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
MingapWidthreal1E-09 - 1Minimum idle time to be used as the gap threshold.

Specifies the bit rate width, if enabled with SBUS<sb>:NRZU:BITRate:SELect. The setting applies to serial bus indexes 1 to 4.

NameTypeValuesDescription
BitrateWidthreal300 - 2000000000Bit rate width to set.

Sets the vertical position of the NRZ unclocked signal.

NameTypeValuesDescription
Position--

Sets the vertical scale of the NRZ unclocked signal.

NameTypeValuesDescription
Scale--

Inverts the current state of the frame and error types, so enabled frames and error types become disabled and vice versa.

Sets the bit state of a field that only consists of one bit.

NameTypeValuesDescription
Frame--
Field--
Bitinteger-

Defines the error type to be filtered on.

NameTypeValuesDescription
enableenumOFF | ONEnables or disables the error type filter.

Defines whether the selected error type is enabled for filtering. It is an asynchronous setting-only command.

NameTypeValuesDescription
enableenumOFF | ONEnables or disables the error filter state.

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe.

NameTypeValuesDescription
Frame--
Field--
Data--

Specifies the data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
Frame--
Field--
Data--

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
frame--Index of the frame.
field--Index of the field.
operator--

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe. You can set the operator with :SBUS<sb>:NRZU:FILTer:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
Data_Maxreal-

Specifies the data pattern, or sets the start value of a data pattern range. It is used for the selected field of the selected frame.

NameTypeValuesDescription
Data_Minreal-Start value of a data pattern range.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
Frame--Selected frame index.
Field--Selected field index.
EnablerenumOFF | ONTurns the checking condition off or on.
enum - Returns the enabler state.

Enables or disables the filtering on NRZ unclocked frames. Only the frames that match the selected filter conditions are displayed.

NameTypeValuesDescription
enableenumOFF | ONControls whether filtering is enabled for the selected frame.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
condenablerenumOFF | ONEnables or disables the checking condition.

Enables the filtering on NRZ unclocked frames. Only the frames that match the selected filter conditions are displayed.

NameTypeValuesDescription
frame--
enablerenumOFF | ONTurns filtering for the selected frame off or on.

Sets the end value of an index range if the operator is set to INRange.

NameTypeValuesDescription
datareal-End value of an index range.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
Index_OperatorenumEQUal | INRange | RANGeOperator for the index in the selected field of the selected frame.

Specifies the index, or sets the start value of an index range.

NameTypeValuesDescription
Index_Mininteger1 - 65535Index value to set as the start of an index range.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
index_operatorenumEQUal | INRange | RANGeOperator for the index in the selected field of the selected frame.

Loads the specified XML file with a list of frame descriptions for the NRZ Unclocked frame format. This is a setting-only asynchronous command.

NameTypeValuesDescription
FileName--

Erases from the volatile memory all field and frame format descriptions that you have created for a specific custom protocol.

Returns the number of frame format descriptions available for a specific custom protocol.

integer - Number of frame format descriptions available for the specified custom protocol.

Saves the current list of frame descriptions to an XML file with the specified name.

NameTypeValuesDescription
FileNamestring-Name of the XML file to save the frame descriptions to.

Specifies the color for the frame description of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
Colorinteger0 - 4294967295Use 32-bit RGB encoding in decimal format.

Returns the number of field descriptions available for the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - Number of field descriptions available for the selected frame.

Specifies the name for the frame description of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
Name--

Specifies the color for the selected field of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
Colorinteger0 - 4294967295Use 32-bit RGB encoding in decimal format.

Specifies a user-defined condition operator for the selected field of the selected frame. The implemented conditions are the equal and array operators.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
Conditionenumequal | arrayCondition operator for the selected field.

Specifies the numerical format for the condition value of the selected field in the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
numeric_formatenumHEX | OCT | BIN | ASCII | ASCii | SIGN | USIG | DEC | LIST | FLINNumeric format for the condition value.

Specifies the order in which the algorithm evaluates the bits of the condition value of the selected field in the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
BitOrderenumLSBF | MSBFBit order to use when evaluating the field condition value.

Specifies which result column of the decode table displays the selected field of the selected frame. The result can be hidden or shown in columns 1 through 3.

NameTypeValuesDescription
ColumnenumNONE | COL1 | COL2 | COL3Result column used to display the selected field.

Queries the software trigger time in seconds. If no software trigger event has occurred, the command returns Invalid.

real (s) - The software trigger time in seconds, or Invalid if no software trigger event has occurred.

Queries the software trigger index for the selected serial bus. If no software trigger event has occurred, the command returns Invalid.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
integer - Index of the software trigger event, or Invalid if no software trigger event has occurred.

Returns the number of decoded frames.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
integer - Returns the number of decoded frames.

Returns the number of fields available in the selected frame. Query only; asynchronous command.

integer - The number of fields available in the selected frame.

Returns the label of the specified field in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
fl--Index of the field.
string - The label of the specified field in the specified frame.

Returns the overall state of the specified field in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
fl--Index of the field.
enum - Overall state of the specified field in the specified frame.

Returns the formatted value of the specified field in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
returns - Returns the formatted value of the specified field in the specified frame.

Returns the value of the specified field in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
fl--Index of the field.
returns - The value of the specified field in the specified frame.

Returns the overall state of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
enum - The overall frame state: OK indicates the frame is valid; LENGth, CRC, PARity, or INComplete indicate an error condition.

Returns the start time of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (s) - Start time of the specified frame.

Returns the end time of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (s) - End time of the specified frame.

Returns the type of the specified decoded frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
returns - The frame type is undefined, because it is custom specific.

Returns the content of the 1st result column of the specified decoded frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
string - Content of the 1st result column of the specified decoded frame.

Query only. Returns the content of the 3rd result column of the specified decoded frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
string - Content of the 3rd result column of the specified decoded frame.

Returns the content of the 2nd result column of the specified decoded frame. Query only and asynchronous.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
string - Content of the 2nd result column of the specified decoded frame.

Sets the polarity for the data line.

NameTypeValuesDescription
DataPolarityenumMANC | MANTSelects the data line polarity.

Selects the channel used as the data signal source for the serial bus Manchester decode. The bus index is given by <sb>.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
DATASourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8Channel for the data signal source.

Sets the phase for the data line. Captures data bits on the clock's first (rising) edge with FEDGe or second (falling) edge with SEDGe.

NameTypeValuesDescription
DataPhaseenumFEDGe | SEDGeSelects which clock edge is used to capture data bits.

Sets the polarity for the enable line.

NameTypeValuesDescription
EnablePolarity--Polarity setting for the enable line.

Selects the channel for the enable signal source.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
DATASourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8Channel for the enable signal source.

Enables checking the minimum idle time between two frames during decoding.

NameTypeValuesDescription
MingapSelectenumOFF | ONEnables or disables checking of the minimum idle time between two frames during decoding.

Sets the minimum duration of the idle time used for gap detection between two frames during decoding. Any inactivity greater than this time is interpreted as a gap and causes resynchronization to the signal.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
mingapwidthreal1E-09 - 1Minimum idle time; inactivity longer than this value is interpreted as a gap.

Enables setting the bit rate. To set the bit rate, use the command SBUS<sb>:MANCh:BITRate:WIDTh.

NameTypeValuesDescription
<BitrateSelect>enumOFF | ONSelects whether bit rate setting is enabled.

Sets the bit rate for the selected serial bus channel. The bus index suffix <sb> is 1 to 4.

NameTypeValuesDescription
BitrateWidthinteger300 - 2000000000Bit rate value to set.

Sets the lower threshold for the data channel.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
DatThresLow--

Sets the vertical position of the Manchester signal.

NameTypeValuesDescription
positionreal-Vertical position of the Manchester signal.

Sets the vertical scale of the Manchester signal.

NameTypeValuesDescription
Scale--

Inverts the current state of the frame and error types for the selected serial bus. Enabled frames and error types are disabled, and vice versa.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Sets the bit state of a field that only consists of one bit.

NameTypeValuesDescription
Frame--
Field--
Bit--

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe. The operator is set with SBUS<sb>:MANCh:FILTer:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
data--End value of the data pattern range.

Defines the error type to be filtered on.

NameTypeValuesDescription
error_namestring-Error type to be filtered on.
enableenumOFF | ONEnables or disables the selected error filter.

Enables filtering of a specific error type in the Manchester filter. The command uses the error index <n> and an ON/OFF enable value.

NameTypeValuesDescription
enableenumOFF | ONEnables or disables the selected error filter.

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe. The operator can be set with SBUS<sb>:MANCh:FILTer:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
data_maxstring-End value of the data pattern range.

Specifies the data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
frame--
field--
data--

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
frameinteger1 - 4Index of the serial bus.
fieldinteger-Index of the field.
operatorenumEQUal | NEQual | LTHan | LEThan | GTHan | GEXHan | INRange | OORangeOperator for the data pattern.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
Frame--
Field--
Enabler--

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LEThan | GTHan | GEThan | INRange | OORangeOperator used for the data pattern.

Enables filtering on the specified Manchester frame. Only frames that match the selected filter conditions are displayed.

NameTypeValuesDescription
enableenumOFF | ONTurns frame filtering off or on.

Sets the end value of an index range when the operator is set to INRange. The start/end range behavior is controlled by the related IOPerator command.

NameTypeValuesDescription
index_maxinteger1 - 65535End value of the index range.

Enables filtering on Manchester frames. Only frames that match the selected filter conditions are displayed.

NameTypeValuesDescription
frame--
enablerenumOFF | ON

Sets the end value of an index range when the operator is set to INRange. The operator is set with `SBUS<sb>:MANCh:FILTer:FRAMe<fr>:FLD<fl>:IOPerator`.

NameTypeValuesDescription
index_Maxinteger1 - 65535End value of the index range.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
Frame--
Field--
OperatorenumEQUal | INRange | RANGe
SBUS<sb>:MANCh:FILTer:FRAMe<fr>:FLD<fl>:IOPerator <Index_Operator>

Specifies the index, or sets the start value of an index range.

NameTypeValuesDescription
Frame--
Field--
Data--
returns

Sets the operator for the index in the selected field of the selected frame. The related query returns the operator.

NameTypeValuesDescription
Frameinteger1 - 4Index of the serial bus.
Fieldinteger-Index of the field.
OperatorenumEQUal | INRange | RANGeIndex operator to apply.

Appends a new frame description to the frame list.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Loads the specified XML file containing a list of frame descriptions. It is a setting-only asynchronous command.

NameTypeValuesDescription
FileNamestring-Name of the XML file to load.

Saves the current list of frame descriptions to an XML file with the specified name.

NameTypeValuesDescription
FileNamestring-

Loads the specified XML file containing a list of frame descriptions. Setting only, asynchronous command.

NameTypeValuesDescription
FileNamestring-Name of the XML file to load.

Saves the current list of frame descriptions to an XML file with the specified name.

NameTypeValuesDescription
FileNamestring-Name of the XML file to save the frame description list to.

Erases from volatile memory all field and frame format descriptions created for a specific custom protocol.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Appends a new field description to the selected frame description.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.

Specifies the color for the frame description of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
colorreal0 - 4294967295Color in 32-bit RGB encoding in decimal format.

Specifies the name for the frame description of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
Name--

Specifies the bit count, hence length, of the selected field in the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
fl--Index of the field.
BitCountinteger1 - 64Bit count of the selected field.

Specifies the color for the selected field of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
Colorinteger0 - 4294967295Use 32-bit RGB encoding in decimal format.

Specifies a user-defined condition operator for the selected field of the selected frame. The numeric format of the condition is set by the field format command.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
Conditionnumeric-User-defined condition operator for the selected field.

Specifies the order in which the algorithm evaluates the bits of the condition value of the selected field in the selected frame.

NameTypeValuesDescription
bitorderenumLSBF | MSBFBit evaluation order for the condition value.

Specifies the numerical format for the condition value of the selected field in the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
fl--Index of the field.
NumericFormatenumHEX | OCT | BIN | ASCII | ASCii | SIGN | USIG | DEC | LIST | FLINNumerical format for the condition value.

Specifies which result column of the decode table displays the selected field of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
ColumnenumNONE | COL1 | COL2 | COL3Result column selection.

Queries the software trigger time in seconds. If no software trigger event has occurred, the command returns Invalid.

real (s) - Returns the software trigger time in seconds, or Invalid if no software trigger event has occurred.

Queries the software trigger index. If no software trigger event has occurred, the command returns Invalid.

integer - Returns the software trigger index, or Invalid if no software trigger event has occurred.

Returns the number of decoded frames.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
integer - The number of decoded frames.

Returns the bit rate of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (bps) - Returns the frame bit rate.

Returns the label of the specified field in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus
fr--Index of the frame
fl--Index of the field
returns - Returns the label of the specified field in the specified frame.

Returns the state of the specified field in the specified frame of the specified serial bus Manchester protocol decode. The result indicates whether the field is valid or has a length, CRC, parity, or completeness error.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
fl--Index of the field.
enum - State of the field. OK means the field has no error and is valid; LENGth, CRC, PARity, and INComplete indicate the respective error conditions.

Returns the value of the specified field in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
returns - Returns the value of the specified field in the specified frame.

Returns the formatted value of the specified field in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
string - Formatted value of the specified field in the specified frame.

Returns the overall state of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
enum - Overall state of the specified frame.

Returns the content of the 1st result column of the specified decoded frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
returns - The content of the 1st result column of the specified decoded frame.

Returns the end time of the specified decoded frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (s) - End time of the specified frame.

Returns the type of the specified decoded frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
string - The frame type is undefined, because it is custom specific.

Returns the content of the 3rd result column of the specified decoded frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
string - Content of the 3rd result column of the specified decoded frame.

Returns the content of the 2nd result column of the specified decoded frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
returns - The content of the 2nd result column of the specified decoded frame.

Selects the CAN-High or CAN-Low line for the serial bus. Both lines are required for differential signal transmission used by CAN.

NameTypeValuesDescription
SignalTypeenumCANL | CANHSelects the CAN line to use.

Sets the source channel to which the line is connected.

NameTypeValuesDescription
sourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8Source channel selection; available choices depend on whether digital channels are present and whether SIC or FAST mode is used.

Sets a threshold value for the SIC transceiver mode.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
SIC Threshold--

Selects the transceiver mode for CAN decoding.

NameTypeValuesDescription
modeenumSIC | FASTTransceiver mode.

Sets the bit rate of the data phase for the CAN XL frame.

NameTypeValuesDescription
XLBitratereal10000 - 30000000Bit rate of the CAN XL data phase.

Sets the position of the sample point within the bit in percent of the nominal bit time for the CAN XL frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
XLSamplePointreal30 - 90Sample point position within the bit, in percent of the nominal bit time.

Activates the symbol list used for CAN decoding. The command applies to the selected serial bus index <sb> from 1 to 4.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
ShowSymbolsenumOFF | ONTurns the symbol list off or on.

Returns the enable state of the CAN filter error condition for the selected serial bus. The query takes an `<Enabler>` qualifier.

NameTypeValuesDescription
enablerenumOFF | ONSelects the enable state to query.
enum - Returns the enable state of the CAN filter error condition.

Enables or disables the checking condition for the selected field of the selected frame. The command applies to CAN filter frame/field enable control for the specified serial bus.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frame--Selected frame.
field--Selected field.
enablerenumOFF | ONEnables or disables the checking condition.
enum - Returns the current checking-condition enable state.

Queries whether CAN frame filtering is enabled for the selected serial bus. The manual lists the corresponding setter and indicates that filtering displays only frames matching the selected filter conditions.

NameTypeValuesDescription
EnablerenumOFF | ONEnables or disables filtering on CAN frames.
enum - Returns the filter enable state.

Sets the end value of an index range if the operator is set to INRange. The related frame/field form uses the same function for a specific frame field.

NameTypeValuesDescription
Frame--
Field--
Data--

Sets the operator for the selected index in the selected field of the selected frame. The query returns the operator for the selected field index.

enum - Returns the operator for the selected field index.

Queries the software trigger time in seconds. If no software trigger event has occurred, the command returns Invalid.

real (s) - Returns the software trigger time in seconds. If no software trigger event has occurred, the command returns Invalid.

Queries the software trigger index for the selected serial bus. If no software trigger event has occurred, the command returns Invalid.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
integer - Index of the software trigger event, or Invalid if no software trigger event has occurred.

Sets the number of transmitted bits per second for the LIN bus. The maximum bit rate for LIN is 20 kbit/s.

NameTypeValuesDescription
Bitratereal20 kbit/sBit rate to transmit in bits per second.

Sets the waveform of the data line.

NameTypeValuesDescription
DataSourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8

Defines the idle state of the LIN bus. The idle state is the recessive state and corresponds to a logic 1.

NameTypeValuesDescription
polarityenumIDLLow | IDLHighIdle state polarity of the bus.

Sets the vertical position of the LIN signal.

NameTypeValuesDescription
LIN position--

Sets the vertical scale of the indicated LIN signal.

NameTypeValuesDescription
LIN scale--

Selects the version of the LIN standard for the specified serial bus.

NameTypeValuesDescription
StandardenumV1X | V2X | J2602 | AUTOLIN standard version to use.

Loads a symbol list file.

NameTypeValuesDescription
FileNamestring-String parameter with path and file name.
SBUS1:LIN:NEWList
'/home/storage/userData/Protocol/LIN.csv'
SBUS1:LIN:SYMBols ON

Activates the symbol list to be used for decoding.

NameTypeValuesDescription
ShowSymbolsenumOFF | ONSelects whether the symbol list is used for decoding.
SBUS1:LIN:SYMBols ON

Inverts the current state of the frame and error types for the selected serial bus filter: enabled items become disabled and vice versa.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Frame--
Field--
Bitinteger-

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe.

NameTypeValuesDescription
Frame--
Field--
Data--

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
BitStateenumONE | ZEROOperator state for the data pattern bit.

Presets the state of the selected frames and error types. It is a setting-only asynchronous command.

Specifies the data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
Frameinteger-Frame index.
Fieldinteger-Field index.
Data--Data pattern or start value of a data pattern range.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Frame--
Field--
OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator used for the data pattern in the selected field of the selected frame.

Defines the error type to be filtered on.

NameTypeValuesDescription
enableenumOFF | ONEnables or disables the error filter.
enum - Returns the enable state of the error filter.

Defines the error type to be filtered on.

NameTypeValuesDescription
enableenumOFF | ONEnables or disables filtering for the selected error type.

Specifies the data pattern, or sets the start value of a data pattern range for a LIN filter frame field. The command variant shown for the selected frame and field sets the minimum data value.

NameTypeValuesDescription
Data_Minreal-Start value of a data pattern range.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
data_operatorenumEQUal | NEQual | LTHan | LEThan | GTHan | GEThan | INRange | OORangeOperator for the data pattern.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
Frame--Selected frame index.
Field--Selected field index.
EnablerenumOFF | ONEnables or disables the checking condition for the selected field.

Enables the filtering on LIN frames. Only the frames that match the selected filter conditions are displayed.

NameTypeValuesDescription
enableenumOFF | ON

Enables or disables filtering on LIN frames. Only frames that match the selected filter conditions are displayed.

NameTypeValuesDescription
Frameinteger1 - 4Index of the serial bus.
EnablerenumOFF | ONTurns frame filtering off or on.

Sets the end value of an index range if the operator is set to INRange.

NameTypeValuesDescription
datainteger-End value of the index range.

Specifies the index, or sets the start value of an index range, for the selected LIN filter frame field.

NameTypeValuesDescription
Index_Mininteger1 - 65535Start value of the index range.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
index_operatorenumEQUal | INRange | RANGeOperator used for the index in the selected field of the selected frame.

Specifies the index, or sets the start value of an index range.

NameTypeValuesDescription
Index_Mininteger1 - 65535Index value to specify or use as the start of an index range.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
Frame--Frame index.
Field--Field index.
OperatorenumEQUal | INRange | RANGeOperator for the index in the selected field of the selected frame.

Queries the software trigger index. If no software trigger event has occurred, the command returns Invalid.

integer - Returns the software trigger index, or Invalid if no software trigger event has occurred.

Returns the number of decoded frames.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
integer - Total number of decoded frames.

Queries the software trigger time in seconds. If no software trigger event has occurred, the command returns Invalid.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
real (s) - Software trigger time in seconds; returns Invalid if no software trigger event has occurred.

Returns the primary bit rate for the specified LIN frame. This is a query-only asynchronous command.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (bps) - The primary bit rate.

Returns the state of the specified field.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
enum - State of the specified field.

Returns the formatted value of the specified field.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
returns - Returns the formatted value of the specified field.

Returns the label of the fields in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
returns

Returns the value of the specified field.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
integer - Returns the value of the specified field.

Returns the data bytes of the specified frame in comma-separated values.

string - Returns the frame data as a count followed by comma-separated byte values.
BUS:LIN:FRAMe4:DATA?

Returns the value of the identifier parity bits of the selected frame.

returns - Returns the value of the identifier parity bits of the selected frame.

Returns the identifier value of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - Identifier value of the selected frame.

Returns the start time of the selected frame. Query only and asynchronous.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (s) - Returns the start time of the selected frame.

Returns the overall state of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
returns

Returns the symbolic label of the specified frame if the label list is enabled.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
returns - Symbolic label of the specified frame if the label list is enabled.

Sets the transmitter-specific nominal clock period (clock tick).

NameTypeValuesDescription
ClockPeriodreal-Nominal clock period for the SENT transmitter.

Specifies a tolerated deviation of the clock for the selected serial bus.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
ClockTolerancereal1 - 25Tolerated deviation of the clock.

Selects the method for CRC calculation.

NameTypeValuesDescription
CRCCalculationenumSAEJ | TLECRC calculation method.

Sets the source of the data line.

NameTypeValuesDescription
DataSourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8Source used for the data line.

Sets the threshold for the data channel. The serial bus index suffix <sb> selects bus 1 to 4.

NameTypeValuesDescription
SDATA Thresreal-Threshold value for the data channel.

Sets the number of data units in a single transmission sequence.

NameTypeValuesDescription
DataNibblesinteger1 - 6Number of data units in a single transmission sequence.

Selects the operating mode of the SENT protocol. If Legacy is selected, the SPC mode is not supported; if SPC mode is selected, the SENT protocol allows calibration by sending a single calibrated value.

NameTypeValuesDescription
modeenumLEGacy | SPCOperating mode of the SENT protocol.

Loads a symbol list file for the selected serial bus SENT protocol.

NameTypeValuesDescription
FileNamestring-String parameter with path and file name.
SBUS1:SENT:NEWList '/home/storage/userData/Protocol/SENT.csv'
SBUS1:SENT:SYMBols ON

Sets the vertical position of the SENT signal.

NameTypeValuesDescription
SENT position--

Specifies the fixed SENT frame length in ticks. This requires the pause pulse setting SBUS<sb>:SENT:PPULse to be PPFL.

NameTypeValuesDescription
FrameLengthinteger104 - 922Fixed frame length in ticks.

Determines whether a pause pulse is transmitted after the checksum nibble. A PPFL setting transmits a pause pulse to achieve a fixed frame length specified by SBUS<sb>:SENT:PPFLength.

NameTypeValuesDescription
PausePulseenumNPP | PP | PPFLSelects whether no pause pulse, a pause pulse, or a pause pulse for fixed frame length is transmitted.

Selects the results to be displayed.

NameTypeValuesDescription
RessDispSelenumTRSQ | SMSG | ALLSelects the displayed results: transmission sequence, serial messages, or all.

Sets the vertical scale of the indicated SENT signal.

NameTypeValuesDescription
SENT scale--

Selects if serial messages are enabled or disabled.

NameTypeValuesDescription
SerialMessagesenumENABled | DISabledSelects whether serial messages are enabled or disabled.

Activates the symbol list to be used for decoding.

NameTypeValuesDescription
ShowSymbolsenumOFF | ONSelects whether the symbol list is shown for decoding.

Enables the filter for all available frames and error types.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Disables the filter for all available frames and error types. It is a setting-only asynchronous command for the selected serial bus.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Frame--
Field--
Bit--

Enables filtering on the selected SENT frame. Only the frames that match the selected filter conditions are displayed.

NameTypeValuesDescription
EnableenumOFF | ONEnables or disables frame filtering.

Enables filtering on SENT frames. Only frames that match the selected filter conditions are displayed.

NameTypeValuesDescription
Frameinteger1 - 4Index of the frame.
EnablerenumOFF | ONEnables or disables the filter for the selected frame condition.

Inverts the current state of the selected SENT frame and error-type filters, turning enabled items off and disabled items on.

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe. You can set the operator with SBUS<sb>:SENT:FILTer:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
frameinteger1 - 4Index of the serial bus.
field--Index of the field.
data--

Specifies the data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
Data_Min--

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe. Use SBUS<sb>:SENT:FILTer:FRAMe<fr>:FLD<fl>:DOPerator to set the operator.

NameTypeValuesDescription
Data_Max--

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Frame--
Field--
OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORange

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
Frame--
Field--
EnablerenumOFF | ON

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator used for the data pattern in the selected field.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
enablerenumOFF | ONEnables or disables the checking condition.

Sets the end value of an index range if the operator is set to INRange. The operator is set with the corresponding IOPerator command.

NameTypeValuesDescription
index_maxinteger-End value of the index range.

Sets the end value of an index range if the operator is set to INRange. The operator is set with the corresponding IOPerator command.

NameTypeValuesDescription
datainteger-End value of an index range.

Specifies the index, or sets the start value of an index range.

NameTypeValuesDescription
Index_Mininteger1 - 65535Index value for the selected field of the selected frame.

Specifies the index, or sets the start value of an index range.

NameTypeValuesDescription
index_mininteger1 - 65535Start index value.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
Index_OperatorenumEQUal | INRange | RANGeOperator for the index in the selected field of the selected frame.

Defines the error type to be filtered on for the specified serial bus. The bus is selected with the <sb> suffix, which ranges from 1 to 4.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
ErrorName--Error type to be filtered on.
EnablerenumOFF | ONEnables or disables the filter.

Defines the error type to be filtered on for the specified serial bus error index. The command uses suffix <sb> for the serial bus index and <n> for the error index.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
ninteger-Index of the error.
enableenumOFF | ONEnables or disables the error filter.

Queries the software trigger time in seconds. If no software trigger event has occurred, the command returns Invalid.

real (s) - Software trigger time in seconds, or Invalid if no software trigger event has occurred.

Queries the software trigger index. If no software trigger event has occurred, the command returns Invalid.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
integer - Returns the software trigger index, or Invalid if no software trigger event has occurred.

Returns the bit rate of the selected serial bus frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (bps) - The bit rate of the frame.

Returns the data of the specified frame.

integer - Comma-separated sequence of integer values (N, D1, D2,…, DN), where N is the number of nibbles in the frame and D1…DN are the nibble values.

Returns the formatted status communication for the selected frame. This is a query-only asynchronous command.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
string - Formatted status communication.

Returns the formatted symbolic data values. Query only and asynchronous.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
string - Formatted symbolic data values.

Returns the identifier type of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
enum - Identifier type of the selected frame.

Returns the number of fields in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
returns - Returns the number of fields in the specified frame.

Returns the identifier value of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
integer - Identifier value of the selected frame.

Returns the duration of the main trigger pulse (MTP) for the selected frame. The MTP is an additional pulse added to the frame when running in "Short PWM Code" (SPC) mode.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
real (s) - Duration of the main trigger pulse.

Returns the formatted value of the specified field.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
returns - Returns the formatted value of the specified field.

Returns the state of the specified field.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
enum - State of the specified field.

Returns the value of the specified field.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
integer - The value of the specified field.

Returns the label of the specified field.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
returns - Returns the label of the specified field.

Returns the number of the pulse pause clock ticks for the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
integer - Returns the pulse pause clock tick count.

Returns the value of the status/communication pulse.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - Returns the value of the status/communication pulse.

Returns the stop time of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (s) - Stop time of the selected frame.

Returns the symbolic label of the specified frame if the label list is enabled.

string - The symbolic label of the specified frame.
BUS:SENT:FRAMe:SYMBol?

Returns the time of the synchronization pulse.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (s) - Time of the synchronization pulse.

Sets the wire on which the bus signal is measured, normal or inverted. The setting affects digitization of the signal.

NameTypeValuesDescription
polarityenumNORMal | INVertedSets the bus signal polarity.

Sets the channel for the signal source.

NameTypeValuesDescription
SourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8Channel used as the MIL-STD signal source.

Sets the lower threshold level of the signal. The setting applies to the specified serial bus bus signal measurement.

NameTypeValuesDescription
threshold high--

Loads a symbol list file.

NameTypeValuesDescription
FileNamestring-String parameter with path and file name.
SBUS1:MILStd:NEWList
'/home/storage/userData/Protocol/MILStd.csv'
SBUS1:MILStd:SYMBols ON

Sets the vertical scale of the indicated MIL-1553 signal.

NameTypeValuesDescription
Scale--

Activates the symbol list used for decoding on the specified serial bus. The bus index is given by <sb> and the setting is OFF or ON.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
ShowSymbolsenumOFF | ONTurns the symbol list used for decoding off or on.

Enables the filter for all available frames and error types.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Inverts the current state of the frame and error types: all enabled frames and error types are disabled and vice versa.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Presets the state of the selected frames and error types.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Frame--
Field--
BitenumONE | ZEROBit state to set for the selected field of the selected frame.

Returns the selected bit index for the selected MIL-STD data pattern field.

integer - Returns the selected bit index.

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe. The operator is set with the corresponding DOPerator command.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
datainteger-

Defines the error type to be filtered on.

NameTypeValuesDescription
ErrorName--
EnablerenumOFF | ON
enum - Returns the enabled state for the selected error filter.

Enables or disables filtering for a specific error type in the MIL-STD filter. The error index is selected with <n>.

NameTypeValuesDescription
enableenumOFF | ONTurns filtering for the selected error type off or on.

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe. You can set the operator with SBUS<sb>:MILStd:FILTer:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
data_max--End value of a data pattern range.

Specifies the data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
Frame--
Field--
Datanull-

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LEThan | GTHan | GEThan | INRange | OORangeOperator used for the data pattern in the selected field of the selected frame.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator for the data pattern in the selected field of the selected frame.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
operatorenumEQUal | NEQual | LTHan | LEThan | GTHan | GEThan | INRange | OORangeOperator used for the data pattern in the selected field.

Sets the end value of an index range if the operator is set to INRange.

NameTypeValuesDescription
Index_Maxinteger-End value of an index range.

Sets the end value of an index range if the operator is set to INRange.

NameTypeValuesDescription
index_maxreal-End value of the index range.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
index_operatorenumEQUal | INRange | RANGeOperator for the index in the selected field of the selected frame.

Specifies the index, or sets the start value of an index range.

NameTypeValuesDescription
Index_Mininteger1 - 65535Start value of an index range.

Specifies the index, or sets the start value of an index range.

NameTypeValuesDescription
Data--
returns - Returns the index value for the selected frame and field.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
index_operatorenumEQUal | INRange | RANGeOperator for the index in the selected field of the selected frame.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
operatorenumEQUal | INRange | RANGeIndex operator for the selected field of the selected frame.

Enables or disables the checking condition for the selected field of the selected frame. Only the specified frame field is affected.

NameTypeValuesDescription
Frameinteger1 - 4Index of the serial bus frame.
Fieldinteger-Index of the field within the frame.
EnablerenumOFF | ONEnables or disables the condition check.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frameinteger-Index of the frame.
fieldinteger-Index of the field.
enablerenumOFF | ONEnables or disables the checking condition for the selected field.
enum - Returns the enable state of the selected field condition.

Enables the filtering on the selected MIL-1553 frame. Only frames that match the selected filter conditions are displayed.

NameTypeValuesDescription
EnableenumOFF | ONTurns frame filtering on or off.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
enableenumOFF | ONEnables or disables the checking condition for the selected field of the selected frame.

Enables the filtering on MIL-1553 frames. Only the frames that match the selected filter conditions are displayed.

NameTypeValuesDescription
Frameinteger1 - 4Index of the serial bus frame to filter.
EnablerenumOFF | ONTurns frame filtering off or on.

Enables the filtering on MIL-1553 frames. Only the frames that match the selected filter conditions are displayed.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frameinteger-Index of the frame.
enablerenumOFF | ONEnables or disables frame filtering.
enum - Returns the enabled state for the selected filter.

Queries the software trigger time in seconds. If no software trigger event has occurred, the command returns Invalid.

real (s) - Returns the software trigger time in seconds; if no software trigger event has occurred, returns Invalid.

Queries the software trigger index. If no software trigger event has occurred, the command returns Invalid.

integer - Returns the software trigger index, or Invalid if no software trigger event has occurred.

Returns the bit rate of the specified MIL-STD word.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
winteger*Index of the word.
integer (bps) - The primary bit rate of the word.

Returns the data pattern of the specified word.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
wstring-Index of the word.
string - List of comma separated values.

Returns the info value for the specified word.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
winteger-Index of the word.
returns

Returns the RT address for the selected word.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
winteger*Index of the word.
integer - RT address for the selected word.

Returns the start time of the selected word.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
winteger* - *Index of the word.
real (s) - The start time of the selected word.

Returns the stop time of the selected word. This is a query-only asynchronous command.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
winteger-Index of the word.
real (s) - Stop time of the selected word.

Returns the symbolic label of the specified word if the label list is enabled.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
winteger-Index of the word.
string - Returns the symbolic label of the specified word if the label list is enabled.

Returns the type of the specified word.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
winteger* - *Index of the word.
enum - Type of the specified word.

Selects the channel for the data signal source.

NameTypeValuesDescription
Sourceenum-Channel for the data signal source.

Selects the source for the strobe signal.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
StrobeSourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8

Sets the minimum duration of a gap. Any inactivity greater than this time is interpreted as a gap and leads to a resynchronization to the signal.

NameTypeValuesDescription
MinGapTimereal1E-09 - 1Minimum gap duration.

Sets the minimum duration of a gap. Any inactivity greater than this time is interpreted as a gap and leads to a resynchronization to the signal.

NameTypeValuesDescription
MinGapTimereal1E-09 - 1Minimum gap duration.

Sets the vertical position of the SpaceWire signal.

NameTypeValuesDescription
positionreal1E-09 - 1Vertical position of the SpaceWire signal.

Sets a value for the hysteresis of the strobe signal.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
StrobeHyst--

Sets the vertical scale of the SpaceWire signal.

NameTypeValuesDescription
scale--

Sets the bit state of a field that only consists of one bit.

NameTypeValuesDescription
Frameinteger1 - 4Index of the serial bus frame.
Fieldinteger-Field selector for the bit state command.
Bitinteger-Bit value to set.

Returns the bit state of a field that only consists of one bit.

NameTypeValuesDescription
bitinteger0 - 1Bit index to query.
enum - Returns the bit state of the selected field bit.

Enables the filter for all available frames and error types.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Disables the filter for all available frames and error types.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Sets the bit state of a field that only consists of one bit.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Frame index.
flinteger-Field index.
BitStateenum-Bit state to set.

Inverts the current state of the frame and error types: all frames and error types that were enabled are disabled and vice versa.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Sets the end value of a data pattern range when the operator is set to INRange or OORANGe. The operator is set with SBUS<sb>:SWIRe:FILTer:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
frameinteger-Index of the frame.
fieldinteger-Index of the field.
data--

Returns the end value of a data pattern range. This query corresponds to the frame/field DMAX setting used when the operator is set to INRange or OORANGe.

real - The end value of the data pattern range.

Queries the enable state of the serial bus sweep filter error filter for a specified bus and error name. The slice shows the query form with a single <Enabler> argument, but does not document the returned value.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe. The operator can be set with the corresponding DOPerator command.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
data_max--End value of the data pattern range.

Specifies the data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
Data_Min--
returns

Specifies the data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
Data--

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
frame--
field--
operatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator used for the data pattern comparison.
enum - Returns the current data-pattern operator.

Sets the operator for the data pattern in the selected field of the selected frame. The query returns the current operator setting.

enum - Returns the operator for the data pattern in the selected field of the selected frame.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
Frame--Frame index.
Field--Field index.
Enabler--Enables or disables the checking condition.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
Enabler--
returns

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
data_operatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator used for the data pattern.
SBUS<sb>:SWIRe:FILTer:FRAMe<fr>:FLD<fl>:DOPerator <Data_Operator>
SBUS<sb>:SWIRe:FILTer:DOPerator? <Operator>

Enables filtering on the specified SpaceWire frame. Only frames that match the selected filter conditions are displayed.

NameTypeValuesDescription
<fr>integer1 - 4Index of the frame.
<Enable>enumOFF | ONEnables or disables frame filtering for the selected frame.

Enables filtering on SpaceWire frames. Only frames matching the selected filter conditions are displayed.

NameTypeValuesDescription
frameinteger1 - 4Index of the serial bus frame.
enablerenumOFF | ONTurns frame filtering off or on.

Enables filtering on SpaceWire frames. Only frames that match the selected filter conditions are displayed.

NameTypeValuesDescription
<Frame>integer1 - 4Index of the frame.
<Enabler>enumOFF | ONEnables or disables the filter on the selected frame.
enum - Returns the enable state for the selected SpaceWire frame filter.

Queries the software trigger time in seconds. If no software trigger event has occurred, the command returns Invalid.

real (s) - The software trigger time in seconds; returns Invalid if no software trigger event has occurred.

Queries the software trigger index. If no software trigger event has occurred, the command returns Invalid.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
integer - Software trigger index; returns Invalid if no software trigger event has occurred.

Returns the parity state of the specified SpaceWire frame field, or the frame status if no such field is present.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
enum - Returns the parity/status code for the frame: OK if the parity check passes, PARity if the parity is incorrect, LENGth if the frame length is not as expected, AMBiguous if the frame is ambiguous, or INComplete if the frame is incomplete.

Returns the data of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
block - Frame data for the specified frame.

Returns the parity state of the Data field in the specified serial bus frame, or frame status codes if the frame has no Data field. This is a query-only asynchronous command.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
enum - Parity state of the Data field, or the frame status if the Data field is absent.

Returns the parity state of the ESC field for the specified serial bus frame. If the frame has no ESC field, it can return a frame status such as length, ambiguous, or incomplete.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
enum - Returns the parity state of the ESC field, or a frame status if the frame has no ESC field.

Returns the overall state of the specified field in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
fl--Index of the field.
enum - Overall state of the specified field in the specified frame.

Returns the label of the specified field in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
fl--Index of the field.

Returns the formatted value of the specified field in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
string - Formatted value of the specified field in the specified frame.

Returns the value of the specified field in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
fl--Index of the field.
returns - Returns the value of the specified field in the specified frame.

Returns the start time of the specified frame. Query only; asynchronous command.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (s) - Start time of the specified frame.

Returns the end time of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (s) - End time of the specified frame.

Returns the type of the specified decoded frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
enum - Type of the specified decoded frame.

Sets a value for the group sub index. Available only if SBUS<sb>:SPMI:GSIDenable is ON.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
GSIDinteger0 - 15Group sub index value.

Enables the use of the group sub ID (GSID). The GSID value is set with SBUS<sb>:SPMI:GIDValue.

NameTypeValuesDescription
UseGSIDenumOFF | ONEnables or disables use of the group sub ID (GSID).

Enables or disables the glitch filter for the selected serial bus. A glitch filter can help filter out short-duration voltage spikes/glitches on the communication line.

NameTypeValuesDescription
UseGlitchFilterenumOFF | ONTurns the glitch filter off or on.

Sets the glitch width. Any signal transitions with a duration smaller than this value are considered a glitch and filtered out. This command is available only if SBUS<sb>:SPMI:GTCHenable is ON.

NameTypeValuesDescription
GlitchFilterWdtreal1E-15 - 1000000000Glitch filter width threshold.

Sets the source of the SPMI clock line.

NameTypeValuesDescription
ClockSourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8

Sets a value for the hysteresis for the source data channel.

NameTypeValuesDescription
SDATA Hyst--Value for the hysteresis for the source data channel.

Sets the source of the data line.

NameTypeValuesDescription
DataSourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8Selects the source signal for the data line.

Sets the vertical position of the SPMI signal.

NameTypeValuesDescription
SPMI position--

Sets the vertical position of the SPMI signal.

NameTypeValuesDescription
SPMI scale--

Sets a user-defined threshold value for the source data line.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
SDATA Thres--

Activates the symbol list to be used for decoding.

NameTypeValuesDescription
ShowSymbolsenumOFF | ONTurns the symbol list used for decoding off or on.

Loads a symbol list file. The command is setting-only and asynchronous.

NameTypeValuesDescription
FileNamestring-String parameter with path and filename.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Frameinteger1 - 4Selected frame index.
Fieldinteger-Selected field index.
Bitinteger-Selected bit.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Bitinteger-

Enables the filter for all available frames and error types.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
<sb>integer1 - 4Index of the serial bus.
<fr>--Selected frame.
<fl>--Selected field.
<BitState>--Operator for the data pattern.

Inverts the current state of the frame and error types: all frames and error types that were enabled are disabled and vice versa.

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe. You can set the operator with SBUS<sb>:SPMI:FILTer:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
data_maxstring-

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe. The operator is set with `SBUS<sb>:SPMI:FILTer:FRAMe<fr>:FLD<fl>:DOPerator`.

NameTypeValuesDescription
Data_Maxreal-End value of a data pattern range.
real - Returns the end value of a data pattern range.

Specifies the data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
Data_Min--

Specifies the data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
data_minstring-The data pattern or start value of the data pattern range.
string - Returns the specified data pattern or start value.

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe. The operator is set with :SBUS<sb>:SPMI:FILTer:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
Data_Maxstring-End value of a data pattern range.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Frameinteger1 - 4Index of the serial bus.
Fieldinteger-Index of the field.
OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator for the data pattern.

Returns the operator for the data pattern in the selected field of the selected frame.

enum - Returns the current data-pattern operator for the selected field of the selected frame.

Defines the error type to be filtered on.

NameTypeValuesDescription
enableenumOFF | ONEnables or disables the selected error filter.

Defines the error type to be filtered on.

NameTypeValuesDescription
enablerenumOFF | ONEnables or disables the error filter.
enum - Returns the enabled state of the error filter.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator for the data pattern.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
Frameinteger1 - 4Index of the serial bus.
Fieldinteger-Index of the field.
Index_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator for the index in the selected field of the selected frame.

Sets the operator for the index in the selected field of the selected frame.

enum - Returns the index operator for the selected field of the selected frame.

Sets the end value of an index range if the operator is set to INRange. The operator can be set with SBUS<sb>:SPMI:FILTer:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
Index_Maxinteger1 - 65535End value of the index range.

Sets the end value of an index range if the operator is set to INRange. The operator can be set with SBUS<sb>:SPMI:FILTer:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
Index_Maxinteger1 - 65535End value of the index range.

Sets or queries the end value of an index range when the filter operator is set to INRange. The related field- and frame-specific form is noted in the slice.

NameTypeValuesDescription
Datastring-No parameter description is provided in the slice.
integer - Returns the current end value of the index range.

Specifies the index, or sets the start value of an index range.

NameTypeValuesDescription
frame--Frame index placeholder from the syntax.
field--Field index placeholder from the syntax.
data--Data placeholder from the syntax.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
Frameinteger1 - 4Index of the serial bus frame.
Fieldinteger-Index of the field.
EnablerenumOFF | ONEnables or disables the checking condition.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
<Frame>--
<Field>--
<Enabler>enumOFF | ON
enum - Returns the enabled state for the selected field check condition.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
enableenumOFF | ONEnables or disables the checking condition for the selected field of the selected frame.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
CondEnablerenumOFF | ONEnables or disables the checking condition for the selected field.

Enables the filtering on SPMI frames. Only the frames that match the selected filter conditions are displayed.

NameTypeValuesDescription
Frame--
EnablerenumOFF | ON

Enables the filtering on SPMI frames. Only the frames that match the selected filter conditions are displayed.

NameTypeValuesDescription
EnablerenumOFF | ONEnables or disables the filter condition.
enum - Returns the filter enable state.

Queries the software trigger time in seconds. If no software trigger event has occurred, the command returns Invalid.

real (s) - Returns the software trigger time in seconds, or Invalid if no software trigger event has occurred.

Queries the software trigger index. If no software trigger event has occurred, the command returns Invalid.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
integer - Software trigger index, or Invalid if no software trigger event has occurred.

Returns the number of decoded frames in the acquisition.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
integer - Number of decoded frames in the acquisition.

Returns the address of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - Frame address.

Returns the BC of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - BC of the specified frame.

Returns the formatted value of the specified byte.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
returns - Formatted value of the specified byte.

Returns the label of the fields in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
string - Returns the label of the fields in the specified frame.

Returns the overall state of the specified frame field. This is a query-only asynchronous command.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
fl--Index of the field.
enum - Overall state of the frame field.

Returns the data value of the specified field.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frnull-Index of the frame.
flnull-Index of the field.
integer - The value of the specified field.

Returns the start time of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (s) - The start time of the selected frame.

Returns the frame type.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
enum - Returns the frame type.

Returns the symbolic label of the specified frame if the label list is enabled.

string - String with the symbolic name of the address or the symbolic name of the frame.

Sets the source of the RFFE clock line for the selected serial bus.

NameTypeValuesDescription
<SourceClock>enumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8Selects the clock source signal.

Sets the source of the data line.

NameTypeValuesDescription
SourceDataenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8Selects which channel, memory, or reference signal drives the data line.

Enables or disables the glitch filter on the SCLK and SDATA lines to improve decode accuracy.

NameTypeValuesDescription
glitchfilterenumOFF | ONTurns the glitch filter off or on.

Sets the maximum glitch width to be ignored.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
GlitchFilterWdtreal1E-11 - 1E-08Maximum glitch width to be ignored.

Sets the vertical position of the RFFE signal.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
RFFE position--

Selects whether the standard or synchronous read mode is used.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
ReadModeenumSTRD | SREADRead mode to use: standard or synchronous read.

Sets the vertical scale of the indicated RFFE signal.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
RFFE scale--Vertical scale of the indicated RFFE signal.

Enables filtering on RFFE frames. Only frames that match the selected filter conditions are displayed.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
enableenumON | OFFTurns frame filtering on or off.

Enables filtering on RFFE frames so that only frames matching the selected filter conditions are displayed.

NameTypeValuesDescription
Frameinteger1 - 4Index of the frame.
EnablerenumON | OFFEnables or disables frame filtering.

Inverts the current state of the frame and error types, so enabled items become disabled and vice versa.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Frameinteger1 - 4Index of the frame.
Fieldinteger-Index of the field.
Bitinteger-Bit index.

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe. The operator is set with SBUS<sb>:RFFE:FILTer:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
frame--
field--
data--End value of the data pattern range.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
frameinteger1 - 4Index of the frame.
fieldinteger-Index of the field.
bitstateenumONE | ZEROBit state for the field bit operator.

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe. The operator is set with SBUS<sb>:RFFE:FILTer:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
fl--Index of the field.
Data_Max--

Specifies the data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
frame--
field--
data--Data pattern or start value of a data pattern range.
returns

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
operatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator used for the data pattern comparison.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
Frame--
Field--
Enabler--

Specifies the data pattern, or sets the start value of a data pattern range, for the selected field of the selected frame.

NameTypeValuesDescription
Data_Minreal-Start value of a data pattern range.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LEThan | GTHan | GEThan | INRange | OORangeOperator used for the data pattern comparison.

Sets the end value of an index range if the operator is set to INRange. The operator is set with SBUS<sb>:RFFE:FILTer:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
index_maxinteger1 - 65535End value of an index range.

Sets the end value of an index range if the operator is set to INRange. The operator is set with SBUS<sb>:RFFE:FILTer:FRAMe<fr>:FLD<fl>:IOPerator.

NameTypeValuesDescription
frame--Frame index.
field--Field index.
data--Data value.

Specifies the index, or sets the start value of an index range.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frame--Index of the frame.
field--Index of the field.
data--
Index_Mininteger1 - 65535

Enables or disables filtering on a specific error type for the selected serial bus. The command is part of the RFFE filter settings and uses an error index suffix.

NameTypeValuesDescription
enableenumOFF | ONTurns the selected error filter off or on.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
index_operatorenumEQUal | INRange | RANGeOperator for the index in the selected field.

Sets the operator for the selected field in the selected frame of the serial bus RFFE filter. A query form is documented for reading back the operator.

NameTypeValuesDescription
Frameinteger1 - 4Index of the serial bus.
Fieldinteger-Index of the field.
OperatorenumEQUal | INRange | RANGeOperator to set for the selected field.
enum - Returns the current operator.

Queries the software trigger time in seconds. If no software trigger event has occurred, the command returns Invalid.

real (s) - Returns the software trigger time in seconds, or Invalid if no software trigger event has occurred.

Queries the software trigger index. If no software trigger event has occurred, the command returns Invalid.

integer - Returns the software trigger index, or Invalid if no software trigger event has occurred.

Returns the parity address zero of the selected frame. This is a query-only asynchronous command.

enum - Parity address zero status of the selected frame.

Returns the parity address one of the selected frame. This is a query-only asynchronous command.

enum - Returns the parity address one of the selected frame.

Returns the number of decoded frames in the current acquisition.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
returns - Number of decoded frames in the current acquisition.

Returns the register address of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - Returns the register address of the selected frame.

Returns the byte count of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - The byte count of the selected frame.

Returns the data value of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
returns

Returns the formatted value of the specified field in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
returns - Formatted value of the specified field in the specified frame.

Returns the overall state of the specified field in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
enum - Overall state of the specified field in the specified frame.

Returns the data value of the specified field in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
integer - Data value of the specified field in the specified frame.

Returns the address of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - The address of the specified frame.

Returns the start time of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (s) - The start time of the selected frame.

Returns the overall state of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
enum - Overall state of the selected frame.

Returns the write bit rate of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
real (bps) - Write bit rate of the specified frame.

Returns the symbolic label of the specified sequence if the label list is enabled.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
string - String with symbolic name of the frame.

Activates the symbol list to be used for decoding.

NameTypeValuesDescription
ShowSymbolsenumOFF | ONEnables or disables the symbol list used for decoding.

Sets the vertical position of the 10BASE-T1S signal.

NameTypeValuesDescription
position--Vertical position of the indicated 10BASE-T1S signal.

Sets the vertical scale of the indicated 10BASE-T1S signal.

NameTypeValuesDescription
TENBTO scale--

Selects the source channel for the 10BASE-T1S protocol. Only analog channels can be used.

NameTypeValuesDescription
sourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8Source channel for the 10BASE-T1S protocol.

Activates the symbol list to be used for decoding.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
ShowSymbolsenumOFF | ONTurns the symbol list off or on.

Loads a symbol list file.

NameTypeValuesDescription
FileNamestring-String parameter with path and file name.
SBUS1:TNOS:NEWList
'/home/storage/userData/Protocol/TNOS.csv'
SBUS1:TNOS:SYMBols ON

Sets a lower threshold for the selected serial bus. The bus is selected by the <sb> suffix, which is an index from 1 to 4.

NameTypeValuesDescription
Lower Thresreal-Lower threshold value to set.

Sets an upper threshold for the selected serial bus.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
Upper Thresreal-Upper threshold value.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Frameinteger1Index of the frame.
Fieldinteger1Index of the field.
Bitinteger-

Disables the filter for all available frames and error types.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
BitStateenum-Operator for the data pattern in the selected field.

Inverts the current state of the frame and error types; all frames and error types that were enabled are disabled and vice versa.

Presets the state of the selected frames and error types.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe. The operator is set with SBUS<sb>:TNOS:FILTer:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
Frame--
Field--
Data--

Specifies the data pattern, or sets the start value of a data pattern range. It applies to the serial bus filter field data minimum value.

NameTypeValuesDescription
Data_Min--

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe. The operator is set with SBUS<sb>:TNOS:FILTer:FRAMe<fr>:FLD<fl>:DOPerator.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
Data_Maxnumeric-

Sets the operator for the data pattern in the selected field of the selected frame. The reset value is EQUal.

NameTypeValuesDescription
Frame--
Field--
OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORange

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
errorname--
enablerenumOFF | ON

Enables or disables the selected error filter entry.

NameTypeValuesDescription
enableenumOFF | ONTurns the selected error filter entry off or on.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
Frame--Index of the frame.
Field--Index of the field.
EnablerenumOFF | ONEnables or disables the checking condition.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator used for the data pattern check.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
enablerenumOFF | ONEnables or disables the checking condition.

Sets the end value of an index range if the operator is set to INRange. Use :SBUS<sb>:TNOS:FILTer:FRAMe<fr>:FLD<fl>:IOPerator to set the operator.

NameTypeValuesDescription
Index_Maxinteger1 - 65535End value of an index range.

Specifies the index, or sets the start value of an index range. The value range is 1 to 65535.

NameTypeValuesDescription
Index_Mininteger1 - 65535Start value or index to specify.

Sets the end value of an index range if the operator is set to INRange. It is part of the serial bus TNOS filter command group and is associated with frame/field index range editing.

NameTypeValuesDescription
Frame--Frame index selector.
Field--Field index selector.
Data--Data value for the filter setting.

Specifies the index, or sets the start value of an index range.

NameTypeValuesDescription
Index_Mininteger1 - 65535Start index value for the filter range.

Enables filtering on the selected 10Base-T1S frame. Only frames that match the selected filter conditions are displayed.

NameTypeValuesDescription
EnableenumOFF | ONTurns frame filtering off or on.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
Index_OperatorenumEQUal | INRange | RANGeOperator for the index in the selected field of the selected frame.

Enables filtering on 10Base-T1S frames so only frames matching the selected filter conditions are displayed.

NameTypeValuesDescription
enableenumOFF | ONTurns frame filtering off or on.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
frame--
field--
operatorenumEQUal | INRange | RANGeSelects the index operator for the selected field.

Queries the software trigger time in seconds. If no software trigger event has occurred, the command returns Invalid.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
real (s) - Software trigger time in seconds; returns Invalid if no software trigger event has occurred.

Queries the software trigger index. If no software trigger event has occurred, the command returns Invalid.

integer - The software trigger index, or Invalid if no software trigger event has occurred.

Returns the number of decoded frames.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
integer - Total number of decoded frames.

Returns the number of fields in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - Total number of fields.

Returns the formatted value of the specified field.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
returns - Formatted value of the specified field.

Returns the label of the specified field. Query only; asynchronous command.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
string - Label of the specified field.

Returns the state of the specified field.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
enum - Returns the field state. The possible values are OK, EPRMble, ESFD, EESD, ECRC, ELEN, and INComplete.

Returns the CRC for the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - CRC for the specified frame.

Returns the value of the specified field.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
integer - The value of the specified field.

Returns the data for the specified frame in comma-separated values.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
string - Comma-separated values for the specified frame, including the count and values.

Returns the destination symbol for the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
string - Destination symbol for the specified frame.

Returns the formatted CRC for the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frnull-Index of the frame.
returns

Returns the formatted data for the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
string - Formatted data for the specified frame.

Returns the formatted destination address of the specified frame. This is a query-only asynchronous command.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
returns - Formatted destination address of the specified frame.

Returns the source address for the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - Source address for the specified frame.

Returns the frame of the source symbol.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
returns - Returns the frame of the source symbol.

Returns the state of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
enum - The frame state.

Returns the stop time of the specified frame. Query only; it is an asynchronous command.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (s) - Stop time of the specified frame.

Returns the type length for the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - Type length of the specified frame.

Returns the type of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
enum - Returns the frame type.

Sets the source channel to which the D- data line is connected.

NameTypeValuesDescription
DMinusSourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8Source channel to which the D- data line is connected.

Sets the source channel to which the D+ data line is connected.

NameTypeValuesDescription
DPlusSourceenumC1 | C2 | C3 | C4 | C5 | C6 | C7 | C8 | D0 | D1 | D2 | D3 | D4 | D5 | D6 | D7 | D8 | D9 | D10 | D11 | D12 | D13 | D14 | D15 | M1 | M2 | M3 | M4 | M5 | M6 | M7 | M8 | R1 | R2 | R3 | R4 | R5 | R6 | R7 | R8

Selects the low speed or full speed variant of the USB 1.1 protocol.

NameTypeValuesDescription
ModeenumUSBLow | USBFull

Sets the vertical position of the USB 1.1 signal.

NameTypeValuesDescription
Positioninteger1 - 100Vertical position of the USB 1.1 signal.

Sets the vertical scale of the USB 1.1 signal.

NameTypeValuesDescription
Scaleinteger1 - 5Vertical scale of the USB 1.1 signal.

Enables the filter for all available frames and error types.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Disables the filter for all available frames and error types.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Inverts the current state of the frame and error types for the selected serial bus USB filter. Frames and error types that were enabled become disabled, and vice versa.

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe. The operator is set with `SBUS<sb>:USBone:FILTer:FRAMe<fr>:FLD<fl>:DOPerator`.

NameTypeValuesDescription
frameinteger1 - 4Index of the frame.
fieldinteger-Index of the field.
datanull-End value of the data pattern range.

Enables or disables the specific frame to be filtered on.

NameTypeValuesDescription
enableenumOFF | ONEnables or disables the selected frame for filtering.

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe. The operator is set with `SBUS<sb>:USBone:FILTer:FRAMe<fr>:FLD<fl>:DOPerator`.

NameTypeValuesDescription
data_maxnull-End value of the data pattern range.

Enables or disables the specific frame to be filtered on.

NameTypeValuesDescription
frame--Frame index.
enablerenumOFF | ONEnables or disables the frame filter state.

Specifies the data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
Framelist of comma-separated values-Frame index and field data used by the filter.
Fieldlist of comma-separated values-Field index and field data used by the filter.
Datalist of comma-separated values-Data pattern values.
list of comma-separated values - Returns the current data pattern.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
operatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator used for the data pattern in the selected field of the selected frame.

Defines the error type to be filtered on.

NameTypeValuesDescription
ErrorNamestring-Error type to be filtered on.
Enablerstring-Enables or disables the filter for the selected error type.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Data_OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator for the data pattern.

Sets the end value of an index range if the operator is set to INRange. The command applies to a specific serial bus, frame, and field.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
imaxinteger1 - 65535End value of an index range.

Specifies the index, or sets the start value of an index range. It is used when the operator is set to INRange.

NameTypeValuesDescription
index_mininteger1 - 65535Start value of an index range or the index value to specify.

Sets the end value of an index range if the operator is set to INRange. The operator is set with `SBUS<sb>:USBone:FILTer:IOPerator`.

NameTypeValuesDescription
frame--
field--
data--

Specifies the index, or sets the start value of an index range.

NameTypeValuesDescription
Frame--
Field--
Data--

Sets the bit state of a field that only consists of one bit.

NameTypeValuesDescription
Frame--
Field--
BitenumONE | ZERO

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
Frame--
Field--
EnablerenumCondEnabler

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
CondEnablerenumON | OFFEnables or disables the checking condition for the selected field of the selected frame.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
operatorenumEQUal | INRange | RANGeOperator for the index in the selected field of the selected frame.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
Frameinteger1 - 4Index of the frame.
Fieldinteger-Index of the field.
OperatorenumEQUal | INRange | RANGeOperator for the index in the selected field.

Queries the software trigger time in seconds. If no software trigger event has occurred, the command returns Invalid.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
real (s) - Software trigger time in seconds. Returns Invalid if no software trigger event has occurred.

Queries the software trigger index. If no software trigger event has occurred, the command returns Invalid.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
integer - Software trigger index; returns Invalid if no software trigger event has occurred.

Returns the number of decoded frames for the selected serial bus.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
integer - Total number of decoded frames.

Returns the device address value of the indicated frame, as shown in the decoded cells and decode results table.

integer - Device address value of the indicated frame.

Returns the bit rate of the specified serial bus frame. This is a query-only asynchronous command.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (bps) - Returns the bit rate of the specified frame.

Returns a comma-separated list of values for the specified frame. The first value is the number of data bytes in the frame, followed by that many data bytes.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
block - A comma-separated list of values where the first value is the number of data bytes in the frame and the remaining values are the data bytes.
SBUS:USB:FRAMe2:DATA?

Returns the number of fields in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - Number of fields in the specified frame.

Returns the start time of the specified USB frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (s) - Start time of the specified frame.

Returns the overall state of the specified USB serial bus frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
enum - Overall state of the frame: OK means valid; PID indicates a packet identifier error; CRC indicates cyclic redundancy check failed; LENGth indicates wrong frame length; BTST indicates a bit stuffing error; INComplete indicates the frame is not completely contained in the acquisition.

Returns the formatted value of the specified field for the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
returns - Formatted value of the specified field for the selected frame.

Returns the name (label) of the specified field in the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
string - The field label of the specified field in the selected frame.

Returns the end time of the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (s) - End time of the specified frame.

Returns the overall state of the specified field for the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
enum - Overall state of the specified field.

Returns the data value of the specified field.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
integer - Data value of the specified field.

Returns the cyclic redundancy check (CRC) checksum of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - The CRC checksum of the selected frame.

Returns the endpoint for the selected serial bus and frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
returns

Returns the formatted CRC value for the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
string - Formatted CRC value for the selected frame.

Returns the frame number for the selected serial bus and frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
integer - Frame number for the selected serial bus and frame.

Returns the packet identifier for the selected serial bus and frame.

integer - Packet identifier for the selected serial bus and frame.

Returns the name of the packet identifier for the selected serial bus and frame.

returns - The packet identifier name for the selected serial bus and frame.

Selects the differential signal source to which the data line is connected.

NameTypeValuesDescription
Sourceenum-Differential signal source to which the data line is connected.

Returns the frame type of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
enum - Frame type of the selected frame.

Sets a hysteresis value for the differential signal on the data line.

NameTypeValuesDescription
hysteresisreal-Hysteresis value for the differential signal on the data line.

Sets the upper threshold value for the differential signal on the data line.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
Upper Thres--

Sets the vertical scale of the USB 2.0 signal.

NameTypeValuesDescription
Scaleinteger1 - 5Vertical scale setting for the USB 2.0 signal.

Enables the filter for all available frames and error types.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Disables the filter for all available frames and error types.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Inverts the current state of the frame and error types for the selected serial bus filter, so enabled items become disabled and vice versa.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Presets the state of the selected frames and error types.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe. The operator is set with `SBUS<sb>:USBHspeed:FILTer:FRAMe<fr>:FLD<fl>:DOPerator`.

NameTypeValuesDescription
frame--Frame index.
field--Field index.
datanull-Data pattern value.

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe. The operator is set with the corresponding DOPerator command.

NameTypeValuesDescription
datablock-List of comma-separated values.
block - Returns the end value of a data pattern range.

Specifies the data pattern, or sets the start value of a data pattern range.

NameTypeValuesDescription
Frame--
Field--
Data--

Specifies the data pattern, or sets the start value of a data pattern range.

string - Returns the data pattern or start value of the data pattern range.

Sets the end value of a data pattern range if the operator is set to INRange or OORANGe. You can set the operator with the corresponding DOPerator command.

NameTypeValuesDescription
data_maxstring-List of comma-separated values.

Enables or disables the specific frame to be filtered on.

NameTypeValuesDescription
enableenumOFF | ONEnables or disables the specific frame filter.

Enables or disables the specific frame to be filtered on.

NameTypeValuesDescription
enablerenumOFF | ONEnables or disables the specific frame to be filtered on.
enum - Returns the filter enable state for the specified frame.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
Frame--
Field--
OperatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORange

Returns the operator for the data pattern in the selected field of the selected frame.

enum - Returns the data operator used for the selected field of the selected frame.

Defines whether the selected USB high-speed filter error type is enabled. The command applies to error index <n> in the selected serial bus <sb>.

NameTypeValuesDescription
EnableenumOFF | ONEnables or disables the error type to be filtered on.

Sets the operator for the data pattern in the selected field of the selected frame.

NameTypeValuesDescription
frinteger-Index of the frame.
flinteger-Index of the field.
data_operatorenumEQUal | NEQual | LTHan | LETHan | GTHan | GETHan | INRange | OORangeOperator for the data pattern.

Sets the end value of an index range if the operator is set to INRange.

NameTypeValuesDescription
datainteger-End value of an index range.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
index_operatorenumEQUal | INRange | RANGeOperator for the index.

Sets the start value of an index range for the selected USB high-speed filter field. The command is asynchronous.

NameTypeValuesDescription
index_mininteger1 - 65535Start value of the index range.

Returns the index minimum for the selected USB high-speed filter field. The query form is documented as returning <Data>.

integer - The index minimum value for the selected frame and field, returned as <Data>.

Sets the operator for the index in the selected field of the selected frame.

NameTypeValuesDescription
index_operatorenumEQUal | INRange | RANGeOperator used for the index in the selected field of the selected frame.

Sets the bit state of a field that only consists of one bit. The query returns the bit state of the selected field bit.

NameTypeValuesDescription
Frame--Frame index.
Field--Field index.
Bit--Bit index.
returns - Returns the bit state of the selected field bit.

Queries the bit state of a one-bit field in the selected USB high-speed serial bus frame. The command is part of the USBHspeed filter setup and uses the selected bus suffix.

NameTypeValuesDescription
bitinteger-Bit index to query.
enum - Returns the bit state of the specified field.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
Frameinteger-Index of the frame.
Fieldinteger-Index of the field.
EnablerenumOFF | ONEnables or disables the checking condition for the selected field.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
enablerenumOFF | ONCondition enabler for the selected field.
enum - Returns the condition enabler state.

Sets the bit state of a field that only consists of one bit. The command also has a query form that returns the bit value.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
BitStateenumONE | ZEROBit state to set.
enum - Returns the bit state of the selected field bit.

Enables or disables the checking condition for the selected field of the selected frame.

NameTypeValuesDescription
CondEnablerenumOFF | ONEnables or disables the checking condition for the selected field of the selected frame.

Queries the software trigger time in seconds. If no software trigger event has occurred, the command returns Invalid.

real (s) - Returns the software trigger time in seconds, or Invalid if no software trigger event has occurred.

Queries the software trigger index. If no software trigger event has occurred, the command returns Invalid.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
integer - The software trigger index. If no software trigger event has occurred, returns Invalid.

Returns the number of decoded frames for the selected serial bus.

integer - Total number of decoded frames.

Returns the device address value of the indicated frame, as shown in the decoded cells and decode results table.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - Address value of the indicated frame.

Returns the number of fields in the specified frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - Total number of decoded frames.

Returns the calculated bit rate of the specified USBH frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
real (bps) - Calculated bit rate of the specified frame.

Returns a comma-separated list of values for the specified USBH frame. The first returned value is the number of data bytes in the frame, followed by those data bytes.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - First value is the number of data bytes to follow, with the remaining returned values being the data bytes themselves.
SBUS:USBH:FRAMe3:DATA?
5,24,156,75,62,230

Returns the overall state of the specified USB high-speed frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
enum - Overall state of the frame.

Returns the formatted value of the specified field for the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
string - The formatted value of the specified field.

Returns the name (label) of the specified field in the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
fr--Index of the frame.
fl--Index of the field.
string - The name (label) of the specified field in the selected frame.

Returns the overall state of the specified field for the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
enum - Overall state of the specified field.

Returns the cyclic redundancy check (CRC) checksum of the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
string - The CRC checksum of the selected frame.

Returns the data value of the specified field.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
flinteger-Index of the field.
integer - Data value of the specified field.

Returns the endpoint for the selected serial bus and frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - The endpoint for the selected serial bus and frame.

Returns the formatted address for the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
returns - Formatted address for the selected frame.

Returns the formatted CRC value for the selected frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
returns - Returns the formatted CRC value for the selected frame.

Returns the frame number for the selected serial bus and frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - Frame number for the selected serial bus and frame.

Returns the packet identifier for the selected serial bus and frame.

integer - Packet identifier for the selected serial bus and frame.

Returns the name of the packet identifier for the selected serial bus and frame.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
string - Formatted packet identifier name.

Returns the content of the "SPLIT packet transaction Endpoint Type" field of the selected frame. This 2-bit field tells the USB hub which split-transaction-state machine to use.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - Returns the Split ET field value for the selected frame; 0=Control, 1=Isochronous, 2=Bulk, 3=Interrupt.

Returns the content of the "SPLIT packet transaction E/U" field (end or unused) of the selected frame. This single-bit field, combined with the S field, describes the payload continuation encoding used for start-split packets; for complete-split packets, it is unused.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - Content of the SPLIT packet transaction E/U field for the selected frame.

Returns the content of the "SPLIT packet transaction Port" field of the selected frame. This field is a 7-bit word that contains the port number of the target hub for which the split transaction is destined.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - The SPLIT packet transaction Port field of the selected frame.

Returns the content of the "SPLIT packet transaction S" field of the selected frame. This single-bit field indicates the speed for the interrupt or control transaction announced by the split packet: 0 = Full speed, 1 = Low speed.

NameTypeValuesDescription
sbinteger1 - 4Index of the serial bus.
frinteger-Index of the frame.
integer - Content of the "SPLIT packet transaction S" field of the selected frame.
SBUS<sb>:USBHspeed:FRAMe<fr>:SPS?

Returns the content of the "SPLIT packet transaction Start / Complete" field of the selected frame. This single-bit field indicates whether the SPLIT token is a start-split transaction or a complete-split transaction.

integer - Returns the Split SC field as a numeric value representing the selected frame's SPLIT packet transaction Start / Complete status.

Deprecated legacy command listed under decode-results replacements. The manual shows it as a legacy command replaced by the corresponding CAN frame decode query.

Deprecated legacy command for Manchester decode result 1. It is replaced by :SBUS<sb>:MANCh:FRAMe<fr>:CONE? since FW 2.9.

returns

Legacy/deprecated query command for NRZ clocked decode results. The manual indicates it is replaced by :SBUS<sb>:NRZC:FRAMe<fr>:CONE?.

Legacy NRZ unclocked decode-result query. The manual marks it as a deprecated command and shows it replaced by the corresponding :CONE? query since FW 2.9.

null - Deprecated query for a decode result; the page slice does not document the returned data format or value type.

Sets the bus arbitration increment base value for the selected serial bus. The command is part of the SBUS<sb>:ARINc subsystem.

NameTypeValuesDescription
valuereal-Base value for bus arbitration increment.

Clears the bus arbitration input filter. This command is part of the SBUS<n>:ARINc:FILTer subsystem.

Enables or disables the specified ARINC filter error entry for the selected bus.

NameTypeValuesDescription
ninteger0 - MAXimumError filter index.

Enables or disables the selected field filter for the specified frame in the SBUS ARINC filter configuration. The command belongs to the frame/field filter branch of the serial bus trigger setup.

NameTypeValuesDescription
sbinteger-Subsequent SBUS index.
frinteger-Frame index.
flinteger-Field index.
stateenumON | OFFTurns the field filter enable state on or off.

Enables or disables the front-end filter for the selected serial bus interface.

NameTypeValuesDescription
stateenumON | OFFTurns the front-end filter on or off.

Sets the maximum gap bits for the serial bus archive. The command is part of the SBUS archive configuration hierarchy.

NameTypeValuesDescription
bitsinteger0 - 255Maximum gap in bits.

Creates a new list of I2C arbitration items for the specified bus. This command has no parameters.

Sets the scale factor for the acquired serial-bus signal. This command belongs to the serial bus acquisition settings.

Sets the source used for serial bus arming on the selected bus. The command belongs to the SBUS arming subsystem.

Returns the switch time for the selected arbitration item. The value is reported for the currently selected arbitration word or list element.

real (s) - The switch time of the selected arbitration item.

Sets the high threshold used for serial bus ARINC triggering.

NameTypeValuesDescription
thresholdreal-High threshold value for ARINC trigger detection.

Sets the lower threshold for the selected ARINC bus input. The command is part of the SBUS<sb>:ARINc threshold settings group.

NameTypeValuesDescription
thresholdreal-Lower threshold value.

Returns the number of words in the current serial bus input signal.

integer - The number of words in the current serial bus input signal.

Returns the data field of the selected serial bus word in the arithmetic decoder.

string - The data field of the selected serial bus word.

Returns the pattern for the specified word in the arbitration field.

string - Returns the pattern for the specified word in the arbitration field.

Returns the SDI field of the selected serial bus word.

string - Returns the SDI field value for the specified word.

Returns the state of the selected word in the selected serial bus. This query is part of the SBUS ARInc WORD subsystem.

enum - Returns the current state of the selected word.

Sets the CAN bus bit rate.

NameTypeValuesDescription
bitratereal0 - 1000000Bit rate for the CAN bus.

Sets the CAN fast threshold hysteresis. The command belongs to the SBUS CAN fast threshold configuration group.

NameTypeValuesDescription
hysteresisreal-Hysteresis value for the CAN fast threshold.

Sets the CAN fast threshold. The command is part of the SBUS CAN FAST subsection.

NameTypeValuesDescription
thresholdreal-Threshold value to set.

Returns the frame count for the selected CAN bus channel.

integer - The frame count for the selected CAN bus channel.

Sets the CAN filter bit. The command is part of the CAN filter configuration under the SBUS subsystem.

Sets the maximum allowed CAN filter data value for the selected filter. This command is part of the CAN filter setup for bus trigger/searching.

Sets the minimum CAN filter value for the selected subordinate bus filter. The command belongs to the CAN filter configuration group.

Sets the DOPerator for a CAN filter field. The command belongs to the SBUS CAN filter frame field configuration subtree.

Enables or disables an extended CAN filter for the selected bus. The manual slice only lists the command in the index and does not provide syntax details, parameters, or return information.

Enables or disables the CAN filter field. The command belongs to the CAN filter configuration tree.

NameTypeValuesDescription
stateenumON | OFFTurns the filter field on or off.

Enables or disables the CAN frame filter field. The command belongs to the CAN frame filter configuration under the selected serial bus.

NameTypeValuesDescription
stateenumENABled | DISabledTurns the filter field on or off.

Enables or disables the selected CAN frame field. The field is addressed by the SBUS, frame, and field indices in the command path.

NameTypeValuesDescription
stateenumON | OFFTurns the field enable state on or off.

Sets the minimum value of the selected CAN frame field. The command belongs to the CAN frame field configuration group.

NameTypeValuesDescription
sbinteger1 - 16Bus number.
frinteger1 - 64Frame number.
flinteger1 - 64Field number.
valuereal-Minimum field value.

Sets the logical operator used to combine field conditions in the CAN frame filter. The command is part of the SBUS CAN frame field filter subsystem.

NameTypeValuesDescription
operatorenum-Logical operator for the field filter conditions.

Enables or disables the CAN filter frame enable state for the specified subbus channel. This command is part of the SBUS<CAN> filter configuration group.

NameTypeValuesDescription
sbinteger-Subbus index.

Sets the maximum input amplitude for the CAN filter. This command belongs to the SBUS CAN filter configuration subtree.

Sets the minimum input filter value for CAN signal bus filtering. This command is part of the SBUS<sb>:CAN:FILTer subsystem.

NameTypeValuesDescription
valuereal-Minimum filter value.

Inverts the CAN filter logic. Use this command to enable or disable inversion of the configured filter condition.

NameTypeValuesDescription
stateenumON | OFFTurns filter inversion on or off.

Sets the logical operator used for CAN filter evaluation.

NameTypeValuesDescription
operatorenum-Logical operator used for the filter.

Returns the acknowledgement state for the selected CAN frame.

enum - Indicates whether the frame acknowledgement is set or not set.

Returns the CAN frame acknowledgment value for the specified frame.

integer - The acknowledgment value of the specified CAN frame.

Returns the current condition state of the CAN frame checksum state. This is a query command.

enum - Returns the current checksum state of the CAN frame.

Returns the CAN frame checksum value. The value is associated with frame <fr> in the selected bus <sb>.

integer - Returns the checksum value for the specified CAN frame.

Returns the bit rate used for the CAN data frame. The frame is identified by the optional frame index <fr> under the selected bus <sb>.

real (bit/s) - The CAN data bit rate for the specified frame.

Returns the BRS state for the specified CAN frame's FD data field. The command is a query on frame field data BRS status.

enum - Returns the BRS state.

Returns the ESI state of the selected CAN FD data field in the specified frame.

enum - Returns the ESI state for the selected CAN FD data field.

Returns the value of the selected field within a CAN frame. The field is identified by the frame and field indices in the command path.

string - The value of the selected field.

Returns the identifier state for the selected CAN frame.

enum - Indicates the current state of the frame identifier setting.

Returns the nominal bit rate of the CAN frame. The command belongs to the CAN frame configuration subtree.

real (bit/s) - The nominal bit rate of the CAN frame.

Returns the CAN signal binding count for the selected frame.

integer - The number of signal bindings in the selected CAN frame.

Returns the start state of the specified CAN frame. The frame is addressed by the <fr> node.

enum - Indicates whether the frame is started or not.

Returns the symbol associated with the selected CAN frame. The frame is specified by the `<fr>` node.

string - The symbol associated with the selected CAN frame.

Returns the CRC status for the specified CAN extra data object. The command is a query for the FCRC setting under SBUS CAN frame extra data.

enum - Returns the FCRC state for the specified extra data object.

Returns the CRC for the selected XDATA CAN frame subfield. The command is a query on the XDATa<o> node within a CAN frame.

integer - The CRC value for the selected XDATA CAN frame subfield.

Returns the virtual CAN ID value for the specified extended data frame object.

real - The virtual CAN ID value of the selected extended data frame object.

Creates a new CAN list in the selected serial bus. The command belongs to the SBUS<sb>:CAN subsystem.

Sets the CAN scale. The command is listed in the CAN bus subsystem command index.

Sets the hysteresis for the CAN SIC signal. The command is listed under CAN signal-integrity controls alongside the corresponding threshold command.

Sets the frequency of the waveform.

NameTypeValuesDescription
Frequencyreal0.001 - 50E+6 or lowerFrequency of the waveform.

Sets the pulse width, the pulse duration of the generated pulse waveform.

NameTypeValuesDescription
PulseWidthreal1.65E-08 - 90000Pulse width of the generated pulse waveform.

Sets the symmetry of a ramp waveform, expressed as the percentage of time the waveform is rising. By changing the symmetry, you can create triangular waveforms.

NameTypeValuesDescription
RampSymmetryreal0 - 100Percentage of time the ramp waveform is rising.

Selects the type of waveform to be generated.

NameTypeValuesDescription
FunctionTypeenumSINusoid | SQUare | RAMP | DC | PULSe | SINC | CARDiac | GAUSs | LORNtz | HAVer | EXPRise | EXPFall | ARBitraryType of waveform to be generated.

Selects the user load, i.e. the load of the DUT at its connection.

NameTypeValuesDescription
loadreal0.01 - 99.99User load value.

Sets the period of the pulse waveform when WGENerator<wg>:FUNCtion[:SELect] is set to PULSe.

NameTypeValuesDescription
Periodreal1E-08 - 1000Period of the pulse waveform.

Presets the waveform generator to a default setup. The default includes function type Sine, frequency 1 MHz, and amplitude 1 Vpp.

Selects the operation mode of the waveform generator.

NameTypeValuesDescription
OperationMode--

Sets the DC signal level. This command is only valid when WGENerator<wg>:FUNCtion[:SELect] is set to DC.

NameTypeValuesDescription
DCLevelreal-5 - 5Level for the DC signal.

Sets the high signal level of the output waveform.

NameTypeValuesDescription
highreal-5.99 - 6High signal level of the output waveform.

Inverts the waveform at the offset level.

NameTypeValuesDescription
InversionenumOFF | ONSpecifies whether waveform inversion is off or on.

Sets the low signal level of the output waveform.

NameTypeValuesDescription
Lowreal-6 - 5.99Low signal level of the output waveform.

Sets the vertical offset of the generated waveform.

NameTypeValuesDescription
offsetreal-5.995 - 5.995Vertical offset of the generated waveform.

Sets the amplitude of the waveform.

NameTypeValuesDescription
amplitudereal? - ?Amplitude of the generated waveform.

Enables the function generator.

NameTypeValuesDescription
stateenumOFF | ONEnables or disables the function generator.

Sets the duty cycle for a square waveform. The duty cycle specifies the percentage of the period during which the waveform is active, i.e. the signal state is high.

NameTypeValuesDescription
SquareDutyCyclereal10 - 90Duty cycle for the square waveform.

Sets the modulation depth, expressed as the percentage of the amplitude range used for AM modulation.

NameTypeValuesDescription
Depthreal0 - 100Modulation depth percentage.

Sets the frequency of the modulation waveform for AM frequency modulation.

NameTypeValuesDescription
Frequencyreal0.001 - 1000000Frequency of the modulation waveform.

Sets the symmetry for the AM ramp modulation waveform, i.e. the percentage of time the waveform is rising.

NameTypeValuesDescription
RampSymmetryreal0 - 100Percentage of time the AM ramp waveform is rising.

Selects the type of the modulating signal for AM modulation.

NameTypeValuesDescription
SignalTypeenumSINusoid | SQUare | RAMPType of the modulating signal used for AM modulation.

Sets the duty cycle for a square waveform. The duty cycle expresses what percentage fraction of the period the waveform is active (signal state high).

NameTypeValuesDescription
SquareDutyCyclereal10 - 90Duty cycle value for the square waveform.

Sets the frequency deviation, the maximum difference between the FM modulated signal and the carrier signal.

NameTypeValuesDescription
Deviationreal? - ?Frequency deviation value.

Sets the frequency of the modulating waveform for FM modulation.

NameTypeValuesDescription
Frequencyreal0.001 - 1000000Frequency of the modulating waveform.

Sets the symmetry for the FM ramp modulation waveform, the percentage of time that the waveform is rising.

NameTypeValuesDescription
RampSymmetryreal0 - 100Sets the symmetry for the FM ramp modulation waveform.

Selects the type of the modulating signal for FM modulation.

NameTypeValuesDescription
signal typeenum-Type of the modulating signal for FM modulation.

Sets the frequency of the first signal in an FSK modulated signal. The <wg> suffix is irrelevant and must be omitted.

NameTypeValuesDescription
frequency1real0.001 - 50E+6Frequency of the first signal.

Sets the frequency of the second signal in an FSK modulated signal. The <wg> suffix is irrelevant and must be omitted.

NameTypeValuesDescription
frequency2real0.001 - 50E+6Frequency of the second signal.

Sets the frequency at which the signal switches between WGENerator<wg>:MODulation:FSK:FONE and WGENerator<wg>:MODulation:FSK:FTWO.

NameTypeValuesDescription
ratereal0.001 - 1000000Frequency at which the FSK signal switches between the first and second tones.

Sets the DC noise level when the waveform generator function is set to DC.

NameTypeValuesDescription
leveldcreal0.001 - 1000000DC noise level.

Sets the level of the noise in percentage of the set "Amplitude" output of the signal.

NameTypeValuesDescription
LevelPctreal0 - 100Noise level as a percentage of the set Amplitude output.

Sets the duty cycle for a square waveform. The duty cycle expresses what percentage fraction of the period the waveform is active, i.e. the signal state is high.

NameTypeValuesDescription
SquareDutyCyclereal10 - 90Duty cycle of the square waveform.

Sets the modulation depth, the percentage of the amplitude range used for PWM modulation.

NameTypeValuesDescription
Depthreal0 - 100Percentage of the amplitude range used for PWM modulation.

Sets the frequency of the modulating waveform for PWM modulation.

NameTypeValuesDescription
frequencyreal0.001 - 1000000Frequency of the modulating waveform.

Sets the symmetry for the PWM ramp modulation waveform, the percentage of time that the waveform is rising.

NameTypeValuesDescription
RampSymmetryreal0 - 100Percentage of time that the waveform is rising.

Sets the file path and file name for an arbitrary waveform, if WGENerator<wg>:ARBGen[:SOURce] is set to ARBitrary.

NameTypeValuesDescription
folder--

Selects the modulation type, which defines how the carrier signal is modified.

NameTypeValuesDescription
modulation typeenumNONE | AM | FM | PWM | ASK | FSK

Sets the duration for which the signal of the arbitrary generator will be output after the trigger event.

NameTypeValuesDescription
RunModeenumSINGle | REPetitiveSelects whether the arbitrary generator runs once or repeatedly after the trigger event.

Sets the sample rate for the arbitrary waveform.

NameTypeValuesDescription
SampleRatereal1 - 312500000Sample rate for the arbitrary waveform.

Selects the source of the arbitrary waveform.

NameTypeValuesDescription
WaveformSourceenumARBitrarySource of the arbitrary waveform.

Enables or disables the waveform generator sweep function.

NameTypeValuesDescription
sweepenumOFF | ONTurns sweeping off or on.

Sets the stop frequency of the sweep signal.

NameTypeValuesDescription
stopFrequencyreal0.001 - 50E+6Stop frequency of the sweep signal.

Sets the start frequency of the sweep signal.

NameTypeValuesDescription
StartFrequencyreal0.001 - 50E+6Start frequency of the sweep signal.

Sets the type of the sweep, specifying whether the frequency changes linearly or logarithmically.

NameTypeValuesDescription
TypeenumLIN | LOGSweep type.

Reads out the CONDition part of the WGENerator overload questionable-status bit. It is listed in the status reporting section for the questionable register.

integer - Returns the condition value of the status bit.

Returns the condition part of the WGENerator temperature questionable status register.

integer - Returns the condition part of the WGENerator<wg>:TEMPerature:STATus questionable status register.

Returns the contents of the EVENt part of the status register to check if an event has occurred since the last reading. Reading an EVENt register deletes its contents.

string - Bit dependent, comma-separated values; see Source values for STATus:QUEStionable:...:[:EVENt] and STATus:QUEStionable:...:[:ENABLe] for a list of the return values.

Returns the contents of the EVENT part of the status register for the temperature status of the waveform generator. Reading the EVENT register deletes its contents.

bit dependent, comma-separated values - Returns the contents of the EVENT part of the status register; bit-dependent, comma-separated values. See source values for status-questionable event registers for the list of return values.

Sets the ENABle part of the questioned status register for the waveform generator overload status. It controls whether true conditions in the corresponding EVENt part are reported as a positive transition in the summary bit.

NameTypeValuesDescription
Value--Bit-dependent, comma-separated values that enable reporting of true events.

Sets the ENABle part that allows true conditions in the EVENt part to be reported for the summary bit in the status byte. If a bit in the ENABle part is 1 and the corresponding EVENt bit is true, a positive transition occurs in the summary bit.

NameTypeValuesDescription
Value--Bit-dependent, comma-separated values.

Sets the negative transition filter for the generator overload status register. If a bit is set, a transition from 1 to 0 in the condition part causes an entry to be made in the corresponding bit of the EVENT part of the register.

NameTypeValuesDescription
Value--Source dependent, comma-separated values.

Sets the positive transition filter for the WGENerator overload status register. A set bit causes a 0-to-1 transition in the condition part to be entered in the corresponding EVENt bit.

NameTypeValuesDescription
valuestring-Source dependent, comma-separated values.

Sets the negative transition filter for the temperature status register. If a bit is set, a transition from 1 to 0 in the condition part causes an entry to be made in the corresponding bit of the event part of the register.

NameTypeValuesDescription
Value--Source dependent, comma-separated values

Sets the positive transition filter for the WGENerator temperature status register. If a bit is set, a transition from 0 to 1 in the condition part causes an entry in the corresponding bit of the EVENT part of the register.

NameTypeValuesDescription
Valuenumeric-Source dependent, comma-separated values.

Obsolete waveform generator modulation command. The manual lists it as a deprecated legacy command with no replacement, obsolete since FW 2.8.

Returns whether acquisition is available on the instrument. The query reports the current acquisition availability state.

enum - Returns the acquisition availability state.

Returns the average acquisition state.

enum - Indicates whether acquisition averaging is enabled or disabled.
:ACQuire:COUNtset / query

Sets the acquisition count. The instrument uses this value for the number of acquisitions in the current acquisition mode.

NameTypeValuesDescription
countinteger1 - MAXimumNumber of acquisitions to average or accumulate, depending on the acquisition mode.

Returns the current acquisition state.

enum - Returns the current acquisition state.

Returns the acquisition resolution. This is a query command.

real - Returns the acquisition resolution.

Sets the current acquisition history record. This command belongs to the ACQuire:HISTory subsystem.

Returns the ISO date for the acquisition history time stamp.

string - The acquisition history time stamp as an ISO date string.

Starts playback of the acquisition history. The command is part of the ACQuire:HISTory subsystem.

Replays a record from acquisition history. This command is listed under ACQuire:HISTory along with related history playback and timestamp commands.

Starts acquisition history. The command is part of the ACQuire:HISTory subsystem.

Sets the time position of acquired history data. This command belongs to the ACQuire:HISTory subsystem.

NameTypeValuesDescription
time position acquisitionreal-Time position for history data acquisition.

Returns the absolute time stamp for the current history acquisition entry.

real (s) - The absolute time stamp of the current history acquisition entry.

Returns the history timestamp as a date string. The date format is determined by the :ACQuire:HISTory:ISODate setting.

string - Returns the history timestamp as a date string.

Returns the reference timestamp used for history records. The value is reported as a relative time reference.

real (s) - Reference timestamp for history records, expressed as a relative time reference.

Returns the relative timestamp of the current history acquisition record.

real (s) - Relative time of the current history acquisition record.

Enables or disables acquisition history mode. The optional state node is documented as part of the command mnemonic.

NameTypeValuesDescription
stateoptenumON | OFFTurns acquisition history mode on or off.

Turns acquisition interpolation on or off.

NameTypeValuesDescription
stateenumON | OFFEnables or disables acquisition interpolation.

Returns the current acquisition point-to-average rate.

real - The current acquisition point-to-average rate.

Returns the default acquisition point value. The command is listed as a query in the command index.

returns - Returns the default acquisition point value.

Returns the maximum number of acquisition points the instrument can use. This is a limit command in the ACQuire:POINts subsystem.

Sets the acquisition points mode. The command determines how the instrument interprets the acquisition points setting.

NameTypeValuesDescription
modeenum-Acquisition points mode.

Sets the number of acquired points. The command also appears in the manual as ACQuire:POINts[:VALue].

NameTypeValuesDescription
valueinteger1 - MAXimumNumber of acquired points.

Returns the amount of memory available for the acquisition system.

integer - The amount of acquisition memory available.

Returns the current acquisition resolution setting.

enum - Returns the current acquisition resolution setting.

Returns the real-time acquisition record length setting.

real - Returns the real-time acquisition record length.

Controls the roll-mode oscilloscope capture setting. The command is listed as a setter with no further syntax or parameter details on this page.

Returns the number of acquisition points used in roll mode.

integer - Returns the number of acquisition points used in roll mode.

Sets the maximum number of segmented acquisitions. This command belongs to the segmented acquisition subsystem.

Enables or disables segmented acquisition.

NameTypeValuesDescription
stateenumON | OFFTurns segmented acquisition on or off.

Returns the minimum available acquisition sample rate. This is a read-only command in the acquisition subsystem.

real - The minimum acquisition sample rate.

Sets the acquisition sample-rate mode. The command is part of the ACQuire:SRATe subsystem and selects how the sample rate is determined.

NameTypeValuesDescription
modeenum-Sample-rate mode.

Sets the acquisition sample rate. The value may be specified directly, or the command may use an optional value node.

NameTypeValuesDescription
valueoptreal-Acquisition sample rate value.

Returns the real acquisition sample rate. The query reports the current sample-rate value used for acquisition.

real (Hz) - The real acquisition sample rate.
:ACQuire:TYPEset / query

Sets the acquisition type.

NameTypeValuesDescription
typeenumNORMal | AVERages | PEAK | HRESolutionAcquisition mode.

:AUToscaleset / query

Performs an automatic setup of the instrument waveform display and settings.

Displays the cursor values on the screen. This command has no parameters documented in the provided slice.

Sets the center frequency of the FFT cursor. This command is part of the cursor FFT setup functions.

Sets the cursor function for the selected cursor pair. The manual slice lists the command in the cursor command index but does not provide additional syntax details here.

Sets the label for the specified horizontal cursor.

NameTypeValuesDescription
ninteger1 - 2Horizontal cursor number.
labelstring-Label text for the cursor.

Sets or queries the label for the selected cursor. The manual page lists this command but does not provide syntax details or parameter definitions in the supplied slice.

Moves the selected cursor to the left maximum peak. This command is one of the cursor maximum search commands.

Moves the cursor to the rightmost of the maximums. This command is part of the cursor maximum functions.

Sets the cursor percent excursion. The command is part of the cursor subsystem.

Selects the source for the specified cursor. The command belongs to the CURSor<cu> subsystem.

Enables or disables screen-based cursor measurement mode for the specified cursor set.

Sets the cursor state on or off for the specified cursor. The command is available for cursor group `<cu>`.

NameTypeValuesDescription
stateenumON | OFFTurns the cursor state on or off.

Sets the cursor style. The available styles are listed on the command's documentation page.

Turns cursor tracking on or off for the selected cursor. The optional state node is documented in the command list entry.

NameTypeValuesDescription
stateoptenumON | OFFEnables or disables cursor tracking.

Sets the source used for user-defined cursors.

NameTypeValuesDescription
sourceenum-Source for the user-defined cursors.

Sets the label for a vertical cursor. The command applies to the specified cursor and vertical cursor number.

NameTypeValuesDescription
cuinteger-Cursor identifier.
ninteger-Vertical cursor number.

Sets the X1 envelope cursor position. The command belongs to the cursor control subsystem.

Sets the X1 position of the cursor. The command is part of the CURSor subsystem for cursor positioning.

NameTypeValuesDescription
positionreal-X1 position value.

Sets the cursor X coupling. The command is part of the cursor configuration subsystem.

NameTypeValuesDescription
couplingenum-X coupling mode.

Returns whether the X delta cursors are inverted.

enum - Returns the inverse state of the X delta cursors.

Returns the XDelta cursor value. The optional VALue node is accepted in the query mnemonic.

real - The XDelta cursor value.

Sets the Y1 cursor position. This command is part of the cursor position controls for cursor <cu>.

NameTypeValuesDescription
positionreal-Y1 position value.

Sets the coupling for the Y direction of the selected cursor pair. The command is part of the cursor subsystem.

NameTypeValuesDescription
couplingenumAC | DCCoupling mode for Y cursor measurements.

Returns the Y delta cursor value. The optional :VALue node may be included in the query syntax.

real - The Y delta cursor value.

Returns whether the digital data is displayed with a header. This is a query command for the selected digital channel.

enum - Indicates the current header display state.

Sets the hysteresis of the specified digital channel or marker. The command belongs to the DIGital<m> subsystem.

NameTypeValuesDescription
hysteresisreal-Hysteresis value for the digital input.

Sets the label for a digital channel indicator. The command is part of the DIGital<m> node.

NameTypeValuesDescription
labelstring-Label text to assign to the digital channel.

Returns whether probing is enabled for the specified digital channel. This is the query form of the digital probe enable command.

enum - Returns the probe enable state.
:DIGital<m>:SIZEset / query

Sets the size of digital channels. The command is documented in the command list for digital channel settings.

:DIGital<m>:SKEWset / query

Sets the skew for the specified digital channel. The command is part of the DIGital<m> command group.

NameTypeValuesDescription
minteger1 - 16Digital channel number.

Selects whether the specified digital channel is displayed as a digital input channel. The command is indexed by channel number <m>.

NameTypeValuesDescription
stateenumON | OFFTurns the digital channel display state on or off.

Sets the threshold level for a digital input channel. The command is part of the DIGital<m> group.

:FORMat[:DATA]set / query

Selects the data format used for waveform transfers and other data output. The command belongs to the FORMat subsystem.

:GATE<g>:CURSorset / query

Sets the cursor position for gate <g> in the gate diagram. The command is part of the gate cursor controls.

Sets the gate group coupling for gate <g>.

NameTypeValuesDescription
ginteger-Gate group index.

Sets the relative start position for gate <g>. The command belongs to the gate controls and is paired with the corresponding relative stop setting.

:GATE<g>:SHOWset / query

Sets whether the gate is shown on the display.

Shows the gate in a Z diagram. The command is listed as a gate command for gate <g>.

:GENerator:SAVset / query

Saves the generator setup to nonvolatile memory. Use this command to store the current generator settings for later recall.

Returns the number of histogram counts for the specified histogram. The histogram is selected by the <m> index.

integer - The number of counts in the specified histogram.

Sets the mode for histogram m. The command is listed in the histogram command group, but this page slice does not show the syntax or allowed values.

Removes histogram data for the specified histogram object.

NameTypeValuesDescription
minteger-Histogram object number.

Selects the source used by histogram m. The manual page only lists the command in the command index and does not provide syntax or parameter details on this slice.

Enables or disables the histogram window. The command belongs to the HISTogram<m>:WINDow subsystem.

NameTypeValuesDescription
stateenumON | OFFTurns the histogram window on or off.

Sets the relative horizontal position of the histogram window. This command is part of the histogram window horizontal positioning controls.

NameTypeValuesDescription
positionreal-Relative horizontal position of the histogram window.

Sets the absolute vertical position of the histogram window. The command belongs to the histogram window vertical absolute settings.

NameTypeValuesDescription
positionreal-Absolute vertical position of the histogram window.

Sets the histogram window vertical mode. The slice only lists the command in the command index and does not provide syntax details or a value description.

Sets the relative vertical position of the histogram window. The command is listed in the command index; this slice does not provide syntax details, parameter ranges, or a return value.

Sets or queries the relative vertical span of the histogram window for the specified histogram. The command belongs to the histogram window vertical relative position/span group.

Sets or selects the active layout. The layout is identified by the <ly> index.

NameTypeValuesDescription
lyinteger-Layout index.

Returns the number of layouts currently defined in the selected layout subsystem.

integer - The number of layouts currently defined.

Returns the number of diagram objects in the specified layout. This is a query command.

integer - The number of diagram objects in the specified layout.

Enables or disables a diagram in the selected layout. The command is documented as an optional-node form of the diagram enable command.

Sets or queries the label for the specified layout diagram. The command is part of the LAYout subsystem and uses the layout index <ly>.

Returns the ID of the content item at the specified child position under a node in a layout. The command belongs to the layout node/children/content hierarchy.

string - The ID of the specified content item.

Returns the number of nodes in the specified layout.

integer - The number of nodes in the specified layout.

Sets the ratio for a layout node. The command belongs to the layout node hierarchy and is documented without additional parameter details on this page slice.

Sets the type of the specified layout node.

NameTypeValuesDescription
lyinteger-Layout index.
nointeger-Node number.
typeenum-Node type.

Sets the horizontal ratio for a result layout. The command is listed as part of the LAYout subsystem for result panes.

NameTypeValuesDescription
ninteger-Result layout index.
ratioreal-Horizontal ratio value.

Sets the vertical ratio for a layout result. The command is part of the LAYout tree and applies to a specific result index.

NameTypeValuesDescription
lyinteger-Layout index.
ninteger-Result index.

Sets the relative position for the specified layout. The command belongs to the LAYout subsystem and uses the <ly> layout index.

Returns the number of zoom windows in the specified layout.

integer - The number of zoom windows in the specified layout.

Returns the diagnostic state of the specified zoom layout. This is a query for a layout zoom node.

NameTypeValuesDescription
zointeger-Zoom node index.
enum - Returns whether the zoom is in diagnostic mode or not.

Selects the source used for the zoomed display. The command is part of the layout zoom subsystem.

Enables or disables the zoom display for the selected layout. This command is listed as the optional enable node of the layout zoom command.

NameTypeValuesDescription
stateenumON | OFFTurns zoom display on or off.

Enables or disables the specified layout. This command is also listed in the manual as part of the layout command tree under :LAYout<ly>.

NameTypeValuesDescription
stateenumON | OFFTurns the layout enable state on or off.

:METer:BANDwidthset / query

Sets or queries the bandwidth used for meter measurements.

Enables or disables the AC RMS function for the selected digital voltmeter meter. This command applies to meter <m>.

NameTypeValuesDescription
stateenumON | OFFTurns AC RMS on or off.

Enables or disables the DVMeter DC measurement for meter <m>.

NameTypeValuesDescription
stateenumON | OFFTurns the DC DVMeter measurement on or off.

Returns the RMS result of the specified DC meter. The meter is identified by the meter index <m>.

real - The RMS result of the specified DC meter.

Enables or disables the DVMeter measurement. The query returns the present enable state.

enum - Returns whether the DVMeter measurement is enabled or disabled.
:METer:MTIMeset / query

Sets the meter MTI time. This command is part of the METer subsystem.

Returns the count of added tests for the specified test set.

integer - The count of added tests for the specified test set.

Creates a diagram for the selected test. The command is part of the MTESt test subsystem for test list item <m>.

Opens an imported/exported test result file for the selected MTESt subsystem. The command is part of the MTESt<m>:IMEXport command group.

Configures what the meter test does when a violation occurs. This command controls the screenshot action for violation handling.

Sets the waveform-saved action for on-violation handling in the MTESt subsystem. This command is listed alongside the related :MTESt<m>:ONViolation:TRIGgerout command.

:MTESt<m>:REMoveset / query

Removes a test module. The command is part of the MTESt subtree for managing test configurations.

Returns the count of fast waveform results for the selected measurement.

integer - Returns the number of fast waveform results for the selected measurement.

Returns the count of positive waveform results for the selected math test measurement.

integer - Returns the number of positive waveform results.

Returns the number of waveforms in the measurement result count for the selected measurement context.

integer - The number of waveforms in the measurement result count.

Returns the frame rate of the selected MTEST result.

real - The frame rate of the selected MTEST result.

Queries the result of the selected MTEST result function. If the optional [:RESult] node is omitted, the basic result is returned.

string - Returns the selected MTEST result value.

Adds a segment to the selected MTEST waveform memory. The command is part of the MTESt<m>:SEGMent<n> subtree.

Returns the number of segments in the specified test mode. This is a query command.

integer - The number of segments in the specified test mode.

Returns the number of points in the specified segment point. This is a query command for a test memory segment point.

integer - The number of points in the specified segment point.

Sets the X coordinate of a point in a test segment.

NameTypeValuesDescription
xreal-X coordinate value for the point.

Sets the Y value of a point in a measurement segment. The command belongs to the MTES test segment/point hierarchy.

NameTypeValuesDescription
minteger1Measurement index.
ninteger1Segment index.
ointeger1Point index.

Removes a segment from the specified MTESt test. The command is part of the segment-management subsystem for MTESt<m>.

NameTypeValuesDescription
minteger-MTESt test index.
ninteger-Segment index to remove.

Returns whether the specified MTEST segment is valid.

enum - Indicates whether the specified segment is valid.
:MTESt<m>:SOURceset / query

Sets the measurement source for the selected test. The command belongs to the MTESt<m> subsystem.

Controls whether the selected MTESt memory trace is visible. This command is part of the MTESt<m> subsystem.

Returns the number of available harmonics for a three-phase measurement. The command is part of the P3PHase harmonics subsystem.

integer - The number of available harmonics.

Sets the upper harmonic display stop value for the selected phase. The command name appears with an optional ":VALue" node in the command list.

NameTypeValuesDescription
valueoptreal-Stop value for the harmonic display.

Enables or disables the waveform display for total power harmonics. This command is listed as an optional-node form of the total power waveform display setting.

NameTypeValuesDescription
stateoptenumENABleTurns the total power waveform display on or off.

Enables or disables the harmonics frequency function for the selected 3-phase measurement. The command belongs to the P3PHase<m>:HARMonics subsystem.

NameTypeValuesDescription
stateenumEN | DISTurns the harmonics frequency function on or off.

Returns the active power measurement for a harmonic power channel, optionally using the actual value form.

real - The active power measured for the specified harmonic power channel. If [:ACTual] is used, the actual value is returned.

Sets the user-defined power factor setting for harmonic measurements. This command is part of the P3Phase harmonic functions subsystem.

Returns the header information for the power harmonic data of the specified three-phase harmonic power channel.

string - Header information for the power harmonic data.

Returns the harmonic power data values for a specified phase and harmonic power trace. The command also supports the optional `:VALues` node in the syntax.

NameTypeValuesDescription
minteger0Phase index.
ninteger0Harmonic power trace index.
block - Harmonic power data values.

Sets the absolute lower level for the harmonic reference level. The command is part of the P3PHase harmonic reference-level configuration for the specified <rl> index.

NameTypeValuesDescription
levelreal-Absolute lower level value.

Sets the upper relative harmonic reference level for the selected phase measurement. The command is part of the P3 phase harmonics reference-level configuration.

NameTypeValuesDescription
minteger-Phase measurement index.
rlreal-Upper relative harmonic reference level.

Sets the harmonic source line selection for the specified three-phase measurement. The command is part of the P3PPhase harmonic source configuration.

NameTypeValuesDescription
minteger-Measurement index.
W2V2aenumW2V2aLine selection.

Sets the voltage source used for the three-phase harmonics wiring option W3VN3a.

NameTypeValuesDescription
sourceenumW3VN3aVoltage source selection for wiring option W3VN3a.

Returns the header for total power data. The command is listed as a query and provides the data header for the three-phase harmonics total power dataset.

string - Returns the total power data header.

Sets the vertical offset for the total power display. This command is part of the three-phase harmonics total power vertical scaling controls.

NameTypeValuesDescription
offsetreal-Vertical offset value.

Enables or disables display of the current peak measurement for a selected phase. The optional node ":ENABle" controls the display state.

NameTypeValuesDescription
ninteger-Measurement index.
stateoptenumENABleEnables the current peak display.

Controls whether the current RMS quality display item is enabled. The command applies to the specified phase and current channel.

NameTypeValuesDescription
ninteger-Current RMS display item index.
enableoptenumENABleEnables the display item.

Controls whether the power-factor display for a specific power quantity is enabled. This command is listed as an optional ":ENABle" node under the power display command.

NameTypeValuesDescription
ninteger-Power display item index.

Enables or disables display of the phase power measurement for a selected power entry. The command is listed as an optional enable node under the phase-power display subsystem.

NameTypeValuesDescription
ninteger-Power entry number.
enableoptenumENABleEnables the display item.

Turns the waveform display for a selected power-quality measurement on or off.

NameTypeValuesDescription
ninteger1 - nSelects the power measurement instance.
enableoptenumON | OFFEnables or disables the waveform display.

Enables or disables the display of the voltage peak quality measurement for the specified phase and index. The optional `ENABle` node controls whether the item is shown.

NameTypeValuesDescription
minteger-Phase selection index.
ninteger-Peak measurement index.
enableoptenumENABleOptional enable node.

Controls whether the displayed RMS voltage quality measurement for a given index is enabled. The command belongs to the P3PHase quality display group and the optional ENABle node may be included or omitted in the syntax listing.

NameTypeValuesDescription
minteger1 - 3Phase group index.
ninteger-Measurement index for the displayed RMS voltage entry.
stateoptenumENABleOptional enable node.

Sets or queries the quality frequency value for the selected three-phase measurement. The command is part of the P3PHase quality subsystem and the value is documented on the referenced manual page.

real - Returns the quality frequency value.

Sets the gate setting for the three-phase quality measurement. The slice lists the command in the quality subsystem command index but does not provide further syntax details or a return value.

Sets or queries the maximum current value used for the phasor display in the quality measurement group. The command belongs to the P3Phase<m>:QUALity:PHASor:CURRent subtree.

Enables or disables the phasor display for the selected phase system. The command is listed as optional-node form under the Quality:PHASor subsystem.

NameTypeValuesDescription
enableoptenumENABleOptional enable node in the command path.

Sets the vertical scale mode for the P3 phase quality power measurement display. The manual lists this command in the command index only, without further parameter details on the shown page.

Sets the absolute upper level reference for the P3 phase quality reference level. This command is part of the absolute reference-level settings under :P3PHase<m>:QUALity:REFLevel<rl>:ABSolute.

Sets the line mode for the reference level used in three-phase quality measurements.

NameTypeValuesDescription
rlinteger1 - 4Reference level index.
modeenumABSolute | RELativeLine mode for the reference level.

Sets the lower relative reference level for the selected phase quality reference level. The command belongs to the RELative reference-level subsystem for phase quality measurements.

Sets the reference level mode for the relative P3 phase quality measurement. The command belongs to the relative reference-level subtree.

NameTypeValuesDescription
modeenum-Relative reference level mode.

Sets the upper relative reference level for the selected phase quality measurement. The command is part of the relative reference-level group under P3PHase<m>:QUALity:REFLevel<rl>.

Returns the current peak result for the specified phase measurement. The optional :ACTual node is part of the query syntax.

NameTypeValuesDescription
minteger-Measurement index selector.
ninteger-Result index selector.
actualoptboolean-Selects the actual value form when included.
returns - Returns the current peak result for the selected measurement and result index.

Returns the current RMS value for the selected current result. The optional :ACTual suffix requests the actual measured value.

NameTypeValuesDescription
ninteger-Result index.
actualoptenumACTualRequests the actual measured value.
real - Current RMS value.

Returns the active power result for the specified phase and index. The optional ACTual node is included in the query mnemonic as shown.

NameTypeValuesDescription
minteger-Phase index.
ninteger-Result index.
real - Active power value for the specified phase and result index.

Returns the apparent power result for the selected phase measurement. The optional :ACTual suffix indicates the actual value form of the result.

real - The apparent power result for the selected phase, in the instrument's power units.

Returns the phase result for the selected power result number. The optional ACTual suffix indicates the actual measured value.

real - The phase result for the selected power result number, reported as the actual measured value when :ACTual is present.

Returns the reactive power result for the specified phase and channel. The optional actual suffix indicates the actual value.

NameTypeValuesDescription
minteger-
ninteger-
actualoptboolean-Selects the actual value.
real - Reactive power result for the specified phase and channel.

Returns the total current crest factor for the selected phase, optionally for the actual value. This is a query form of the quality result readout.

NameTypeValuesDescription
minteger1 - ?Phase index selector.
ninteger1 - ?Result index selector.
actualoptenumACTualSelects the actual value.
real - Returns the total current crest factor.

Returns the total current RMS result for the specified phase measurement. The optional actual suffix indicates the actual measured value form.

NameTypeValuesDescription
ninteger-Measurement index.
actualoptenumACTualSelects the actual value form.
real - Total current RMS result for the specified measurement.

Returns the apparent power result for the specified phase or total measurement. The optional :ACTual suffix indicates the actual value variant.

NameTypeValuesDescription
ninteger1 - 3Result index.
actualoptenumACTualSelects the actual value variant.
real - Apparent power result value.

Returns the total power factor for the selected phase. The optional ":ACTual" suffix returns the actual measured value.

NameTypeValuesDescription
ninteger-Phase index.
real - Total power factor value for the selected phase; with ":ACTual", the actual measured value is returned.

Returns the reactive power result for the selected phase, with the optional actual-value form available. This is a query command.

NameTypeValuesDescription
minteger-Phase index.
ninteger-Result index.
actualoptenumACTualRequests the actual value form.
real - Reactive power result value for the selected phase.

Returns the total voltage crest factor for the selected phase measurement. The optional :ACTual suffix is shown in the syntax line.

NameTypeValuesDescription
minteger1 - 4Measurement index.
ninteger-Result index.
real - Total voltage crest factor for the selected result.

Returns the peak voltage result for the specified phase-quality result channel. The optional :ACTual node is documented in the command mnemonic.

NameTypeValuesDescription
ninteger-Result channel index.
real - Returns the peak voltage result for the specified channel.

Returns the total RMS voltage result for the selected phase quality measurement. The optional ACTual suffix is shown in the syntax line.

NameTypeValuesDescription
minteger-Phase measurement index.
ninteger-Result index.
real - Returns the total RMS voltage result.

Returns the actual crest factor of the phase voltage result for the specified measurement. The optional :ACTual suffix is documented as part of the query syntax.

NameTypeValuesDescription
minteger-Phase measurement index.
ninteger-Result index.
real - Actual crest factor of the phase voltage result.

Returns the peak voltage result for the selected phase-quality measurement, with an optional actual-value suffix.

NameTypeValuesDescription
minteger-Measurement index.
ninteger-Result index.
actualoptenumACTualOptional actual-value suffix.
real - Peak voltage result value.

Returns the voltage RMS result for the specified phase-quality result number, with an optional actual-value suffix. The result is a query in the phase quality measurement results group.

real - Returns the voltage RMS result value for the selected result number. The optional :ACTual suffix indicates the actual result form.

Selects the current source for the P3PHase quality measurement. The command is part of the QUALity:SOURce subtree and is documented as a setter with no associated query in this slice.

Selects the current-source configuration for a three-phase quality measurement. This command is part of the P3PHase<m>:QUALity:SOURce:CURRent group and the listed setting is W3VN3a.

Selects the quality source input configuration for the P3PHase measurement. The command is listed without additional syntax or parameter details in this slice.

Selects the line configuration used as the source for phase 3 quality measurements. The command is part of the P3PhaSE<m>:QUALity:SOURce subsystem.

Selects the voltage source wiring configuration for the power quality measurement source. The command belongs to the P3Phaise quality source settings and corresponds to the W3VN3a wiring option.

Returns the values of the total power data. The command can also be used as :P3PHase<m>:QUALity:TOTal:POWer:DATA:HEADer? to return the header.

string - Returns the total power data values.

Sets the vertical offset for total power display. The listed command is part of the P3Phase quality total-power vertical settings.

NameTypeValuesDescription
offsetreal-Vertical offset value.
:P3PHase<m>:TYPEset / query

Sets the type for the selected P3PHase measurement. The command applies to the specified measurement instance <m>.

NameTypeValuesDescription
minteger-Measurement instance index.
typeenum-Type selection for the P3PHase measurement.

Enables or disables the P3P phase measurement for the specified measurement instance.

NameTypeValuesDescription
enableenum0 | 1 | OFF | ONTurns the P3P phase measurement on or off.

Sets the label for a specified bit of a parallel bus. The label is used to identify the bit in bus displays and listings.

NameTypeValuesDescription
pbinteger0Parallel bus number.
ninteger0Bit number within the bus.
labelstring-Bit label text.

Sets the skew of a specific bit in a parallel bus. The command belongs to the PBUS bit configuration group.

NameTypeValuesDescription
pbinteger-Parallel bus number.
ninteger-Bit number within the selected bus.

Sets or queries the state of bit n of the specified parallel bus. The command is part of the PBUS bit configuration group.

NameTypeValuesDescription
stateoptenumON | OFFEnables or disables the selected bit.
:PBUS<pb>:CLEarset / query

Clears the specified bus. The command is listed as a bus command with no additional syntax details on this page slice.

NameTypeValuesDescription
pbinteger-Bus index placeholder in the mnemonic.
:PBUS<pb>:CLOCkset / query

Sets the protocol bus clock settings for the selected bus. The command is listed in the PBUS command group, but this page slice does not provide syntax details or parameter ranges.

:PBUS<pb>:CLONset / query

Clones the settings of the selected bus to another bus. The command belongs to the PBUS bus setup group.

NameTypeValuesDescription
pbinteger-Bus index.

Sets the bus clock slope for the selected parallel bus. The command is listed without additional parameter details in this slice.

Sets the data format used for PBUS trace data. The command appears in the PBUS command list and has no additional syntax details on this page.

Returns the values in the bus data list. The optional VALues node is included in the query mnemonic but does not change the returned data described in this slice.

string - Returns the bus data values.

Returns the number of entries in the bus decode table.

integer - Number of entries in the bus decode table.

Returns the data table for the selected protocol bus decoder.

string - Returns the decoder data table.

Sets the vertical scale for the digital signals display within a bus. The command appears in the list of PBUS digital-signal display commands.

Sets the display visibility of digital bus information for the selected bus. The command is listed under the PBUS bus display subsystem.

:PBUS<pb>:SCALeset / query

Sets the vertical scale for the selected power bus. The command syntax also indicates an optional value form, :SCALe[:VALue], in the command list.

:PBUS<pb>:SKEWset / query

Sets the skew of the specified bus bit. This command belongs to the PBUS bit-level configuration group.

NameTypeValuesDescription
pbinteger-Bus number.
:PBUS<pb>:STATeset / query

Sets the state of the specified bus bit. The command can be used with or without the optional node name in brackets.

NameTypeValuesDescription
stateenum-State of the bus bit.

Sets or queries the technology associated with the specified bus. The slice lists the command in the PBUS subsystem command index but does not provide further syntax or data details.

Sets the threshold coupling for the selected bus. It is documented as part of the PBUS bus setup commands.

NameTypeValuesDescription
pbinteger0Bus index.

Sets the threshold for bus <pb> bit <n>. The command is documented in the PBUS subsystem command list.

Returns the probe part number for the specified probe channel.

string - The part number of the probe identified by <ch>.

Returns the probe production date for the specified channel.

string - The probe production date for the selected probe channel.

Returns the probe serial number for the specified channel. The command is part of the probe identification group.

string - The probe serial number.

Returns the software version of the probe identified by the channel number.

string - The probe software version.

Returns the common probe meter results for the selected probe channel.

string - Common probe meter results for the selected probe channel.

Returns the positive PMETER results for the specified probe channel.

string - Returns the positive PMETER results for the specified probe channel.

Returns the single PMEter result for the selected probe channel.

string - The single PMEter result for the selected probe channel.

Sets the probe coupling setup. The command is part of the PROBe<ch>:SETup group and controls the coupling configuration for the specified probe channel.

Sets the unit used for probe setup advanced measurements. The command belongs to the probe channel setup advanced group.

Executes the gain alignment routine for the specified probe channel. The command writes the resulting alignment information.

NameTypeValuesDescription
chinteger-Probe channel number.

Sets the manual probe attenuation for the specified channel.

NameTypeValuesDescription
chinteger-Probe channel number.
attenuationreal-Manual attenuation value.

Returns the probe attenuation setting. The optional AUTO form queries the automatic attenuation state.

enum - Returns the probe attenuation setting or, with AUTO, the automatic attenuation state.

Returns the probe bandwidth setting for the specified channel.

real - The probe bandwidth setting for the specified channel.

Returns the probe capacitance setup value for the specified channel.

real - Returns the probe capacitance setup value.

Returns the maximum DC range for the specified probe channel.

real - The maximum DC range for the specified probe channel.

Returns the minimum DC range setting for the specified probe channel.

real - The minimum DC range setting for the specified probe channel.

Performs probe degaussing for the specified probe channel.

NameTypeValuesDescription
chinteger-Probe channel number.

Controls whether the probe setup display is differential. The command is listed in the probe setup command index, but this slice does not provide syntax details or parameter/return documentation.

Returns the automatic gain setting for the specified probe channel setup.

enum - Returns the automatic gain state.

Returns the probe input impedance setting for the selected probe channel.

enum - Returns the current probe impedance setting.

Returns the laser state of the selected probe channel.

enum - Returns the laser state of the selected probe channel.

Sets the probe setup mode for the specified channel.

NameTypeValuesDescription
chinteger-Probe channel number.
modeenum-Probe setup mode.

Returns the probe setup name for the specified channel.

string - The probe setup name for the specified channel.

Sets the probe offset for the specified probe channel to the meter-top offset mode. This command is part of the probe setup offset commands.

Selects whether the probe offset uses autozero. This command belongs to the probe setup offset functions.

NameTypeValuesDescription
stateenumUSEautozeroEnables autozero use for probe offset.

Adjusts the probe offset zero adjustment for the selected probe channel. The command appears in the probe setup command list, but this page slice does not provide syntax details, parameters, or a return value.

Sets the probe offset for the specified channel. This command is part of the probe setup subsystem.

NameTypeValuesDescription
chinteger-Channel number.

Returns the probe setup state for the specified channel.

enum - Returns the current probe setup state for the specified channel.

Adjusts the probe termination setup. The command belongs to the probe setup tree for channel-specific probe configuration.

Returns the measured termination condition for the selected probe channel. The command is a query and yields the termination measurement result.

real - Measured termination value for the probe channel.

Sets the probe terminal mode for the specified channel.

NameTypeValuesDescription
chinteger-Probe channel number.
modeenum-Probe terminal mode.

Returns the probe type for the specified channel.

enum - The probe type of the specified channel.

Sets or queries the X-axis mode for a reference curve. The optional XMODe node selects the axis mode used for the referenced curve.

Clears the specified reference curve.

NameTypeValuesDescription
rcinteger-Reference curve number to clear.

Sets the home mode for the reference curve specified by <rc>.

NameTypeValuesDescription
<rc>integer-Reference curve index.

Sets the horizontal scale factor for the selected reference curve. The command is part of the REFCurve<rc>:RESCale hierarchy.

NameTypeValuesDescription
factorreal-Horizontal scale factor.

Sets the horizontal offset for a reference curve rescaling operation.

NameTypeValuesDescription
offsetreal-Horizontal offset value for the reference curve scale.

Turns the horizontal rescale state for the reference curve on or off.

NameTypeValuesDescription
stateenumON | OFFEnables or disables horizontal rescaling for the reference curve.

Restores the referenced curve from saved data. The slice only lists the command name in the command index and does not provide syntax or detailed behavior.

Saves the current reference curve. The command stores the active reference curve data for the selected reference curve index <rc>.

Sets the vertical mode of the reference curve. The command belongs to the REFCurve<rc> subsystem.

Sets the absolute lower limit level for the selected reference level. The command belongs to the REFLevel<rl>:ABSolute group.

NameTypeValuesDescription
LLEVelreal-Absolute lower limit level value.

Sets the absolute middle level for the reference level <rl>.

NameTypeValuesDescription
middle levelreal-Absolute middle level for the reference level <rl>.

Sets the upper level for absolute reference level measurements.

NameTypeValuesDescription
levelreal-Upper absolute reference level.

Sets the relative mode for the reference level. The available mode choices are documented on the command’s detail page, but they are not shown in this slice.

Sets the upper relative reference level for the selected reference level. The command belongs to the REFLevel<rl>:RELative subtree.

:RUNset / query

Starts acquisition in continuous run mode.

:RUNContset / query

Starts continuous run mode. This command is listed alongside RUN and RUNSingle in the command index.

:RUNSingleset / query

Starts a single acquisition sequence. The instrument acquires one trigger event and then stops.

RUNSingle

Updates the current setup reference configuration in the saved setup configuration. This command is listed alongside SAVeset:CONFig:PREView.